Partial Coherence Interferometry Failure Rate in a Teaching Hospital Leslie A. Wei 1,2, BA, Nickolaus P. Katsoulakis 2, MD, Theodoros Filippopoulos 3,

Slides:



Advertisements
Similar presentations
Comparison of Endothelial Cell Count by Manual and Automated Methods in Normal Corneas and in Fuchs' Endothelial Dystrophy Somasheila I Murthy, Debarun.
Advertisements

Comparison of surgically induced astigmatism after phacoemulsification trough 3.2, 2.2 and 1.8 clear corneal incision. Luis Izquierdo Jr MD. PhD. Maria.
Presented by: Ray T. Oyakawa, M.D. Sharper Vision Centers A MEDICAL GROUP Comparison of PalmScan and IOL Master axial lengths measurements No financial.
Accuracy of Predicted Refractive Error in Resident-Performed Cataract Surgery Using Partial Coherence Interferometry Nickolas P. Katsoulakis, M.D., Paul.
A simple and accurate method of alignment for toric intraocular lens implantation using anterior segment optical coherence tomography (OCT). Kazuno Negishi,
Long-Term Follow-Up of Toric Intraocular Lens Implantation for Keratoconus Alejandro Navas, MD, MSc, Martha Jaimes, MD, Jesús Cabral, MD, Arturo Ramirez.
V. S. Liarakos, K. van Dijk, L. Ham, L. Baydoun and G.R.J. Melles Anterior Chamber vs Posterior Chamber IOL in DMEK for Pseudophakic Bullous Keratopathy.
Development of an Apple iPhone IOL Calculator Application for Cataract Surgery Rony Gelman, MD 1, Richard E. Braunstein, MD 1 1 Department of Ophthalmology,
Joshua Ki Hu Vanderbilt Eye Institute Ophthalmology, PGY-4 DATE Vanderbilt Eye Institute.
Visual outcome & subjective visual symptoms of the Tecnis ZM900 multifocal intraocular lens in Asian eyes Dr Colin S.H. Tan MBBS, MMed (Ophth), FRCSEd.
Anupama Kotha 1, Simar J. Singh 1, William B. Trattler 1,2, Carlos Buznego 1,2 The authors have no financial interest in the subject matter of this poster.
Eltutar, Kadir; Akcetin, Tulay A.; Ozcelik, N. Demet Istanbul Education and Research Hospital Department of Ophthalmology The authors state that they have.
Progressive Multifocal Intraocular Lens G. Rubiolini M.D. Italy Disclosure of finanacial interest Author's research is partially funded.
INTRA-OCULAR LENS POWER CALCULATION IN POST- REFRACTIVE SURGERY ASIAN EYES USING THE HAIGIS-L FORMULA Dr Daphne Han, FRCS Dr Wei-Han Chua, FRCS Dr Peter.
M. Allison Roensch, MD, Preston H. Blomquist, MD, Nalini K Aggarwal, MD, James P. McCulley, MD Department of Ophthalmology University of Texas Southwestern.
Placement of Toric Intraocular Lens and the Long-term Change in the Axis of Corneal Astigmatism after Sutureless Cataract Extraction by Phacoemulsification.
The Effect of the Restor Multifocal IOL on Frequency Doubling Perimetry Elizabeth Yeu, MD1, Elizabeth Woznak, BS2, Nicole Kesten, BS2, Steven VL Brown,
Contrast-Enhanced Corneal Wound Imaging by Optical Coherence Tomography Preeya K. Gupta, MD Justis P. Ehlers, MD Terry Kim, MD Duke Eye Center, Durham,
Somasheila I. Murthy, Prashant Garg, Pravin K. Vaddavalli
Changes of Presenile Cataracts that had been Operated over 10 years in Korea Sung Kun Chung, M.D. ; Eun-Jung Jun, M.D. Hyun Seung Kim. Department of Ophthalmology.
Dept. of Ophthalmology, Friedrich-Alexander University Erlangen-Nuremberg, Erlangen, Germany. Outcome after big-bubble deep anterior lamellar keratoplasty.
G. Jacob 1,2, C. Bouchard 2, S. Kancherla 1. Edward Hines, Jr. VA Hospital, Hines, IL, Department of Ophthalmology 1. Loyola University Medical Center,
Phacoemulsification in eyes with previous anterior chamber phakic IOL surgery Walton Nosé, MD, PhD 1,2 Adriana dos Santos Forseto, MD 1 Mariana Ávila,
Hong Kong Eye Hospital Ms Frenchy Chiu Dr Victoria Wong IOL master
Neeti Parikh, MD Fuxiang Zhang, MD Department of Ophthalmology Henry Ford Hospital A Comparison Of Patient Satisfaction With Modified Monovision Versus.
Purpose: Introduction:  At initial evaluation: For post-op day # 0 patients: Pre-op VA was 20/50.6 (0.395 ± 0.198); Post-op VA was 20/102.0 (0.196 ± 0.162);
Biometric Accuracy in High Hypermetropes and Myopes
Occurrence of Retinal Detachment After Cataract Surgery Peter Jeppesen MD, PhD and Thomas K. Olsen, MD, DMSc Department of Ophthalmology Århus University.
Glaucoma and Penetrating Keratoplasty : Incidence, Risk Factors, and Outcomes Sonika Gupta Consultant Ophthalmology Max Eye Care New Delhi, India Author.
INTRAOCULAR LENS POWER CALCULATION BY IMMERSION A-SCAN BIOMETRY VERSUS CONTACT A-SCAN BIOMETRY MEASUREMENTS BEFORE CATARACT SURGERY Burak Bilgin**, M.D.,
Anterior Segment Optical Coherence Tomography and In Vivo Confocal Microscopy Findings in Femtosecond Laser- Assisted Keratoplasty Kurt H. Kelley, MD;
Sonia Yoo, MD 1 Fernanda Piccoli, MD 1 Artur Schmitt, MD 1 Takeshi Ide, MD 1 Tsontcho Ianchulev, MD 2 Authors have no financial interest in this subject.
