New Flowcell – How Easy? 1
2
SenCell 3
Small, robust, easy to handle Stepless variable spacing 4
SenCell - Specifications Working electrode GC 2mm Reference electrode Salt Bridge - blue ISAAC - green HYREF – black 5
Sensitivity – wall jet Flow cell design A. thin-layer B. wall-jet C. porous flow-through 6
Cotrell Equation 7 ECD Cottrell equation i lim = 1.47 n F (F m ) 1/3 (DA/b) 2/3 C i lim = C spacer thickness A= WE area A b = spacer thickness C = concentration
Variable spacing – more signal Stepless variable spacing Smaller spacing Higher signal 8
Spacing 9
Sensitivity – more signal 100um vs 12 um spacing 10
Sensitivity – more signal Stepless variable spacing Smaller spacing Higher signal Spacing SenCell (µm) S/N of 100 nM standard DA 5-HT
Sensitivity LOD 50 pM DA (0.23 fmol) 65 pM 5-HT (0.25 fmol) 100 pmole/L standards in Hac + Ringer, 5 uL injection 12
Pressure Higher flow rate due to UHPLC Smaller spacing down to 10um 13
Sencell Max Pressure Leak tight ring Rating: 25 bar 14
UHPLC compatible High speed analysis Fast response Noise level time (min) [nA] Current DA 5-HT 100 nmole/L standards in Hac, 5 uL injection 15
UHPLC compatible time (min) [nA] Current DA 5-HT 100 nmole/L standards in Hac, 5 uL injection 16
Conclusion SenCell – The most Easy to use flowcell Wall jet – performance Variable spacing - more signal Less noise by improved fluid dynamics UHPLC compatible 17