Vacuum II G. Franchetti CAS - Bilbao 30/5/2011G. Franchetti1.

Slides:



Advertisements
Similar presentations
General Characteristics of Gas Detectors
Advertisements

Introduction to Plasma-Surface Interactions Lecture 6 Divertors.
Ion Sources Some characteristics of ion sources (especially in high precision work): It should have high efficiency in generating ions of the element of.
Chapter 2 Introduction to Heat Transfer
CHAPTER 5: Mass and Energy Analysis of Control Volumes
Overview of Vacuum Systems Tejas Deshpande 24 July, 2014.
427 PHC.  Atomic emission spectroscopy (AES) is based upon emission of electromagnetic radiation by atoms.
Center for Materials for Information Technology an NSF Materials Science and Engineering Center Vacuum Fundamentals Lecture 5 G.J. Mankey
TEST GRAINS AS A NOVEL DIAGNOSTIC TOOL B.W. James, A.A. Samarian and W. Tsang School of Physics, University of Sydney NSW 2006, Australia
1 Introduction to Plasma Immersion Ion Implantation Technologies Emmanuel Wirth.
L.B. Begrambekov Plasma Physics Department, Moscow Engineering and Physics Institute, Moscow, Russia Peculiarities, Sources and Driving Forces of.
1 NEG films: recent R&D progress Paolo Chiggiato (for the EST-SM-DA section) Vacuum Issues of the LHCb Vertex Detector 28 November NEG films: choice.
Mass Spectroscopy Mass Spectrometry ä Most useful tool for molecular structure determination if you can get it into gas phase ä Molecular weight of.
Vacuum equipment. Pumps Rotary vane pumps Rotary piston pumps Roots pumps Turbo molecular pumps Diffusion pumps Getter pumps –Sublimation pumps.
Molecular Mass Spectroscopy Chem Introduction In Mass Spectroscopy (MS), atomic and molecular weights are generally expressed in terms of atomic.
Vacuum systems Electron beam – mean free path: Gun – column - sample Signal detection – electron and X- ray collection: Scattering of emitted electrons.
Basics of Vacuum Technology Unit: Pa = N/m 2 = J/m 3 UnitPa or N/m 2 bar10 5 mbar100 atm= 760 torr x 10 5.
Mass Spectrometry.
Viscosity. Average Speed The Maxwell-Boltzmann distribution is a function of the particle speed. The average speed follows from integration.  Spherical.
Kinetic Theory The kinetic theory is an explanation of how particles in matter behave.
Photoelectron Spectroscopy Lecture 5 – instrumental details –General spectrometer design –Vacuum generation and measurement.
High-Vacuum Technology Course
MECHANISMS OF HEAT TRANSFER
Solids, Liquids, and Gases. Kinetic Theory The kinetic theory is an explanation of how particles in matter behave. The three assumptions of the kinetic.
Dry Etching, General Principles Dr. Marc Madou, Fall 2012 Class 7.
Pressure 1 atmosphere ~ 1 bar ~ 760 mm Hg ~ 760 torr ~ 100,000 Pa Ion gauges read in mbar i.e. 1x mbar = 1x atm. Sometimes ion gauges read.
Atomic Emission Spectroscopy
the Ionosphere as a Plasma
Current and Direct Current Circuits
Introduction to Plasma- Surface Interactions G M McCracken Hefei, October 2007.
F.M.H. Cheung School of Physics, University of Sydney, NSW 2006, Australia.
PC4250 Secondary Ion Mass Spectrometry (SIMS). What is SIMS? SIMS is a surface analysis technique used to characterize the surface and sub-surface region.
20-21 January 2009, RAL Joint DL-RAL Accelerator Workshop 1 INVESTIGATION OF NON- EVAPORABLE GETTER FILMS O. B. Malyshev, K.J. Middleman, A. Hannah and.
