Lecture 3-Building a Detector (cont’d) George K. Parks Space Sciences Laboratory UC Berkeley, Berkeley, CA.

Slides:



Advertisements
Similar presentations
Topic 8. Gamma Camera (II)
Advertisements

1 Continuous Scintillator Slab with Microchannel Plate PMT for PET Heejong Kim 1, Chien-Min Kao 1, Chin-Tu Chen 1, Jean-Francois Genat 2, Fukun Tang 2,
GM COUNTER.
X-Ray Astronomy Lab X-rays Why look for X-rays? –High temperatures –Atomic lines –Non-thermal processes X-ray detectors X-ray telescopes The Lab.
Ion Beam Analysis techniques:
Dispersive property of a G-M tube HV - + In the proportional region a G-M tube has dispersive properties tube voltage.
Chapter 8 Planar Scintigaraphy
Basic Principles of X-ray Source Detection Or Who Stole All Our Photons?.....
Scintillators. When radiation interacts with certain types of materials, it produces flashes of light (scintillation) Materials that respond this way.
Photomultipliers. Measuring Light Radiant Measurement Flux (W) Energy (J) Irradiance (W/m 2 ) Emittance (W/m 2 ) Intensity (W/sr) Radiance (W/sr m 2 )
Time-of-Flight at CDF Matthew Jones August 19, 2004.
Emre Ertuğrul Emin Şahin Seçkin Gökçe KMU 396 Material Science and Technology.
Semiconductor Light Detectors ISAT 300 Foundations of Instrumentation and Measurement D. J. Lawrence Spring 1999.
Principles of Radiation Detection
Lesson 17 Detectors. Introduction When radiation interacts with matter, result is the production of energetic electrons. (Neutrons lead to secondary processes.
1 Detectors RIT Course Number Lecture Single Element Detectors.
Detectors. Measuring Ions  A beam of charged particles will ionize gas. Particle energy E Chamber area A  An applied field will cause ions and electrons.
Characterization of Detectors NEP= noise equivalent power = noise current (A/  Hz)/Radiant sensitivity (A/W) D = detectivity =  area/NEP IR cut-off maximum.
Transducers Converts one type of energy into another. Light  Electrical (current, voltage, etc.) What characteristics should we look for in a transducer?
High Energy Detection. High Energy Spectrum High energy EM radiation:  (nm)E (eV) Soft x-rays X-rays K Soft gamma rays M Hard gamma.
Radiation Safety level 5 Frits Pleiter 02/07/2015radiation safety - level 51.
Main detector types Scintillation Detector Spectrum.
Lecture 2-Building a Detector George K. Parks Space Sciences Laboratory UC Berkeley, Berkeley, CA.
Photoelectron Spectroscopy Lecture 7 – instrumental details –Photon sources –Experimental resolution and sensitivity –Electron kinetic energy and resolution.
Instruments for Radiation Detection and Measurement
Photomultiplier Tube. What is it? Extremely sensitive detector of light in the ultraviolet, visible and near infrared Multiplies the signal produced by.
Gamma Spectroscopy HPT TVAN Technical Training
Radiation Detection and Measurement II IRAD 2731.
Scintillation Detectors
Lecture 11  Production of Positron Emitters, Continued  The Positron Tomograph.
Ion Collectors and Detectors
Instruments for Radiation Detection and Measurement Lab # 3 (1)