Phoebe D. Lenhart, M.D. 1, Amy K. Hutchinson, M.D. 1, Michael J. Lynn, M.S. 2, Scott R. Lambert M.D. 1 1 Department of Ophthalmology, Emory University,
Khairidzan Mohd Kamal MD
Metallic Foreign Body Embedded in the Posterior Lens Capsule Helen R. Moreira, MD; Michele S. Todman, MD; Paul J. Botelho, MD Division of Ophthalmology,
Endothelial Keratoplasty in Patients With an Anterior Chamber Intraocular Lens: A Montreal Experience Georges M. Durr, MD 1,2 Johanna Choremis, MD, FRCSC.
Prospective, Randomized, Double-Blind Study Comparing Nepafenac 0.1% (Nevanac™) and Bromfenac Sodium 0.09% (Xibrom™) in Post-Op Cataract Patients Hon-Vu.
Mean Keratometry Measurement Post Penetrating Keratoplasty Jacky Yeung MSc MD, Stephanie Baxter MD FRCS(C) Department of Ophthalmology, Hotel Dieu Hospital,
The authors have no financial interest in the subject matter of this poster. FINANCIAL DISCLOSURES.
Adriana S. Forseto1, MD Walton Nosé1,2, MD
IOL Calculations Based on Partial Biometry in Humanitarian Missions Joseph Schmitz, MD Kimberly Davis, MD, FACS Scott McClatchey, MD The authors have no.
Sherman W. Reeves, MD, MPH 1,3,4 ; Jacob A. Kozisek, OD 1,2 ; Noumia Cloutier-Gill, OD 1,2 ; David R. Hardten, MD 1,2,3,4 Accuracy of Scheimpflug Imaging.
Visual and IOP Outcomes after PRK in Pigment Dispersion Syndrome [Poster Number: P190] Kraig S. Bower, Denise A. Sediq, Charles D. Coe, Keith Wroblewski,
Comparison of Central Corneal Thickness, Anterior Chamber Depth, and Central Corneal Power Measurements between Two Scheimpflug Imaging Systems Yuichi.
Faik Orucov*, MD, Joseph Frucht-Pery, MD, David Landau, MD, Eyal Strassman, MD, Abraham Solomon, MD Clinical outcome of thin corneas after laser in situ.
Comparison of visual function following piggyback implantation of Acrysof ReSTOR intraocular lenses with Tecnis multifocal ZM900 intraocular lenses. Rodrigo.
Comparing Factors Affecting Surgically Induced Astigmatism
COMPARISON AND EVALUATION OF OCULAR BIOMETRY USING A NEW NON-CONTACT OPTICAL LOW COHERENCE REFLECTOMETER David Goldblum 1,3, Kaspar Rohrer 1, Rudolf Waelti.
Simulated Experiments on IOL Power Calculation Using Anterior Segment OCT Dong Hyun Jo, M.D., 1,2 Mee Kum Kim, M.D., 1,2 Won Ryang Wee, M.D. 1,2 1 Department.
F.I. Camesasca, MD Zeiss Invent ZO Aspheric IOL: Long-Term Results of Refractive and Aberrometric Analysis F. I. Camesasca* P. Vinciguerra.
Liquifaction Method and Extent of Posterior Capsule Opacification: Two-Year Follow-up Marie Kalfertova, Mariya Burova, Pavel Rozsival, Nada Jiraskova Nada.
Poster# P36 Comparison of Techniques of Measuring Anterior Chamber Depth and Corneal Curvature in Indian Eyes Sujata Das, MS, FRCS Anjula Kumari, B.Optom.
Corneal CXL in Pediatric Patients with Progressive Keratoconus Stephanie Wise, Christian Diaz, Karolien Termote, Paul J. Dubord, Martin McCarthy, Sonia.
Intraocular Lens Outcomes: Comparison of Technologies and Formulas Carolina Eyecare Physicians, LLC Research Assistant Professor of Ophthalmology Storm.
Changes of Axial Length After LASIK Surgery: A 3-Year Follow-Up Study
Suppressive effect of combined treatment of orthokeratology and 0
Comparison of Endothelial Cell Loss After Phacoemulsification Performed by Third Year Residents and Anterior Segment Surgeons Alexandra Braunstein, MD.
Corneal Pachymetry in Prediction of Refraction After Cataract Surgery
From: Key Developments for Partial Coherence Biometry and Optical Coherence Tomography in the Human Eye Made in Vienna Invest. Ophthalmol. Vis. Sci ;57(9):OCT460-OCT474.
Refractive outcomes of intraoperative wavefront aberrometry versus optical biometry alone for intraocular lens power calculation Zina Zhang MD1, Logan.
Aug 11, 2017 Comparison of predictive accuracy on Partial coherence interferometry (PCI) and swpt-source optical coherence tomographry (SS-OCT) Choun-Ki.
Effect of Axial Length Measurement Method on Refractive Outcomes of Cataract Surgery: Real World Comparison of Partial Coherence Interferometry and Immersion.
Opacification of the optic of an Akeos Adapt® intraocular lens
Early Experience with Descemet’s Stripping Automated Endothelial Keratoplasty Combined with Phacoemulsification: Clinical and Refractive Outcome University.
Sarosh A. Janjua MD1, Sandra L. Cremers MD FACS1
Poster Number: P90 Category: Intraocular Surgery (Cataract and Refractive) Optimization of IOL Power Calculation Constants: By Unit or by Surgeon? Nathaniel.
Roxana Ursea, MD Matthew T. Feng, MD Ovette Villavicencio, PhD
Results of corrective surgery: secondary lens implantation at a cataract surgery training centre Mehul Shah,shreya shah, adway appalware,pramod upadhyay,
Khairidzan Mohd Kamal MD
Michael Goodman, Alexandra Paul and Andrew Hsu
Presentation transcript:

Partial Coherence Interferometry Failure Rate in a Teaching Hospital Leslie A. Wei 1,2, BA, Nickolaus P. Katsoulakis 2, MD, Theodoros Filippopoulos 3, MD, Paul B. Greenberg 1,2, MD 1 Ophthalmology Section, Veterans Affairs Medical Center, Providence, RI 2 Division of Ophthalmology, Warren Alpert Medical School of Brown University, Providence, RI 3 Cornea and Refractive Surgery Service, New York Eye and Ear Infirmary, New York, NY 4 Private Practice, Athens, Greece Sources of financial support: none Author financial and proprietary interests: The authors have no financial interests or proprietary interests to disclose.

Purpose To investigate the PCI biometry failure rate using the IOLMaster (Carl Zeiss Meditec, Dublin, CA) in a cohort of patients with visually significant cataract seen in a resident ophthalmology clinic at a Veterans Affairs (VA) medical center.

Background  IOLMaster (Carl Zeiss Meditec AG)  Infrared optical biometry system using partial coherence interferometry for axial length measurement  Accuracy within 20 mm  Ultrasound is accurate to 100 mm  Superior to applanation ultrasound biometry  Non-contact  IOLMaster (Carl Zeiss Meditec AG)  Infrared optical biometry system using partial coherence interferometry for axial length measurement  Accuracy within 20 mm  Ultrasound is accurate to 100 mm  Superior to applanation ultrasound biometry  Non-contact

Background  Partial coherence interferometry (PCI) has an 8- 22% failure rate due to a variety of causes such as mature cataracts, inability to fixate secondary to macular degeneration, tremor, keratopathy, nystagmus, and vitreous hemorrhage 1,2,3,4.  The new composite IOLMaster Version 5 software has demonstrated a failure rate of 7.4%, mostly due to posterior subcapsular cataracts 5.  There have been few studies on the efficacy of PCI in a teaching hospital setting.  Partial coherence interferometry (PCI) has an 8- 22% failure rate due to a variety of causes such as mature cataracts, inability to fixate secondary to macular degeneration, tremor, keratopathy, nystagmus, and vitreous hemorrhage 1,2,3,4.  The new composite IOLMaster Version 5 software has demonstrated a failure rate of 7.4%, mostly due to posterior subcapsular cataracts 5.  There have been few studies on the efficacy of PCI in a teaching hospital setting.

Methods  After obtaining IRB approval, the biometric measurements of 147 consecutive patients (165 eyes) who had resident-performed cataract surgery at a VA Medical Center from September 2007 to June 2008 were examined.