Vacuum science.
Methods and Tehniques in Surface Science Prof. Dumitru LUCA “Alexandru Ion Cuza” University, Iasi, Romania.
Vacuum Fundamentals 1 atmosphere = 760 mm Hg = kPa 1 torr = 1 mm Hg vacuum range pressure range low 760 ~ 25 torr medium 25~ high ~ 10.
Techniques of Vacuum and Basics of High Voltage (2/3) Pauli Heikkinen Jyväskylä University.
Why plasma processing? (1) UCLA Accurate etching of fine features.
Buffer Gas Cooling of atomic and molecular beams Wenhan Zhu Princeton University 11/06/2007.
Vacuum Fundamentals 1 atmosphere = 760 mm Hg = kPa 1 torr = 1 mm Hg vacuum range pressure range low 760 ~ 25 torr medium 25~ high ~ 10.
CHAPTER 5: Mass and Energy Analysis of Control Volumes
Chapter 21: Molecules in motion Diffusion: the migration of matter down a concentration gradient. Thermal conduction: the migration of energy down a temperature.
High-Vacuum Technology Course
Center for Materials for Information Technology an NSF Materials Science and Engineering Center Vacuum Systems Lecture 6 G.J. Mankey
GPS S8P1. Students will examine the scientific view of the nature of matter. b. Describe the difference between pure substances (elements and compounds)
Sputter deposition.
Vacuum Workshop on Vacuum Systems for Synchrotron Light Sources, Barcelona, Vacuum Technology, 12 th and 13 th September 2005 L. Schulz Introduction Pumps.
Ionization Detectors Basic operation
Vacuum Technology.
Analytical Separations
Chapter 21: Molecules in motion Diffusion: the migration of matter down a concentration gradient. Thermal conduction: the migration of energy down a temperature.
Rutherford’s Model: Conclusion Massive nucleus of diameter m and combined proton mass equal to half of the nuclear mass Planetary model: Electrons.
-Plasma can be produced when a laser ionizes gas molecules in a medium -Normally, ordinary gases are transparent to electromagnetic radiation. Why then.
1 Vacuum chambers for LHC LSS TS Workshop 2004 Pedro Costa Pinto TS department, MME group Surface Characterization & Coatings Section.
Physical Science Heat and Matter. Matter Anything that occupies space and has mass Ex. Air Law of Conservation of Matter Matter is neither created or.
Heat Transfer by Convection
 Has fixed volume  Has fixed shape  Molecules are held in specific locations  by electrical forces  vibrate about equilibrium positions  Can be.
Jari Koskinen 1 Thin Film Technology Lecture 2 Vacuum Surface Engineering Jari Koskinen 2014.
SPES Target Group Data…… INFN-CISAS-CNR collaboration The Ablation Ion Source for refractory metal ion beams A preliminary design.
Vacuum Systems for Electron Microscopy Constraints on Specimens Specimens placed in the electron microscope must be able to withstand very high vacuum.
DOE Plasma Science Center Control of Plasma Kinetics
PVD & CVD Process Mr. Sonaji V. Gayakwad Asst. professor
UNIT - 4 HEAT TRANSFER.
Mass Spectroscopy. Mass Spectroscopy Mass Spectrometry Most useful tool for molecular structure determination if you can get it into gas phase Molecular.
V. Measurement pressure gauges, flow meters
VISUAL AIDS for instruction in VACUUM TECHNOLOGY AND APPLICATIONS
Ionization detectors ∆
VIII. Production of Vacuum
1.6 Glow Discharges and Plasma
Molecular Mass Spectroscopy
Presentation transcript:

Vacuum II G. Franchetti CAS - Bilbao 30/5/2011G. Franchetti1

Index 30/5/2011G. Franchetti2 Creating Vacuum (continuation) Measuring Vacuum Partial Pressure Measurements

Diffusion Ejector pump 30/5/2011G. Franchetti3 Laminar flow Inlet Outlet Cold surface Boiling oil Schematic of the pump operating pressure: – mbar

30/5/2011G. Franchetti4 Laminar flow Inlet Outlet Cold surface Boiling oil PiPi PoPo Pump principle: The vacuum gas diffuses into the jet and gets kicked by the oil molecules imprinting a downward momentum The oil jets produces a skirt which separate the inlet from the outlet

30/5/2011G. Franchetti5 Inlet Outlet Diffusion

Problems 30/5/2011G. Franchetti6 Laminar flow Outlet Cold surface Boiling oil Pi Po Inlet Back streaming Back-Migration

Cures 30/5/2011G. Franchetti7 Laminar flow Outlet Cold surface Boiling oil Pi Po Inlet Baffle (reduces the pumping speed of 0.3) Cold surface Cold Cap

30/5/2011G. Franchetti8 with The pumping speed S m is proportional to the area of the inlet port 100 mm diameter  S m = ~ 250 l/s for N 2 LEYBOLD (LEYBODIFF)

Capture Vacuum Pumps 30/5/2011G. Franchetti9 Getter Pumps (evaporable, non-evaporable) Sputter ion Pumps Cryo Pumps Principle Capture vacuum pumps are based on the process of capture of vacuum molecules by surfaces

Getter Pumps 30/5/2011G. Franchetti10 Solid Bulk Surface Gas

11 Heating in vacuum Oxide dissolution -> activation T = RT T = T a T = RT NEGs pump most of the gas except rare gases and methane at room temperature Native oxide layer -> no pumping Pumping Getters are materials capable of chemically adsorbing gas molecules. To do so their surface must be clean. For Non-Evaporable Getters a clean surface is obtained by heating to a temperature high enough to dissolve the native oxide layer into the bulk. Definition of NEG P. Chiggiato

30/5/2011G. Franchetti12 C.Benvenuti, CAS 2007 Sorption Speed and Sorption Capacity

13 energy carrier: noble gas ions substrate to coat: vacuum chamber target material: NEG (cathode) driving force: electrostatic Choice of the coating technique for thin film: sputtering NEG composition  The trend in vacuum technology consists in moving the pump progressively closer to the vacuum chamber wall.  The ultimate step of this process consists of transforming the vacuum chamber from a gas source into a pump.  One way to do this is by “ex-situ” coating the vacuum chamber with a NEG thin film that will be activated during the “in situ” bakeout of the vacuum system.

Dipole Coating Facility Quadrupole Coating Facility 30/5/2011G. Franchetti14 M.C. Bellachioma (GSI)

Sputter ion pumps 30/5/2011G. Franchetti15 B E Anode Cathode E P1 P2 Titanum

Sputtering process 30/5/2011G. Franchetti16 + The adsorbing material is sputtered around the pump trapped electrons

30/5/2011G. Franchetti17 A glimpse to the complexity Adsorbed ions Trapped electron ionization process ion bombardment with sputtering

30/5/2011G. Franchetti18 Example of Pumping speed J.M. Laffterty, Vacuum Science, J. Wiley & Son, 1998

Cryo Pumps 30/5/2011G. Franchetti19 m v1v1 Stick to the Wall !! Dispersion forces between molecules and surface are stronger then forces between molecules Cold Wall

30/5/2011G. Franchetti20 cold Vessel nwnw pwpw ncnc pcpc Pump P w = pressure warm P c = pressure in the cold I w = flux Vessel  Pump I c = flux Pump  Vessel molecules stick to the cold wall Schematic of a cryo pump

30/5/2011G. Franchetti21 Now By using the state equation In the same way Ifno pumping although Thermal Transpiration Relation between pressures

30/5/2011G. Franchetti22 When the two pressures breaks the thermal transpiration condition a particle flow starts We find We define and P c depends on the capture process Cryocondensation:P w is the vapor pressure of the gas at T c

Summary on Pumps 30/5/2011G. Franchetti23 N. Marquardt

Gauges 30/5/2011G. Franchetti24 Liquid Manometers MacLeod Gauges Viscosity Gauges Thermal conductivity Gauges Hot Cathode Gauges Alpert-Bayard Gauges Penning Gauges

Liquid Manometers 30/5/2011G. Franchetti25 P2P2 P1P1 h Relation between the two pressure issue: measure precisely “h”by eye +/- 0.1 mm High accuracy is reached by knowing the liquid density but with mercury surface tension depress liquid surface Mercury should be handled with care: serious health hazard

McLeod Gauge 30/5/2011G. Franchetti26 Δh Closed h0h0 Quadratic response to P 2 First mode of use: The reservoir is raised till the mercury reaches the level of the second branch (which is closed) P2P2 Second mode of use: keep the distance Δh constant and measure the distance of the two capillaries  linear response in Δh

Viscosity Gauges 30/5/2011G. Franchetti27 Gas molecule ω φ It is based on the principle that gas molecules hitting the sphere surface take away rotational momentum The angular velocity of the sphere decreases R ρ = sphere density α = coefficient of thermal dilatation T = temperature v a = thermal velocity

30/5/2011G. Franchetti28 (K. Jousten, CAS 2007) F.J. Redgrave, S.P. Downes, “Some comments on the stability of Spinning Rotor Gauges”, Vacuum, Vol. 38,

Thermal conductivity Gauges 30/5/2011G. Franchetti29 hot wire A hot wire is cooled by the energy transport operated by the vacuum gas Accommodation factor By measuring dE/dt we measure P

Energy Balance 30/5/2011G. Franchetti30 V W c /2 WRWR  Energy loss by gas molecules  Energy loss by Radiation  Energy loss by heat conduction When the energy loss by gas molecule is dominant P can be predicted with contained systematic error ε = emissivity

30/5/2011G. Franchetti31 The Gauge tube is kept at constant temperature and the current is measured So that Through this value  E Pirani Gauge Example of power dissipated by a Pirani Gauge vs Vacuum pressure.

Ionization Gauges Principle 30/5/2011G. Franchetti32 electrons accelerating gap L V +

30/5/2011G. Franchetti33 V + IONIZATION positive ions new electrons i + = current proportional to the ionization rate E i-i-

30/5/2011G. Franchetti34 Electrons ionization rate Sensitivity

Hot Cathode Gauges 30/5/2011G. Franchetti35 Grid Anode Hot Cathode 30 V + 180V + Electric field

Hot Cathode Gauges 30/5/2011G. Franchetti36 Grid Anode Hot Cathode 30 V + 180V + In this region the electrons can ionize vacuum particles A current between grid and anode is proportional to the vacuum pressure i+

Limit of use 30/5/2011G. Franchetti37 Upper limit:it is roughly the limit of the linear response Lower limit:X-ray limit Grid Anode new electron X-ray The new electron change the ionization current P PmPm X-ray limit

Alpert-Bayard Gauge 30/5/2011G. Franchetti38 Grid Anode X-ray Hot Cathode the probability that a new electron hits the anode is now very small The X-ray limit is easily suppressed by a factor ~

30/5/2011G. Franchetti39 Schematic of the original design (K. Jousten, CAS 2007)

Penning Gauge (cold cathode gauges) 30/5/2011G. Franchetti40 B = magnetic field E = electric field Electrons motion anode cathode

30/5/2011G. Franchetti One electron is trapped Under proper condition of (E,B) electrons get trapped in the Penning Gauge

30/5/2011G. Franchetti One electron is trapped one vacuum neutral gas enters into the trap

Ionization process 30/5/2011G. Franchetti43 Neutral vacuum atom Electron at ionization speed Before CollisionCollision  Ionization Charged vacuum atom Electron at ionization speed New electron

30/5/2011G. Franchetti Ionization Charged vacuum atom Electron at ionization speed New electron this ion has a too large mass and relatively slow velocity, therefore its motion is dominated by the electric field and not by the magnetic field

30/5/2011G. Franchetti Ionization Charged vacuum atom New electron Motion of stripped ion

30/5/2011G. Franchetti Negative space charge formation More electrons are formed through the ionization of the vacuum gas and remains inside the trap

30/5/2011G. Franchetti When the discharge gets saturated each new ionization produce a current I

30/5/2011G. Franchetti48 The time necessary for the discharge formation depends upon the level of the vacuum lower pressure  longer time of formation Sensitivity of a SIP (the same as for the Penning Gauge). J.M. Lafferty, Vacuum Science,p. 322

Summary on Gauges 30/5/2011G. Franchetti49 N. Marquardt

Partial Pressure Measurements 30/5/2011G. Franchetti50 These gauges allow the determination of the gas components Partial pressure gauges are composed Ion Source Mass Analyzer Ion current detection System data output