Materials Considerations in Photoemission Detectors S W McKnight C A DiMarzio.
1 Light Collection  Once light is produced in a scintillator it must collected, transported, and coupled to some device that can convert it into an electrical.
Photon detection Visible or near-visible wavelengths
References Hans Kuzmany : Solid State Spectroscopy (Springer) Chap 5 S.M. Sze: Physics of semiconductor devices (Wiley) Chap 13 PHOTODETECTORS Detection.
1 Performance of multi-anode PMT employing an ultra bi-alkali photo-cathode and rugged dynodes Takahiro Toizumi Tokyo Institute of Technology S. Inagwa.
Peter Križan, Ljubljana March 17, 2006 Super B Workshop, Frascati Peter Križan University of Ljubljana and J. Stefan Institute Aerogel RICH and TOP: summary.
Chapter 12 Atomic X-Ray Spectroscopy
EDS Energy Dispersive Spectroscopy
SCINTILLATION COUNTER. PRINCIPLE When light radiations strike fluorescent material it produces flashes of light called scintillations. These are detected.
Scanning Electron Microscope (SEM)
Photodetection EDIT EDIT 2011 N. Dinu, T. Gys, C. Joram, S. Korpar, Y. Musienko, V. Puill, D. Renker 1 Micro Channel plate PMT (MCP-PMT) Similar to ordinary.
Detection plan for STEIN, telescope of the space mission CINEMA 1 August 3td 2011-Diana Renaud.
References Hans Kuzmany : Solid State Spectroscopy (Springer) Chap 5 S.M. Sze Physics of semiconductor devices (Wiley) Chap 13 PHOTODETECTORS.
Final Presentation By Matthew Lewis 17 th March 2006 “To Determine the Accuracy that GOES True Numbers can Reproduce the Full X-ray Spectrum of the Sun”
Experimental set-up for on the bench tests Abstract Modeling of processes in the MCP PMT Timing and Cross-Talk Properties of BURLE/Photonis Multi-Channel.
Nuclear Medicine: Planar Imaging and the Gamma Camera Katrina Cockburn Nuclear Medicine Physicist.
Medical Image Analysis Interaction of Electromagnetic Radiation with Matter in Medical Imaging Figures come from the textbook: Medical Image Analysis,
SEM- Schematic Overview. Electron Detection Tungsten Filament Electron Source.
Techniques for Nuclear and Particle Physics Experiments By W.R. Leo Chapter Eight:
Chapter V Radiation Detectors.
Lecture 12  Last Week Summary  Sources of Image Degradation  Quality Control  The Sinogram  Introduction to Image Processing.
3/2003 Rev 1 II.3.5 – slide 1 of 23 IAEA Post Graduate Educational Course Radiation Protection and Safe Use of Radiation Sources Session II.3.5 Part IIQuantities.
Radiation detectors Ion chamber 2. Geiger Muller counter (GM).
MoNA detector physics How to detect neutrons. Thomas Baumann NSCL.
SWIA Front End Peer Review
Scintillation Detectors
Origin of Quantum Theory
Chapter 3. Components of Optical Instruments
PAN-2013: Radiation detectors
On behalf of the GECAM group
Gamma-Gamma Correlations in Na-22
Scintillation Counter
Lecture 1—Basic Principles of Scintillation
HE instrument and in-orbit performance
Photomultiplier (PMT) Tubes
X-ray detectors single photon detectors
PHYS 3446 – Lecture #16 Monday ,April 2, 2012 Dr. Brandt
PHYS 3446 – Lecture #17 Wednesday ,April 4, 2012 Dr. Brandt
Presentation transcript:

Lecture 3-Building a Detector (cont’d) George K. Parks Space Sciences Laboratory UC Berkeley, Berkeley, CA

Summary of Lecture 1 and 2

Photocathode

Photoelectron Emission Process

Photocathodes vs Spectral Emission of scintillators

Transmittance of window material of PMT

Dark Current <1 photo-e - 2 photo-e - 1 photo-e -

Dark current (cont’d)

Temperature Characteristics of Dark Current

Linearity Non-Linearity starts when anode current exceeds A.

Spatial Uniformity

Magnetic field effects Magnetic field deflects electrons in PMT. To reduce magnetic effect, shield PMT with  -metal. Why peak not at 0? Unit (magnetic field mT)

Incident photons and PMT output

Pulse Height Distribution (Energy Spectra) ~5.9 keV ~ 662 keV

Energy Resolution E dN/dE

Energy Resolution for Scintillation Detector

Quantum Efficiency (Bi-Alkali)

Summary (Scintillation + PMT)

High energy charged particles

Comparison NaI(Tl) vs Ge(Li) Pulse height Spectrum of Ag

Band Model

Solid state detectors

Solid State Detectors (cont’d)

Operation of semiconductor detector

Calibration of Detectors The amount of energy required to produce electron-ion pair in Si detectors is 3.5 eV. Hence, if we know how many electron-ion pairs are produced, we obtain the energy of the particle. Detectors are biased so electrons and ions are collected separately at anode and cathode. A detector is calibrated with known beam energies. The size of the pulse measured is directly related to the original particle energy. The distribution of pulse-height vs energy gives the differential energy spectrum

Detector resolution vs energy

Energy Resolution of Semiconductor Detectors MaterialZ  BandgapIon E (e-h)Energy Resolution Si (77 o K) eV3.61 eV keV (77K) keV Ge(77K) keV (0.35%) keV keV CdTe(300K) ~ keV keV (2.8%) HgI 2 (300K) keV (2.6%) 5.96 keV

Early design of quadrispheric analyzer ESA designs include cylindrical, spherical and quadrispherical shaped plates. ESAs are basically capacitors with voltage applied across the plates. + and – charges are deflected in opposite directions. Advantages of curved plates include reducing HV (analyzer constant) and UV rejection Once E/q selected, particle is recorded by an electron detector CEM).

Low Energy Proton and Electron Differential Energy Analyzer (LEPEDEA)

Degradation and Noise

MicroChannel Plates (MCP)

Operation of MCP (schematic)

Degradation of MCPs

BURLE, Long-life MCP test

Background counts in detectors

A scintillation detector Incident photon h stops in scintillator, generates scintillation photons. Scintillation photons propagate to photocathode and produce photoelectrons Photoelectrons multiplied at subsequent dynodes and collected at the anode Measured current from Anode to ground directly proportional to photoelectron flux generated at photocathode

Schematic diagram of a working Balloon-borne X-ray Detector

Summary Discussed briefly how to build detectors to measure Photons and Charged Particles Basic principle is relatively simple. There are only a few components for particle and photon detection. However, there are many details that can affect the measurements. Innovative ways to use these components to design one of a kind instrument to enhance science goals. SSTs can be used to detect Energetic Neutral Atoms (ENA). ESA + TOF + MCP can be used to determine M/q. Pin hole, coded aperture, modulation collimators yield 2D information. Stereoscopic view (2 SC) can yield 3D information.

Optimization

Particle measurements (Reminder)

Issues to consider in instrument design - Energy resolutions: More science with high energy resolution. - Detector efficiency: Maintain as high as possible. - Pulse height defect: Difference between true and apparent energy - Channeling: effect of crystal orientation - Dead layer: low energy threshold - Radiation damage: degrades energy resolution and counting efficiency - UV rejection (MCPs): spurious counts contamination - Pulse pile up: Loss of true counts. - Leakage current: - Detector Noise: - Changes with bias voltage: Effect if bias voltage is low - Temperature sensitivity: loss of detector performance - Cleanliness: produce noise in system. - Micro-acoustic sensitivity:

The End

Temperature Characteristics of Photocathodes Temperature change is large near the long wavelength cutoff. Unit is in % / o C

Life Characteristics Damage from the last dynode due to heavy electron bombardment

Summary of PMT Quantum efficiency Collection efficiency Window material Photocathode material Gain Dark Current Spatial Uniformity Temperature Charactristics Magnetic field effect Life characteristics Electronic noise Dynamic range Linearity Time response Decide photon counting or measure current How to select photocathode material (eg, S-13, etc..) Less than single photon spectrum; single and two photon spectra. Important for single photon counting. Determine HV for operation (900, 1100, 1200, etc..) Determine quantum efficiency of photocathode

Channel Electron Multipliers ESAs usually combine with CEMs or MCPs. Straight tubes produce large ion feedback current. Ion feedback can be reduced by curving the channeltrons Evans (1965)