Relocating corneal endothelial area by non-contact specular microscopy Tomoyuki Kunishige, M.D., Hisaharu Suzuki, M.D., Hideaki Oharazawa, M.D., Toshihiko.

Slides:



Advertisements
Similar presentations
TEMPLATE DESIGN © Learning Effect With Repeated Use of the DynaVision D2 Visual Motor Evaluation William P. McCormack,
Advertisements

Comparison of Visual Outcomes With Toric IOL and With Limbal Relaxing Incisions in Cataract Surgery Eriko Fukuyama, MD Fukuyama Eye Clinic Fukuoka, Japan.
Endothelial Cell Changes as an Indicator for Upcoming Allograft Rejection Following Descemet Membrane Endothelial Keratoplasty Jack Parker Jr, MD; Lamis.
Comparison of Endothelial Cell Count by Manual and Automated Methods in Normal Corneas and in Fuchs' Endothelial Dystrophy Somasheila I Murthy, Debarun.
بسم الله الرحمن الرحيم IN THE NAME OF GOD. Implantation of an Artisan phakic intraocular lens for the correction of high myopia, high hyperopia, aphakia.
Katsuya Yamazoe, MD, Takefumi Yamaguchi, MD, Kazuki Hotta, MD, Yoshiyuki Satake, MD, Kenji Konomi, MD, Seika Den, MD, Jun Shimazaki, MD Presented by: Abdulrahman.
Relationship Between Donor Graft Cell Count and Visual Outcome in DSAEK Patients Authors Silvin Bakalian MD PhD, Johanna Choremis MD FRCSC, Michele Mabon.
Non-Invasive Keratograph Tear Break-Up Time versus Invasive Fluorescein-Associated Tear Stability Test: Experimental Validation Study Nizar S Abdelfattah,
In-the-Bag Toric IOL for Correction of Astigmatism in Keratoconus and after Corneal Surgery Bart T.H. van Dooren, M.D., Ph.D Ilse E.M.A. Mol, M.D. Department.
USE OF THE KONAN NONCON-ROBO SPECULAR MICROSCOPE IN CLINICAL RESEARCH
Phacoemulsification with Goniosynechialysis in the Management of Refractory Acute Angle-closure Glaucoma Ghasem Fakhraie*, MD, Mahmoud Jabbarvand, MD,
Implantation of a single-piece acrylic intraocular lens using an anterior chamber maintainer Tomoyuki Kunishige, Hisaharu Suzuki, Toshihiko Shiwa, Hiroshi.
David Zadok MD, Isaac Avni MD
Clinical Specular Microscopy Corneal Endothelial Cell Morphology
Outcome of refractive surgery in keratoconus suspect eyes
University of Verona Eye Clinic Department of Neurological and Motor Science Head: Prof. Giorgio Marchini Prevalence of Corneal Ectasia in Patients With.
Modified Postoperative Regimen to Increase Retention of Keratoprostheses Juan Carlos Abad, MD Private practice, Medellín - Colombia Carolina Gómez Calle,
Aaron Mulheren, Kathryn Wilson, MSN, APN, Milind Deogaonkar, MD
Kendall R.B. Dobbins, MD Geisinger Medical Center
Comparison of Orbscan and Ultrasound Pachymetry Measurements Faik Orucov, MD, Abraham Solomon, MD,Ziv Caspi, David Landau, MD, Eyal Strassman, MD, and.
Ahmad Kheirkhah MD, Rodrigo Muller MD, Deborah Pavan-Langston MD, Andrea Cruzat MD, Pedram Hamrah MD Ocular Surface Imaging Center, Massachusetts Eye and.
Effects of IOP Lowering Agents on Myopic Regression after Refractive Surgery Lim, Taehyung M.D., Hong, So Jin M.D., Cho, Beom Jin M.D., Ph.D. Chung Kyu-Hyung.
TRANSPALPEBRAL TONOMETER APPLICATION DURING INTRAOCULAR PRESSURE EVALUATION IN THE PATIENTS WITH REFRACTION ANOMALY BEFORE AND AFTER KERATOPHOTOREFRACTIVE.
Evaluation of Epithelial Changes in Limbal Stem Cell Deficiency Using in Vivo Confocal Microscopy ERIC CHAN, Luxia Chen, Sophie X. Deng Cornea and Uveitis.
Evaluation of Interface Reflectivity and Corneal Aberrations following DSAEK Hamid Khakshoor, MD Professor of Ophthalmology Mashhad University of Medical.
The Effects of Deep Brain Stimulation on the Motor Symptoms of Parkinson’s Disease Aaron Mulheren Kathryn Wilson, MSN, APN Milind Deogaonkar, MD Science.
Hong Kong Eye Hospital Ms Frenchy Chiu Dr Victoria Wong IOL master
Ten-year Clinicopathological Review Of Ocular Surface Squamous Neoplasia In An Ophthalmological Center In Mexico City Lucero Pedro-Aguilar, M. D. Alvarez-Melloni.
Seyed Javad Hashemian MD Eye Research Center Rasoul Akram Hospital Tehran University of Medical Sciences No financial interest
Purpose: Introduction:  At initial evaluation: For post-op day # 0 patients: Pre-op VA was 20/50.6 (0.395 ± 0.198); Post-op VA was 20/102.0 (0.196 ± 0.162);
Glaucoma and Penetrating Keratoplasty : Incidence, Risk Factors, and Outcomes Sonika Gupta Consultant Ophthalmology Max Eye Care New Delhi, India Author.
Evaluation of refractive error measurements obtained by three different aberrometers Radha Ram, BA Li Wang, MD, PhD Mitchell P. Weikert, MD, MS Disclosure:
© 2010 Delmar, Cengage Learning 1 © 2011 Delmar, Cengage Learning PowerPoint Presentation to Accompany.
