An evaluation of the isotope ratio performance of an axial TOF - ICP - mass spectrometer F. Vanhaecke, L. Moens, R. Dams, L. Allen* and S. Georgitis* University.

Slides:



Advertisements
Similar presentations
FC-MS from Teledyne Isco CombiFlash ® a Name You Can Rely On.
Advertisements

Advanced Higher Unit 3 Mass Spectrometry. Mass spectrometry can be used to determine the accurate molecular mass and structural features of an organic.
Atomic Mass Spectrometry Yongsik Lee Introduction ► Atomic mass spectrometry  Versatile and widely used tool  All elements can be determined ►
Module 2: Analytical methods and separation techniques.
Inductively Coupled Plasma Mass Spectrometry Dr. Lloyd Allen and Dr
Ion Sources Some characteristics of ion sources (especially in high precision work): It should have high efficiency in generating ions of the element of.
1 Atomic Absorption Spectroscopy Atomic Emission Spectroscopy Lecture 18.
Inductively Coupled Plasma Mass Spectrometry or ICP-MS
IDDepth δ 13 CCarbonNitrogenδ 13 C Depth: (cm)(‰)(%) (‰)(cm) Gr. 3-5A %0.17% Gr %0.12% Gr %0.09%
1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission.
Chemical Analysis Qualitative Analysis Quantitative Analysis Determination “Analyze” a paint sample for lead “Determine” lead in a paint sample.
Mass Spectroscopy Mass Spectrometry ä Most useful tool for molecular structure determination if you can get it into gas phase ä Molecular weight of.
Instrumental Chemistry Chapter 11 Atomic Mass Spectrometry.
Atomic Mass Spectrometry
Mass spectrometry 8/23/12 What are the principles behind MS?
Isotope Ratio Performance of an Axial Time of Flight ICP-MS Stuart Georgitis 1, Lloyd Allen 1, and Janos Fucsko 1, Frank Vanhaecke 2 1 LECO Corporation.
Types of Analyses Single Collector –Uses only one collector or detector –This is the case with a Daly or EM Advantages: –Need only one detector –Peaks.
V. V. Parkhomchuk, S.A. Rastigeev BINP, Novosibirsk, Russia. ION SELECTION IN ACCELERATOR MASS SPECTROMETER BINP SB RAS.
4. Mass Spectrometry Objectives:
Agilent 7500 ICP-MS Date: 28JUL2005.
TOF Mass Spectrometer &
Chapter 8 – Mass Spectrometry. Mass Spectrometry The mass spectrometer can be used for: – Quantitative analysis – as a sophisticated and very sensitive.
Secondary ion mass spectrometry (SIMS)
Mass spectrometry Ions are analyzed on the basis of their m/z Chlorine has 2 isotopes, 35 Cl and 37 Cl, in the approximate ratio of 3 :1. Electrons are.
Define Problem Select Appropriate Methods Obtain and store sample Pre-treat sample Perform required measurements Compare results with standards Apply necessary.
INF380 - Proteomics-61 INF380 – Proteomics Chapter 6 – Mass Spectrometry – MALDI TOF The MALDI-TOF instruments are the simplest MS instruments suitable.
Effect of Sample Introduction on ICP-MS Detection of Nonmetal Analytes in a Simulated HPLC Matrix Adam E. Dill and Jon W. Carnahan Department of Chemistry.
Hiden Analytical Excellence in plasma diagnostics An Introduction to the Hiden EQP Excellence in Plasma Diagnostics An Introduction to the Hiden EQP
Mass Spectroscopy Mass spectroscopy is a powerful tool in organic chemistry that can be used to provide information about the molecular formula and structural.
MASS SPECTROMETRY. CONTENTS Prior knowledge Background information The basic parts of a mass spectrometer The four stages of obtaining a spectrum How.
Overall efficiency and speed of analysis – key factors are ions per sample molecule desorbed and ions produced per laser pulse Precision and accuracy of.
MS Calibration for Protein Profiles We need calibration for –Accurate mass value Mass error: (Measured Mass – Theoretical Mass) X 10 6 ppm Theoretical.
Chemistry Topic: Atomic theory Subtopic : Mass Spectrometer.
7-1 Atomic Mass Spectroscopy General Features Mass Spectrometers §Sources Atomic mass spectroscopy determines elements by mass §Mass to charge ratio §Lower.
Detection Limits N no longer >> N B at low concentration What value of N-N B can be measured with statistical significance? Liebhafsky limit: Element is.
Mass spectrometry (Test) Mass spectrometry (MS) is an analytical technique that measures masses of particles and for determining the elemental composition.
Advanced Analytical Chemistry – CHM 6157® Y. CAIFlorida International University Updated on 9/26/2006Chapter 3ICPMS Interference equations Isobaric.
Instrumental Methods: Intro
Validation Defination Establishing documentary evidence which provides a high degree of assurance that specification process will consistently produce.
LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY
1 Chapter 12 Mass Spectrometry ( 質譜 ) Organic Chemistry, 7 th Edition L. G. Wade, Jr. 藥學系 藥物分析科 許秀蘊 教授
Chapter 29 Mass Spectrometry. 29 A Principles of mass spectrometry In the mass spectrometer, analyte molecules are converted to ions by applying energy.
Atomic Structure Topic The mass spectrometer Describe and explain the operation of a mass spectrometer Describe how the mass spectrometer.
Yonsei Proteome Research Center Peptide Mass Finger-Printing Part II. MALDI-TOF 2013 생화학 실험 (1) 6 주차 자료 임종선 조교 내선 6625.
1 In-house Training on ICP-MS. ICP-MS??? Inductively Coupled Plasma Mass Spectrometry or ICP-MS is an analytical technique used for elemental determinations.
Metabolomics Part 2 Mass Spectrometry
THE MASS SPECTROMETER WHAT IS A MASS SPECTROMETER
M. Fernández, E. Conde, A.I. Barrado Department of Technology
Atomic Structure Topic 2.
Chemical Ideas 6.5 Mass spectrometry.
S. A. Rastigeev , A. R Frolov, A. D. Goncharov, V. F. Klyuev, E. S
Acknowledgements Slides and animations were made by Dr. Jon Karty Mass Spectrometry Facility Indiana University, Bloomington.
Chem. 133 – 4/13 Lecture.
Tandem MS.
Experiments at LCLS wavelength: 0.62 nm (2 keV)
Starter Complete the quiz and hand in..
Instrumental Methods: Intro
Metabolomics Part 2 Mass Spectrometry
Mass Spectroscopy. Mass Spectroscopy Mass Spectrometry Most useful tool for molecular structure determination if you can get it into gas phase Molecular.
Strategies for Eliminating Interferences in Optical Emission Spectroscopy Best practices to optimize your method and correct for interferences to produce.
Chem. 133 – 2/16 Lecture.
Instrumental Chemistry
Instrumental Chemistry
Molecular Mass Spectroscopy
Satish Pradhan Dnyanasadhana College, Thane. Department of Chemistry S
Isotope Ratio Mass Spectrometers (IRMS)
Suppose we make a new compound containing of C,H, and N
Presentation transcript:

An evaluation of the isotope ratio performance of an axial TOF - ICP - mass spectrometer F. Vanhaecke, L. Moens, R. Dams, L. Allen* and S. Georgitis* University of Ghent, Laboratory of Analytical Chemistry, Proeftuinstraat 86, B-9000 Gent, Belgium. * LECO Corporation, 3000 Lakeview Ave., St. Joseph, MI USA.

1. Instrumentation LECO Renaissance TOF-ICPMS instrument 123 Acceleration Extraction Skimmer Sampler Detector Ion Mirror ICP Torch Vacuum Stages Flight Tube

simultaneous extraction of ions from ICP  automatic correction for instabilities in ion source & during ion sampling process on-axis acceleration up to full spectra / s modified ETP discrete dynode electron multiplier  use of analog detection mode to avoid pulse counting saturation  no correction for detector dead time required Meinhard type C concentric nebulizer & double pass cyclonic spray chamber (PTFE)  free aspiration of sample solution 1. Instrumentation (2)

2. Data processing Accurate background correction  application of 1 mV offset value

 use of signal intensity at m/z = 210 as internal reference (no ionic species present)  correction for signal drift & instrument instability  improvement of both precision & accuracy Isotope ratio precision expressed as RSD% for n = Data processing (2)

3.Influence of acquisition time per replicate & analyte concentration on precision

3.Influence of acquisition time per replicate & analyte concentration on precision(2)

Despite use of analog mode, trends in agreement with Poisson counting statistics  average improvement 50  500  g/L = 3.5   10 = 3.2 (A)  average improvement 1  30 s = 5.2   30 = 5.5 (A)  average improvement 10  30 s = 1.8   3 = 1.7 (B) No further improvement on increasing acquisition time > 30 s !! 3.Influence of acquisition time per replicate & analyte concentration on precision (3)

4.Typical isotope ratio precision at 30 s acquisition time / replicate (RSD% for n = 10)

5.Isotope ratio precision for transient signals (RSD% for n = 5) - FI signals, FWHM = 6 s Isotope ratio precision at 500 µg/L still comparable to the best values ever reported for quadrupole-based ICPMS

6. Mass discrimination 1% per mass unit at mid-mass (Cd)  comparable to quadrupole-based or sector field instrumentation 7. Accuracy The accuracy attainable was evaluated by comparing Pb isotopic results for a Merck Titrisol ('natural') Pb standard solution obtained by (1) TOF- ICPMS and (2) single collector TIMS. For TOF-ICPMS, 'NIST Common lead isotopic standard' was used as an external standard.

For TOF-ICPMS, the result is the average of 3 subsequent determinations,. Each determination consisted of 5 replicate measurements (30 s) of the 'sample' solution and 5 replicate measurements of the NIST standard solution. For TIMS, the result is the average of 5 determinations. For both techniques, the uncertainties indicated are 95% confidence intervals. 7. Accuracy (2)

At sufficiently high signal intensities and for a moderate acquisition time, axial TOF-ICPMS yields a typical isotope ratio precision  0.05% RSD. This precision can be obtained for many ratios simultaneously. The mass discrimination is comparable to that observed with quadrupole-based and sector field instruments and when appropriately corrected for, accurate results can be obtained. 8. Conclusions