Non-Invasive Blood Pressure Device for Use in fMRI Imaging Applications February 18, 2005 Jose Alvarado Benjamin Huh Sanjeet Rangarajan Dr. André Diedrich.

Slides:



Advertisements
Similar presentations
Week 6 – Exercise Physiology Gen Bio II. Homeostasis Maintenance of a relatively steady physiological state Maintenance of a relatively steady physiological.
Advertisements

Blood Pressure.
Monitoring The Emergency Patient ABDULRAHMAN SINDI ( Emergency Medicine Resident)
Unit 7: Fitness testing for sport and exercise
Hydraulics.
Section III: Concept 07 Cardiovascular Fitness
Presentation Package for Concepts of Fitness and Wellness 6e Section III: Concept 08 Cardiovascular Fitness All rights reserved.
Control of the Internal Environment
Michael Shadrin 4/25/11 Comp 089.  The evolution of modern technology played a crucial role in advancing medicine.  Computer technology has become an.
P /20093 Composites Autoclave Brandon Allen (EE) Matt Brady (ME) Brian Cario (ME) Matt DiFrancesco (ME) John Mink (EE) ‏ Chris DeMinco.
Physical Education Studies
Presentation Package for Concepts of Physical Fitness 12e
Ventricle Chamber Valve Imaging Chamber ECG Machine Aortic Compliance Chamber Buffer Chamber Secondary Compliance Chamber Flow Meter Pressure Sensors DAQ.
Johns Hopkins Motor “Pneu Step” The Surgical hand of the future.
DEPARTMENT OF ENGINEERING SCIENCE INSTITUTE OF BIOMEDICAL ENGINEERING Marco F Pimentel, Graduate Student, Centre for Doctoral Training in Healthcare Innovation,
Blood pressure. Blood pressure Preparation for measurement.
Blood Pressure Basics.
Control of Heartbeat and Monitoring Blood Pressure
WHAT’S A TECHNOLOGY SYSTEM? A technology system is a machine that processes digital data. A computer is a technology system. A computer installation is.
FDR Presentation Personal Health Monitoring Device Bassam Noaman Dina El-Eissa Mentor: Prof Zhenyu Guo Instructor: Prof J Silverman 4/12/00.
Blood Pressure— The driving force Stephen Hales 1733 Blood pressure (hydrostatic pressure) is the force exerted by the blood against any unit area of vessel.
Blood Pressure Assessment
Chapter 2.  Types of Network  Circuit Switched & Packet Switched  Signaling Techniques  Baseband & Broadband  Interference  Transmission Medium.
Copyright © 2009 Wolters Kluwer Health | Lippincott Williams & Wilkins Comprehensive Medical Assisting, 3 rd Ed Unit Four: Fundamentals of Clinical Medical.
Ruslan Masinjila Aida Militaru.  Nature of the Problem  Our Solution: The Roaming Security Robot  Functionalities  General System View  System Design.
Methodology:  Plethysmogram is a measure of optical impedance -- Obtained from tail cuff with photo-sensor (Figure 1) -- Blood pressure α blood volume.
Sport Books Publisher1 The Heart and Lungs at Work Chapter 7.
Physiologic Control Algorithms for Rotary Blood Pumps using Pressure Sensor Input Edward Bullister, Ph.D. Sanford Reich, Ph.D. APEX Medical, Inc. ISRP.
Erik Simpanen September 27 th, 2011 BME 281. Heart Disease  A generic term for referring to numerous different forms of sicknesses that effect the heart.
VITAL SIGNS BLOOD PRESSURE (BP).
Objectives Understand the importance of fiber-optic technologies in the information society Identify the fundamental components of a fiber-optic cable.
Blood Pressure Required to move blood and all its constituents throughout the body.
VIP4Air - VIP4Air/Slave Minimal Air/Oil Micro-lubrication Product Features and Operation Dave Stoyanoff - Dropsa USA.
TEAM MEMBERS: STAN WEED (TEAM LEADER), PAVEL KUZNETSOV, ERIC-THOMAS ARROYO PROJECT ADVISORS: DR. JOHN R. LACOURSE, PROF. FRANCIS C. HLUDIK Wireless Activity.
General Schemaatic of HR Control
Non-Invasive Blood Pressure Waveform Monitoring in Mice By Andrew McClellan Laura Miller.
CPAP Remote Alarm System BME 272: Oral Report Three Group 15 Kathleen Grunder Jessica Paulsen Molly Rice.
Electromyography (EMG)
BME 181 March 4, 2013 Presented by: Corey Gomes. J.A. McWilliams Late 1800’s Electrical impulses John Hopps Radio frequencies Mechanical and electrical.
Isolation Lock-Out Room Johns River Quarry Site Contact: Blake Ardrey, Quarry Manager, BCM, Johns River Quarry Phone:
Intra-Aortic Balloon Pump What it is and what it does
Measuring Monkey Eye Movement in the MRI Team Members: Josh Anders Betsy Appel Bryan Baxter Alyssa Walsworth Client: Luis Populin,Ph. D. Advisor: Justin.
State of Engineering in Precision Agriculture, Boundaries and Limits for Agronomy.
Leg Compression Device to Assist in Ultrasound Testing Design Team: Mark Rawls and Jordan Winston Advisors: Dr. Raul Guzman and Dr. Paul King.
CM27 – 8 th July 2010 LH2 System Progress and Future Plans M Hills T Bradshaw M Courthold I Mullacrane P Warburton.
Non-Invasive Blood Pressure Device for Use in fMRI Imaging Applications November 18, 2004 Jose Alvarado Ben Huh Sanjeet Rangarajan Dr. André Diedrich Dr.
Non-Invasive Blood Pressure Device for Use in fMRI Imaging Applications February 18, 2005 Jose Alvarado Benjamin Huh Sanjeet Rangarajan Dr. André Diedrich.
PRiME (Physiological Recording in MRI Environment) John Kakareka Signal Processing and Instrumentation Section, Division of Computational Bioscience, Center.
A Cost Effective Centralized Single parameter Patient Monitoring System Abstract Lack of Medical monitoring equipment's in rural areas of underdeveloped.
Abstract Client Expectations References Acknowledgments Richard Bamberg, Dustin Gardner, Matt Kudek, Chelsea Wanta Advisor: Willis Tompkins, Ph.D. Client:
By: Mason Jellings, Sarah Reichert, John Byce, and Justin Gearing Advisor: Prof. Tom Yen Client: Dr. Joshua Medow Instrument Controlled Microscopy for.
PRESENTED BY : HATIM DINI.  INTRODUCTION  INDIRECT METHOD  GENERAL FACTS  PULPATORY METHOD  AUSCULATORY METHOD  ULTRASOUND METHOD  OSCILLOMETRIC.
Electrocardiography. Anatomy of the Heart Circulation.
Do Now 11/5/14 1.Which chambers of the heart act as pumps? 2.Where does blood go after it leaves the right ventricle? Through what valve does it pass?
Blood Pressure. Blood pressure is the pressure exerted by circulating blood upon the walls of blood vessels, and is one of the principal vital signs.
Date of download: 7/3/2016 Copyright © 2016 SPIE. All rights reserved. Static autoregulation. The classic static autoregulation curve shows a plateau of.
Chapter 6 Vital Signs.
Heart.
A System View of Optical Fiber Communication
Data practice: CPR survival rates. Get started early to save a life!
Customised Foot Orthoses Design
Motor Drive Prof. Ali Keyhani. Modern Variable Speed System A modern variable speed system has four components: 1. Electric Motor 2. Power Converter -
A System View of Optical Fiber Communication prt.2
Self-Adjusting Jacket ECE 445 SP2017
Topic 6.2 – The blood system
January 21, 2004 Students Advisors Jose Alvarado Ben Huh
Arch205 building construction vertical elevators
Chapter 19: Physiology of the Cardiovascular System
Portable Digital Blood Pressure Monitor
Blood Pressure (BP) -practical
Presentation transcript:

Non-Invasive Blood Pressure Device for Use in fMRI Imaging Applications February 18, 2005 Jose Alvarado Benjamin Huh Sanjeet Rangarajan Dr. André Diedrich Dr. John Gore Dr. Richard Shiavi StudentsAdvisors

Baroreflex Failure Efferent autonomic activity controlling blood pressure is determined at the level of medullary brainstem nuclei.The input information is integrated in these brainstem nuclei, where it serves to determine the balance of efferent parasympathetic and sympathetic activity. An essential portion of the input information comes from the baroreflexes, whose function is to maintain arterial pressure within a narrow range. The baroreflex as a system can potentially be damaged at any site (e.g. aortic arch, carotid arteries). Failure of the baroreflex at any point produces a characteristic clinical syndrome of volatility of blood pressure and heart rate The incidence rate of this disorder is extremely small.

Background Functional Magnetic-Resonance Imaging (fMRI) has recently allowed novel insights into the function of individual brain sites. Patients with baroreflex failure have extremely labile blood pressure due to loss of buffering function of blood pressure control. Higher centers of the brain stem and cortical structures may have potentiating effects on changes in autonomic outflow. Measuring blood pressure continuously during fMRI procedures could provide numerous benefits to the study of autonomic disorders.

The Penaz Measurement Technique If an externally applied pressure is equal to arterial pressure at all times, the arterial walls are unloaded and the photoplethysmogram will be constant. The Finometer uses two methods to determine the set point, which represents the unloaded blood volume.

The Penaz Measurement Technique This set point is used in a servo-loop, so as the measured photoplethysmogram varies from the set point a servo-valve is driven to increase and decrease the cuff pressure to maintain the set point. A fast servo-valve allows cuff pressure to equal arterial pressure throughout each cycle. The cuff pressure is measured with a transducer and the resulting signal is displayed as the arterial pressure.

Existing Schematic

Set Point Criteria Two methods Servo start-up adjustment method Steps cuff pressure upward and interprets magnitude and shape of photoplethysmogram at each step. When cuff pressure is between systolic and diastolic pressures the pulsations of the photoplethysmogram become maximal. Servo self-adjustment method Provides a fine-tuning of the set point and corrects for slowly changing physiologic conditions in the finger. Occurs periodically throughout measuring time as often as every ten beats, and as infrequent as 70 beats.

The Problem Current commercial devices are able to continuously measure blood pressure but not in the presence of magnetic fields. The electrical sensor system for the finger cuff as well as the pneumatic pump interfere with the highly sensitive fMRI magnet leading to distortion of the MRI images. Also, some components of the cuff are made up of ferromagnetic materials are dangerous when placed in close proximity to an fMRI machine.

The Problem (Setup)

Market Analysis Use for only in research and hospital settings. Large market potential because of applications in other MRI and fMRI studies. An optical continuous non-invasive blood pressure measuring device could be used in conjunction with electromagnetic trackers which are used in image guided surgery. Profit will come from selling the design of the fMRI compatible finger cuff to pre-existing companies that manufacture the cuffs. Perhaps too specific of an item to be able to market the device independently and for this reason it could be a better to sell the design to a current company.

Our Solution Retrofit finger cuff blood pressure devices to use optical transmission techniques instead of electrical transmission techniques. Replace electrical cables and sensors with optical components: Fit cuff with an optical cable terminator. Develop an fiber-optic interface compatible with existing commercially available electrical systems. Design fMRI compatible shielding. Extend length between cuff and electrical components without losing pneumatic function. Keep all electrical components that may cause potential interference in the observation room.

Hydraulic System

Estimated Design Cost ItemQuantityCost per ItemTotal USD Fiber Optic Cable15 feet$1/foot$15.00 LEDs4$5-$10$40.00 Air Piping15 feet$0.50/foot$7.50 Photodetector1$25$25.00 Rubber Insulation2 square feet$3/foot 2 $6.00 Inflation Bag for Cuff1$5$5.00 Hydraulic SuppliesVaries$10 Total Cost: $108.50

Completed Tasks Tested pneumatic extension with Finometer and found a maximal extension length of 6 ft using an air-only system. Dissected finger cuff. Visited 3T fMRI scanning facility and measured required extension length (13 feet, 3 inches). Met with Dr. Jansen to discuss strategy for fiber optic transmission. Finished and refined hydraulic system twice, ready for final testing.

Current Work Complete and test fiber optic transmission system. Construct mounts for the hydraulic system that will keep the system at a consistent height to eliminate hydrostatic pressure. Test the whole system in the fMRI lab for accuracy. Finish writing final paper and construct the poster presentation.

Future Work Test the whole system in the fMRI lab for accuracy. Finish writing final paper. Construct the poster presentation.

Questions Questions?