Ageing of large CsI photocathodes exposed to ionizing radiation in a gaseous RICH detector A. Braem, G. De Cataldo, A. Di Mauro, A. Franco, A. Gallas Torreira,

Slides:



Advertisements
Similar presentations
GEM Chambers at BNL The detector from CERN, can be configured with up to 4 GEMs The detector for pad readout and drift studies, 2 GEM maximum.
Advertisements

GM COUNTER.
Drift velocity Adding polyatomic molecules (e.g. CH4 or CO2) to noble gases reduces electron instantaneous velocity; this cools electrons to a region where.
Radiation Detectors / Particle Detectors
1 First observations on CsI -coated ThGEM in Trieste presented by Gabriele Giacomini phone meeting 06/08/2008.
Peter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute The HERA-B RICH counter.
Aging, High Rate and Shielding L. Lopes Lip-Coimbra.
Characterization of primed state of CVD diamond by light and alpha particles C. Manfredotti Experimental Physics Department University of Torino INFN-
Detector R&D: J-PARC-E16 K. Ozawa (Univ. of Tokyo) for the E16 collaboration.
The problem: The latest theoretical and experimental results suggest investigating a physics domain for p t higher than the actual one covered by the HMPID.
Proportional Counters
Detectors. Measuring Ions  A beam of charged particles will ionize gas. Particle energy E Chamber area A  An applied field will cause ions and electrons.
Outer Tracker Comments Aging Review Report: Ulrich Uwer University of Heidelberg LHCb Week May 27, 2004.
1 Alhussein Abuhoza, S. Biswas, U. Frankenfeld, J. Hehner, Ch. Schmidt GSI, Darmstadt, Germany and H.R. Schmidt University of Tübingen, Germany Study of.
CsI Photocathode Production and Testing
Catania VLVnT09 Athens, Greece 1/14 Catania Performances of four super bialkali large area photomultipliers with respect to.
Cube Measurements Tent Crew. Scintillation BNL 241 Am Semi- collimated  Spectralon Diffuse UV Reflector SBD  -Trigger Scint. Light Poisson.
Preliminary Results from a Trial Beam Test of the Small HBD Prototype at LEGS Bob Azmoun BNL HBD Working Group Meeting May 10, 2005.
1 First observations on CsI -coated ThGEM in Trieste presented by Gabriele Giacomini phone meeting 06/08/2008.
Pylos RICH2002 F.Piuz CERN 1 RING IMAGING CHERENKOV SYSTEMS BASED ON GASEOUS PHOTO-DETECTOR TRENDS AND LIMITS F. PIUZ CERN Geneva, Switzerland.
First results of an aging test of a full scale MWPC prototype for the LHCb muon system Workshop on “Aging Phenomena in Gaseous Detectors”, October 2-5.
This study has been performed with an apparatus consisting of two identical standard MWPCs, a clean gas system, an outgasing box, a Copper X-ray generator.
Test Beam the μ-RWELL Test Beam the μ-RWELL G. Bencivenni (a), R. de Oliveira (b), M. Gatta (a), G. Felici (a), G. Morello (a), M. Poli Lener (a) (a),
1 NA58 RICH Test Beam T10 sept.2014 Fulvio Tessarotto – Stefano Levorato Partecipants: Alessandria, Aveiro, Budapest, Calcutta, Freiburg, Liberec, Prague,
DEVELOPMENT OF A THGEM-BASED PHOTON DETECTOR FOR CHERENKOV IMAGING APPLICATIONS Silvia Dalla Torre INFN - Trieste On behalf of an Alessandria-CERN-Freiburg-Liberec-Prague-Torino-TriesteCollaboration.
CsI-TGEM vs. CsI-MWPC photodetector Some part of this work was performed in collaboration between CERN and Breskin group.
Study of UV absorption and photoelectron emission in RPC (Resistive Plate Counters) detector with an UV source Carlo Gustavino (INFN-LNGS) RPC and their.
work for PID in Novosibirsk E.A.Kravchenko Budker INP, Novosibirsk.
Ionization Detectors Basic operation
1 5-9 October th ICATPP, Como, Italy S. Maltezos NITROGEN MOLECULAR SPECTRA OF AIR FLUORESCENCE EMULATOR USING A LN 2 COOLED CCD S. Maltezos, E.
Andrea Di Simone Andrea Di Simone – INFN Roma2 Andrea Di Simone CERN PH/ATC and INFN-CNAF On behalf of ATLAS RPC groups: Lecce, Napoli, Protvino, Roma2.
R & D at BHU B.K. Singh (On behalf of HEP Group).
GEM basic test and R&D plan Takuya Yamamoto ( Saga Univ. )
The 21st International Conference on Ultrarelativistic nucleus-nucleus collisions, March 30 – April 4, Knoxville, TN Results from cosmics and First LHC.
Catania 11 ICATPP october, 2009 Como 1/12 Catania Comparative measurements of the performances of four super bialkali large.
The HMPID performance during the LHC run period Detector stability and PID performance (with the contribution of: G. Volpe, S. Bansal, A Franco,
Jyly 8, 2009, 3rd open meeting of Belle II collaboration, KEK1 Charles University Prague Zdeněk Doležal for the DEPFET beam test group 3rd Open Meeting.
Study of TGEMs and RETGEMs for the possible ALICE upgrade By Himank Anand and Isha Shukla (CERN summer students) Supervisor :Vladimir Peskov.
The RICH Detectors of the LHCb Experiment Carmelo D’Ambrosio (CERN) on behalf of the LHCb RICH Collaboration LHCb RICH1 and RICH2 The photon detector:
Itzhak Tserruya RICH 2004, Playa del Carmen 1 RICH 2004 Playa del Carmen, Mexico Nov. 30 – Dec. 5, 2004 Itzhak Tserruya Short report from RICH 2004 at.
1 P.Rebecchi (CERN) “Monitoring of radiation damage of PbWO 4 crystals under strong Cs 137  irradiation in GIF-ECAL” “Advanced Technology and Particle.
Results from the CMS-RPC tests at CERN Gamma Irradiation Facility Roberto Guida CMS-RPC (Bari, Beijing, Napoli, Pavia, Sofia) Seoul, October 2005.
Mauro Iodice INFN Roma (Italy) Hall A Collaboration Meeting - JLAB May 18, 2004.
Cold PM test at Indiana final measurements, September 9 – 24, 2007 PM under test: Hamamatsu R7725 serial # ZK3692 (tube with Pt underlayer) Hans-Otto Meyer.
TCPD test measurement 1 TCPD (TGEM CCC Photon Detector) test measurement ELTE, MTA KFKI RMKI, REGARD Group (Budapest, Hungary): Levente Kovács G. Hamar,
Update on the HBD Craig Woody BNL DC Meeting May 11, 2005.
1 Two-phase Ar avalanche detectors based on GEMs A. Bondar, A. Buzulutskov, A. Grebenuk, D. Pavlyuchenko, Y. Tikhonov Budker Institute of Nuclear Physics,
1 HBD R&D: Update Itzhak Tserruya (for A. Kozlov, I. Ravinovich and L. Shekhtman) Weizmann Institute, Rehovot DC meeting Feb. 14, 2003.
Quality-evaluation of Cesium Iodide photocathodes for the ALICE/High Momentum Particle Identification detector by means of a VUV-Scanner system Herbert.
Radiation study of the TPC electronics Georgios Tsiledakis, GSI.
1 A two-phase Ar avalanche detector with CsI photocathode: first results A. Bondar, A. Buzulutskov, A. Grebenuk, D. Pavlyuchenko, R. Snopkov, Y. Tikhonov.
1 Fulvio TESSAROTTO GDD meeting, CERN, 01/10/2008 Trieste THGEM news New THGEM test in the COMPASS hall New THGEM test in the COMPASS hall Preparation.
Update on THGEM project for RICH application Elena Rocco University of Eastern Piedmont & INFN Torino On behalf of an Alessandria-CERN-Freiburg-Liberec-
1 HMPID Project Assembly & Tests of CsI QE monitors: VUV-scanner and ASSET Milestone: December 2002.
R&D on Hadron Blind detector, recent results Issues addressed: - gain limits in CF 4 with heavily ionizing particles - operation.
Mg Films Grown by Pulsed Laser Deposition as Photocathodes: QE and surface adsorbates L. Cultrera INFN – National Laboratories of Frascati.
XI Workshop on Resistive Plate Chambers and Related Detectors, INFN, 5-10 February, Aging test of high rate MRPC Wang Yi Department of Engineering.
Pulsed Laser Deposition and Quantum Efficency of Mg films University of Lecce L. Cultrera.
1 RICH 2013 Long term experience with CsI photocathodes in gas photon detectors S. Dalla Torre on behalf of the Trieste-COMPASS group.
RD_FA : WP - 4 C. Chatterjee1, S. Dalla Torre, S. Dasgupta2, S. Levorato, F. Tessarotto, A. Valentini1 1 – new 2 – formally he cannot be list.
Study of TGEMs and RETGEMs for the possible ALICE upgrade
DOE Plasma Science Center Control of Plasma Kinetics
IBF in THGEM-based PHOTON DETECTORS, an update
Long irradiation at the side
Production of Cesium Iodide Photocathodes for
Instrumentation for Colliding Beam Physics 2017
Beam Tests of Ionization Chambers for the NuMI Neutrino Beam Monitoring System MINOS.
First observations on CsI-coated ThGEM in Trieste - ADDENDUM -
MWPC’s, GEM’s or Micromegas for AD transfer and experimental lines
Gain measurements of Chromium GEM foils
Presentation transcript:

Ageing of large CsI photocathodes exposed to ionizing radiation in a gaseous RICH detector A. Braem, G. De Cataldo, A. Di Mauro, A. Franco, A. Gallas Torreira, H. Hoedlmoser, P. Martinengo, E. Nappi, F. Piuz, E. Schyns Bari INFN, CERN Attending the conference

1 RICH DETECTOR BASED ON CsI-PHOTOCATHODE The cathode of a gaseous detector is coated with a thin layer of CsI converting photons of WL < 210nm. The single electron emitted is in turn amplified in the detector. Usually, the coated cathode is segmented into pads allowing for a 2D-localization of the Cherenkov photons. CsI-RICH systems under operation DRAWBACK OF THE MWPC OPEN GEOMETRY: THE CsI-PC IS BOMBARDED BY THE IONIC AVALANCHES Fig. 1 Table 1

2 SOME HINTS ABOUT LARGE CsI-PHOTOCATHODES Chemical affinity: CsI strongly hygroscopic, however QE (quantum efficiency) recovers from short exposure to ambient atmosphere (10-15% hygr.) (see H. Hoeldmoser’s talk at this conference) no reaction with gases: O 2,noble gases,CO 2, CH 4,etc (NIM A461,2001,584) Photocathode production to achieve a reproducible QE performance over a large series (42 PCs), follow a strict procedure. Main steps are: - specific elaboration and cleaning of the PCb-pad substrate - thermal evaporation under vacuum followed by an heat treatment under vacuum - protection of the PC in a protective vessel flushed with Argon - during any transfer or storage, keep the PC under argon flow,( H 2 O, O 2 <10ppm) PC QE PERFORMANCE KEPT STABLE OVER SEVERAL YEARS (NIM A ) Fig. 2 CsI-QE differential plot for several large PCs obtained from Cherenkov events analysis. Performance has been improved from 1998 production (green curve)

3 AGEING induced at the CsI surface, driven by physisorption mechanisms such as surface photolysis or electrolysis. Ageing is observed under: - direct exposure to high photon flux (≈ ph/s,cm 2,no elecrical field) (not relevant in most of RICH application) - ion sputtering from avalanches in the wire chamber under gas gain STATUS OF AGEING STUDIES - in the 90ies, several lab studies, mixing photon irradiation and ion sputtering, and also exposure to air (NIM A364,1995,243; A387,1997,154; A454,2000,365) - HEP experiments at SPS test beams, NA44 and STAR using low level irradiation (NIM A502, 2003,76) - COMPASS: exposed to high level irradiation; some damaged spots observed (S. Dalla Torre, this conference) - HADES-GSI GOAL OF THIS WORK - study the ageing of the CsI-PC under the only ion sputtering of avalanches - using the final CsI-PCs and detector modules to be installed at the ALICE experiment - CsI-PCs never exposed to ambient atmosphere EXPERIMENTAL METHOD - irradiate several spots of the PC with Sr90-electrons - record the accumulated charge deposited by the ion avalanches - correlate this charge with the measurement of the CsI-QE at these spots and at non irradiated reference points

4 Sr90 C6F14 Test beam CsI - PC annular Cherenkov fiducial zone irradiation spot Sr90 CsI - PC annular Cherenkov fiducial zone irradiation spot EXPERIMENTAL LAYOUT - RICH detector equipped with a 60x48cm 2 CsI-PC anode wires running horizontal (Proto-3) - small C 6 F 14 radiator defining a Cherenkov fiducial zone (blue) on the CsI-PC - 3 positions for locating a collimated Sr-90 source irradiating spots of 32mm Ø overlapping the fiducial Cherenkov ring Only 8 wires, overlapping the irradiated spots, are raised at full amplification voltage. The ageing will be effective along a stripe, 64 cm long, 35mm wide. C 6 F 14 radiator Fig. 3Fig. 4 Proto-3

5 MEASUREMENTS during the irradiation - I a = anodic current of the 8 wires raised to HV (to check possible anode wire ageing) - I 20pads = cathodic current of the 20 pads covered by the direct irradiation cone. (Pad size: 8.0x8.4mm 2 ) - anode wire rate and pulse height spectra - charge at each cathode pad using the analog RO Gassiplex As seen in the fig. 6, the charge profile can be measured per pad rows (and pad columns) These elements provide the distribution of the ionic charge accumulated on the CsI-cathode (mC/cm 2,s) to be correlated with the expected QE variation Sr-90 source collimator Along the X-coord, the charge profiles extend largely out of the direct irradiated area (blue lines), due to the multiple scattering of the electrons in the methane and the brehmstralung photons emitted from the thin mylar window Fig. 5 Fig. 6

6 LOCAL MEASUREMENTS OF THE CsI QUANTUM EFFICIENCY (QE) 2 methods are used: -THE VUV-SCANNER A XY-stepping system allows to scan in vacuum a CsI-PC with a VUV-beam, Ø3-5mm, D 2 -lamp. The photocurrent is recorded at every pad and normalized to a reference PM. A pad mapping of a quantity proportionnal to the QE integrated over the VUV-beam spectrum is shown in Fig. 7 - Cherenkov single photon counting (see layout in Fig. 3) Exposing the detector at a pion beam, Cherenkov rings that overlap the irradiated spots are recorded. Further analysis provide the differential QE plot at selected location in thr Cherenkov fiducial area. (see results in fig. 18, 19) Fig. 7 Fig. 8 Position 2 Position1 Position 3

7 IRRADIATIONS SEQUENCE position Date start Source MBq HV V Duration hours I 8wires nA I 20pads nA Dose mC/cm 2 Dose rste mC/cm 2,s E E E Dose = (I 20pad x duration[s]) / (20 pads area = 13.44cm 2 VUV-SCANS SEQUENCE positionEnd irradiationVUV scanDays after irrad position position Position Position Position position Position Table 2 Table 3 The expected dose integrated over 10 years operation in ALICE is evaluated at 0.5 mC/cm 2 FOR COMPARISON WITH ALICE EXPOSURE

8 anode wires VUV-scans after irradiation at 3 positions (see table 2) X Y “Inefficiency” profiles taken along pad columns (Y-coord) at different X values. The sharp edges fit the stripe of the 8 anode wires under voltage making the ageing effect well restricted to the area where ions impinge the CsI-PC Fig. 9 Fig. 10 Fig. 11

9 NOVEL BUT UNDESIRED EFFECT: IT HAS BEEN OBSERVED THAT THE DEGRADATION OF THE QE INCREASES WITH THE TIME ELAPSED AFTER THE IRRADIATION IN ABSENCE OF GAS GAIN OR FURTHER IRRADIATION. THIS BEHAVIOUR HAS BEEN FOUND AT THE 3 POSITIONS OF IRRADIATION WHILE NON-IRRADIATED SPOTS WERE CHECKED TO BE STABLE Fig. 12

10 EVOLUTION OF THE INEFFICIENCY WITH THE TIME ELAPSED AFTER THE IRRADIATION X-PROFILES ALONG X-COORDY-PROFILES ALONG Y-COORD POSITION1: TIME ELAPSED after irrad: 52,142,155 DAYS (table 2 POSITION 2:TIME ELAPSED after irrad: 3,118,208,220 DAYS (table 2) Fig. 13 Fig. 14

11 HOW THE CsI-QE DECREASE IS CORRELATED TO THE IONIC DOSE DEPOSITED BY THE IRRADIATION ? AS SEEN IN FIG. 6,13 AND 14, BOTH, THE CHARGE AND CsI-QE DECREASE ARE MEASURED ALONG THE SAME PAD ROW. THEN, FIG. 15 SHOWS THE VARIATION OF THE CsI-QE DECREASE VS CHARGE ALONG THE X-COORD. THE DEPENDANCE IS NOT FOUND LINEAR. A LOGNORMAL FUNCTION IS ACTUALLY CHOSEN TO FIT THE DATA, SUGGESTING THAT THE LOSS OF PHOTOEMISSIVE SITES IS PROPORTIONAL TO THE SITE DENSITY Fig. 15

12 Fig. 16 Fig. 17 THE DEPENDANCE OF THE CsI-QE DECREASE VS CHARGE CAN ALSO BE OBTAINED FROM THE MEASUREMENTS MADE AT DIFFERENT DOSES AND POSITIONS (SEE FIG. 12) WHEN TAKEN AT THE SAME TIME ELAPSED AFTER IRRADIATION AS SHOWN IN FIG. 16, THE FITTING FUNCTIONS ARE NOT THE SAME AT DIFFERENT TIMES A GOOD AGREEMENT (FIG. 17) WITH THE PREVIOUS METHOD (FIG. 15) IS ONLY OBTAINED FOR THE SAME ELAPSED TIME, OF 129 DAYS IN THIS PLOT. SUCH A BEHAVIOUR MIGHT BE CORRELATED WITH THE INCREASE OF THE AGEING EFFECT WITH TIME expected dose for 10 y ALICE operation (0.50 mC/cm 2 )

SIMULATION x (pads) y (pads) Test beam with PROTO 3 in agreement with QE decrease measure by VUV scanner The differential CsI-QE curve has to scaled down as seen in Fig. 18 to obtain a simulation close to the data (Fig. 19) COMPARISON BETWEEN THE VUV-SCANS AND THE CHERENKOV PHOTON COUNTING Fig. 19 Fig. 18

14 COMMENTS THESE PRELIMINARY RESULTS ARE RAISING SEVERAL QUESTION MARKS WHY THE CsI-QE DECREASE EVOLVE AFTER IRRADIATION ? IS THE RATE OF AGEING INCREASE WITH TIME CORRELATED WITH THE INITIAL DOSE ? IS THE AGEING DRIVEN BY THE PARTICLE RATE OR BY THE INTEGRATED CHARGE ? IS IT STILL AN INFLUENCE OF RESIDUAL ADSORBED WATER ?