Quantifying Cardiac Deformation by strain (-rate) imaging Hans Torp NTNU, Norway Hans Torp Department of Circulation and Medical Imaging Norwegian University.

Slides:



Advertisements
Similar presentations
Special Topic In Cardiac MR Imaging Medical Imaging & Image Processing Lab MIIP Center for Informatics Science Nile University, Egypt Abdallah G. Motaal.
Advertisements

Chapter3 Pulse-Echo Ultrasound Instrumentation
TV/TVM Tissue Doppler … Introducing Tissue Doppler What is it ?
MEDT8002 Blood flow measurement by Doppler technique Hans Torp Institutt for sirkulasjon og medisinsk bildediagnostikk Hans Torp NTNU, Norway.
MEDT8007 Simulering av ultralydsignal fra spredere i bevegelse Hans Torp Institutt for sirkulasjon og medisinsk bildediagnostikk Hans Torp NTNU, Norway.
Doppler Echocardiography
Doppler Echocardiography
Mina Emad Azmy Research Assistant – Signal and Image Processing Lab.
Foundations of Medical Ultrasonic Imaging
Doppler echocardiography & tissue doppler Dolly mathew.
Lecture 23 Exemplary Inverse Problems including Earthquake Location.
Non-linear acoustics in Ultrasound Imaging Hans Torp Department of Physiology and Biomedical Engineering NTNU NTNUNorway Hans Torp NTNU, Norway.
There are three major types of Blood flow Doppler patterns: 1) Continuous Wave 2) Pulse Wave 3) Color Flow DOPPLER ULTRASOUND.
Mostafa Eladawy MD,AFSA Basics of Echocardiography.
Hospital Physics Group
Artifact cancellation and nonparametric spectral analysis.
Lagrangian measurements using Doppler techniques: Laser and Ultrasound Nicolas Mordant (Ecole Normale Supérieure de Paris) Romain Volk, Artyom Petrosyan,
Artificial Learning Approaches for Multi-target Tracking Jesse McCrosky Nikki Hu.
Master thesis by H.C Achterberg
Dept. Elect. Eng. Technion – Israel Institute of Technology Ultrasound Image Denoising by Spatially Varying Frequency Compounding Yael Erez, Yoav Y. Schechner,
Tracking a maneuvering object in a noisy environment using IMMPDAF By: Igor Tolchinsky Alexander Levin Supervisor: Daniel Sigalov Spring 2006.
DOPPLER SONOGRAPHY By Dr/ Dina Metwaly.
THE ULTRASOUND IMAGE: GENERATION AND DISPLAY
Ultrasonography.
11/18/02Technical Interchange Meeting Progress in FY-02 Research RDA –Capability to collect time series data –Control of phase shifter Phase coding –Sigmet’s.
Introduction to estimation theory Seoul Nat’l Univ.
TISSUE DOPPLER BASICS DR BIJILESH U.
Resident Categorical Course
1 Ultrasonic Elasticity Imaging. 2 Elasticity Imaging Image contrast is based on tissue elasticity (typically Young’s modulus or shear modulus).
Speckle Correlation Analysis1 Adaptive Imaging Preliminary: Speckle Correlation Analysis.
Review of Ultrasonic Imaging
HighlY Constrained Back PRojection (HYPR) Thank you to Oliver Wieben!!
Forward-Scan Sonar Tomographic Reconstruction PHD Filter Multiple Target Tracking Bayesian Multiple Target Tracking in Forward Scan Sonar.
Progress in identification of damping: Energy-based method with incomplete and noisy data Marco Prandina University of Liverpool.
Doppler Ultrasound Dr Mohamed El Safwany, MD.. Introduction The Doppler Effect refers to the change in frequency that results when either the detector/observer.
1SBPI 16/06/2009 Heterodyne detection with LISA for gravitational waves parameters estimation Nicolas Douillet.
December 9, 2014Computer Vision Lecture 23: Motion Analysis 1 Now we will talk about… Motion Analysis.
Spatio-Temporal Free-Form Registration of Cardiac MR Image Sequences Antonios Perperidis s /02/2006.
Systolic Dysfunction in children with ESRD Conventional EchoTissue dopplerSpeckle Tracking Echo.
Saudi Board of Radiology: Physics Refresher Course Kostas Chantziantoniou, MSc 2, DABR Head, Imaging Physics Section King Faisal Specialist Hospital &
Digital Beamforming. Beamforming Manipulation of transmit and receive apertures. Trade-off performance/cost to achieve: –Steer and focus the transmit.
TTK4165 Color Flow Imaging Hans Torp Department of Circulation and Medical Imaging NTNU, Norway Hans Torp NTNU, Norway.
Evaluating Perceptual Cue Reliabilities Robert Jacobs Department of Brain and Cognitive Sciences University of Rochester.
Ultrasonic Imaging Parameters Student: Mei-Ru Yang Wei-Ning Lee Advisor: Pai-Chi Li.
Ultrasound Basis Michel Slama Amiens.
MEDT8002 Ultralyd bildediagnostikk Faglærer: Hans Torp Institutt for sirkulasjon og bildediagnostikk Hans Torp NTNU, Norway.
Performance Issues in Doppler Ultrasound 1. 2 Fundamental Tradeoffs In pulsed modes (PW and color), maximum velocity without aliasing is In pulsed modes,
Motion tracking TEAM D, Project 11: Laura Gui - Timisoara Calin Garboni - Timisoara Peter Horvath - Szeged Peter Kovacs - Debrecen.
2D Strain/Motion Anaylsis Program: An In Vitro Study
Doppler Spectral Analysis
Lecture on Ultrasonography
of Segmental Dysfunction in Myocardial Ischemia
ABSTRACT Background: Although 2D strain imaging has shown promise to define myocardial dysfunction, it provides only subjective assessment about the extent.
and Strain Rate: Validation Against Sonomicrometry
Ventricular Pacing Alters Twisting Synchrony of the Left Ventricle
No relationships to disclose:
Clutter filtering in color flow imaging
MTI RADAR.
Validation Against Sonomicrometry
Spatially Varying Frequency Compounding of Ultrasound Images
2-Dimensional Strain Studies in Open-Chest Instrumented Pigs
Meihua Zhu, MD, PhD; Muhammad Ashraf, MD
Fundamentals Data.
Timing Synchronization with Band Edge Filters
Review of Ultrasonic Imaging
Digital Beamforming.
Segmentation of cardiac MRI using particle filters
Propofol reduces tissue-Doppler markers of left ventricle function: a transthoracic echocardiographic study  J.R. Larsen, P. Torp, K. Norrild, E. Sloth 
Nome Sobrenome. Time time time time time time..
Presentation transcript:

