Study of Flap Thickness Precision with a Femtosecond Laser Shinagawa LASIK Center Tatsuya Yonekawa,MSc ; Minoru Tomita,MD.PhD ; Youhei Iida,CE ; Yuko Inada,CE.

Slides:



Advertisements
Similar presentations
Femtosecond Laser–Assisted Sutureless Anterior Lamellar Keratoplasty
Advertisements

INTRODUCTION TO iLASIK TM. It’s Time For The iLASIK ™ Procedure Using a unique combination of the most advanced technology, the iLASIK procedure is fast,
Case report The features of the femtosecond laser system flap creation process on the eye after radial keratotomy. O.F. Ziiatdinova, L. N. Safiullina Eye.
Authors have no financial interest
Steve Schallhorn MD Myth Busters in Patient Selection for LASIK Steve Schallhorn MD.
Swept Source Optical Coherence Tomography for Evaluation of Posterior Corneal Changes after Refractive Surgery Dr. Tommy Chung Yan Chan Dr. Vishal Jhanji.
Comparison between Dynamic contour tonometry, Goldmann applanation tonometry and Ehlers-corrected Goldmann applanation tonometry in eyes after Laser In.
By Kelsey Matthews. History  LASIK stands for laser in situ keratomileusis  Developed in 1991 by a greek doctor Ioannis Pallikaris  Purpose to correct.
Outcome of refractive surgery in keratoconus suspect eyes
PRK Enhancement with Mitomycin - C after LASIK - a case series
Hemi-Automated Lamellar Keratoplasty (HALK) Leonard Yuen, MD MPH MRCOphth Jodhbir Mehta, FRCS FRCOphth Li Lim, FRCOphth Donald Tan, FRCS FRCOphth SINGAPORE.
Dept. of Ophthalmology, Friedrich-Alexander University Erlangen-Nuremberg, Erlangen, Germany. Outcome after big-bubble deep anterior lamellar keratoplasty.
Comparison of Orbscan and Ultrasound Pachymetry Measurements Faik Orucov, MD, Abraham Solomon, MD,Ziv Caspi, David Landau, MD, Eyal Strassman, MD, and.
Jérôme C. VRYGHEM, M.D. Brussels Eye Doctors Brussels, Belgium
INTRA OPERATIVE ONLINE PACHY METRY –A SAFETY TOOL Dr. KUMAR J DOCTOR DOCTOR EYE INSTITUTE MUMBAIINDIA NO FINANCIAL INTEREST.
Effects of IOP Lowering Agents on Myopic Regression after Refractive Surgery Lim, Taehyung M.D., Hong, So Jin M.D., Cho, Beom Jin M.D., Ph.D. Chung Kyu-Hyung.
TRANSPALPEBRAL TONOMETER APPLICATION DURING INTRAOCULAR PRESSURE EVALUATION IN THE PATIENTS WITH REFRACTION ANOMALY BEFORE AND AFTER KERATOPHOTOREFRACTIVE.
Femto-Lasik after corneal transplantation
Hong Kong Eye Hospital Ms Frenchy Chiu Dr Victoria Wong IOL master
Comparison of Wavefront and Corneal Aberration Changes after Advanced Corneal Surface Ablation and Femtosecond Thin Flap LASIK Tahra AlMahmoud, MBBS 1,
Evaluation of Zonal Equivalent Keratometry Readings After LASIK Timmy Kovoor, MD Orkun Muftuoglu, MD V.Vinod Mootha, MD Steven Verity, MD R. Wayne Bowman,
Authors: Jaroslav Madunicky, MD Katerina Buusova Smeckova, MD, MBA Col. Assoc.Prof. Jiri Pasta, MD, PhD. Iveta Nemcova, MD, PhD. Eva Vyplasilova, MD Kristina.
Topical Cyclosporine for Postoperative use in PRK and LASIK: Preliminary Results LCDR David D. Hessert, M.D. Ophthalmology Chief Resident Naval Medical.
Custom Flap Formation: Advantages of Elliptical Flaps for LASIK Patients Custom Flap Formation: Advantages of Elliptical Flaps for LASIK Patients Henry.
Natalie Stanciu, MD Richard M. Awdeh, MD Takeshi Ide, MD Sonia Yoo, MD Bascom Palmer Eye Institute ASCRS 2009: San Francisco.
Khairidzan Mohd Kamal MD
Assessment of VisuMax Femtosecond Laser Accuracy and Precision of Flap Thickness and Centration Dan Z Reinstein MD MA(Cantab) FRCSC 1,2,3,4 Timothy J Archer,
Mohamed A Guenena, MD Helga P Sandoval, MD, MSCR, Kerry D Solomon, MD Magill research center for vision correction Storm Eye Institute Medical University.
Comparison of Central Corneal Thickness and Peripheral Corneal Thickness using Sheimpflug system, Optical Coherence Tomography and Ultrasound Pachymetry.
Refractive Accuracy of LASIK Using the IntraLase and Zyoptix BACKGROUND Wavefront guided customized lasik procedures are designed to correct both lower.
Myoung Joon Kim, MD / Sara Yoon, MD Tracy Purcell, PhD / David J Schanzlin, MD L aser In Situ Keratomileusis versus Photorefractive Keratectomy for the.