Assessment of macular thickness and function by optical coherence tomography and multifocal electroretinogram in uncomplicated cataract surgery Yang Li,
Khairidzan Mohd Kamal MD
Assessment of Incisional Wounds Before and After Intraocular Lens Insertion In Microincision Cataract Surgery Akimi Kizawa1), Shuichiro Hayashi2), Daijiro.
Phototoxic Maculopathy After Scleral Sutured IOL Implantation NC CHO, DW LEE, EY KWEON, MJ KIM Department of Ophthalmology, Chonbuk National University.
Pattern of Pain and Analgesic use after Collagen Crosslinking for Progressive Keratoconus Ramon Coral Ghanem, MD, PhD. Renan Ferreira Oliveira, MD. Vinicius.
Cincinnati Children’s Hospital Medical Center
1 Non-contact Specular Microscopy for Evaluation of Corneal Endothelium in Early Fuchs’ Endothelial Corneal Dystrophy Jianyan Huang 1, MD, PhD; Tudor Tepelus.
REFRACTIVE OUTCOMES WITH TORIC ICL IMPLANTS CHIEF AUTHOR: Dr. D.RAMAMURTHY CO – AUTHOR: Dr. R.CHITRA The authors have no financial interest in the subject.
Naomi R. Goldberg, MD PhD Kenneth J. Wolf, MD Eric J. Wolf, MD FACS The authors have no financial interest in the subject matter of this poster. Comparison.
Outcomes of Transscleral Sulcus Fixation of Intraocular Lenses through a 2.4-mm Incision with an Injector System: 1-Year Follow-Up Akiko Masai, MD, Tomoichiro.
Outcomes of Transscleral Sulcus Fixation of Intraocular Lenses through a 2.4-mm Incision with an Injector System: 1-Year Follow-Up Akiko Masai, MD, Tomoichiro.
Risk Factors and Outcomes of Donor Lenticule Dislocation Following DSEK Pravin K Vaddavalli MD, Suntia Chaurasia MD, Muralidhar Ramappa MD, Prashant Garg.
Long-Term Longitudinal Change in Keratometry After Pediatric Cataract Surgery Rupal H. Trivedi, MD MSCR M. Edward Wilson, MD Osman Melih, MD Dipankar Bandyopadhyay,
No author has any financial or proprietary interest in any materials or methods mentioned Seung Hyun Kim M.D. ; Tae Hoon Oh M.D. Department of Ophthalmology.
Thomas John, MD Clinical Associate Professor Loyola University at Chicago Maywood, Illinois Ritika Patel Chicago Medical School.
Etiological and decision making factors for repeat DSAEK or PK in cases of failed Descemet stripping automated endothelial keratoplasty First and Presenting.
DSEK for the treatment of endothelial disease in India -Initial Experience in 80 eyes- Authors have no financial interest Dr Ashish Nagpal MD, FRCS Dr.
Diminished Endothelial Cell Counts in Corneas with Krukenberg Spindles Authors: Tod M. Haller1, Chad B. Haller2, Melvin L. Haller1, Judah Beck3, Keith.
Rio de Janeiro Corneal Tomography and Biomechanics Study Group CONCOMITANT KERATOCONUS AND FUCHS DYSTROPHY: TWO DISEASES IS BETTER THAN JUST ONE? Dr. Ambrósio.
The authors have no conflicting interest in the subject matter of this poster. In Vivo Evaluation of DSAEK Interface with Scanning- Laser Confocal Microscopy.
First 25 Descemet Membrane Endothelial Keratoplasty (DMEK) Outcomes for an Experienced DSAEK Surgeon Authors: Victoria Epstein, MS.,DO, Josh Wallsh, MS-III,
Liquifaction Method and Extent of Posterior Capsule Opacification: Two-Year Follow-up Marie Kalfertova, Mariya Burova, Pavel Rozsival, Nada Jiraskova Nada.
Long-term results of Phakic Refractive Lens (PRL™) implantation in high myopic eyes. Ioannis G. Pallikaris 1, 2, MD, PhD, Maria I. Kalyvianaki 1, MD, PhD,
Corneal Endothelial Cell Changes After LASIK Using a Femtosecond Laser Minoru Tomita, MD, PhD, Masaru Aoyama, MD, Minoru Tomita, MD, PhD, Masaru Aoyama,
John P Berdahl M.D.* David R. Hardten M.D., F.A.C.S.* *No relevant financial disclosures.
D.r Nishant Nawani, MS Dr. Surinder Singh Pandav, MD Dr. Amit Gupta, MD Dr. Sushmita Kaushik, MD Advanced Eye Centre PGIMER, Chandigarh The authors have.
Poster# P36 Comparison of Techniques of Measuring Anterior Chamber Depth and Corneal Curvature in Indian Eyes Sujata Das, MS, FRCS Anjula Kumari, B.Optom.
CORNEAL ENDOTHELIAL PROFILE AFTER FERRARA RING IMPLANTATION Leonardo Torquetti, MD, PhD Paulo Ferrara, MD, PhD Paulo Ferrara Eye Clinic Belo Horizonte.
FreeVis LASIK Zentrum Universitätsklinikum Mannheim
Comparison of Endothelial Cell Loss After Phacoemulsification Performed by Third Year Residents and Anterior Segment Surgeons Alexandra Braunstein, MD.
Study of Death to Preservation Time and
Diminished Endothelial Cell Counts in Corneas with Recurrent Erosions
Anil Vedula MD Financial Disclosures: None Richard M. Awdeh MD
Lorelei L. Vicente, M.D. Brian S. Boxer Wachler, M.D.
Yumiko Ban,1,2,3) Seika Den,2) Jun Shimazaki,2,3)
Jianjiang Xu, Wenqing Zhu, Jiaxu Hong
Jonathan M. Davidorf, MD Los Angeles, CA ASCRS Annual Meeting
Presentation transcript:

Relocating corneal endothelial area by non-contact specular microscopy Tomoyuki Kunishige, M.D., Hisaharu Suzuki, M.D., Hideaki Oharazawa, M.D., Toshihiko Shiwa,M.D., Hiroshi Takahashi, M.D. Authors have no finacial interest. Nippon Medical School Hospital

Backgrounds The corneal endothelial damages evaluated by specular microscopy are often considered to indicate invasiveness of cataract surgery. However, the small area of the specular microscopy limits its reliability of analysis, because the microscopic field may not necessarily trace the same area in every examination.

Purpose To examine whether the same corneal area can be located by non-contact specular microscopy in serial observations.

Methods ・ Subjects : ・ Normal Volunteers: 15 eyes of 15 volunteers mean age = 39 years, male : female = 4 : 11 ・ Pseudophakic patients (1 month postop): 15 eyes of 15 patients mean age = 73 years, male : female = 9 : 6 ・ Observation period : April, 2009 ~ January, 2010

A non-contact specular microscope (noncon Robo CA, Konan Medical, Japan) ・ Analysis device ・ Evaluation time intervals Repeated three times at 1-minute intervals and 1month later ・ Evaluation of fixation Fixation status of all subjects, i.e. whether they could fix the target for 10 seconds, was confirmed by phoriascope before each examination.

・ Marker cells and confirmation of their relocation An unique pattern of a cluster of cells was used as a marker. Whether the same pattern can be located in repeated examinations was assessed.

Results In all subjects including normal volunteers and pseudophakic patients, the marker cells could be located in serial specular microscopic observations.

Representative photographs of a normal volunteer The beginning 1 minute later 2 minutes later 1 month later

Representative photographs of a pseudophakic patient The beginning 1 minute later 2 minutes later 1 month later

Discussion Reliable analysis of morphological alterations by non- contact specular microscopy should be based on the fact that the observation objects are the same. However, since the area of non-contact specular microscopy is small, it has not been clarified whether the usual examination by non- contact specular microscopy actually captures the same area or not. In this study, in subjects who can maintain their fixation at least 10 seconds, the same endothelial area can be located by non-contact specular microscopy in serial observations. The results suggest that usual examination by non-contact specular microscopy seems to be reliable enough in analyzing the morphological alterations in corneal endothelium.

Conclusions The same corneal endothelial area can be located by non-contact specular microscopy in repeated examinations.