Quantifying Cardiac Deformation by strain (-rate) imaging Hans Torp NTNU, Norway Hans Torp Department of Circulation and Medical Imaging Norwegian University of Science and Technology Norway

Quantifying Cardiac Deformation by strain (-rate) imaging Describe deformation by strain and strain rateDescribe deformation by strain and strain rate Ultrasound methods for strain rateUltrasound methods for strain rate –speckle tracking versus Doppler methods –Clutter noise and thermal noise –angle dependency Frame rate issuesFrame rate issues Visualization of strain and strain-rateVisualization of strain and strain-rate Hans Torp NTNU, Norway

Velocity gradient – strain rate V2 – V1 L = Rate of deformation = Strain Rate V1 V2 Velocity gradient: L (Myocardial) velocity gradient is an instantaneous property

Integrated velocity gradient versus strain ”Growth function” Strain = exp{ IVG } - 1

Integrated velocity gradient versus strain

Envelope RF

What is the best velocity estimator? s1s1 s2s2 Autocorrelation method is optimal (Maximum likelihood estimator) velocity ~ angle(R) RF signals IQ signals x1x1 x2x2 R=x1*  x2

What is the best velocity estimator? R=x1*  x2 v=c/4piT angle(R)

Estimation error is minimum when correlation is maximum Hans Torp NTNU, Norway Correlation magnitude Velocity estimate

Estimate of velocity gradient (strain rate) depth range [mm] Velocity [m/sec] Linear regression Weighted Linear regression (Maximun likelihood)

Simulation experiment Strain rate estimators Simulation no Strain rate Linear regression Weighted Linear regression (Maximun likelihood)

Clutter noise

bias towards zero for velocity measurements increased variance for strain rate Clutter filter helps when tissue velocity is high limited effect in apical region Second harmonic (octave) imaging reduces clutter independent of tissue velocity

Fundamental and second harmonic signal separated by filter FundamentalSignal from septum Noise from LV cavity 2. harmonic Hans Torp NTNU, Norway

Second Harmonic TDI Fundamental, f=1.67MHz Second harmonic, f=3.33MHz Fundamental and second harmonic calculated from the same data setFundamental and second harmonic calculated from the same data set No significant noise differenceNo significant noise difference Second harmonic TDI gives more aliasing.Second harmonic TDI gives more aliasing.

Second Harmonic SRI Fundamental, f=1.67MHz Second harmonic, f=3.33MHz Fundamental and second harmonic calculated from the same data setFundamental and second harmonic calculated from the same data set Significant noise reduction when using the second harmonic frequency bandSignificant noise reduction when using the second harmonic frequency band Aliasing is not a problem due to small velocity differencesAliasing is not a problem due to small velocity differences

Frame rate issues in tissue velocity and strain rate imaging Packet acquisition frames/sec Packet acquisition tissue interleaving frames/sec Continuous acquisition tissue interleaving frames/sec - TVI aliasing - TVI aliasing + Offline spectral Doppler (Work in progress) Image sector: 70 deg. Parallell beams :2

Packet acquisition - continuous acquisition

Myocardial velocity and strain rate with 300 frames/sec Velocity v1v1 v2v2 Strain rate Time Hans Tarp NTNU, Norway

Lateral movement

Angle corrected strain

Summary 1 Strain rate from Tissue Doppler is possible for motion along the ultrasound beam with high temporal resolutionStrain rate from Tissue Doppler is possible for motion along the ultrasound beam with high temporal resolution Weighted linear regression gives minimum estimation errorWeighted linear regression gives minimum estimation error Second harmonic Tissue Doppler reduce clutter noise artefacts in strain rate imagingSecond harmonic Tissue Doppler reduce clutter noise artefacts in strain rate imaging

Summary 2 Integrated strain is improved by tracking material pointsIntegrated strain is improved by tracking material points 2D speckle-tracking gives angle- independent strain, with reduced temporal resolution2D speckle-tracking gives angle- independent strain, with reduced temporal resolution A combination of high frame rate tissue Doppler and lower frame rate speckle tracking is probably the best solution for strain imagingA combination of high frame rate tissue Doppler and lower frame rate speckle tracking is probably the best solution for strain imaging 3D reconstruction of strain (-rate) covering the left ventricle can be obtained from 3 standard apical views3D reconstruction of strain (-rate) covering the left ventricle can be obtained from 3 standard apical views