A New Technique for Precise, Predictable SBK Surgery Using the B&L Zyoptix XP Microkeratome David R. Shapiro, MD Shapiro Laser Eye Center Ventura, California.
Naomi R. Goldberg, MD PhD Kenneth J. Wolf, MD Eric J. Wolf, MD FACS The authors have no financial interest in the subject matter of this poster. Comparison.
Adriana S. Forseto1, MD Walton Nosé1,2, MD
Visual and IOP Outcomes after PRK in Pigment Dispersion Syndrome [Poster Number: P190] Kraig S. Bower, Denise A. Sediq, Charles D. Coe, Keith Wroblewski,
Effect of an irregular anterior surface on a lamellar femtosecond laser cut Robert E. Fintelmann, MD Michele M. Bloomer, MD Bennie H. Jeng, MD Authors.
Comparison of Central Corneal Thickness, Anterior Chamber Depth, and Central Corneal Power Measurements between Two Scheimpflug Imaging Systems Yuichi.
Faik Orucov*, MD, Joseph Frucht-Pery, MD, David Landau, MD, Eyal Strassman, MD, Abraham Solomon, MD Clinical outcome of thin corneas after laser in situ.
Augenabteilung am St. Franziskus Hospital Münster Lasik Flap Thickness using the Zyoptix XP Microkeratome Susanne Stottmeister, Suphi Taneri * * Travel.
Long Term Results of Femtosecond Laser-Assisted Sutureless Anterior Lamellar Keratoplasty “FALK” Mohamed Abou Shousha, MD, Sonia H. Yoo, MD, William Feuer,
COMPARISON AND EVALUATION OF OCULAR BIOMETRY USING A NEW NON-CONTACT OPTICAL LOW COHERENCE REFLECTOMETER David Goldblum 1,3, Kaspar Rohrer 1, Rudolf Waelti.
Simulated Experiments on IOL Power Calculation Using Anterior Segment OCT Dong Hyun Jo, M.D., 1,2 Mee Kum Kim, M.D., 1,2 Won Ryang Wee, M.D. 1,2 1 Department.
Case report No financial interest The features of the Femtosecond Laser system flap creation process on the eye after radial keratotomy O.F. Ziiatdinova,
Corneal Biomechanical Changes Following Surface Keratorefractive Surgery Teeravee Hongyok, MD, Christopher J. Rapuano, MD, Ajoy Virdi, MBBS, Elisabeth.
Comparison of Varying the Ablation Diameter in LASIK and LASEK:The Effects on Anterior Chamber Depth (ACD) and Regression at 3 months Minoru Tomita, MD,
Corneal Endothelial Cell Changes After LASIK Using a Femtosecond Laser Minoru Tomita, MD, PhD, Masaru Aoyama, MD, Minoru Tomita, MD, PhD, Masaru Aoyama,
Changes of Axial Length After LASIK Surgery: A 3-Year Follow-Up Study
Mumbaieyecarecornealasik.com.
Femtosecond Laser Good, Bad & Ugly
William W. Culbertson, MD Bascom Palmer Eye Institute, Miami, FL
Measurement of 60 kHz Femtosecond LASIK Flaps with Anterior Segment Ocular Coherence Tomography & Ultrasonic Subtraction Pachymetry Edward E. Manche, MD.
Maria A Henriquez MD Luis Izquierdo MD Jackeline Parraga MD
DSAEK using corneas with previous LASIK
Dr. James A. Davies, M.D. is the Medical Director of Davies Eye Center and The Surgical Eye Care Center. As a surgical pioneer, Dr. Davies has been recognized.
Comparison of Single-Scheimpflug and Dual-Scheimpflug Pachymetry Measurements and Effect of Decentration World Cornea Congress VI April 7-9, 2010.
Crosslinking with Simultaneous Implantation of Intrastromal Corneal Ring Segments in Keratoconus: Safe and Efficacy Author: Luiz Antonio de Brito Martins.
Model of Hyperopic Shift After DSEK
Clinical outcome of thin corneas after laser in situ keratomileusis
University of Ulsan College of Medicine
Asociación Para Evitar La Ceguera En México
Three-Year Follow-up after LASIK in Eye with Extremely Thin Corneal Bed Hidemasa Torii, MD, Kazuno Negishi, MD, Murat Dogru, MD, Takefumi Yamaguchi, MD,
Optical Coherence Pachymetry with the ESIRIS DOCTOR EYE INSTITUTE
Mohamed Abdelrahman Awadalla,FRCS Magrabi Eye Hospital Cairo - Egypt
Long term PCO preventive effect of capsular bending ring
The authors have no financial interest
The Effect of Corneal Thickness on Ocular Drug Penetration
Comparing Endothelial Cell Density after Sub-Bowman’s Keratomileusis and Photorefractive Keratectomy for the Treatment of Myopia Ryan T. Smith, MD Daniel.
Retrospective Study of 4 Cases That Developed Ectasia After LASIK (Intralase flaps) with Low-Risk Based on Ectasia Risk Score System Allan Luz, MD Colin.
Refractive Surgery Outcomes: Corneas Thinner than 500 microns Vs
Effectiveness of Accuracy of Eye Trackers in Laser Refractive Surgery
Presentation transcript:

Study of Flap Thickness Precision with a Femtosecond Laser Shinagawa LASIK Center Tatsuya Yonekawa,MSc ; Minoru Tomita,MD.PhD ; Youhei Iida,CE ; Yuko Inada,CE ; Sachiko Imada,BSc ; Mika Morizumi,BSc Masashi Shimizu,BSc ; Hiroyuki Nomura,CE ; Naoyuki Kurokawa, MD Yutaka Makisima, MD ; Nobuo Nakamura, MD. PhD ; Tadayuki Nakamura, MD Masaru Aoyama, MD ; Tadahiko Tsuru, MD.PhD

Introduction Recently, many authors have reported advantages of corneal flap preparation using Femtosecond laser(FS 60™;AMO). FS 60™ requires Patient Interface (PI) for connection with the patient. FS 60™ can make corneal flaps with precise thickness (Planar- Flap) by the applanation of the cornea with PI. applanation Planar-Flap

Purpose During operation, we often encounter occasions that flap thickness differs from the input value. May it have an individual error in PI? Isn't it a possibility of the effect of PI that the flap thickness is different from the input value? Therefore, we measured PI by the measuring device. Error value corrected in correlation to PI regulation. Corrected value is inputted (manually correction) into FS60 ™. (Shinagawa method) We studied whether we could make a flap without errors.

Subjects and methods The subjects were 240 eyes of 120 patients (mean age years old, mean cornea thickness ±23.91um) who underwent LASIK using measured PI with FS 60™ at our clinic from July 2007 to September The thickness of the flaps was calculated by the following equation: Subtraction pachymetry method (pre-operative cornea thickness - post lifting bed thickness). The corneal thickness was determined by a contact ultrasonic pachymater (SP3000™;Tomey). The subjects were divided into two groups by flap thickness: 90 um group (98 eyes) and 100 um group (152 eyes). IntraLase FS60™ Raster flap mode (Power 0.80uJ) 9.0 mm – 8.8 mm diamater Hinge angle was 60 degree (superior type) Side-cut angle was 70 degree (Power 0.80uJ)

Method of PI measurement Sterilized clinical engineer picked PI (n=531) randomly under the clean room and examined for measurement right before the operation. Using an precise measuring device (IDH0530™; Mitutoyo), we measured five spots without damaging the PI lens surface. We limited the measurement range to be ±10μm, within the range of official PI error by the manufacturer. Utilized the PI with lens tilt of ±5um or less. PI error was determined by the difference between the value from the lens center and the PI regulation by the manufacturer.

Result of PI measurement n=153 Mean ( μm ) SD ( μm ) Min ( μm ) Max ( μm ) Error None9.0 Tilt None8.0

Input for Correction PI does have few errors. PI thickness can differ a little and it effects the flap thickness. We want to make a flap according to the corrected value, thus less errors. Therefore, we measured PI thickness before surgery to make corneal flaps with more correct thickness. Our method utilizes a measured PI Input for correction (Shinagawa method) is necessary! A flap thickness of 100um is made at the time of 100um input in PI of ±0um As for PI of a + error, a flap is made thinner As for PI of a - error, a flap is made thicker We defined PI correction value to be the difference between the PI standard value of ±0um and the value from the lens center, which was reflected onto the system of FS60 ™ laser. PI regulations → Thin !!Thick !!

Flap Results ① Flap Thickness ( μm ) Number of Eyes Mean Flap Thickness ( μm ) SD ( μm ) Min ( μm ) Max ( μm ) Mean(μm)SD(μm) Min ( μm ) Max ( μm ) 90μm Right μm Left μm Right μm Left

Flap Results ② measured PI 90μmMeasured PI 100μm p Value

Summary The mean flap thickness was 92.18±4.11um (Range:82-99um) in 90 um group, and ±4.93um (Range:91-112um) in 100 um group. In the flap, a right and left correlation was not seen.

Conclusions We could reduce the errors for flap thickness by measuring PI value and making correction manually. Creating more reliable and better flaps are possible by implementing this method, measured PI value, for all operations. When making thin flap is needed, by controlling input values for less errors, foreseeable accidents and troubles can be avoided. The flap with good accuracy was able to be achieved in reproducibility good by applying this measured PI to all the operations. The present study shows that the use of measured PI enables to make flaps with more precise thickness than the method using unmeasured PI as reported by Binder et al: J Cataract Refract Surg 2004; 30:1-9. We will classify the combination with anterior segment OCT into a field in the future.

References Flap dimensions created with the Intralase pulsion laser. Binder et al : J Cataract Refract Surg 2004 ; 30:1-9. Anterior Segment OCT Analysis of Thin IntraLase Femtosecond Flaps. Jason E.Stahl et al : Journal of Refract Surgery 2007 ; Vol23: Flap thickness reproducibility in laser situ keratomileusis with a femtosecond laser : Optical coherence tomography measurement. Jin-Hyoung Kim et al : J Cataract Refract Surg 2008 ; 34: