Type IV Cryomodule (T4CM) Vibration Analysis Update January 23, 2007 Mike McGee 3 rd T4CM Meeting at INFN-Milan
Courtesy of DESY DESY Cryomodule Measurements Mike McGee, 23 Jan rd T4CM Meeting at INFN-Milan
Courtesy of DESY ANSYS Model Validation Mike McGee, 23 Jan 2007 Applying Harmonic Study (involving measured frequencies & displacements) using CMS Method Using DESY measurement (as input) and comparing ANSYS output to corresponding (transfer function) DESY measurement 3 rd T4CM Meeting at INFN-Milan
Courtesy of DESY DESY Measurement Transfer Functions Mike McGee, 23 Jan ) Vacuum Vessel Top vs HeGRP 1) Ground vs Vacuum Vessel Top 3) HeGRP vs Quad 3 rd T4CM Meeting at INFN-Milan
Courtesy of DESY TTF Single ANSYS Model Applied Mike McGee, 23 Jan rd T4CM Meeting at INFN-Milan Example: Transfer function between HeGRP and Quad Consider the DESY vertical measurement, by applying sine wave input with displacement (amplitude) at specific frequencies.
Courtesy of DESY Vertical Excitation & Response Mike McGee, 23 Jan rd T4CM Meeting at INFN-Milan Correlation Area Two signals can be correlated:
Courtesy of DESY ANSYS TTF Model Results Mike McGee, 23 Jan rd T4CM Meeting at INFN-Milan ANSYS model predicts the DESY measurement to within ~ 30% in most cases 24% 28% 34% Resonant frequencies found using modal analysis
Courtesy of DESY TTF Modal Shapes Mike McGee, 23 Jan rd T4CM Meeting at INFN-Milan
Courtesy of DESY Transverse Excitation & Response Mike McGee, 23 Jan 2007 Correlation Area 3 rd T4CM Meeting at INFN-Milan
Courtesy of DESY Modal Shapes of Resonant Frequencies Mike McGee, 23 Jan rd T4CM Meeting at INFN-Milan A very weak correlation transversely
Courtesy of DESY Model Improvement and Further Measurement Mike McGee, 23 Jan rd T4CM Meeting at INFN-Milan Slight damping initially considered in ANSYS model Promising prediction of actual DESY vibration measurements have been realized, however continued measurements are needed Ringing measurements at DESY are needed to define system damping estimate
Modal Analysis (fixed-fixed) Summary Mike McGee, 23 Jan rd T4CM Meeting at INFN-Milan Lowest (fundamental) frequency vertical with longitudinal mode shape is not present. T4CMTTF f (Hz)Modal Shapef (Hz)Modal Shape Vertical w/ Longitudinal 12.7Transverse Pendulum12.4Transverse Pendulum 16.3Transverse S-shape15.7Transverse S-shape 17.9Transverse S-shape19.2Transverse S-shape
Courtesy of DESY 3-in-Series w/ ends: Fixed-Fixed Mike McGee, 23 Jan 2007 TTF ModelT4CM Model 12.7 Hz12.4 Hz 3 rd T4CM Meeting at INFN-Milan
Courtesy of DESY Mike McGee, 23 Jan 2007 TTF ModelT4CM Model 3-in-Series w/ ends: Fixed-Fixed 15.7 Hz16.3 Hz 3 rd T4CM Meeting at INFN-Milan
Courtesy of DESY Mike McGee, 23 Jan 2007 TTF ModelT4CM Model 3-in-Series w/ ends: Fixed-Fixed 19.2 Hz17.9 Hz 3 rd T4CM Meeting at INFN-Milan
Courtesy of DESY Fermilab Website Mike McGee, 23 Jan rd T4CM Meeting at INFN-Milan ANSYS Results given in picture and movie files ANSYS Input Code files Measurements
Future Work Perform more specific vibration measurement on TTF cryomodule at DESY Begin Sensitivity Studies using T4CM model Mike McGee, 23 Jan rd T4CM Meeting at INFN-Milan