Ion Sources 30/5/2011G. Franchetti51 Vacuum gas is ionized via electron-impact  the rate of production of ions is proportional to each ion species Electron-impact ionization process e v inelastic scattering: kinetic energy transfer from the electron to the molecule meme M Before Impact After impact e v meme M+M+ e v meme

30/5/2011G. Franchetti52 The minimum energy to ionize M is called “appearance potential” Electron energy (eV) 15 IP = 15 eVM + + e - At the appearance potential the production rate is low appearance potential A. von Engel, Ionized Gases, AVS Classics Ser., p. 63. AIP Press, 1994

Ion source 30/5/2011G. Franchetti53 ionization region electrons source acceleration gap VfVf VgVg electrons ions ion collector E E VcVc Open source Bayard-Alpert gauge V g < V c

30/5/2011G. Franchetti A schematic F = ion transmission factor Mass analyzer Ion Detection

30/5/2011G. Franchetti55  Faraday Cup  Secondary electron Multiplier (SEM) Idea: an ion enters into the tube, and due to the potential is accelerated to the walls. At each collision new electrons are produced in an avalanche process # electrons output # ion input G = Gain L d V 0 = applied voltage V = initial energy of the electron K = δ V c where δ = secondary emission coefficient V c = collision energy J.Adams, B.W. Manley IEEE Transactions on Nuclear Science, vol. 13, issue 3, p. 88

Mass Analyzers 30/5/2011G. Franchetti Ion equation of motion U, V constant r0r0 define:

30/5/2011G. Franchetti57 By rescaling of the coordinates the equation of motion becomes Mathieu Equation Stability of motion horizontalvertical q=0, a>0unstablestable q=0, a<0stableunstable The presence of the term q, changes the stability condition Development of stable motion The ion motion can be stable or unstable Stable or unstable motion is referred to a channel which is infinitely long, but typically a length correspondent to 100 linear oscillation is considered enough Example: For N 2 at E k = 10 eV  v s = 8301 m/s For a length of L = 100 mm 100 rf oscillation f = 8.3 MHz typically f ~ 2 MHz

30/5/2011G. Franchetti a q Stable in y a q Stable in x Stability Chart of Mathieu equation

30/5/2011G. Franchetti a q a Stability region in both planes q a q 0 =0.706 a 0 =0.237

30/5/2011G. Franchetti60 Stability region in both planes q a (0.706; 0.237) Given a certain species of mass M 1 there are two values U 1, V 1 so that q=q 0 and a=a 0 By varying V and keeping the ratio V/U constant, the tip of the stability is crossed and a current is measured at V=V 1 V i+i+

30/5/2011G. Franchetti61 Magnetic Sector Analyzer Example: A 90 0 magnetic sector mass spectrometer B is varied and when a current is detected then E z is the ions kinetic energy Resolving power W source = source slit width W collector = collector slit width J.M. Lafferty, Vacuum Science, p. 462

Omegatron 30/5/2011G. Franchetti62

30/5/2011G. Franchetti63 The revolution time is independent from ion energy If the frequency of the RF is 1/τ a resonant process takes place and particle spiral out collector B E Resolving power

Conclusion 30/5/2011G. Franchetti64 Creating and controlling vacuum will always be a relevant part of any accelerator new development. These two lectures provides an introduction to the topic, which is very extensive: further reading material is reported in the following bibliography THANK YOU FOR YOUR ATTENTION

30/5/2011G. Franchetti65 Kinetic theory and entropy,C.H. Collie, Longam Group, 1982 Thermal Physics, Charles Kittel, (John Wiley and Sons, 1969). Basic Vacuum Technology (2 nd Edn), A Chambers, R K Fitch, B S Halliday, IoP Publishing, 1998, ISBN Modern Vacuum Physics, A Chambers, Chapman & Hall/CRC, 2004, ISBN The Physical Basis of Ultrahigh Vacuum, P. A. Redhead, J. P. Hobson, E. V. Kornelsen, AIP, 1993, ISBN Foundation of Vacuum Science, J.M. Lafferty, Wiley & Sons, 1998 Vacuum in accelerators, CERN Accelerator School, CERN June 2007 Vacuum technology, CERN Accelerator School, CERN Acknowledgements: Maria Cristina Bellachioma