X-Ray Intensifying Screens George David Associate Professor Medical College of Georgia Department of Radiology.

Slides:



Advertisements
Similar presentations
Analog Imaging II Intensifying Screens
Advertisements

Technical Aspects: the machine the image
Advanced Biomedical Imaging Lecture 3
Radiographic Film & Intensifying Screens
IMAGE FORMATION  Introduction  The Invisible and Visible Image  Image Characteristics.
Components of Radiographic Image Quality
Resident Physics Lectures
X-ray radiography Seminar The "geometrical" unsharpness of X-ray images is limited by finite dimension of the source, i.e. the window on.
Chapter 15 Radiographic Intensifying Screens And Cassettes
Chapter 9 Film & Film Holders.
معاونت درمان امور مامایی اردیبهشت 90. برای ثبت اطلاعات در برنامه نرم افزاری نظام مراقبت مرگ پریناتال ابتدا لازم است برنامه نرم افزار info-path وپرنیان.
BME 560 Medical Imaging: X-ray, CT, and Nuclear Methods
فاکتورهای مهم در ایجاد یک مقاله علمی
Definition. قانون بیز P(b|a) = P(a|b)P(b)/P(a) P(b|a) = P(a|b)P(b)/P(a) این قانون برای استنتاج آماری استفاده می شود. این قانون برای استنتاج آماری استفاده.
RAD 254 Chapter 11 Radiographic Film/Receptor
Image Formation Differential Absorption of X-rays by the Patient  Absorption of non-attenuated x-rays by screen  Emission of light  Formation of Film.
 QC testing of screen speed should occur on acceptance and then yearly.  Evaluate first whether similar cassettes marked with the same relative speed.
Most of the images recorded during conventional radiography are obtained with film/screen combination image receptors. Which in lessens the patient dose.
Intensifying Screens Kyle Thornton DMI 50B.
RAD 354 Chapt. 13 Intensifying Screens
FILM CASSETTES & INTENSIFYING SCREENS WEEK 9
Resident Physics Lectures Christensen, Chapter 6Filters George David Associate Professor Department of Radiology Medical College of Georgia.
Chapter 7 Dental X-Ray Film.
INSTRUMENTATION. ELECTRICAL REQUIREMENTS AND EFFICIENCY  Effective voltage  High frequency inverter technology (over 10,000Hz)  Constant potential.
A-LEVEL PHYSICS 15/09/ :49 Medical Physics.
1 Network Address Translation (NAT). 2 Private Network شبکه خصوصی شبکه ای است که بطور مستقیم به اینترنت متصل نیست در یک شبکه خصوصی آدرس های IP به دلخواه.
Dr Mohamed El Safwany, MD.
Resident Physics Lectures
Image Receptor Systems
Image Quality Radiographic Resolution.
Abab presents today. A closer look at the production process of a movie soundtrack. What elements are affecting recorded sound quality ?
Chapter 6 Image Receptors Radiology. Introduction To further understand radiography, one must know how a permanent record is produced using x-rays. To.
Photon Tissue Interactions
Physical Characteristics of X- Ray Film & Film Processing
Sounds of Old Technology IB Assessment Statements Topic 14.2., Data Capture and Digital Imaging Using Charge-Coupled Devices (CCDs) Define capacitance.
RAD 254 Chapter 13 Intensifying Screens Physical purpose: to convert x-ray photons into light photons (done at the phosphor layer)
Analog Imaging III By Professor Stelmark.
FILM CASSETTES & INTENSIFYING SCREENS WEEK 9
Grids George David Associate Professor Department of Radiology
IMAGE RECEPTORS. Follow the Sequence- Film System Tube Tube Patient Patient (bucky) or non-bucky (bucky) or non-bucky Cassette Cassette INTENSIFYING SCREEN.
Image Receptors Film Construction, Film Handling, Cassette Construction, DarkRoom, Digital Imaging 1.
Film Imaging Lavin: Chapter 5 CTVT: Ronnie’s snood
Image Receptor Unsharpness By Professor Stelmark.
Analog X-ray Imaging Recommended Book: Walter Huda, REVIEW OF RADIOLOGIC PHYSICS By: Maisa Alhassoun
Barium & Iodine Imaging Physics
INTENSIFYING SCREENS.
Image Receptors.
Resident Physics Lectures
Image Receptors.
[c.
Imaging on Film RVT: Chapter 5
Reproduction Techniques in Animals
نظریه رفتار برنامه ريزي شده Theory of Planned Behavior
به نام خدا.
SY800 router mode [AD-14-TB ].
Address: Times New Roman, size 34
Quality Control Testing of Screen Speed
روش های نگهداری مواد غذایی
بررسی قطبش در لیزر های کاواک عمودی گسیل سطحی(vcsel)
SSO Single Sign-on Systems
Resident Physics Lectures (Year 1)
Resident Physics Lectures
دینامیک سیستمهای قدرت مدرس: دکتر مهدی بانژاد
نظریه رفتار برنامه ريزي شده Theory of Planned Behavior
Reproduction Techniques in Animals
میکروسکوپ های الکترونی عبوری فاطمه احمدپور، محمد هادی مقیم
Angular Kinetics مهدی روحی بهار 85.
فصل ششم مدارهای ترتیبی.
کتابخانه دانشکده پرستاری و مامایی دانشگاه علوم پزشکی شهید بهشتی
Resident Physics Lectures (Year 1)
Presentation transcript:

