Limulus Amebocyte Lysate (LAL) Test Methods

Slides:



Advertisements
Similar presentations
BL TIER 3 TIER 3 Identify suitable experimental procedures for measuring rates of reactions Identify the factors affecting the rate of a reaction Calculate.
Advertisements

PURPOSE OF RESEARCH  Project as a whole: By using several methods of measuring fungal biomass and speciating fungi, we wish to compare existing and modified.
CHEMISTRY ANALYTICAL CHEMISTRY Fall
Spectrophotometers and Concentration Assays
Part Three Basic Test Methods
Chapter: Solutions, Acids, and Bases
Acid – Base Titrations.
General Approach of Haemostasis Prothrombin Time (Continue) Lecture 5:
Experiment 6 Amount of Dye in a Sports Drink. Goal To make a Beer’s Law standard curve To use the standard curve (and spectrophotometry) to determine.
Enzyme-linked Immunosorbent Assay
In the name of God 1. Summer School Influenza Unit, Pasteur Institute of Iran summer
Chapter 14 Principles of Neutralization Titrations
One Stage Factor V Assay
Chapter 13 Titrimetric Methods
Redox Titrations Introduction 1.) Redox Titration
The effect of inhibitor (Inorganic phosphate & Sodium fluoride) on the rate of an enzyme catalyzed reaction.
Strength of Acids and Bases The strength of an acid (or base) is determined by the amount of IONIZATION This is not the concentration If the acid (or base)
MLAB Coagulation Keri Brophy-Martinez
Acid-Base Titrations.
Spectrophotometric methods for determination of proteins concentration
Spectrophotometers and Concentration Assays
Spectrophotometric methods for determination of proteins
QUALITY CONTROL MICROBIOLOGY
Enzyme-Linked Immunosorbent Assay
Introduction to Analytical Chemistry Dr M. Abd-Elhakeem Faculty of Biotechnology General Chemistry Lecture 7.
for Endotoxin Detection
CHAPTER 25 Lesson 2 Strengths of Acids and Bases.
1 Spectroscopic ANALYSIS Part 5 – Spectroscopic Analysis using UV-Visible Absorption Chulalongkorn University, Bangkok, Thailand January 2012 Dr Ron Beckett.
 It is a solution that does not change the pH of a solution.  Consists of a mixture of a weak acid and its conjugate base or a weak base and its conjugate.
Enzyme Linked Immunosorbent Assay (ELISA). ELISA Enzyme Linked Immunosorbent Assay (ELISA) Term Was Coined By Engvall and Pearlmann in 1971 Different.
Chapter 24: Solutions, Acids, and Bases
Topic 6 Kinetics Rates of reaction Collision theory.
The Enzyme Linked Immunosorbent Assay (ELISA).. Capture ELISAs Antigen Capture: In this, more specific approach, a capturing Ab is adsorbed onto the solid.
Acid/Base Titration Buffers. Buffers A mixture composed of a weak acid and its conjugate base (acidic buffer) OR weak base and its conjugate acid (basic.
Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function.
Section 6.3—Acidity, pH How does concentration of acid affect the pH of a sports drink?
Titration and pH Curves..   A titration curve is a plot of pH vs. volume of added titrant.
ENDOTOXINS Richard Marchand MD Associate professor University of Montreal.
Ch. 15 Titration And pH. Ionization of Water _____________________: two water molecules produce a hydronium ion and a hydroxide ion by transfer of a proton.
Basic Immunologic Procedures Part 1
1 Chapter 2 Steps in a chemical analysis Plan of analysis Before doing any quantitative analysis, the following questions should be answered: 1-
Radiopharmaceutical Production
Titration  Lab technique commonly utilized to determine an UNKNOWN concentration of a chemical compound with a KNOWN concentration of another chemical.
P.Sci. Unit 11 Cont. Solutions, Acids, and Bases Chapter 8.
Introduction Charles River Laboratories revolutionary FDA licensed endotoxin detection system, the Endosafe Portable Test System (PTS), is a rapid point.
Kinetics. Rate of Reaction Reaction kinetics is the study of rates of reaction. The rate of a reaction is defined as the change in concentration per unit.
Imon Rahman. Introduction During the early days of the pharmaceutical industry it was noticed that some solutions when injected into the bloodstream induced.
Luminescence.
Toxinometer™ ET-6000 Endotoxin Detection System System contains : analysis module (1), Toximaster® QC7 software (pre-loaded and separate disk), personal.
Snake and scorpion bites represent a serious public health problem around the world. Antitoxic sera preparations are the only treatment options. These.
Absorption spectrum and spectrophotometric determination of concentration Experiment 1.
Molecular Fluorescence Spectroscopy
Titration and pH Curves..   A titration curve is a plot of pH vs. volume of added titrant.
One-Stage Quantitative
Kinetics analysis of β-fructofuranosidase enzyme
ELISA for mAb detection and Quantification
Titration and pH Curves.
LAL: Choice of Test Method
Quantitative tests for proteins
Limulus Amebocyte Lysate (LAL) Test Methods
Introduction to Solutions
Quantitative tests for proteins
Limulus Amebocyte Lysate (LAL) Test Methods
Quantitative tests for proteins
Titration of Mixtures of Acids Using a Strong Base For two acids to titrate separately, the ka of the stronger one should be at least 104 times greater.
Titration Basics Titration = addition of a measurable volume of a known solution (titrant) to an unknown solution until it is just consumed Use the stoichiometry.
Quantitative tests for proteins
Quantitative tests for proteins
Enzyme Linked  Immuno sorbent  Assay  
Urine Analysis Protein. Protein: Urine normally contains a scant amount of protein, which derives both blood and urinary tract itself. Proteins present.
Presentation transcript:

Limulus Amebocyte Lysate (LAL) Test Methods

LAL Test Methods The gel-clot method The kinetic turbidimetric method The chromogenic methods (kinetic and endpoint)

Gel-Clot Turbidimetric Chromogenic

Biochemical Reaction Turbidimetric Endotoxin Factor C Activated Factor C b-Glucan Factor B Activated Factor(s) B/G Factor G Clotting Enzyme Activated Clotting Enzyme Coagulogen Coagulin Gelation Turbidimetric Modified from Iwanaga et al., 1985

The Gel-Clot Method Simplest and most widely used The USP referee method The labeled gel-clot reagent sensitivity (l) is the least concentration of endotoxin to cause a solid clot under standard conditions

Reading the Gel-Clot Test positive cloudy negative

Turbidimetric Methods As coagulin molecules coalesce forming particles, the reaction mixture becomes turbid The rate of increase in turbidity is a function of endotoxin concentration

Turbidimetric Methods light DETECTOR

Kinetic Data - OD vs time optical density time threshold OD 0.25 EU/ml 0.125 EU/ml 0.0625 EU/ml 0.03125 EU/ml 0.5 EU/ml

Kinetic Turbidimetric Method The threshold OD (onset OD) is used as a point of reference for data collection The greater the endotoxin concentration, the shorter the time taken to reach the onset OD

Kinetic Turbidimetric Method The onset time is the time that it takes (in seconds) for the reaction to reach the onset OD Standard curves are constructed by plotting log10(onset time) on log10(endotoxin concentration)

Standard Curve (Kinetic Test)

Kinetic Turbidimetric Method Calculate sample endotoxin content by comparing with standards Take sample onset time and reference against standard curve to determine its endotoxin content

Interference Most samples, at some concentration, interfere with the LAL reaction Interference is caused by sample interaction with the LAL reagent sample interaction with endotoxin

Inhibition Inhibition is a reduction in sensitivity of the assay which causes an underestimation of the concentration of endotoxin Inhibition controls (PPC’s) prevent misinterpretation of negative results

Enhancement Enhancement is an increase in the sensitivity of the assay which causes an overestimation of the concentration of endotoxin Positive product controls (PPC’s) prevent misinterpretation of positive results in the photometric methods

False Positives Enhancement is not a false positive! A false positive test is a positive in the absence of endotoxin False positives are rare trypsin (all methods) activated serine proteases (chromogenic) beta-glucans (suspected, all methods)

Positive Product Control All LAL tests must have a control to demonstrate that the sample itself does not cause a false negative result A known quantity of endotoxin is added to a portion of the sample under test to provide an inhibition or positive product control (PPC)

Remove Interference Dilute with LRW first Use a more sensitive LAL reagent or method to increase the MVD Reconstitute LAL with Pyrosol (strongly buffered products outside the pH range, highly concentrated electrolytes, or for sample/endotoxin interactions)

Maximum Valid Dilution The maximum valid dilution (MVD) is the greatest possible dilution at which the limit can be detected This is the dilution used for the pass/fail test The MVD increases with increasing test sensitivity

Maximum Valid Dilution If l is 0.125 EU/mL and the unknown has an endotoxin limit of 2 EU/mL, calculate the MVD: Limit in EU/mL 2 EU/mL = 16 = l in EU/mL 0.125 EU/mL Limit in EU/mL 2 EU/mL = 64 = l in EU/mL 0.03125 EU/mL

Select a Sensitivity Sensitivity is lowest point on curve Consider the endotoxin limit and MVD Perform preliminary tests If interference cannot be overcome without exceeding the MVD of a product, go to a more sensitive reagent or method