High resolution X-ray analysis of a proximal human femur with synchrotron radiation and an innovative linear detector M.Bettuzzi, R. Brancaccio, F.Casali,

Slides:



Advertisements
Similar presentations
CT Scanning: Dosimetry and Artefacts
Advertisements

Image Reconstruction.
IMAGE QUALITY.
Digital Radiography.
Chapter 8 Planar Scintigaraphy
Advanced Biomedical Imaging Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University Lecture 6 Basic physical principles of.
RAD 354 Chapt. 26 Digital Imaging Many types/names for the digital imaging to come – Types CR: Barium fluorohalide PSP SPR (scan projection RAD): Nal scintillator/photodiode.
RAD309 Patient Dose.
SCANCOMEDICAL Computed Tomography SCANCO User Meeting 2005 Dr. Bruno Koller SCANCO Medical AG
Image reconstruction and analysis for X-ray computed microtomography Lucia Mancini 1, Francesco Montanari 2, Diego Dreossi 3 1 Elettra - Trieste 2 A.R.P.A.
First DQE measurment of 500 µm thick Si hybrid pixel sensor with Photon counting readout for X-ray medical imaging M.Caria, J.Chaput, L.Sarry, B.Surre.
IMAGE QUALITY NOISE LINEARITY CROSS-FIELD UNIFORMITY IMAGE ARTIFACTS.
Special Imaging Techniques Chapter 6 Bushong. Dynamic Computed Tomography (DCT) Dynamic scanning implies 15 or more scans in rapid sequence within one.
tomos = slice, graphein = to write
X-Rays and CT Scans.
Conventional and Computed Tomography
LEC ( 2 ) RAD 323. Reconstruction techniques dates back to (1917), when scientist (Radon) developed mathematical solutions to the problem of reconstructing.
Seeram Chapter 11: Image Quality
Basic principles Geometry and historical development
COMPUTED TOMOGRAPHY I – RAD 365 CT - Scan
8/18/2015G.A. Fornaro Characterization of diffractive optical elements for improving the performance of an endoscopic TOF- PET detector head Student: G.
Quality Control Rad T 110.
Dott. Dario Tresoldi CNR IPCF ME
DIGITAL FLUOROSCOPY.
Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.
Frontiers in 3D scanning Prof Phil Withers Manchester X-ray imaging Facility University of Manchester.
Mineral grains from Ivuna ( ) Short summary 4 samples removed from the Ivuna meteorite were imaged at beamline BL47XU of SPring-8 on June 16,
Department of Metallurgy and Materials Engineering Materials Performance and Non-Destructive Testing Optimization and validation of micro-CT for the characterization.
Test of the proposed method Introduction CCD Controller CCD Illuminator gel Filter 585nm Assembling the phantom before its irradiation. The phantom, ready.
Medical Image Analysis Medical Imaging Modalities: X-Ray Imaging Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan,
Computed Tomography Physics, Instrumentation, and Imaging
A-LEVEL PHYSICS 15/09/ :49 Medical Physics.
ITAM CAS, IEAP CTU IWORID 2004Glasgow, th July IWORID 2004 OPTIMIZATION OF X ‑ RAY DYNAMIC DEFECTOSCOPY USING MEDIPIX-2 FOR HIGH FRAME RATE READ-OUT.
Image Quality Radiographic Resolution.
Integral University EC-024 Digital Image Processing.
Computed Tomography Q & A
Computed Tomography Physics, Instrumentation, and Imaging
Factors affecting CT image RAD
LOGO ภาควิชาอิเล็กทรอนิกส์ คณะวิศวกรรมศาสตร์ สถาบันเทคโนโลยีพระจอมเกล้าเจ้าคุณทหารลาดกระบัง.
Performance limits of a 55  m pixel CdTe detector G.Pellegrini, M. Lozano, R. Martinez, M. Ullan Centro Nacional de Microelectronica, Barcelona, 08193,
: Chapter 11: Three Dimensional Image Processing 1 Montri Karnjanadecha ac.th/~montri Image.
Assessing Single Crystal Diamond Quality
THE EFFECT OF NON-IDEAL DETECTORS ON ENERGY WEIGHTED SPECTRA USED IN X-RAY MEDICAL IMAGING George D. Patatoukas 1, Panagiotis F. Liaparinos 1, Anastasios.
Historical Review and Modalities in Diagnostics
Principles and Practice of Radiation Therapy
Computed Tomography Diego Dreossi Silvia Pani University of Trieste and INFN, Trieste section.
CT QUALITY MANAGEMENT. SPATIAL RESOLUTION CONTRAST RESOLUTION NOISE IMAGE ARTIFACTS RADIATION DOSE.
R. Pani Department of Experimental Medicine and Pathology University of Rome La Sapienza-Italy. Flat Panel PMT: advances in position sensitive photodetection.
Part No...., Module No....Lesson No
Modulation Transfer Function (MTF)
Fluoroscopy. Real-time imaging Most general-purpose fluoroscopy systems use TV technology, operating at 30 frames/sec May be recorded (barium swallow.
L2036H1 (031205c). Imaging No.: c Sample: IDP (L2036H1) Beamline: BL47XU X-ray photon energy: 7 keV Exposure time for each projection: 1.5 sec.
Neutron Imaging for the Hydrogen Economy Neutrons see material differently than x-rays The fine details of the water in this Asiatic Lily are clear to.
Electron Spectrometer: Status July 14 Simon Jolly, Lawrence Deacon 1 st July 2014.
IAEA International Atomic Energy Agency General Radiography Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 1(1)
Copyright 2009 CSEM | PhD Seminar, 4 th June 2009| V. Revol | Page 0 X-ray Phase Contrast Imaging Vincent Revol Zürich,
Beam Profile Monitor for Spot-Scanning System Yoshimasa YUASA.
Conclusions References 1. A. Galimberti et al., Nucl. Instrum. Meth. A 477, (2002). 2. F. Capotondi et al., Thin Solid Films 484, (2005).
Thickness of CZT detector 110 MeV140 MeV DETECTOR A (1 mm CZT + 5 mm CZT) DETECTOR B (1 mm CZT + 10 mm CZT) DETECTOR C (1 mm CZT + 15 mm CZT) A. Generation.
DESCRIPTION OF PIXIRAD  The PIXIRAD Imaging Counter is an INFN-Pisa Spin-off  It works in photon counting  Pulse discrimination with two thresholds.
Computed Tomography Computed Tomography is the most significant development in radiology in the past 40 years. MRI and Ultrasound are also significant.
History of digital radiology 1980 Kinos L.T. ET all developed a portable radiographic X-ray camera made of tantalum and aluminum. Digital radiography.
Computed tomography. Formation of a CT image Data acquisitionImage reconstruction Image display, manipulation Storage, communication And recording.
Computed Tomography Basics
Computed Tomography Data Acquisition
Imaging Characteristics
Basic principles Geometry and historical development
Computed Tomography (C.T)
Computed Tomography (C.T)
Computed Tomography (C.T)
Presentation transcript:

High resolution X-ray analysis of a proximal human femur with synchrotron radiation and an innovative linear detector M.Bettuzzi, R. Brancaccio, F.Casali, S. Cornacchia, N.Lanconelli, A.Miceli, M. P.Morigi, A. Pasini, D. Romani, A.Rossi Department of Physics, University of Bologna and INFN, Section of Bologna IEEE NSS – MIC, Rome 21st October 2004

The experiment was carried out in the framework of a project, powered by the University of Bologna, concerning the characterization of the human bone tissue by means of physical techniques Ostheoporosis Project

Good results were already obtained with the 3D micro-CT technique on small bone samples With our micro-CT system we obtained images with a pixel size from 20 to 30 microns (less than 20 microns with high magnification µ–focus source) A good agreement with results achieved by other systems (i.e. Skyscan) was found micro-CT slice

However, such a technique allows to investigate only small samples (1-2 cm maximum size), that means you have to physically cut the bone into slices and make out a carrot of one of them The aim of the present experiment was to verify the possibility of performing the same kind of analysis over a complete human bone (i.e. a femur) without altering its structure by any mechanical treatment Looking forward, this should be a first step toward the hard task of building an X-ray CT system for the structural bone analysis “in vivo”

Trabeculae size in a human bone vary from 0.1 to 0.3 mm while spacing from 0.2 to 2 mm We had to build a new detector with the proper spatial resolution and a sufficiently wide field of view The synchrotron radiation source was choose because it provides a high flux of mono-energetic photons Thus it is the best quality source for this experiment, though it’s not the most suitable one for routine analysis

A new linear intensified detector was designed at the University of Bologna ( EU patented ) The detector is based on two main components: a digital intensified CCD camera (the EBCCD) a special coherent fiber optics adapter

The Electron Bombarded CCD camera Produced by Geosphaera (Moscow)

Scheme of the EBCCD tube

The fiber-optics guide permits the conversion of the geometry from linear to rectangular format Concept Realization From 1024x512 to 5600x50 effective pixels Produced by the Vavilov Institute (S.Petersburg)

Scintillating screen stripe Gd 2 O 2 S:Tb Fiber optics light guide129×1.45 mm 2 to 18.4×10.6 mm 2 Multialkali input window 24.5 mm Back-thinned frame transfer CCD1024×512 Intensifier tube voltage 6 kV Image size (multislice)5600×50 pixel Detector assembly scheme

Mounted detector

66  m 129  m SPLINE interpolation of missing data into the spacings between neighbouring FO ribbons

Microfocus X-ray tube Controlled axis Fiber optics fan adapter EBCCD Scheme of the CT system (microfocus source)

Radiograph of a calibrated bar-pattern direct measurement of the spatial resolution 5 lp/mm (100 µm)

Modulation Transfer Function microfocus X-ray tube measurements (focal spot 5 µm) Spatial resolution of the detector (at 5% measured MTF value) 60 µm

Noise and DQE

Experiments at Elettra SRF in collaboration with Rizzoli Institute, Bologna August April 2004

ELETTRA Syrmep Beamline

Beam size width:120 mm (our detector is 130 mm long) height:4 mm (our detector is 1.5 mm high) Beam energy 34 keV max

Experimental set-up at the SYRMEP beamline beam detector sample proximal femur (pig)

Proximal femur of a pig Maximum size 7,41 x 5,42 cm 2 CT parameters Energy 34 keV Number of projections 500 Angle 180° Voxel side 23  m position of the CT slices 1 2 Each CT section is made of about 35 slices, one pixel thick (23 µm) Original projections size 5600 x 35 pixel Exposure Time: 50 ms Number of averaged frames : 16 Total exposure time for 500 projections: 400 s Total CT time: 8000 s (2 hours e 13 minutes)

1

2

The trabeculae structure becomes visible, thought, spatial resolution is not enough at this level to define it properly

The use of a high resolution linear detector with synchrotron light allowed the definition of the trabecular structure in a CT slice of a complete proximal femur bone of a pig The image quality was too low and it was not suitable for a proper segmentation and for the calculation of the isomorphometric parameters of the bone structure

Improvements better alignment of the detector collection of a large number of projection recalibration and accurate reconstruction human femur analysis

Human femur CT Parameters of the CT 1400 projections over 180 degrees 16 frame averaged for each projection 25 ms exposure time for each frame 2 hours 15 minutes total scanning time

Radiography of a human femur

Micro-CT slice

A high resolution computed tomography of a complete human femur was performed Thus, it is possible to study the trabeculae structure of the complete human femur The high quality of the images obtained after the CT reconstruction allows the calculation of the histomorphometric parameters of the bone

Further steps: Replicate the experiment with a microfocus X-ray tube Optimize the source-detector system Optimize the number of projections for a good image quality with a reduced dose

THANK YOU FOR YOUR ATTENTION