Results  Thirty-nine patients (42 eyes) were excluded due to a lack of preoperative cataract history and physical in the electronic medical record.  One hundred and eight patients  One hundred and eight patients (123 eyes) met the inclusion criteria: 106 patients were male (98.1%); mean age was 72.4 (range 50-91).  Thirty-nine patients (42 eyes) were excluded due to a lack of preoperative cataract history and physical in the electronic medical record.  One hundred and eight patients  One hundred and eight patients (123 eyes) met the inclusion criteria: 106 patients were male (98.1%); mean age was 72.4 (range 50-91).

37.4%  Forty-six eyes (37.4%) were documented PCI failures that required immersion A-scan for axial length determination. version 5  The subgroup of eighty-eight eyes that had IOLMaster version 4 readings had a failure rate of 43.2% while the subgroup of thirty-five eyes with IOLMaster version 5 readings had a significantly lower failure rate of 22.9% (OR = 0.38, p < 0.05). 37.4%  Forty-six eyes (37.4%) were documented PCI failures that required immersion A-scan for axial length determination. version 5  The subgroup of eighty-eight eyes that had IOLMaster version 4 readings had a failure rate of 43.2% while the subgroup of thirty-five eyes with IOLMaster version 5 readings had a significantly lower failure rate of 22.9% (OR = 0.38, p < 0.05). Results

 All PCI failures were due to dense cataract; the type of cataract was not specified.  Eyes that successfully underwent PCI had significantly better mean preoperative best corrected visual acuity (BCVA) of 0.48 (20/60) than those requiring immersion A-scan, which had a mean preoperative BCVA of 1.4 (<20/400) (p = 0.01).  All PCI failures were due to dense cataract; the type of cataract was not specified.  Eyes that successfully underwent PCI had significantly better mean preoperative best corrected visual acuity (BCVA) of 0.48 (20/60) than those requiring immersion A-scan, which had a mean preoperative BCVA of 1.4 (<20/400) (p = 0.01).

Limitations  Retrospective nature  Variability in measurements secondary to different resident physicians performing PCI  Small sample size  Large number of exclusions due to incomplete electronic medical records  Inability to analyze the PCI failures in terms of type of cataract because of lack of documentation.  Retrospective nature  Variability in measurements secondary to different resident physicians performing PCI  Small sample size  Large number of exclusions due to incomplete electronic medical records  Inability to analyze the PCI failures in terms of type of cataract because of lack of documentation.

Conclusions  PCI biometry failure rate is still significant in the veteran population due to dense cataracts, though the new Version 5 software does appear to decrease this rate.  Poor preoperative vision is a predictor of PCI failure.  We look forward to continued improvements in PCI that will facilitate pre-operative biometry in cases of advanced cataract in teaching hospital settings.  PCI biometry failure rate is still significant in the veteran population due to dense cataracts, though the new Version 5 software does appear to decrease this rate.  Poor preoperative vision is a predictor of PCI failure.  We look forward to continued improvements in PCI that will facilitate pre-operative biometry in cases of advanced cataract in teaching hospital settings.

References  1 Narvaez J, Cherwek DH, Stulting RD, et al. Comparing immersion ultrasound with partial coherence interferometry for intraocular lens power calculation. Ophthalmic Surg Lasers Imaging 2008;39(1):30-4.  2 Haigis W, Lege B, Miller N, et al. Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis. Graef’s Arch Clin Exp Ophthalmol 2000;238:  3 Rajan MS, Keilhorn I, Bell JA. Partial coherence laser interferometry vs conventional ultrasound biometry in intraocular lens power calculations. Eye 2002;16:  4 Ueda T, Taketani F, Ota T, et al. Impact of nuclear cataract density on postoperative refractive outcome: IOL Master versus ultrasound. Ophthalmologica 2007;221:  5 Hill W, Angeles R, Otani T. Evaluation of a new IOLMaster algorithm to measure axial length. J Cataract Refract Surg 2008; 34:  1 Narvaez J, Cherwek DH, Stulting RD, et al. Comparing immersion ultrasound with partial coherence interferometry for intraocular lens power calculation. Ophthalmic Surg Lasers Imaging 2008;39(1):30-4.  2 Haigis W, Lege B, Miller N, et al. Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis. Graef’s Arch Clin Exp Ophthalmol 2000;238:  3 Rajan MS, Keilhorn I, Bell JA. Partial coherence laser interferometry vs conventional ultrasound biometry in intraocular lens power calculations. Eye 2002;16:  4 Ueda T, Taketani F, Ota T, et al. Impact of nuclear cataract density on postoperative refractive outcome: IOL Master versus ultrasound. Ophthalmologica 2007;221:  5 Hill W, Angeles R, Otani T. Evaluation of a new IOLMaster algorithm to measure axial length. J Cataract Refract Surg 2008; 34: