Pacemakers and AICD ’ s. Pacemaker Basics Provides electrical stimuli to cause cardiac contraction when intrinsic cardiac activity is inappropriately.

Slides:



Advertisements
Similar presentations
Appendix E Pacemakers Gail Walraven, Basic Arrhythmias, Seventh Edition ©2011 by Pearson Education, Inc., Upper Saddle River, NJ.
Advertisements

Pacemaker Basics Module 5
Dual Chamber Temporary Pacing Operations & Troubleshooting
ECG TRAINING MODULE 4 BY BRAD CHAPMAN RCT.
Temporary Pacemakers Karim Rafaat, MD.
Basic Overview ECG Rhythm Interpretation
Pacemaker Malfunctions Even less amusing!. Pacemaker Codes (NASPE/BPEG) Position I IIIII Category Chamber(s) Chamber(s) Response to paced sensed sensing.
The Very Basics of Pacing Glenn Estell Medtronic Pribcipal Clinical Specialist.
Modes of Pacing Seoul National University Hospital
Pacemaker for beginners KITA yosuke Iizuka Hospital.
EKG Monitoring.
Emergency Medicine Ryan Ngiam
Pacemaker troubleshooting-single chamber pacemakers
Superconductors A class of materials and compounds whose resistances fall to virtually zero below a certain temperature, T C T C is called the critical.
ECG Rhythm Interpretation
Pacemakers and Implantable Cardioverter-Defibrillators
Appendix E Pacemakers Gail Walraven, Basic Arrhythmias, Sixth Edition ©2006 by Pearson Education, Inc., Upper Saddle River, NJ.
Chapter 2 for 12 Lead Training -RHYTHM PRACTICE-
Pacemakers and Implantable Defibrillators
By Corey Thompson.  Thing I hope you learn  History  What they are  There importance to pt who have them  How they relate to us as EMS professionals.
The Electrical Management of Cardiac Rhythm Disorders Bradycardia Device Course The Electrical Management of Cardiac Rhythm Disorders, Bradycardia, Slide.
Welcome to ASATT Region 7 Educational Meeting
Heart Rhythm Devices in the ER
Juan Camilo Diaz Cardiac Pacemakers.
Sinus, Atrial, Junctional / Nodal, Ventricular, Blocks, others.
Anesthetic Management of Cardiac Rhythm Management Devices N746: Summer 2014 Jennifer Ranieri.
Dr.Gharibzadeh Alaleh Rashidnasab
Pacemaker & its Classification
Pacemakers and Implanted Defibrillators Mike Harlan.
Pacemaker Follow-up Alpay Çeliker MD. Hacettepe University Department of Pediatric Cardiology 3rd International Summer School on Cardiac Arrhythmias, 9-12.
Pacemaker Emergencies Arun Abbi MD Jan 21, Overview Initial approach Pocket Complications Acute complications with placement Nonarrythmic complications.
Single Chamber Temporary Pacing Operations & Troubleshooting.
 Any atrial area may originate an impulse.  Rhythms have upright P waves preceding each QRS complex.  Not as well-rounded  Heart rates usually from.
Elsevier items and derived items © 2006 by Elsevier Inc. Chapter 37 Interventions for Clients with Dysrhythmias.
Basic Concepts—Electricity and Pacemakers Module 3
Pacemakers and Implantable Cardioverter Defibrillators Chapter 10
IABP, VAD, ICD & Pacemaker Therapies Diane E. White RN CCRN PhD.
Lecture 8 Electrical energy EMF Resistors in series and parallel
Lecture Objectives Describe sinus arrhythmias Describe the main pathophysiological causes of cardiac arrhythmias Explain the mechanism of cardiac block.
Noise, Reset, EGM, Magnet, Advanced Hysteresis Noise Göran Mathson, 2005.
Basic EKG Interpretation
AV Blocks Artificial Pacemakers Terry White, RN, EMT-P.
Mar 20, 2008 ECG Rounds Yael Moussadji, R4. Case 1.
Pacemakers.
Device and Antiarrhythmic Drugs: Advantages and Pitfalls Teresa Menendez Hood, M.D.
ECG in Pacemaker Malfunction
NEED FOR CARDIAC PACEMAKER Rhythmic beating of heart originates from SA node If SA node doesn’t function properly results in decreased heart rate & change.
ECG intereptation Abdualrahman ALshehri Lecturer King Saud University
Copyright © 2014 Wolters Kluwer Health | Lippincott Williams & Wilkins Chapter 33 Disorders of Cardiac Conduction and Rhythm.
Supparerk Prichayudh M.D
Cardiovascular Therapeutic Management 2013
Pacemaker ECGs Dr. K Chan Ruttonjee and Tang Shiu Kin Hospitals
22nd April 2009 ECG Recording and Basic Interpretation.
Cardiology for Dr. Pelaez By Sai Kumar Reddy American International Medical University, St.Lucia.
Implantable Cardioverter Defibrillators & End of Life Anna Wilson Cardiac Physiologist Waikato Hospital.
Heart Blocks and Pacing
ECG RHYTHM ABNORMALITIES
Questions included for Critical Care Competency Day
Pacemaker for beginners
Temporary Pacemakers.
Chapter 4 Atrial Rhythms.
Pacemakers and Implantable Cardioverter-Defibrillators
PACEMAKER Yoga Yuniadi
Pacemakers and AICD’s.
Pacemakers.
Dysrhythmias Disorders of formation or conduction (or both) of electrical impulses within heart Can cause disturbances of Rate Rhythm Both rate, rhythm.
Pacemakers and Devices – Interactive Session
EKGs and Pacemakers Cooper University Hospital
Diathermy and its safe use
Presentation transcript:

Pacemakers and AICD ’ s

Pacemaker Basics Provides electrical stimuli to cause cardiac contraction when intrinsic cardiac activity is inappropriately slow or absent Sense intrinsic cardiac electric potentials

ICD Basics Designed to treat a cardiac tachydysrythmia Performs cardioversion/defibrillation Ventricular rate exceeds programmed cut-off rate Ventricular rate exceeds programmed cut-off rate ATP (antitachycardia pacing) Overdrive pacing in an attempt to terminate ventricular tachycardias Overdrive pacing in an attempt to terminate ventricular tachycardias Some have pacemaker function (combo devices)

Pacemaker and ICD Basics Pulse Generators Placed subcutaneously or submuscularly Placed subcutaneously or submuscularly Connected to leads Connected to leads Battery Battery Most commonly lithium-iodide type Life span 5 to 8 years Output voltage decreases gradually Makes sudden battery failure unlikely Makes sudden battery failure unlikely

Pacemaker and ICD Basics Asynchronous Fixed rate Fixed rate Impulse produced at a set rate Impulse produced at a set rate No relation to patients intrinsic cardiac activity No relation to patients intrinsic cardiac activity Susceptible to Torsades if impulse coincides with t wave Susceptible to Torsades if impulse coincides with t wave

Pacemaker and ICD Basics Synchronous Demand mode Demand mode Sensing circuit searches for intrinsic depolarization potential Sensing circuit searches for intrinsic depolarization potential If absent, a pacing response is generated If absent, a pacing response is generated Can mimic intrinsic electrical activity pattern of the heart Can mimic intrinsic electrical activity pattern of the heart

Pacemaker Nomenclature IIIIIIIVV Chamber Paced Chamber Sensed Response to Sensing Rate Modulation, Programmability Anti- tachycardia Features A=AtriumA=AtriumT=TriggeredP=SimpleP=Pacing V=VentricleV=VentricleI=Inhibited M=Multi- programmable S=Shock D=DualD=DualD=Dual R=Rate Adaptive D=Dual O=NoneO=NoneO=NoneC=Communicating O=None

Examples VVI Paces ventricle Paces ventricle Senses ventricle Senses ventricle Inhibited by a sensed ventricular event Inhibited by a sensed ventricular event

Pacing Nomenclature Examples AAT Paces atria Paces atria Senses atria Senses atria Triggers generator to fire if atria sensed Triggers generator to fire if atria sensedDDD Paces atria and ventricle Paces atria and ventricle Senses atria and ventricle Senses atria and ventricle Atrial triggered and ventricular inhibited Atrial triggered and ventricular inhibited EKG – 2 spikes EKG – 2 spikes

Atrial Spike Ventricular Spike DDD

Pacemaker Lead System Endocardial leads placed via central access Placed in right ventricle and/or atria Placed in right ventricle and/or atria Fixed to the endocardium via screws or tines Experimental pacing systems 2 atrial leads (minimize afib) 2 atrial leads (minimize afib) Biventricular pacing Biventricular pacing

Magnet Inhibition Closes an internal reed switch Causes sensing to be inhibited Causes sensing to be inhibited Temporarily turns pacemaker into “ asynchronous ” mode (set rate) Temporarily turns pacemaker into “ asynchronous ” mode (set rate) Does NOT turn pacemaker off Rate can confer info regarding battery life Distinct rates for BOL, ERI, EOL Distinct rates for BOL, ERI, EOL

Pacemaker Indications Absolute indications Sick sinus syndrome Sick sinus syndrome Symptomatic sinus bradycardia Symptomatic sinus bradycardia Tachy-brady syndrome Tachy-brady syndrome Afib with slow ventricular response Afib with slow ventricular response 3 rd degree heart block 3 rd degree heart block Chronotropic incompetence Chronotropic incompetence Inability to increase heart rate to match exercise Prolonged QT syndrome Prolonged QT syndrome

Pacemaker Indications 3 rd Degree heart block

Pacemaker Indications Relative indications Cardiomyopathy CardiomyopathyDilatedHypertrophic Severe refractory neurocardiogenic syncope Severe refractory neurocardiogenic syncope Paroxysmal atrial fibrillation Paroxysmal atrial fibrillation

ICD Indications Generally Used in cases where there was a previous cardiac arrest Used in cases where there was a previous cardiac arrest Or, patients with undetermined origin or continued VT or VF despite medical interventions Or, patients with undetermined origin or continued VT or VF despite medical interventions

Pacemaker Complications EKG abnormalities due to Failure to output Failure to output Failure to capture Failure to capture Sensing abnormalities Sensing abnormalities Operative failures Operative failures

Pacemaker Failure to Output Definition No pacing spike present despite indication to pace No pacing spike present despite indication to paceEtiology Battery failure, lead fracture, break in lead insulation, oversensing, poor lead connection, “ cross-talk ” Battery failure, lead fracture, break in lead insulation, oversensing, poor lead connection, “ cross-talk ” Atrial output is sensed by ventricular lead

Pacemaker Failure to Capture Definition Pacing spike is not followed by either an atrial or ventricular complex Pacing spike is not followed by either an atrial or ventricular complexEtiology Lead fracture or dislodgement, break in lead insulation, elevated pacing threshold, MI at lead tip, drugs, metabolic abnormalities, cardiac perforation, poor lead connection Lead fracture or dislodgement, break in lead insulation, elevated pacing threshold, MI at lead tip, drugs, metabolic abnormalities, cardiac perforation, poor lead connection

Pacemaker Sensing Abnormalities Oversensing Senses noncardiac electrical activity and is inhibited from correctly pacing Senses noncardiac electrical activity and is inhibited from correctly pacing Etiology Etiology Muscular activity (diaphragm or pecs), EMI, cell phone held within 10cm of pulse generator Undersensing Incorrectly misses intrinsic depolarization and paces Incorrectly misses intrinsic depolarization and paces Etiology Etiology Poor lead positioning, lead dislodgement, magnet application, low battery states, MI

Pacemaker Operative Failures Due to pacemaker placement Pneumothorax Pneumothorax Pericarditis Pericarditis Perforated atrium or ventricle Perforated atrium or ventricle Dislodgement of leads Dislodgement of leads Infection or erosion of pacemaker pocket Infection or erosion of pacemaker pocket Infective endocarditis (rare) Infective endocarditis (rare) Venous thrombosis Venous thrombosis

Pacemaker Complications Pacemaker syndrome Patient feels worse after pacemaker placement Patient feels worse after pacemaker placement Presents with progressive worsening of CHF symptoms Presents with progressive worsening of CHF symptoms Due to loss of atrioventricular synchrony, pathway now reversed and ventricular origin of beat Due to loss of atrioventricular synchrony, pathway now reversed and ventricular origin of beat

ICD Complications Similar to pacemaker complications Operative failures Operative failures Same as pacemakers Sensing and pacing failures Sensing and pacing failures Inappropriate cardioversion Inappropriate cardioversion Ineffective cardioversion/defibrillation Ineffective cardioversion/defibrillation Device deactivation Device deactivation

ICD Sensing failures Similar to pacmakers Oversensing Oversensing Undersensing Undersensing Appropriate failure to treat Programmed cut off at 180 bpm Programmed cut off at 180 bpm If V Tach occurs at 160 bpms, appropriately fails to cardiovert If V Tach occurs at 160 bpms, appropriately fails to cardiovert

ICD Inappropriate Cardioversion Most frequent complications Provokes pain and anxiety in pts Consider when Pt is in afib Pt is in afib With ventricular response > programmed cut off Received multiple shocks in rapid succession Received multiple shocks in rapid successionEtiology Afib, T-wave oversensing, lead fracture, insulation breakage, MRI, EMI Afib, T-wave oversensing, lead fracture, insulation breakage, MRI, EMI

ICD Inappropriate Cardioversion Treatment Magnet over ICD inhibits further shocks Magnet over ICD inhibits further shocks Does NOT inhibit bradycardiac pacing Does NOT inhibit bradycardiac pacingNote Some older devices produce beep with each QRS Some older devices produce beep with each QRS If left on for >30 seconds, ICD disabled and continous beep If left on for >30 seconds, ICD disabled and continous beep To reactivate, lift off magnet and then replace for > 30 seconds, beep will return with each QRS To reactivate, lift off magnet and then replace for > 30 seconds, beep will return with each QRS

ICD Failure to Deliver Cardioversion Etiology Failure to sense, lead fracture, EMI, inadvertent ICD deactivation Failure to sense, lead fracture, EMI, inadvertent ICD deactivationManagement External defibrillation and cardioversion External defibrillation and cardioversion Do not withhold therapy for fear of damaging ICD If pt ’ s internal defibrillator activates during chest compressions, you may feel a mild shock (no reports of deaths related to this) Antidysrhymthic medications Antidysrhymthic medications

ICD Ineffective Cardioversion Etiology Inadequate energy output Inadequate energy output Rise in the defibrillation threshold Rise in the defibrillation threshold MI at the lead site MI at the lead site Lead fracture Lead fracture Insulation breakage Insulation breakage Pre-programmed set of therapies per dysrythmia Manufacturer specific Manufacturer specific Once number of attempts reached, will not deliver further shocks until new episode is declared Once number of attempts reached, will not deliver further shocks until new episode is declared

Electromagnetic Interference Can interfere with function of pacemaker or ICD Device misinterprets the EMI causing Rate alteration Rate alteration Sensing abnormalities Sensing abnormalities Asynchronous pacing Asynchronous pacing Noise reversion Noise reversion Reprogramming Reprogramming

Electromagnetic Interference Examples Metal detectors Metal detectors Cell phones Cell phones High voltage power lines High voltage power lines Some home appliances (microwave) Some home appliances (microwave)

Electromagnetic Interference Intensity of electromagnetic field decreases inversely with the square of the distance from the source Newer pacemakers and ICDs are being built with increased internal shielding