W. Wester, Fermilab DES Callibration Meeting Jan 18, 20131 DECal scans update The DECal flat field system is capable of generating system response maps.

Slides:



Advertisements
Similar presentations
1 Progress in understanding flatness measurements Tom Diehl November 24, 2005 Draft as of 11/24/05.
Advertisements

Image reconstruction and analysis for X-ray computed microtomography Lucia Mancini 1, Francesco Montanari 2, Diego Dreossi 3 1 Elettra - Trieste 2 A.R.P.A.
Optical Astronomy Imaging Chain: Telescopes & CCDs.
1Ellen L. Walker ImageJ Java image processing tool from NIH Reads / writes a large variety of images Many image processing operations.
ABRF meeting 09 Light Microscopy Research Group. Why are there no standards? Imaging was largely an ultrastructure tool Digital imaging only common in.
Measures of Dispersion or Measures of Variability
1 Lites FPP-SP Initial Reduct SOT #17 Meeting, NAOJ, April Solar-B FPP Initial Data Reduction for the FPP Spectro- Polarimeter October, 2004 Bruce.
6/9/2015Digital Image Processing1. 2 Example Histogram.
(conventional Cartesian reference system)
DC Characterization of CCD-Based Detectors for Use in Multi-Chip Focal Plane Arrays SDW 2005 June 2005, R. PhilbrickBall Aerospace & Technologies Corp.
Digital Image Processing
Using ESRI ArcGIS 9.3 3D Analyst T I N
Page 1 © J.Paul Robinson, Purdue University BMS 631 – LECTURE007.PPT BMS 602/631 - LECTURE 8 Flow Cytometry: Theory J. Paul Robinson Professor.
Digital Photography Fundamentals Rule One - all digital cameras capture information at 72 dots per inch (DPI) regardless of their total pixel count and.
Lens ALens B Avg. Angular Resolution Best Angular Resolution (deg) Worst Angular Resolution (deg) Image Surface Area (mm 2 )
Different sources of noise in EM-CCD cameras
Measurement of the absolute efficiency,
NICMOS IntraPixel Sensitivity Chun Xu and Bahram Mobasher Space Telescope Science Institute Abstract We present here the new measurements of the NICMOS.
AST3 detector properties
University of Illinois at Chicago Electronic Visualization Laboratory (EVL) CS 426 Intro to 3D Computer Graphics © 2003, 2004, 2005 Jason Leigh Electronic.
Bulk Silicon CCDs, Point Spread Functions, and Photon Transfer Curves: CCD Testing Activities at ESO Mark Downing, Dietrich Baade, Sebastian Deiries, (ESO/Instrumentation.
Spatial Filtering: Basics
1 COROT WEAEK Meudon May The CCD performances by Miss Pernelle Bernardi Mr Tristan Buey and the CCD team on stage Régis Schmidt, Bertrand le.
Quantum Efficiency Measurements of a NIRSpec infrared sensor in the ODL* Peter McCullough, Project P. I. Mike Regan, ODL Lead Kevin Lindsay Eddie Bergeron.
First Results from Cherwell, a CMOS sensor for Particle Physics By James Mylroie-Smith
Graphics Graphics Korea University cgvr.korea.ac.kr 1 Texture Mapping 고려대학교 컴퓨터 그래픽스 연구실.
CCD Detectors CCD=“charge coupled device” Readout method:
A tour of fundamental statistics introducing Basic Statistics.
September 16, 2008LSST Camera F2F1 Camera Calibration Optical Configurations and Calculations Keith Bechtol Andy Scacco Allesandro Sonnenfeld.
First tests of CHERWELL, a Monolithic Active Pixel Sensor. A CMOS Image Sensor (CIS) using 180 nm technology James Mylroie-Smith Queen Mary, University.
Course Website: Digital Image Processing Image Enhancement (Spatial Filtering 1)
CSC508 Convolution Operators. CSC508 Convolution Arguably the most fundamental operation of computer vision It’s a neighborhood operator –Similar to the.
SNAP Calibration Program Steps to Spectrophotometric Calibration The SNAP (Supernova / Acceleration Probe) mission’s primary science.