X-Ray Intensifying Screens George David Associate Professor Medical College of Georgia Department of Radiology

Convert x-rays to light many light photons created per x-ray photon absorbed in screen Light photons have much less energy light from screen exposes film film much more sensitive to light than to x-rays screens substantially reduce patient dose Factor of 100’s screen use virtually universal Photon Light Screen

light tight container for film holds film in tight contact with screens over entire surface gaps drastically increase image unsharpness All non-mammo cassettes use two screens One above film One below film Cassette Screens Film

Two screens produce more light Less radiation required to achieve a given optical density Requires two emulsions on film One above one below Cassette Screens Film

easier to manufacture emulsion shrinks when it dries Having two emulsions minimizes curling photographic advantage faster system two screens used each emulsion optimally captures light produced by “its” screen double emulsion filmscreens

require regularly cleaning Dust, dirt, paper, hair,etc prevent screen light from reaching film Causes white dots on image

Light emitted by crystals inorganic salts called phosphors older phosphor materials calcium tungstate original phosphor material used in radiology emits blue light zinc cadmium sulfide

image tubes cesium iodide film screens –barium strontium sulfate –yttrium –rare earths »gadolinium »lanthanum »tantalates

Advantages over direct film exp. Drastically decreased patient dose (X 100’s) Shorter exposure times Configuration cassette sandwiches film between 2 screens

plastic protective coat phosphor layer reflecting layer base support layer One screen

Protective Layer applied over phosphor made of plastic approximately mils thick Functions prevents static electricity provides physical protection provides surface suitable for cleaning Phosphor Layer contains phosphor crystals approximately mils thick

Reflecting Coat reflects light emitted toward back of screen phosphors emit light in all directions not all screens have reflecting coat Reduces resolution made of white substance (titanium dioxide) 1 mil thick Base Layer Mechanical support cardboard or polyester plastic approximately mils thick

Photon must be absorbed by a screen Screen must emit light Light must reach film Light must expose film Each step above has an associated efficiency

Efficiency of energy conversion from x-rays to light 5% for calcium tungstate 850 light photons per x-ray photon absorbed up to 20% for newer phosphors such as rare earth Can be as high as 45% for direct digital DR systems

ability of light emitted by phosphor to escape screen & expose film typically half of light emitted by screen does not reach film

exposure required without screen exposure required with screen for calcium tungstate intensification factor increases with kVp thicker body parts cause increase filtering raises effective kVp small number of x-ray photons interact directly with film negligible film darkening contribution

phosphor layer thickness thicker screen faster poorer detail because of light spread or diffusion light produced further from film size of phosphor crystals presence or absence of light-absorbing dye dye reduced lateral light diffusion better resolution poorer efficiency (lower speed) phosphor efficiency

Thicker phosphor increases absorption Increases speed Reduces patient exposure Diffusion of light causes unsharpness light travels further from point of origin in screen to film Film Screen

Why use 2 thin emulsions rather than 1 thicker one? light produced closer to emulsion less light spread X-Ray

light from one screen exposes opposite emulsion Top Screen Bottom Screen Bottom Emulsion Top Emulsion Film X-Ray

poorer resolution light travels further, spreads more caused by incomplete absorption of light by adjacent emulsion X-Ray

Maximum number of line pairs (line & space) per millimeter resolved by screen-film system line & space have equal width Typical values Film ~100 line pairs per mm Film / screen systems ~ 10 line pairs per mm maximum

خاصیت فلورسانس صفحات تشدیدکننده در درجه حرارتهای پایین بیشتر است. در درجه حرارتهای بالا این موضوع ممکن است دارای اهمیت کلینیکی باشد. در دمای 35 درجه سانتی گراد جهت ایجاد دانسیته ای مشابه با یک رادیوگرافی که در دمای 21 درجه سانتی گراد انجام شده است، شرایط اکسپوژر را می بایست حدود 25% افزایش داد.

کاستی که در آن صفحات تشدید کننده نصب شده اند، یک محفظه تنگ ضد نور برای فیلم ایجاد می نماید. همچنین کاست فیلم را در تمامی سطحش در تماس محکم با صفحات تشدید کننده نگه می دارد. هنگامی که تماس بین فیلم و صفحه تشدید کننده کامل باشد، یک نقطه نوری در صفحه، نقطه ای نظیر خود را بر روی فیلم ثبت می نماید. در صورتی که این تماس ضعیف باشد، نقطه نوری قبل از رسیدن به فیلم پراکنده و منتشر می شود و درنتیجه تصویر رادیوگرافیک آن ناواضح می گردد.

روش ساده ای برای آزمایش تماس فیلم و صفحه وجود دارد. یک تکه از صفحه سیمی بر روی کاست قرار داده می شود و از آن رادیوگرافی به عمل می آید. میزان وضوح تصویر سیمها در نواحی مختلف فیلم با هم مقایسه می گردد و ناحیه ای که در آن اتصال بین صفحه و فیلم ضعیف می باشد، از طریق ناواضح بودن تصویر سیمها در آن ناحیه مشخص می گردد. صفحه سیمی بایستی از جنس آهن، برنج یا مس باشند، چون صفحات آلومینیومی یا پلاستیکی به میزان کافی پرتو X را جذب نخواهند کرد.

هرگونه ماده خارجی بر روی صفحه مانند کاغذ یا خون، مانع از عبور فوتونهای نوری شده و بسته به اندازه و شکل خود، ایجاد یک ناحیه اکسپوز نشده بر روی فیلم می نماید. نقاط مزبور از عمده ترین منابع پوسیدگی بیش از حد صفحات تشدید کننده می باشند. تمیز نمودن صفحات می بایست با استفاده از ترکیبات آنتی استاتیک و پاک کننده و هم چنین به ارامی و بدون مالش شدید با یک پارچه نرم و بدون پرز صورت گیرد و تا خشک شدنم آنها درب کاست باید باز باشد.

increase thickness of phosphor layer Change to different phosphor material with higher absorption efficiency More absorption for given thickness Change to different phosphor material with higher conversion efficiency More light per absorption

Par Speed Calcium Tungstate 20% High Speed Calcium Tungstate 40% Thicker screen, poorer resolution Rare Earth60%

Atomic Number tungsten of calcium tungstate higher than rare earth, more photoelectric interaction K-Edge tungsten: 69.5 keV Yttrium: 17 keV Barium: 37 keV Lanthanum: 39 keV Gadolinium: 50 keV Lower K-edge greatly increases absorption in diagnostic energy range Photon Energy Absorption

Screen’s light spectrum must match film’s color sensitivity optimize speed by matching film response to screen light

Calcium Tungstate Somewhat continuous blue spectrum (430 nm wavelength) Gd 2 O 2 S:Tb narrower green spectrum (544 nm wavelength) most but not all rare earth screens emit predominantly green light