Our goal is to understand the properties of delocalized vacancies in solid 4 He. Using a triplet of orthogonal diffusion measurements we plan to locate.

Slides:



Advertisements
Similar presentations
Introductory Circuit Analysis Robert L. Boylestad
Advertisements

The L-Network L-networks are used to match the output impedance of one circuit to the input of another. Rsource < Rload, 1< Q < 5 Rsource > Rload, 1
CHAPTER 3: SPECIAL PURPOSE OP-AMP CIRCUITS
MR TRACKING METHODS Dr. Dan Gamliel, Dept. of Medical Physics,
COMMUNICATION SYSTEM EEEB453 Chapter 3 (III) ANGLE MODULATION
Chapter 28. Magnetic Field
Alternating Current Circuits And Electromagnetic Waves Chapter 21.
Magnetic Resonance Imaging
Nuclear Magnetic Resonance
The Electromagnetic Properties of Electric Guitar Pickups Dan Carson, Prof. Steve Errede University of Illinois at Urbana-Champaign Image courtesy of
(random-orientation)
Lecture 2 1 H Nuclear Magnetic Resonance. Gas Chromatograph of Molecular Hydrogen at –100 °C Thermoconductivity Detector 12.
NMR SPECTROSCOPY.
ELEG 479 Lecture #9 Magnetic Resonance (MR) Imaging
The NMR spectrometer Magnet Probe Coils Transmitters
CHAPTER 5: DIFFUSION IN SOLIDS
Integrated Circuits Design for Applications in Communications Dr. Charles Surya Department of Electronic and Information Engineering DE636  6220
Magnetic Resonance Imaging Basic principles of MRI This lecture was taken from “Simply Physics” Click here to link to this site.
Resonance condition. Pulse A coil of wire placed around the X axis will provide a magnetic field along the X axis when a direct current is passed through.
Principles of Electronic Communication Systems
Chapter 6 FM Circuits.
Problem Solving Part 2 Resonance.
Chapter 4 Electrical Principles
Electron Paramagnetic Resonance spectrometer
SPIN 2004 Oct. 14, 2004 W. Kim, S.S. Stepanyan, S. Woo, M. Rasulbaev, S. Jin (Kyungpook National University) S. Korea Polarization Measurements of the.
Physics 121: Electricity & Magnetism – Lecture 11 Induction I Dale E. Gary Wenda Cao NJIT Physics Department.
Principles of MRI Physics and Engineering
Chapter 27 Lecture 12: Circuits.
Electronic Circuits OSCILLATORS.
Alternating-Current Circuits Chapter 22. Section 22.2 AC Circuit Notation.
Magnetic Field Lines for a Loop Figure (a) shows the magnetic field lines surrounding a current loop Figure (b) shows the field lines in the iron filings.
NMR Background Curve Fitting Caitlyn Meditz, on behalf of the Nuclear and Particle Physics Group. Advisor K. Slifer, E. Long Overview Dynamic Nuclear Polarization.
Alternating Current Circuits
Passive components and circuits
Capacitance and Dielectrics
Magnetic Resonance Imaging
STATES OF MATTER Chemistry CP.
Unit-3 Tuned Amplifiers
1 Passive components and circuits - CCP Lecture 4.
Electrical Engineering 348: ELECTRONIC CIRCUITS I Dr. John Choma, Jr. Professor of Electrical Engineering University of Southern California Department.
McGraw-Hill © 2008 The McGraw-Hill Companies, Inc. All rights reserved. Principles of Electronic Communication Systems FM Circuits.
Travelling Wave Tube For Broadband amplifier helix TWTs (proposed by Pierce and others in 1946 ) are widely used For High average power purposes the.
Chapter 8 Alternating Current Circuits. AC Circuit An AC circuit consists of a combination of circuit elements and an AC generator or source An AC circuit.
Oscilloscope The main purposes of the oscilloscope are: Can view the input waveform voltage or current. Can measure instantaneous values of the input waveform.
NMR and The Designing and Manufacturing of an NMR Probe for 109 Ag in Li x Ag 3 V 4 O 11 Cathode Material, for 0.72
R Wyllie, R Wakai, T G Walker University of Wisconsin, Madison Spin-Exchange Relaxation Free Heart signal frequency spectrum from DC-100Hz Adult heart.
Magnetic Resonance Imaging Glenn Pierce, King’s College London, Department of Physics Introduction Edward Purcell and Felix Bloch were both awarded the.
1 Nuclear Magnetic Resonance ANIMATED ILLUSTRATIONS MS Powerpoint Presentation Files Uses Animation Schemes as available in MS XP or MS 2003 versions A.
1 Driven Oscillations Driven Oscillators –Equation of motion –Algebraic solution –Complex algebra solution Phase and amplitude of particular solution –Algebraic.
Crystal Oscillator Circuit and Its Working
CHAPTER 5 DC AND AC BRIDGES.
Principles of Electronic Communication Systems. Chapter 6 FM Circuits.
Electric Pressure Transducer
CLARIDGE CHAPTER 2.2; 3.2.
(Instrument part) Thanundon Kongnok M
Applications of the Canonical Ensemble: Simple Models of Paramagnetism
Amateur Extra Q & A Study Pool
Applications of the Canonical Ensemble:
all Cooper pairs must behave in the same way
10.6 Fourier Transform Mass Spectrometry
10.4 Continuous Wave NMR Instrumentation coherent detection
8.5 Modulation of Signals basic idea and goals
Image construction. Image construction. (A) Step 1: Slice selection. A slice-selecting gradient (GS) is applied at the same time as the excitatory radiofrequency.
MRI: From Protons to Pixels
How MRI Works By Wesley Eastridge, adapted from and with illustrations from The Basics of MRI by Joseph P. Hornak, Ph.D.
Alternating Current Circuits
Nuclear Magnetic Resonance
(4)ELECTRONIC SUPPORT SYSTEM
Presentation transcript:

Our goal is to understand the properties of delocalized vacancies in solid 4 He. Using a triplet of orthogonal diffusion measurements we plan to locate the c-axis of a single crystal sample. This will allow us to measure diffusion as a function of temperature along the c-axis and along the basal plane to look for differences in the activation energy. Positive Ions Diffusion of positive ions on the basal plane takes twice as much energy as on the c-axis. D + c ~ e -E/T ; D + b ~ e -2E/T Hypothesis: Two vacancies are needed to move along the basal plane. Introduction

Why 3 He? Figure 1: Temperature dependence of the diffusion coefficient of impurities in a HCP crystal of 4 He. ●0.75% 3 He, ○2.17% 3 He, □positive ions Previous measurements of 3 He diffusion have assumed isotropy, but the results vary. Activation energies span nearly a factor of 2, which could be explained by the anisotropy hypothesis. 3 He seems to diffuse by the same mechanism as positive ions. In the graph above, the slope of the solid line shows activation energies for both 3 He and positive ion samples. The tails on the right side of the 3 He lines are due to a different diffusion mechanism. Figure 2: Log(D) vs. 1/T from the literature. Suspected angle from c- axis provided.

Cryo-Cooling Figure 3: Phase Diagram for 4 He Figure 4: Schematic of cooling apparatus A single-crystal helium sample requires: temperatures under 2 K 25 atmospheres of pressure 1.Inner Vacuum Can (IVC) isolates the experimental cell 2.IVC immersed in a bath of liquid helium, 4.2 K: hydrogen exchange gas while precooling 3.Thin tube leading to main bath allows liquid helium to enter 1 K pot at about 1 cc/min. Pump attached to 1 K pot activated. Achieves 1.5 K. Thin copper band from 1K plate to 3 He plate provides minimal thermal coupling. 4.Pump for 3 He pot activated. Supply of helium fixed; limited duration run. 3 He plate thermally coupled to sample cell by a copper rod. Cell approximately same temperature as 3 He pot. If 3 He, with lower vapor pressure, were used, the 3 He pot could reach temperatures as low as 0.3 K. For the 1 K lower limit required, 4 He is sufficient.

Resonant Circuit The resonant circuit for NMR is embedded in the cell before crystal formation begins. The crystal grows around the circuit, filling the inductor. As nuclear spins within the solenoid couple to the RF potential across the circuit, the impedance of the inductor changes, producing a signal. Coupling occurs when the precession rate of spins matches the RF frequency. Precession rate is related to the applied magnetic field by B = γω, where γ = 2.04x10 8 / Ts. An inductor and a capacitor in parallel form a primitive resonant circuit with frequency 1/2π(LC) ½. A large resistor (~1 MΩ) in parallel with these and the small coupling capacitor in series with the rest of the circuit match impedance of the circuit with the 50 Ω line impedance at the resonant frequency. Two adjustments are needed to match the real and imaginary components of impedance. R is calculated once and soldered into the cell, while C T is left exposed for easy adjustment. Figure 5: Resonant Circuit with Adjustable Coupling

NMR: Finding a Line RF power is fed into the resonant circuit, producing a reflection. For circuit impedance Z, the reflection is given by: With proper coupling, Z = 50Ω+r, where r is a perturbation caused by aligned spins within the solenoid. Thus, Figure 6: Electrical arrangement for detection of NMR resonance line.

This reflection is amplified and mixed with the original signal to produce a DC signal proportional to the power reflected. Care is taken to match electrical path lengths of the reflected and synchronization signal so that any phase difference is due to a complex impedance in the circuit. The magnetic field in the cell is modulated in the audio range so that any change in reflection is picked up as an audio signal on top of the DC output of the mixer. The signal is then sent through a transformer to strip out the DC component. The remaining audio signal is fed into a lockin detector, where any change in reflection will appear as a signal synchronized with the original audio input. The NMR line is found by slowly sweeping the static field in the cell while watching the lockin detector. Because of the extreme precision required (greater than 1 part in 1000), a smaller coil within the main solenoid is swept through lower currents to pinpoint the exact field of resonance.

Measuring Diffusion NMR requires producing a net dipole moment in the atoms of the sample with a strong magnetic field. In addition, we apply a static field gradient to the sample in the direction to be measured. Three pairs of magnet coils produce gradients in three orthogonal directions. Any gradient desired can be created by linear superposition of these. Only a thin plane normal to the gradient is then in resonance. The width of the slice (W) is inversely proportional to the strength of the gradient. An RF pulse at the Larmour frequency tilts the spins of 3 He atoms in this slice only. A sufficiently strong pulse will randomize the spins in the resonant slice, reducing the reflected signal from the circuit. As the randomized spins diffuse out of the slice and are replaced by preferentially oriented spins, the reflected signal will recover with some characteristic time constant τ. The diffusion coefficient along the gradient is then D = W 2 /τ. Because of the gradient, spins will precess at different rates throughout the slice, necessitating additional compensating techniques.

Locating the C-axis Diffusion along the c-axis and diffusion on the basal plane potentially have different diffusion constants and activation energies. The diffusion constant for a gradient at an angle θ from the c-axis must be given for an HCP lattice by: When measurements are made in three orthogonal directions, cos 2 (θ 1 )+ cos 2 (θ 2 )+ cos 2 (θ 3 )=1. With D c, D b, E c, and E b as fitting parameters, measurements at three different temperatures are sufficient to locate the c-axis. The field gradient can then be oriented along the c-axis and along at least one vector on the basal plane. The diffusion constant in each direction can be assessed at various temperatures. Figure 8: x, y, and z measurements used to locate c-axis.

Data Analysis To determine the activation energy needed for diffusion in a particular direction, the diffusion coefficient must be measured at various temperatures covering the range of interest (in this case from 1 K to nearly 2 K). The energy is extracted using the following prescription. References: S.C. Lau and A.J. Dahm, J. Low Temp Phys. 112, 47 (1998). K.O. Keshishev, JETP 45, 273 (1977). A plot of ln(D) vs. 1/T is fit to a straight line. The slope of the line will be –E.