1/10/2016Collaboration Meeting Feb 2011 Start Counter 1 Start Counter Update W. U. Boeglin L. Guo E. Pooser Measurements on EJ-212 scintillator bars Attenuation.

Slides:



Advertisements
Similar presentations
NDVCS measurement with BoNuS RTPC M. Osipenko December 2, 2009, CLAS12 Central Detector Collaboration meeting.
Advertisements

Collaboration Meeting, Oct : Start Counter Start Counter Eric Pooser W. U. Boeglin L. Guo P. Khetarpal Experimental Setup SiPM Scan Attenuation Lengths.
Micromegas studies using cosmic rays Franck Sabatié May 7th 2009 Saclay cosmic ray bench Data acquisition system and analysis tools MIP detection Position.
R&D of Strip/Block Scintillators E.P.Jacosalem, S.Iba, N.Nakajima, H.Ono, A.L.Sanchez, A.M.Bacala & H.Miyata GLD Calorimeter Group 8 th ACFA Workshop on.
CPR2 Optical Cables Joey Huston Michigan State University …taking optical signals from edge of CPR/crack detectors to PMT boxes mounted on back of wedges.
Silicon Photomultiplier Readout Electronics for the GlueX Tagger Microscope Hall D Electronics Meeting, Newport News, Oct , 2007 Richard Jones, Igor.
Studies of Small Scintillating Cells with Modified Geometries Alexander Dyshkant for NICADD at NIU.
Genova/Pavia/Roma Flavio Gatti, PSI, July 1st, Timing Counter july 2004.
Scintillation Detectors
Forward TOF Prototyping Ryan Mitchell GlueX Collaboration Meeting November 2005.
Trend ground array 1 1.General considerations 2.Study on the prototype detector 3.Inclined ground array simulation 4.Making work plan in 2014.
The Transverse detector is made of an array of 256 scintillating fibers coupled to Avalanche PhotoDiodes (APD). The small size of the fibers (5X5mm) results.
Report on SiPM Tests SiPM as a alternative photo detector to replace PMT. Qauntify basic characteristics Measure Energy, Timing resolution Develop simulation.
Photo-detector and scintillator studies at ITEP Outline 1. Check of casting form 2. Study of MRS APD’s from CPTA 3. MC simulation of light collection in.
May 31, 2008 SuperB PID sessionMarko Starič, Ljubljana Marko Starič J. Stefan Institute, Ljubljana Report on hardware tests and MC studies in Ljubljana.
THE FORWARD PROTON DETECTOR AT DZERO Gilvan Alves Lafex/CBPF 1) MOTIVATION 2) DETECTOR OPTIONS 3) FPD R&D 4) OUTLOOK Lishep 98 Lafex/CBPF Feb 17, 1998.
1 Prototype Calibration System for Forward Hadron Calorimeter L. Almeida, M. Baarmand, L. Caraway, M. Hohlmann, T. Qureshi, I. Vodopianov FLORIDA TECH.
Start Counter Collaboration Meeting September 2004 W. Boeglin FIU.
PANDA Anton A. Izotov, Gatchina A.A.Izotov, Gatchina, PANDA Forward TOF Walls. Side TOF walls in dipole Magnet SiPM/PMT187.
Optimization of Detectors for Time of Flight PET Marek Moszyński, Tomasz Szczęśniak, Soltan Institute for Nuclear Studies, Otwock-Świerk, Poland.
MEG II 実験のための 陽電子タイミングカウンター実機建設 Construction of Positron Timing Counter for MEG II experiment 西村美紀(東大) 他 MEGIIコラボレーション 日本物理学会 2015年 秋季大会 大阪市立大学(杉本キャンパス)
IoP HEPP Lancaster 31/03 – 02/04/08 Gavin Davies Dr. Athanasios Hatzikoutelis Dr. Laura Kormos Lancaster University T2K UK Collaboration  DS-Ecal Design.
Systematic Studies of Small Scintillators for New Sampling Calorimeter E.P.Jacosalem, S.Iba, N.Nakajima, H.Ono, A.L.Sanchez, A.M.Bacala & H.Miyata GLD.
Experimental Nuclear Physics Some Recent Activities 1.Development of a detector for low-energy neutrons a. Hardware -- A Novel Design Idea b. Measure the.
PCAL Strips light transmission properties April 25, 2011April 26, 20111April 27,
Gain stability and the LYSO beam radiation monitor measurements
