Comparison of Central Corneal Thickness, Anterior Chamber Depth, and Central Corneal Power Measurements between Two Scheimpflug Imaging Systems Yuichi.

Slides:



Advertisements
Similar presentations
Asymmetric Keratoconus Study: Novel Tomographic Parameters to enhance sensitivity to detect abnormalities in eyes with normal axial curvature maps from.
Advertisements

Corneal topography orbscan
Swept Source Optical Coherence Tomography for Evaluation of Posterior Corneal Changes after Refractive Surgery Dr. Tommy Chung Yan Chan Dr. Vishal Jhanji.
Comparison of corneal astigmatism measured with 3 devices Mariko Shirayama, M.D, Li Wang, M.D, PhD, Mitchell P. Weikert, M.D, Douglas D. Koch, M.D. Cullen.
Eugene Ng, Arthur B. Cummings, Patrick P. Collins, Alexander V. Goncharov, Diana Bogusevschi, Chris Dainty, Michael C. Mrochen. The authors of this paper.
Agreement Between Corneal Topography with the Dual-Scheimpflug Placido Corneal Topographer, Scheimpflug Corneal Topographer, and Scanning-Slit Topography.
Corneal Biomechanics, Pachymetric Progression Profile and Corneal Volume: Indices for Detecting Ectasia and Screening Refractive Candidates Renato Ambrósio.
Purpose We studied the interdevice agreement of three systems in the different measurements of the corneal and anterior segment biometric parameters. The.
David Zadok MD, Isaac Avni MD
E-Poster ALC Canedo, MD 2010 Rio de Janeiro CornealTomograhyandBiomechanics Study Group Study Group The impact of reference body selection for calculating.
Outcome of refractive surgery in keratoconus suspect eyes
Repeatability of Anterior and Posterior Corneal Higher- Order Aberrations in 4 mm, 6 mm and 8 mm diameters measured by Pentacam System Jadwiga Wojtowicz.
Variations in corneal biomechanical parameters and central corneal thickness during the menstrual cycle. David Zadok, MD, Yakov Goldich, MD, Yaniv Barkana,
Dept. of Ophthalmology, Friedrich-Alexander University Erlangen-Nuremberg, Erlangen, Germany. Outcome after big-bubble deep anterior lamellar keratoplasty.
G. Jacob 1,2, C. Bouchard 2, S. Kancherla 1. Edward Hines, Jr. VA Hospital, Hines, IL, Department of Ophthalmology 1. Loyola University Medical Center,
Comparison of Orbscan and Ultrasound Pachymetry Measurements Faik Orucov, MD, Abraham Solomon, MD,Ziv Caspi, David Landau, MD, Eyal Strassman, MD, and.
Evaluation of Corneal Parameters and Spherical Aberration After DSAEK Measured with Pentacam System Orkun Muftuoglu, Pawan Prasher, R. Wayne Bowman, Steven.
Effects of IOP Lowering Agents on Myopic Regression after Refractive Surgery Lim, Taehyung M.D., Hong, So Jin M.D., Cho, Beom Jin M.D., Ph.D. Chung Kyu-Hyung.
Evaluation of Interface Reflectivity and Corneal Aberrations following DSAEK Hamid Khakshoor, MD Professor of Ophthalmology Mashhad University of Medical.
Evaluation of Zonal Equivalent Keratometry Readings After LASIK Timmy Kovoor, MD Orkun Muftuoglu, MD V.Vinod Mootha, MD Steven Verity, MD R. Wayne Bowman,
Glaucoma Screening With a New Non-Contact Air Impulse Tonometer Pandelis A. Papadopoulos,1,2, Dimitrios P. Bessinis 3, Alexandros A. Papadopoulos 1, Konstantinos.
Relationship between Pachymetry and Ocular Response Analyzer Measurements Alejandro Rivera MD.
The authors have no financial interest in the subject matter of this poster Yinfei Xu, Peter S. Hersh, MD, David S. Chu, MD Institutional Affiliations.
ROTATING SCHEIMPFLUG TOPOGRAPHIC PARAMETERS IMPORTANT IN DISTINGUISHING NORMAL FROM KERATOCONIC CORNEAL MORPHOLOGICAL FEATURES Clayton Falknor, MD, Orkun.
Evaluation of refractive error measurements obtained by three different aberrometers Radha Ram, BA Li Wang, MD, PhD Mitchell P. Weikert, MD, MS Disclosure:
Phoebe D. Lenhart, M.D. 1, Amy K. Hutchinson, M.D. 1, Michael J. Lynn, M.S. 2, Scott R. Lambert M.D. 1 1 Department of Ophthalmology, Emory University,
Khairidzan Mohd Kamal MD
Variations in corneal biomechanical parameters and central corneal thickness during the menstrual cycle. Yakov Goldich, MD, David Zadok MD, Yaniv Barkana,
Rio de Janeiro Corneal Tomography and Biomechanics Study Group Dr. Ambrósio is consultant for Oculus Optikgeräte GmbH (Wetzlar, Germany) Agreement of Artificial.
ELEVATION AND PACHYMETRY VALUES IN NORMAL CORNEAS OBTAINED BY GALILEI ASCRS Boston April Vision Sana Research Group Vision Sana Research Group.
Endothelial Keratoplasty in Patients With an Anterior Chamber Intraocular Lens: A Montreal Experience Georges M. Durr, MD 1,2 Johanna Choremis, MD, FRCSC.
Comparison of Central Corneal Thickness and Peripheral Corneal Thickness using Sheimpflug system, Optical Coherence Tomography and Ultrasound Pachymetry.
Mean Keratometry Measurement Post Penetrating Keratoplasty Jacky Yeung MSc MD, Stephanie Baxter MD FRCS(C) Department of Ophthalmology, Hotel Dieu Hospital,
The Effect of Corneal Anterior Surface Eccentricity on Astigmatism after Cataract Surgery Choul Yong Park MD 1 Sung Jun Lee MD 1 Prabjot Channa MD 2 Roy.
Adriana S. Forseto1, MD Walton Nosé1,2, MD
1 P.O. dei Pellegrini; A.S.L. Napoli 1Centro 2 Dipartimento di Oftalmologia, Seconda Università degli Studi Napoli Scheimpflug topography, Scanning-Slit.
Evaluation of Corneal Tomography in Primary Pterygium Sahil Goel, MD (Presenting Author), Murugesan Vanathi MD *The authors have no financial interests.
Sherman W. Reeves, MD, MPH 1,3,4 ; Jacob A. Kozisek, OD 1,2 ; Noumia Cloutier-Gill, OD 1,2 ; David R. Hardten, MD 1,2,3,4 Accuracy of Scheimpflug Imaging.
Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD Noo Ne Eye Hospital, Seoul,
Study of Flap Thickness Precision with a Femtosecond Laser Shinagawa LASIK Center Tatsuya Yonekawa,MSc ; Minoru Tomita,MD.PhD ; Youhei Iida,CE ; Yuko Inada,CE.
Financial Disclosure: Medical Director–Galilei R&D Consultant, Ziemer Group AG, Port, Switzerland Consultant & Territory Manager for.
Minimizing Risk in Visian ICL Implantation.
COMPARISON AND EVALUATION OF OCULAR BIOMETRY USING A NEW NON-CONTACT OPTICAL LOW COHERENCE REFLECTOMETER David Goldblum 1,3, Kaspar Rohrer 1, Rudolf Waelti.
Simulated Experiments on IOL Power Calculation Using Anterior Segment OCT Dong Hyun Jo, M.D., 1,2 Mee Kum Kim, M.D., 1,2 Won Ryang Wee, M.D. 1,2 1 Department.
Evaluating Post Laser Refractive Surgery Corneal Thickness Using Artemis-2 Very High Frequency Arc-Scanning Ultrasonography Versus Pentacam * † Sharon.
Liquifaction Method and Extent of Posterior Capsule Opacification: Two-Year Follow-up Marie Kalfertova, Mariya Burova, Pavel Rozsival, Nada Jiraskova Nada.
Comparison of Varying the Ablation Diameter in LASIK and LASEK:The Effects on Anterior Chamber Depth (ACD) and Regression at 3 months Minoru Tomita, MD,
Corneal Endothelial Cell Changes After LASIK Using a Femtosecond Laser Minoru Tomita, MD, PhD, Masaru Aoyama, MD, Minoru Tomita, MD, PhD, Masaru Aoyama,
Poster# P36 Comparison of Techniques of Measuring Anterior Chamber Depth and Corneal Curvature in Indian Eyes Sujata Das, MS, FRCS Anjula Kumari, B.Optom.
CORNEAL ENDOTHELIAL PROFILE AFTER FERRARA RING IMPLANTATION Leonardo Torquetti, MD, PhD Paulo Ferrara, MD, PhD Paulo Ferrara Eye Clinic Belo Horizonte.
Changes of Axial Length After LASIK Surgery: A 3-Year Follow-Up Study
Characteristics of Primary Angle-Closure Glaucoma Patients with Normal Intraocular Pressure at the First Visit Won Hyuk Oh1, Bum Gi Kim1, Joo Hwa Lee2.
Measurement of 60 kHz Femtosecond LASIK Flaps with Anterior Segment Ocular Coherence Tomography & Ultrasonic Subtraction Pachymetry Edward E. Manche, MD.
Maria A Henriquez MD Luis Izquierdo MD Jackeline Parraga MD
Comparison of Single-Scheimpflug and Dual-Scheimpflug Pachymetry Measurements and Effect of Decentration World Cornea Congress VI April 7-9, 2010.
Hôpital Ophtalmique Universitaire
Comparison of Central Corneal Thickness measured by
MP Weikert, M Shirayama, L Wang, DD Koch
Mehdi Ghajarnia, MD Mark Mifflin, MD John A. Moran Eye Center
Comparison of corneal powers obtained from four different devices
Mohamed A Guenena, MD Helga P Sandoval, MD, MSCR Kerry D Solomon, MD
Associate Professor of Ophthalmology
Vicente J. Correa-Gomez MD, Guillermo Tapia MD, Oscar Macias Manuel MD, Alejandro Navas Perez MD, Gilberto Islas MD, Tito Ramírez-Luquin MD, Enrique O.
Sun Woong Kim, M.D.1, Hae Jung Sun, M.D.1,
New 3-Dimensional Scheimpflug Analysis of Cataract Density
Bernilla E; Izquierdo Jr. L; Miranda M.
Peter Lee MD, Howard Gimbel MD, Maria Ferensowicz MA
Khairidzan Mohd Kamal MD
The Effect of Corneal Thickness on Ocular Drug Penetration
International Vision Correction Research Centre
Presentation transcript:

Comparison of Central Corneal Thickness, Anterior Chamber Depth, and Central Corneal Power Measurements between Two Scheimpflug Imaging Systems Yuichi Hori, MD, PhD, and Yuzo Deguchi, MD Department of Ophthalmology Toho University Sakura Medical Center, Sakura, Japan Financial disclosure: The authors have no financial interest in any aspect of this presentation.

Background Measurements of corneal thickness and curvature are of paramount importance in corneal refractive surgery. A rotating Scheimpflug analysis system allows acquisition of multiple images of the anterior segment of the eye and calculation of measurements of the central corneal thickness (CCT), corneal curvature, and anterior chamber depth (ACD). Several rotating Scheimpflug analysis systems are available for use in the clinic.

Purpose To compare the preoperative CCT, ACD, and central corneal power measured with the Pentacam HR™ (Oculus Optikgeräte GmbH, Germany) rotating single Scheimpflug camera and the Galilei™ Dual Scheimpflug Analyzer (Ophthalmic Systems AG, Switzerland) in normal corneas.

Subjects 43 right eyes of 43 patients (21 women, 22 men) scheduled to undergo cataract surgery at the Department of Ophthalmology, Toho University Sakura Medical Center Mean age: 60.8 ± 16.5 years No corneal abnormalities as verified by slit-lamp examination Mean spherical equivalent: ± 2.05 diopters (D)

Methods Each patient underwent imaging with both rotating Scheimpflug imaging systems, the Pentacam HR and the Galilei The instruments were used in random order. The CCT, ACD, and central corneal power (front and back axial powers) were compared between the two Scheimpflug imaging systems. The data obtained with the two modalities were presented graphically on Bland-Altman plots. Pentacam HR™Galilei™

CCT Mean (-14.11) Mean - 2 SD Mean +2 SD Average CCT (um) CCT Pentacam – CCT Galilei (um) Mean SD SD25.72 Bland-Altman plot Pentacam HR (um)Galilei (um) Mean Standard Deviation (SD) * P < 0.001, paired t test Results 1 The mean CCT values measured with the Pentacam HR (mean, 542.4±33.6 um) was significantly (p<0.001, paired t test) thinner than that measured with the Galilei (556.0±32.7 um). CCT Pentacam – CCT Galilei (um)

ACD Average ACD (mm) ACD Pentacam – ACD Galilei (mm) Mean SD SD Bland-Altman plot Pentacam HR (mm)Galilei (mm) Mean 2.83 SD * P = 0.24, paired t test Results 2 There was no significant (p=0.24) difference between the ACD measurements obtained using the Pentacam HR (2.83±0.53 mm) and the Galilei (2.83±0.47 mm). Mean +2 SD Mean -2 SD Mean (-0.018) ACD Pentacam – ACD Galilei (mm)

Central Corneal Power (front axial power) Average front central axial power (D) Front axial power: Pentacam –Galilei (D) Mean0.121 SD SD1.035 Bland-Altman plot Pentacam HR (D)Galilei (D) Mean SD * P = 0.133, paired t test Results 3 There was no significant (p=0.133) difference between the central corneal power (front axial power) measurements obtained using the Pentacam HR (44.58±1.09 D) and the Galilei (44.46±1.13 D). Mean +2 SD Mean -2 SD Mean (0.121) Front central axial power: Pentacam –Galilei (D)

Central Corneal Power (back axial power) Average back central axial power (D) Back axial power: Pentacam –Galilei (D) Mean0.2 SD SD0.327 Bland-Altman plot Pentacam HR (D)Galilei (D) Mean SD * P < 0.001, paired t test Results 4 The mean back central axial power measured by the Galilei (-6.58±0.24 D) was significantly (p<0.001) smaller than that measured by the Pentacam HR (-6.38±0.23 D). Mean +2 SD Mean -2 SD Mean (0.2) Back central axial power: Pentacam –Galilei (D)

Summary The CCT measured with the Pentacam HR was significantly thinner (approximately 14 um) than that measured with the Galilei. There was no significant difference between the ACD measurements obtained with the two imaging systems. There was no significant difference between the front central axial power measured by the two systems. However, the back central axial power measured by the Galilei was significantly smaller (approximately 0.2 diopter) than that measured by the Pentacam HR.

Discussion/Conclusion The Galilei is designed to analyze the anterior ocular segment using dual rotating Scheimpflug cameras integrated with a Placido topographer. This design may lead to differences in measurements between the Pentacam HR and the Galilei. We should know the characteristics of the data measured with these systems.

References Holladay JT, et al. Corneal power measurements using scheimpflug imaging in eyes with prior corneal refractive surgery. J Refract Surg (10): Read SA, et al. Corneal topography with Scheimpflug imaging and videokeratography: comparative study of normal eyes. J Cataract Refract Surg (6): Prospero Ponce CM, et al. Central and peripheral corneal thickness measured with optical coherence tomography, Scheimpflug imaging, and ultrasound pachymetry in normal, keratoconus-suspect, and post-laser in situ keratomileusis eyes. J Cataract Refract Surg (6): Menassa N, et al. Comparison and reproducibility of corneal thickness and curvature readings obtained by the Galilei and the Orbscan II analysis systems. J Cataract Refract Surg (10): Lackner B, et al. Repeatability and Reproducibility of Central Thickness Measurement with Pentacam, Orbscan, and Ultrasound. Optometry and Vis Sci (10): Bunce C. Correlation, Agreement, and Bland-Altman Analysis: Statistical Analysis of Method Comparison Studies. Am J Ophthalmol (1):4-6. Quisling S, et al. Comparison of Pentacam and Orbscan II on posterior curvature topography measurements in keratoconus eyes. Ophthalmology (9):