Assessment of MRI Induced Heating of Cardiac Pacing Leads: Relevance to Target Patient Populations Presenter: Daniel G. Hullihen Jr. Director of Business.

Slides:



Advertisements
Similar presentations
Magnetic Resonance Imaging
Advertisements

Sanrad-XinAo MDT.
The importance of MRI, a few numbers  MRI units worldwide in 2003  75 millions scans per year performed  Constant need for over 1000 MRI technologists.
MRI in patients with pacemakers DR PRADEEP SREEKUMAR.
H MRI Marker Active MRI Stents & Vena Cava Filter -brief introduction- Andreas Melzer University of Applied Sciences Gelsenkirchen Institute for Innovative.
Radio Frequency Induced Heating Near Medical Devices During MRI Justin Peterson University of Western Ontario Medical Biophysics March. 23, 2011.
Functional Brain Signal Processing: EEG & fMRI Lesson 12 Kaushik Majumdar Indian Statistical Institute Bangalore Center M.Tech.
Equipment Magnetic resonance imaging (MRI) scan requires the use of a very strong magnetic field. Unlike other devices used in radiology, MR imaging.
ELEG 479 Lecture #9 Magnetic Resonance (MR) Imaging
Electrical eddy currents in the human body: MRI scans and medical implants Brent Hoffmeister Rhodes College Department of Physics.
Topics in Medical Physics Xiaoming Zheng, PhD. School of Dentistry and Health Sciences December 2009, Chengdu.
FMRI: Biological Basis and Experiment Design Lecture 18: Physical practicalities Digression: analysis ICE9: Example for WA8 Safety limits –dB/dt –SAR –Acoustic.
MRI: MAGNETIC RESONANCE IMAGING
Regulatory Approval of New Devices SCMR 2015 Interventional CMR Workshop Technology for Interventional cMRI Steve Wedan.
Amyotrophic Lateral Sclerosis (ALS) Sarah Belair and Hannah McLaughlin.
Magnetic Resonance Imaging Astounding. The MRI magnet and patient table The operator’s console.
The UE directive 2004/40/EC and Magnetic Resonance Imaging a debate Thierry Metens Ph D Université Libre de Bruxelles, Hôpital Erasme Brain awareness week.
M. D. A NDERSON Cancer Center Magnetic Resonance Imaging: 4.7 T, 40 cm Bruker Biospec James A Bankson, Ph.D. and John D Hazle, Ph.D. Small Animal Cancer.
Software Development For Correction of Gradient- Nonlinearity Distortions in MR Images T.S. Lee, K.E. Schubert Computer Science CSUSB R.W. Schulte Radiation.
September, 2003BME 1450 Introduction to NMR 1 Nuclear Magnetic Resonance (NMR) is a phenomenon discovered about 60 years ago.
US FUNDAMENTALS: B-MODE AND DOPPLER MACHINE SETTINGS.
Lecture 4 13 C NMR: DEPT IR Spectroscopy: - How it works - Interpretation of spectra Due: Lecture Problem 2.
Influence of Pacing Lead Design on MRI-Induced Lead Heating Presenter: Stuart MacDonald VP Research and Development Heart Rhythm 2006 Boston, MA May 18,
Human Subjects in fMRI Research Credits: Robert Savoy, Ph.D. Franz Schmitt, Ph.D.
MR Scanner Transmit Gain as an Indicator For Lead Heating November, 2006.
Functional Brain Signal Processing: EEG & fMRI Lesson 17 Kaushik Majumdar Indian Statistical Institute Bangalore Center M.Tech.
OBJECTIVE Without reference, identify at least four out of six principles pertaining to the application of transducers related to patient care.
Anthropometrics.
Active Detuning of Inductively Coupled Surface Coils Jolinda Smith Lewis Center for Neuroimaging University of Oregon.
Pacemakers and MRI An outdated contraindication? Emily Myers Research in Allied Health East Tennessee State University.
Laser Treated Metallic Probes for Cancer Treatment in MRI Systems July 08, Advance Materials Processing and Analysis Center (AMPAC) Department of.
Magnetic Resonance Imaging
BC10 Small Extremity Coil (1.5T or 3T) GE MR System Key Benefits:  Optimized design for both high SNR and image uniformity  Excellent image quality when.
University of Western Ontario
Tom S. Lee, Keith E. Schubert Reinhard W. Schulte
Biophan Technologies, Inc. Jeffrey L. Helfer, Vice President of Engineering ISMRM Workshop on MRI Safety November 6, 2005.
Population of 4D Computational Phantoms for CT Imaging Research and Dosimetry W. Paul Segars, PhD Carl E. Ravin Advanced Imaging Labs Duke University.
201 MHz Cavity Plans Derun Li Center for Beam Physics Lawrence Berkeley National Laboratory February 11, 2011 MAP Friday Video Conference Meeting.
Medical Resonance Imaging MRI. First medical images: X-rays Discovered in 1895 Images of bones What part of the body is this?
Chapter 12 Therapeutic Heating Applications of Radio Frequency Energy C-K. Chou.
CEREBRAL BLOOD FLOW QUANTIFICATION FOR ARTERIAL SPIN LABELED PERFUSION MRI AT 3 TESLA 4° Congresso Annuale dell’Italian Chapter dell’ISMRM Perugia,
Human primary cells These cells taken from living tissue are extremely accurate as they are literally coming from the source and are available from many.
Characterization of proton-activated implantable markers for proton range verification using PET J. Cho1, G. Ibbott1, M. Kerr1, R. A. Amos2, F. Stingo1,
Medical Resonance Imaging
MRI in Medicine Prof. E.S. Yang The Jockey Club MRI Engineering Center
Lesson Plan MRI Scan Experience
MR Scanner Transmit Gain as a Potential Indicator For Lead Heating
The strength of the flux density at the Earth’s surface ranges from less than 30μT in an area including most of South America and South Africa to over.
Sunday Case of the Day Physics
M. Frascaroli, L. Moro. , L. Sibilla, M. Baldi, I. Carne. , D
Murat Bozkurt, M. D. , Mustafa Akkaya, M. D. , Safa Gursoy, M. D
Date of download: 12/24/2017 Copyright © ASME. All rights reserved.
Human Proportions The basic guide for body proportion is in head lengths. The figure of an average adult is 7 to 7.5 head lengths: First head length:
4D eXtended CArdiac-Torso (XCAT) Phantoms
Key Stage 3 Magnetic Resonance Imaging Image bank
Volume 10, Issue 12, Pages (December 2013)
Motor and Nonmotor Circuitry Activation Induced by Subthalamic Nucleus Deep Brain Stimulation in Patients With Parkinson Disease  Emily J. Knight, PhD,
Anthropometrics.
MRI Safety Seyed Ali Hosseini.
Volume 10, Issue 12, Pages (December 2013)
Basics of MRI By Dr. Mohamed Samieh.
Human Subjects in fMRI Research
Lesson Plan MRI Scan Experience
Center for Magnetic Resonance Research
Lesson Plan MRI Scan Experience
Single-finger subcutaneous defibrillation lead and “active can”: a novel minimally invasive defibrillation configuration for implantable cardioverter-defibrillator.
The effect of vertical centering and scout direction on automatic tube voltage selection in chest CT: a preliminary phantom study on two different CT.
Lesson Plan MRI Scan Experience
Superconducting Magnets
Magnetic measurements at ambient temperature on the MQXFBP1
Presentation transcript:

Assessment of MRI Induced Heating of Cardiac Pacing Leads: Relevance to Target Patient Populations Presenter: Daniel G. Hullihen Jr. Director of Business Development RSNA 2006 November 28, 2006

2 Tissue proximal to implanted lead electrodes can experience significant heating during magnetic resonance imaging (MRI) Lead heating depends on many factors MR technologists reposition patients in bore Patients, and MR scanner bores, come in different sizes

3 Patient shoulder width varies and is 36 cm for 5 th percentile adult females Patients can be positioned off center in the bore to optimally image region of interest Pacing system and lead path can vary significantly from patient to patient

4 Three different bipolar pacing lead designs Four different lateral MRI bore positions Determination of MRI induced heating – Full range of clinically relevant positions

5 Left installation Right installation

6 Test Samples Used: Standard bipolar active fixation pacing lead (Commercially Available) Prototype bipolar active fixation pacing lead with resonant circuit in distal tip (Resonant Circuit) Prototype bipolar active fixation pacing lead without resonant circuit in distal tip (Control)

7 Luxtron® model 3100 fluoroptic thermometry system with SSM model optical probes Gelled-saline solution: 5.8 g PAA, 0.8 g NaCl per liter of de-ionized water in head / torso phantom vessel (ASTM F a) GE 1.5 Tesla MR system, body coil, FSE- XL, whole body average SAR: W/kg

8 Lead developed for testing with readily available components Resonant Circuit Components

9 Leads placed in four (4) lateral positions 1.Initial Position: Centerline in bore 2.Second Position: 9.5cm ( ~40%) off centerline in bore 3.Third Position: 17cm ( ~71%) off centerline in bore 4.Fourth Position: On Edge, 24cm off centerline in bore; extreme case

10 Initial Position: Centerline in bore

11 Second Position: 9.5cm ( ~40%) off centerline in bore

12 Third Position: 17cm ( ~71%) off centerline in bore

13 Forth Position: On Edge, 24cm off centerline in bore

14

15 In all positions the lead with the resonant circuit in the distal tip had temperature increases of less than 0.75ºC.

16

17 Shifting a patient laterally in the MR scanner bore significantly affects heating of tissue at the distal tip of pacing leads. Incorporating a resonant circuit into the distal tip of the lead makes the lead less prone to tissue heating with respect to a patient’s lateral position in the scanner’s bore.