R&D of Calorimeter using Strip/Block Scintillators with SiPM
UPDATE ON THE FORWARD PROTON DETECTOR Gilvan Alves Lafex/CBPF Introduction Accelerator Roman Pots Detector Future Plans April 1, 1998 (no kidding)
FT Prototype Test Purpose: test the prototype of the FT Calorimeter and Scintillator Hodoscope Technique: prototype is installed under the Hall B photon.
ECFA-DESY, Praha 16/11/02 S. Nemecek: Tiles & fibres... Tile and fibre measurements (Tile HCAL) J. Cvach, S. Němeček Institute of Physics AS CR, Prague.
Genova/Pavia/Roma Flavio Gatti, PSI, 11 February, Timing Counter: january 2004.
R&D of Calorimeter using Strip/Block Scintillator with SiPM E.P. Jacosalem, S. Iba, N. Nakajima, H. Ono, A.L. Sanchez & H. Miyata Niigata University ILC.
CLAS12 Pre-shower H. Voskanyan (YerPhI)  R&D: Test measurements  Prototype CLAS Collaboration, CLAS12 TWG Meeting February 28, 2007, JLAB.
Gamma Detector of Plastic Scintillator Oct. 2, 2009 IP-BSM Group 1.
G.Villani Aug. 071 CALICE pixel Laser testing update Laser MIP equivalent calibration update Si detector coupled to low noise CA + differentiator (no shaper)
FTOF Status Anton A. Izotov, Frascati
GSI 9Feb09 NMI3 – Integrated Infrastructure Initiative for Neutron Scattering and Muon Spectroscopy, Joint Research Activity (JRA8): MUON-S. Contract:
SiPM Workshop EU-FP7(HP3), Vienna, 16 Feb F.G. H.O. Study of scintillator detectors time resolution with different SiPM readout on T10 test beam.
Study of the MPPC for the GLD Calorimeter Readout Satoru Uozumi (Shinshu University) for the GLD Calorimeter Group Kobe Introduction Performance.
DEVELOPMENT OF THE POLYSTYRENE SCINTILLATOR TECHNOLOGY AND PARTICLE DETECTORS ON THEIR BASES VLADIMIR RYKALIN IHEP, PROTVINO INSTR-14, NOVOSIBIRSK, 24.
P. Lecoq CERN2 February ENVISION WP2 Meeting CERN Group contribution to ENVISION WP2 Paul Lecoq CERN, Geneva.
1 The Scintillation Tile Hodoscope (SciTil) ● Motivation ● Event timing/ event building/ software trigger ● Conversion detection ● Charged particle TOF.
TOF Update GlueX Collaboration Meeting Jefferson Lab May 12, 2009 Paul Eugenio & Alexander Ostrovidov Florida State University.
IFR Detector R&D status
IFR Detector R&D status
Activities at SMI/Vienna in testing the performance of SiPMs
Seoul National University On behalf of J-PARC E18 Collaboration
Activities and Results from PNPI GATCHINA
Wander baldini on behalf of the SHIP Muon Group
Commissioning of the Crystal evaluation setup
Vladimir Rykalin NRC KI IHEP Baksan - 50, Nalchik, June, 2017.
Scintillation Detectors
The Beam Test at Fermilab:
Prototype activities update
大強度
CLAS12 Forward Detector Element PCAL
SCINTILLATING DIGITAL HADRON CALORIMETER
The E835 SCI-FI Detector In Ferrara It has been built a scintillating fibers detector for the tracking system of exp. E835 (Charmonium spectroscopy) a.
4 channels, each as input or output
Timing Counter Sept CSN I, Assisi 2004 Giorgio Cecchet.
2 channels, each as input or output
Scintillator HCal R&D Update
Measured baseline fluctuations and dark counts with Burle-Photonis MCP-PMT's Jean-Francois Genat and Edward May Dec 2009 –Jan 2010.
Deng Ziyan Jan 10-12, 2006 BESIII Collaboration Meeting
MEG実験アップグレードに向けたSiPMを用いた ポジトロン時間測定器の研究開発
Efficiency Study of Prototype Scintillator for INGRID
Particle identification —TOF detector of BESIII
Forward TOF Anton A. Izotov, PNPI.
CLAS 12 Status of the test bench Calculation of the time resolution
Some characteristics of the extruded scintillator.
Presentation transcript:

1/10/2016Collaboration Meeting Feb 2011 Start Counter 1 Start Counter Update W. U. Boeglin L. Guo E. Pooser Measurements on EJ-212 scintillator bars Attenuation Lengths Propagation Time Comparison of PM and SiPM Detectors Time Resolution measurements

EJ-212 Scintillator Bar 1/10/2016Collaboration Meeting Feb 2011 Start Counter 2 Dimensions: 12,15 mm x 600 mm: thicknesses: 2, 3 mm With and without a VM2000 backing, mounted in dark box PMT and SiPM touches end of scintillator Used 90 Sr source Light output measurements: Scintillator Trigger Detector PM or SiPM

1/10/2016Collaboration Meeting Feb 2011 Start Counter 3 Attenuation Length 3 x 15 x 600 (mm) Fit Range (20 – 50 cm) No VM2000 BackingVM2000 Backing

Attenuation Lengths Summary 1/10/2016Collaboration Meeting Feb 2011 Start Counter 4 The attenuation lengths are much improved as compared to the EJ-200 bars Due to better manufacturing and diamond-milled edges On average, with VM2000 placed at the end of the scintillator, we see a factor 1.6 better attenuation length Scintillator Dim.No VM BackingVM Backing 2 x 15 x 600 (mm)81.2 cm135 cm 3 x 12 x 600 (mm)50.8 cm76.2 cm 3 x 15 x 600 (mm)69.6 cm125 cm

O’Scope Images (SiPM): Source Close to Detector 1/10/2016Collaboration Meeting Feb 2011 Start Counter 5 No VM2000 BackingVM2000 Backing VM Backing has factor 1.2 Larger signal size Poor fall time for SiPM ~ 17 – 20 (ns) FMPMT Trigger FMPMT Trigger SiPM

O’Scope Images (SiPM): Source Far From Detector 1/10/2016Collaboration Meeting Feb 2011 Start Counter 6 No VM2000 BackingVM2000 Backing VM Backing has factor 1.6 Larger signal size No VM signal size when close is factor 1.9 larger than when far away VM signal size when close is factor 1.4 larger than when far away FMPMT Trigger SiPM FMPMT Trigger

Propagation Time of Light in the scintillator bars 1/10/2016Collaboration Meeting Feb 2011 Start Counter 7

Comparison of Detector Time Resolutions (PM vs. SiPM) 3 x 15 x 600 (mm) 1/10/2016Collaboration Meeting Feb 2011 Start Counter 8 SiPM PMT No VM2000 BackingVM2000 Backing

Comparison of Detector Time Resolutions (PM vs. SiPM) Summary 1/10/2016Collaboration Meeting Feb 2011 Start Counter 9 Time Resolution gets worse in the 25 – 45 (cm) range when measured with the PM and bars have VM2000 backing  This is not the case when measured with the SiPM σ(SiPM) ≈ 2.3 * σ(PMT) Scint. Dim. (mm) SiPM (No VM) (ps) PM (No VM) (ps) SiPM (VM) (ps) PM (VM) (ps) 2 x 15 x < σ < < σ < < σ < < σ < x 12 x < σ < < σ < < σ < < σ < x 15 x < σ < < σ < < σ < < σ < 320

Time Resolution Comparison (No VM2000 Backing) 1/10/2016Collaboration Meeting Feb 2011 Start Counter 10 PMTSiPM 3x15 2x153x12 & 2x15 3x12 3x15

1/10/2016Collaboration Meeting Feb 2011 Start Counter 11 Time Resolution Comparison (With VM2000 Backing) PMT SiPM 3x15 2x15 3x12 & 2x15 3x15 3x12

Time Resolution Comparison Summary 1/10/2016Collaboration Meeting Feb 2011 Start Counter 12 The 2 x 15 (mm) bar has the worst time resolution with the SiPM The 3 x 12 (mm) and the 3 x 15 (mm) bars have approximately the same time resolutions with SiPM The 3 x 15 (mm) has the best time resolution

Summary 1/10/2016Collaboration Meeting Feb 2011 Start Counter 13 Further measurements need to be made concerning the SiPM Cooling of SiPM so as to reduce noise and improve timing measurements Full wrapping of the 3mm bars with VM2000 (or similar highly reflective material) Begin testing of scintillator coupled to light guide

Attenuation Length 2 x 15 x 600 (mm) Fit Range (20 – 50 cm) 1/10/2016Collaboration Meeting Feb 2011 Start Counter 14 No VM2000 BackingVM2000 Backing

1/10/2016Collaboration Meeting Feb 2011 Start Counter 15 Attenuation Length 3 x 12 x 600 (mm) Fit Range (20 – 50 cm) No VM2000 BackingVM2000 Backing

Comparison of Detector Time Resolutions (PM vs. SiPM) 2 x 15 x 600 (mm) 1/10/2016Collaboration Meeting Feb 2011 Start Counter 16

Comparison of Detector Time Resolutions (PM vs. SiPM) 3 x 12 x 600 (mm) 1/10/2016Collaboration Meeting Feb 2011 Start Counter 17