Ocular Response Analyzer Waveform Analysis in the Ectatic Corneas: Correlation of the New Corneal Biomechanics Parameters and Severity of Keratoconus Kristin.

Slides:



Advertisements
Similar presentations
Asymmetric Keratoconus Study: Novel Tomographic Parameters to enhance sensitivity to detect abnormalities in eyes with normal axial curvature maps from.
Advertisements

Biomechanical Properties of the Cornea in Normal- Tension Glaucoma Authors: Leonidas Traipe Ines Cayuqueo Fabiola Cerfogli Claudia Goya Allister Gibbons.
Statement of Commercial Interest: RDJ: None DRH: honoraria from Reichert for educational lectures on corneal biomechanics Multiple studies have shown ex.
Corneal biomechanical properties and their association with intraocular pressure (IOP) and central corneal thickness (CCT) Dimitrios P. Bessinis 1, Pandelis.
Incidence of Blepharitis in Patients Undergoing Phacoemulsification Jodi Luchs, MD Carlos Buznego, MD William Trattler, MD The authors of this poster have.
Comparison between Dynamic contour tonometry, Goldmann applanation tonometry and Ehlers-corrected Goldmann applanation tonometry in eyes after Laser In.
Comparison of corneal astigmatism measured with 3 devices Mariko Shirayama, M.D, Li Wang, M.D, PhD, Mitchell P. Weikert, M.D, Douglas D. Koch, M.D. Cullen.
Correlation Of Corneal Contour With Higher Order Ocular Aberrations (HOA) In Indian Eyes Manish Chaudhary MS Amit Gupta MS Advanced Eye Centre, Postgraduate.
Long-Term Follow-Up of Toric Intraocular Lens Implantation for Keratoconus Alejandro Navas, MD, MSc, Martha Jaimes, MD, Jesús Cabral, MD, Arturo Ramirez.
Influence of Keratoplasty Technique on Corneal Biomechanical Properties in Keratoconus Hans R Vellara, Noor Q Ali, Charles NJ McGhee, Dipika V Patel Department.
Biomechanical Properties of Healthy and Keratoconic Corneas Evaluated Using an Ultra-High-Speed Scheimpflug Camera nznec Department of Ophthalmology University.
Corneal Biomechanics, Pachymetric Progression Profile and Corneal Volume: Indices for Detecting Ectasia and Screening Refractive Candidates Renato Ambrósio.
E-Poster ALC Canedo, MD 2010 Rio de Janeiro CornealTomograhyandBiomechanics Study Group Study Group Comparative Study of Corneal Biomechanical Properties.
Blepharitis and Dry eyes in Aromatase Inhibitor Users Kiran Turaka, M.D. Kristin M. Hammersmith, M.D. Jennifer M. Nottage, M.D. Christopher J. Rapuano,
Outcome of refractive surgery in keratoconus suspect eyes
Contrast-Enhanced Corneal Wound Imaging by Optical Coherence Tomography Preeya K. Gupta, MD Justis P. Ehlers, MD Terry Kim, MD Duke Eye Center, Durham,
Elvin H. Yildiz, Elisabeth J. Cohen, Ajoy S. Virdi, Kristin M. Hammersmith, Peter R Laibson, and Christopher J. Rapuano Cornea Service, Wills Eye Institute,
Variations in corneal biomechanical parameters and central corneal thickness during the menstrual cycle. David Zadok, MD, Yakov Goldich, MD, Yaniv Barkana,
Blepharitis and Dry eyes in Aromatase Inhibitor Users
A Prospective Trial Comparing Scleral Pneumotonometry to Goldmann Applanation Tonometry Sara Duke, MD, Usiwoma Abugo, BS, Shuchi Patel MD Loyola University.
Effects of IOP Lowering Agents on Myopic Regression after Refractive Surgery Lim, Taehyung M.D., Hong, So Jin M.D., Cho, Beom Jin M.D., Ph.D. Chung Kyu-Hyung.
TRANSPALPEBRAL TONOMETER APPLICATION DURING INTRAOCULAR PRESSURE EVALUATION IN THE PATIENTS WITH REFRACTION ANOMALY BEFORE AND AFTER KERATOPHOTOREFRACTIVE.
Correlation Between Corneal Biomechanics and Wavefront Aberrations Alejandro Rivera, MD Private Practice Mexico City The author has no financial interest.
Blepharitis and Dry eyes in Aromatase Inhibitor Users Kiran Turaka, M.D. Kristin M. Hammersmith, M.D. Jennifer M. Nottage, M.D. Christopher J. Rapuano,
Occurrence of Retinal Detachment After Cataract Surgery Peter Jeppesen MD, PhD and Thomas K. Olsen, MD, DMSc Department of Ophthalmology Århus University.
Relationship between Pachymetry and Ocular Response Analyzer Measurements Alejandro Rivera MD.
Oliver Findl 1,2, Nino Hirnschall 1,2, Alja Crnej 2 1 VIROS Vienna Institute for Research in Ocular Surgery, Hanusch Hospital, Vienna, Austria 2 Moorfields.
World Cornea Congress VI Michael K. Smolek & Panagiotis Kompotiatis Department of Ophthalmology LSU Eye Center of Excellence, New Orleans The authors.
ROTATING SCHEIMPFLUG TOPOGRAPHIC PARAMETERS IMPORTANT IN DISTINGUISHING NORMAL FROM KERATOCONIC CORNEAL MORPHOLOGICAL FEATURES Clayton Falknor, MD, Orkun.
Evaluation of Long-Term Keratoconus Progression Using Scanning-Slit Topography First Author: David Galarreta, MD Contributing Authors: Nestor Cortes, MD,
Khairidzan Mohd Kamal MD
Variations in corneal biomechanical parameters and central corneal thickness during the menstrual cycle. Yakov Goldich, MD, David Zadok MD, Yaniv Barkana,
Preetam Kumar, Shreyas Ramamurthy, Pravin K. Vaddavalli
Severity of Herpes Zoster Ophthalmicus: Onset at Younger Than 60 Years Versus 60 Years or Older Neelofar Ghaznawi MD, Ajoy Virdi MD, Amir Dayan, Christopher.
ELEVATION AND PACHYMETRY VALUES IN NORMAL CORNEAS OBTAINED BY GALILEI ASCRS Boston April Vision Sana Research Group Vision Sana Research Group.
Pattern of Pain and Analgesic use after Collagen Crosslinking for Progressive Keratoconus Ramon Coral Ghanem, MD, PhD. Renan Ferreira Oliveira, MD. Vinicius.
Comparison of Central Corneal Thickness and Peripheral Corneal Thickness using Sheimpflug system, Optical Coherence Tomography and Ultrasound Pachymetry.
Mean Keratometry Measurement Post Penetrating Keratoplasty Jacky Yeung MSc MD, Stephanie Baxter MD FRCS(C) Department of Ophthalmology, Hotel Dieu Hospital,
1 Non-contact Specular Microscopy for Evaluation of Corneal Endothelium in Early Fuchs’ Endothelial Corneal Dystrophy Jianyan Huang 1, MD, PhD; Tudor Tepelus.
Incidence of Cataract Extraction After Corneal Transplantation in Young and Middle-Aged Patients Revathi Naadimuthu, MD Gerald W. Zaidman, MD Brandon Mirochnik,
Evaluation of Efficacy and Safety of Intracorneal Ring Segment (Intacs SK) in Keratoconus Abdulrahman Al-Muammar, MD, FRCSC I have no financial interest.
E-Poster Rio de Janeiro Corneal Tomography and Biomechanics Study Group The study of the correlations of the optic disc size with Cornal Geometry and Biomechanical.
Evaluation of Corneal Tomography in Primary Pterygium Sahil Goel, MD (Presenting Author), Murugesan Vanathi MD *The authors have no financial interests.
Ahmed El-Massry, M.D. Professor of Ophthalmology - Faculty of Medicine University of Alexandria Egypt long-Term Results of Corneal Biomechanical Changes.
Visual and IOP Outcomes after PRK in Pigment Dispersion Syndrome [Poster Number: P190] Kraig S. Bower, Denise A. Sediq, Charles D. Coe, Keith Wroblewski,
Comparison of Central Corneal Thickness, Anterior Chamber Depth, and Central Corneal Power Measurements between Two Scheimpflug Imaging Systems Yuichi.
Authors: Leonidas Traipe Magdalena Rios Edison Villagra Allister Gibbons Marcelo Hidalgo Claudia Goya Authors have no financial interest.
Corneal Hysteresis Role of Corneal Biomechanics in Refractive Surgery Luis Rodriguez, MD Anny Villegas, MD Clínica de Córnea Centro Médico Docente La Trinidad.
Mathematical Description of the Ocular Response Analyzer Applanation Signal in Post-LASIK Ectasia Mujtaba A. Qazi, M.D. 1,2 David A. Luce, Ph.D. 3 Jay.
Poster produced by Faculty & Curriculum Support (FACS), Georgetown University School of Medicine LASIK Combined with Corneal Cross-linking in Eyes with.
BIOMECHANICAL PROPERTIES OF THE CORNEA IN PSEUDO-EXFOLIATION SYNDROME Aylin Kılıç Ertan, MD Özgür Bilen, MD Kudret Eye Hospital, Ankara, Turkey There is.
Corneal Biomechanical Changes Following Surface Keratorefractive Surgery Teeravee Hongyok, MD, Christopher J. Rapuano, MD, Ajoy Virdi, MBBS, Elisabeth.
CORNEAL ENDOTHELIAL PROFILE AFTER FERRARA RING IMPLANTATION Leonardo Torquetti, MD, PhD Paulo Ferrara, MD, PhD Paulo Ferrara Eye Clinic Belo Horizonte.
1. Ankara University School of Medicine, Department of Ophthalmology
Instituto para Preservación de la Visión
CRF variations ( initial CRF – final CRF )
Characteristics of Primary Angle-Closure Glaucoma Patients with Normal Intraocular Pressure at the First Visit Won Hyuk Oh1, Bum Gi Kim1, Joo Hwa Lee2.
Corneal High-Order Aberrations in Keratoconus
JP Deva COMPARISON OF PRELIMINARY RESULTS FROM CROSSLINKED KERATOCONUS AND NORMAL CORNEAE USING ORA JP Deva
Influence of UVA-Riboflavin corneal collagen cross-linking on biomechanical properties of keratoconic eyes David Zadok MD, Yakov Goldich MD, Yaniv Barkana.
Jodi Luchs, MD Carlos Buznego, MD William Trattler, MD
Effects of corneal collagen crosslinking on Ocular Response Analyzer waveform-derived variables in keratoconus and post-LASIK ectasia Spatial treatment.
Associate Professor of Ophthalmology
Federal University of São Paulo, Brazil
Bruno Valbon, Ana Caiado, Ricardo Louzada, Renato Ambrosio Jr *.
Yakov Goldich MD,Yaniv Barkana MD, Isaac Avni MD, David Zadok MD,
Corneal Biomechanics after Penetrating Keratoplasty and Descemet’s Stripping Automated Endothelial Keratoplasty Samer Hamada MRCOphth FRCS(Ed)1,3 Louai.
Hyuck Jin Choi, Joo Youn Oh, Won Ryang Wee, Mee Kum Kim,
Jun Her MD. PhD, Sung-wook Kim MD, Jae-hwan Ahn MD
The Effect of Corneal Thickness on Ocular Drug Penetration
Presentation transcript:

Ocular Response Analyzer Waveform Analysis in the Ectatic Corneas: Correlation of the New Corneal Biomechanics Parameters and Severity of Keratoconus Kristin M. Hammersmith, MD, Peter R. Laibson, MD, Christopher J. Rapuano, MD Teeravee Hongyok, MD, Elisabeth J. Cohen, MD, Kristin M. Hammersmith, MD, Peter R. Laibson, MD, Christopher J. Rapuano, MD Cornea Service, Wills Eye Institute Jefferson Medical College, Thomas Jefferson University, Philadelphia, Pennsylvania, USA The authors have no financial interest in the subject matter for this poster. World Cornea Congress VI, Boston, MA, USA, April 7-9, 2010

Introduction Ocular Response Analyzer (ORA) measurements of biomechanical properties of the cornea may be another helpful tool to help detect early keratoconus and aid in disease classification. Corneal hysteresis (CH) and corneal resistance factor (CRF) are significantly lower in keratoconic eyes compared to normal eyes, but the values overlap and can not distinguish between mild keratoconus and normal, when used alone. 1, 2 The ORA signal waveforms differ from normal waveforms in many ways, such as lower amplitude of applanation peaks in ectatic corneas compared to normals. 3 1 Luce DA. J Cataract Refract Surg 2005;31(1): Kirwan C, et. al. Ophthalmologica 2008;222(5): Kerautret J, J Cataract Refract Surg 2008;34(3):510-3.

New ORA Waveform Parameters With new software (version 2.04), the ORA can mathematically describe waveform morphological characteristics including signal peak, width, slope, area under the curve, noise, and aspect ratio (height/width) of the waveform. We describe the new waveform parameters in figures below and tables in the following slide. Fig.1 * Fig.2 ** Fig.3 ***Fig.4 ***

Parameter NameDescription From 75% of Applanation Peak* From 50% of Applanation Peak** Aindex-“smoothness” or degree of "non- monotonicity" or number of breaks when peak changes direction in peak 1 Bindex-“smoothness” or degree of "non- monotonicity" or number of breaks when peak changes direction in peak 2 P1areaP1area1area under the curve of peak1 (sum of values) P2areaP2area1area under the curve of peak2 (sum of values) Aspect1Aspect11aspect ratio (height/width) of peak1 Aspect2Aspect21aspect ratio (height/width) of peak2 Uslope1Uslope11rate of increase from base to peak value of peak 1 Uslope2***Uslope21***rate of increase from base to peak value of peak 2 (downslope in real time of peak2) *** Dslope1Dslope11rate of decrease from peak to base value of peak 1 Dslope2***Dslope21***rate of decrease from peak to base value of peak 2 (upslope in real time of peak2) *** W1W11width of peak1 at "base" of peak1 region W2W21width of peak2 at "base" of peak2 region Parameter NameDescription From 75% of Applanatio n Peak* From 50% of Applanation Peak** H1H11height (from lowest to highest value) of peak1 H2H21height (from lowest to highest value) of peak2 Dive1-absolute value of monotonic decrease on downslope part of peak1 starting at the peak value Dive2***-absolute value of monotonic decrease on downslope part of peak2 starting at the peak value Path1Path11absolute value of path length around peak1 Path2Path21absolute value of path length around peak2 Mslew1-maximum single step increase in rise of peak1 (longest continuous line without a break) Mslew2***-maximum single step increase in rise of peak2 Slew1-aspect ratio of dive1 (value of dive divided by width of dive region) Slew2***-aspect ratio of dive2 (value of dive divided by width of dive region) Aplhf-high frequency "noise" in region between peaks (normalized by product of average of peak heights times width of region) Waveform Score (WS)The overall score calculated from these parameters using neural network. 15 Higher score represents a better waveform and more reliable measurement. *Data are analyzed from upper 75% of the applanation peak (Fig. 1 ) **Data are analyzed from upper 50% of the applanation peak (Fig. 2) ***The data from the second applanation were analyzed in time-reversed fashion (Fig.3,4)

Purpose To evaluate the correlation between the Ocular Response Analyzer (ORA) signal morphology using new waveform parameters and severity of disease in eyes with keratoconus and pellucid marginal degeneration.

Methods Patients diagnosed with keratoconus and pellucid marginal degeneration on the Wills Eye Institute Cornea Service from March 2007 to April 2008 were prospectively enrolled in a study of patients with glaucoma or suspect glaucoma and age-matched controls. (Cohen EJ.,Trans Am Ophthalmol Soc Dec;107: ) Eyes with previous ocular surgery or hydrops were excluded. Clinical and demographic characteristics were recorded. The corneal biomechanics were measured using the ORA with new software version Corneal curvature and central thickness were measured using Humphrey® Atlas™ 995 Topographer (Carl Zeiss Meditec, Inc., Dublin, Ireland) and AccuPach V (Accutome, Malvern, PA, USA). The correlation between ORA parameters and various disease severity indices were examined using Pearson correlation coefficients and a stepwise logistic regression model.

Results: Patient Characteristics 93 eyes of 62 patients were included in the study. Patients were Caucasian (91%) or African-American (8%). Clinical diagnosis was keratoconus in 86 eyes (92.5%) and pellucid marginal degeneration n 7 eyes (7.5%). Patient Characteristics Mean ± SDUnits Age59.6 ± 10.8years Simulated K46.7 ± 4.4D Maximal topographic K 52.9 ± 6.1D CCT493 ± 64µm

Representative signals from keratoconic (KCN) eyes without scarring The waveform morphology and biomechanical parameters were correlated with the severity of keratoconus. Mild KCN Moderate KCN Severe KCN

Correlation of Central Corneal Thickness and Waveform Parameters There were significant correlations between central corneal thickness (CCT) and a number of ORA parameters including corneal hysteresis (CH), corneal resistance factor (CRF), Aindex, Bindex, W1, Aspect 1, Aspect11, Dslope 11, P1, and P2 (p<0.05). Width of peak 1 (W1) was found to be the most significant predictor of CCT (r=0.29, p=0.006). Plot of CCT and W1 (r=0.29, p=0.006) CCT W1

Correlation of Keratometry and Waveform Parameters Maximal topographic and steep simulated keratometries were significantly correlated with CH, CRF, inward applanation pressure (P1) and waveform score (WS) (see table). Flat and average simulated Ks were not significantly correlated with any ORA parameters (all p>0.05) ORA Waveform Parameters Maximal topographic Keratometry Steep simulated Keratometry rp valuer CH CRF-0.35< P WS-0.49<

Correlation of Waveform Parameters and Other Characteristics Corneal scarring also significantly altered waveform morphology and various parameters including CH, CRF, Aindex, Bindex, Aspect1, Aspect2, Uslope1, Uslope2, H1, H2, Dive1, Dive2, Mslew2, Slew2, Uslope21, Uslope11, Dslope21, H11, H21, W1, P1 and WS (p<0.05). Location of maximal corneal ectasia did not significantly affect ORA parameters. These parameters were not significantly different between keratoconus and pellucid marginal degeneration.

Conclusion We demonstrated that the corneal biomechanical parameters such as CH, CRF and also new ORA waveform parameter such as W1, and waveform score were significantly correlated to disease severity of ectatic corneas. These parameters might be helpful indicators to aid severity classification of keratoconus and pellucid marginal degeneration and monitor progression of the disease beyond the corneal thickness and corneal topography that have been used in the past.