Basic Perspective Projection Watt Section 5.2, some typos Define a focal distance, d, and shift the origin to be at that distance (note d is negative)
Chapter 3 Device Monitor Screen Otasuke GP-EX! Chapter 3 Device Monitor Screen Chapter 3 Device Monitor Screen.
Physics 114: Lecture 8 Measuring Noise in Real Data Dale E. Gary NJIT Physics Department.
How to describe Accuracy And why does it matter Jon Proctor, PhotoTopo GIS In The Rockies: October 10, 2013.
F Don Lincoln, Fermilab f Fermilab/Boeing Test Results for HiSTE-VI Don Lincoln Fermi National Accelerator Laboratory.
CCD Image Processing: Issues & Solutions. CCDs: noise sources dark current –signal from unexposed CCD read noise –uncertainty in counting electrons in.
Understanding Color Correction Outline  Color Casts Defined  Channels Defined  Auto Methods of Color Correction-- Levels  Manual Methods of Color Correction--
Muons at CalDet Introduction Track Finder Package ADC Corrections Drift Points Path Length Attenuation Strip-to-Strip Calibration Scintillator Response.
THE CCD RIDDLE REVISTED: SIGNAL VERSUS TIME – LINEAR SIGNAL VERSUS VARIANCE – NON-LINEAR Mark Downing, Peter Sinclaire. European Southern Observatory ESO.
HBD Transmission Monitor Update III: Noise Analysis B. Azmoun & S. Stoll BNL HBD Working Group Meeting 10/10/06.
XRT SOT Alignment DeLuca With comments from Tarbell & Metcalf 21-Jan-2006.
Non-linear filtering Example: Median filter Replaces pixel value by median value over neighborhood Generates no new gray levels.
Describing Data Week 1 The W’s (Where do the Numbers come from?) Who: Who was measured? By Whom: Who did the measuring What: What was measured? Where:
Binomial Distribution Possion Distribution Gaussian Distribution Lorentzian Distribution Error Analysis.
CCD Calibrations Eliminating noise and other sources of error.
Relative Spectral Response and Flat Fields with Internal Calibration Lamps Luisa M. Lara IAA-CSIC Granada (SPAIN)
NAC flat fielding and intensity calibration
Charge Transfer Efficiency of Charge Coupled Device
CCD Image Processing …okay, I’ve got a bunch of .fits files, now what?
Examinations of the relative alignment of the instruments on SOT
Texture Mapping cgvr.korea.ac.kr.
Common Core Math I Unit 1: One-Variable Statistics Boxplots, Interquartile Range, and Outliers; Choosing Appropriate Measures.
Detective Quantum Efficiency Preliminary Design Review
Image Processing, Lecture #8
Digital Image Processing
T. J. Okamoto (NAOJ/Kyoto Univ.)
Image Processing, Lecture #8
Common Core Math I Unit 2: One-Variable Statistics Boxplots, Interquartile Range, and Outliers; Choosing Appropriate Measures.
Modern Observational/Instrumentation Techniques Astronomy 500
Common Core Math I Unit 1: One-Variable Statistics Boxplots, Interquartile Range, and Outliers; Choosing Appropriate Measures.
Common Core Math I Unit 1: One-Variable Statistics Boxplots, Interquartile Range, and Outliers; Choosing Appropriate Measures.
UVIS Calibration Update
6A Types of Data, 6E Measuring the Centre of Data
Miniature Train.
Adding Surface Detail 고려대학교 컴퓨터 그래픽스 연구실.
Adding Surface Detail 고려대학교 컴퓨터 그래픽스 연구실.
IN5350 – CMOS Image Sensor Design
Presentation transcript:

W. Wester, Fermilab DES Callibration Meeting Jan 18, DECal scans update The DECal flat field system is capable of generating system response maps by scanning projected light of known wavelength and intensity onto a flat screen These scans will be taken on a ~monthly basis during engineering or bad weather time. Scans taken at the end of October and November. DES-doc-6885 (Collaboration Meeting talk has details). (1) Monitor changes in throughput (2) Provide system response curves

Summary plot – all filters plotted W. Wester, Fermilab DES Callibration Meeting Jan 18, Spread is range over amplifiers

Take the “none”-Filter scan and correct the measured throughput to get a measurement of the filter throughput. Initial study for the center of one of the center CCDs. W. Wester, Fermilab DES Callibration Meeting Jan 18, Filter measurements

W. Wester, Fermilab DES Callibration Meeting Jan 18, Study system response vs focal plane position No filter, 700nm, ON – OFF with an overscan correction More counts towards center, but no QE nor gain corrections Each CCD is divided into boxes 18 col and 25 rows (110x162 or pixels per box) edge pixels excluded Plot the for the middle 68.2% x/y coordinates are arbitrary S1

Header gain vs gain from photon transfer curves W. Wester, Fermilab DES Callibration Meeting Jan 18, Header quantities GAINA, GAINB Gain from PTC study (Jiangang Hao) Neither is perfect, but the PTC study numbers show less obvious differences between the two amplifiers. Expect the QE to be the same for the entire CCD. Here vertical Is arbitrary with a correction for the light intensity measured by the DECal photodiode. some radial features no QE correction

W. Wester, Fermilab DES Callibration Meeting Jan 18, Study RMS vs focal plane position Gain corrected RMS in each “box” Some interesting features including “corners” where there is typically 40 counts RMS spread compared with other devices with RMS ~20 counts Not exactly sure of the importance or implications.

W. Wester, Fermilab DES Callibration Meeting Jan 18, Study RMS vs focal plane position Gain corrected RMS in each “box” Some interesting features including “corners” where there is typically 40 counts RMS spread compared with other devices with RMS ~20 counts Not exactly sure of the importance or implications. Stable as tape bumps?

W. Wester, Fermilab DES Callibration Meeting Jan 18, Response (no filter) vs focal plane position Calculate the truncated average of ON-OFF (overscan, gain, and intensity corrected) for median “box” for each wavelength (1 point per amplifier) Use 22 CCDs either near the center or towards the edge (total of 44 response curves) Center / Outer CCDs

W. Wester, Fermilab DES Callibration Meeting Jan 18, Response (no filter) vs focal plane position Normalize the response of the center CCD (N4) at 900nm with an edge CCD (S29). Similar difference as when two center CCDs are compared (N4 and S4). This is just a first look. Need to figure out how to best quantify (e.g. normalize to N4?).

Initial look at filter scans as a function of focal plane position W. Wester, Fermilab DES Callibration Meeting Jan 18, Truncated average response (overscan, gain, intensity corrections) General features comparing CCD to CCD seem to hold but not exactly. (You do have to be careful as color scale is not normalized to mean) g-band 500nm Y-band 1000nm

Initial look at filter scans as a function of focal plane position W. Wester, Fermilab DES Callibration Meeting Jan 18, Truncated average response (overscan, gain, intensity corrections) Maybe u-band is a bit different. Note: each CCD has different QE’s g-band 500nm u-band 390nm

W. Wester, Fermilab DES Callibration Meeting Jan 18, Response (g filter) vs focal plane position Normalize the response of the center CCD (N4) at ~530nm with an edge CCD (S29) and a center CCD (S3). Larger difference comparing nearby devices This is just a first look.

W. Wester, Fermilab DES Callibration Meeting Jan 18, Summary and Next steps Tools exist for looking at DECal data as a function of the focal plane position. Even see polishing/tape marks. Gain corrections do a good job removing differences between the two amplifiers on a CCD. It is thought that quantum efficiency variations account for most of the CCD to CCD variation (a next step is to include these). For no filter, a first look shows no gross systematic variations in the throughput versus focal plane position The focal plane looks a bit different at different wavelengths through the appropriate filters when comparing CCD to CCD. For g-filter, a first look also shows no gross systematics in the throughput as a function of focal plane position.