The PHENIX Hadron Blind Detector Craig Woody PHENIX Focus February 9, 2010.

Slides:



Advertisements
Similar presentations
Thermal Photon Measurement using the HBD in PHENIX :Status and Update Sky RolnickUC RiversideAPS April Meeting 2011.
Advertisements

Benji Lewis APS April Meeting Hadron Blind Detector implementation during PHENIX Run-10 Benji Lewis On behalf of the PHENIX Collaboration.
1 HBD Update Itzhak Tserruya DC meeting, April 9, 2008 April 9, 2008.
Bingchu Huang, USTC/BNL 1 Bingchu Huang (for STAR Collaboration) University of Science and Technology of China (USTC) Brookhaven National Laboratory (BNL)
Sourav Tarafdar Banaras Hindu University For the PHENIX Collaboration Hard Probes 2012 Measurement of electrons from Heavy Quarks at PHENIX.
CsI Photocathode Production and Testing
HBD Operation in Run 9 Thomas K Hemmick for the HBD Crew.
Position Sensitive SiPMs for Ring Imaging Cherenkov Counters C.Woody BNL January 17, 2012.
PHENIX Fig1. Phase diagram Subtracted background Subtracted background Red point : foreground Blue point : background Low-mass vector mesons (ω,ρ,φ) ~
I. Ravinovich Di-electron measurements with the Hadron Blind Detector in the PHENIX experiment at RHIC Ilia Ravinovich for the PHENIX Collaboration Weizmann.
Cube Measurements Tent Crew. Scintillation BNL 241 Am Semi- collimated  Spectralon Diffuse UV Reflector SBD  -Trigger Scint. Light Poisson.
Preliminary Results from a Trial Beam Test of the Small HBD Prototype at LEGS Bob Azmoun BNL HBD Working Group Meeting May 10, 2005.
HBD Performance Estimates Zvi Citron 3 April 2008 HL Meeting.
Hadron Blind Detector:
1 Alexander Milov QM2006, Shanghai Nov 15, 2006 Construction and expected performance of the Hadron Blind Detector for PHENIX experiment at RHIC Alexander.
Jason Kamin May 17 th, Title Slide!! HBD. Jason Kamin May 17 th, Di-lepton Physics Diverse Physics: –Vector Mesons –Dalitz –Correlated semi-
1 Thomas K Hemmick May 17 th, The Gas Electron Multiplier: GEM Thomas K. Hemmick Stony Brook University.
Prototype Tests and Construction of the Hadron Blind Detector for the PHENIX Experiment at RHIC Craig Woody Brookhaven National Lab For the PHENIX Collaboration.
IN PREPARATION FOR PHD CANDIDACY REVIEW SKY D ROLNICK 10/29/08 Measuring Chiral Symmetry Restoration via low-mass e+e- pairs in Heavy Ion Collisions.
Itzhak Tserruya IEEE 2007, October 29, 2007, Hawaii Itzhak Tserruya Weizmann Institute of Science, Rehovot, Israel for the HBD group: for the HBD group:
12/12/09 MPGD 神戸大学 1 Yusuke Komatsu A B. Azmoun B, C. Woody B, K. Ozawa A University of Tokyo A,Brook Haven National Lab. B.
HBD Gas and QE Monitoring Craig Woody BNL HBD Working Group Meeting October 19, 2005.
1 HBD update Itzhak Tserruya DC Upgrades meeting, January 14, 2005 NIM paper II: Generic R&D ~ completed Full scale prototype construction Pending issues.
1 HBD Towards Run-9 Itzhak Tserruya H/L meeting, April 3, 2008 April 3, 2008.
Itzhak Tserruya, July 7,  Towards HBD Proposal  Ongoing activities and plans for the next two months Itzhak Tserruya HBD meeting at BNL, September.
1 Sasha Milov DC meeting... March 7, 2007 HBD status Sasha Milov March 7, 2007.
Thursday, December 17, 2015 Sky D. Rolnick UC Riverside 1 Baseline study for Chiral Symmetry Restoration using the Hadron Blind Detector in the PHENIX.
HBD CDR Gas System and Monitoring Craig Woody BNL DC Upgrades/EC Meeting March 9, 2005.
10/26/06Takao Sakaguchi, BNL1 Hadron Blind Detector for the PHENIX experiment at RHIC Takao Sakaguchi Brookhaven National Laboratory For the PHENIX Collaboration.
M. Muniruzzaman University of California Riverside For PHENIX Collaboration Reconstruction of  Mesons in K + K - Channel for Au-Au Collisions at  s NN.
Status of TPC/HBD for PHENIX Craig Woody BNL DC Upgrades Meeting February 12, 2002.
Measurement of photons via conversion pairs with PHENIX at RHIC - Torsten Dahms - Stony Brook University HotQuarks 2006 – May 18, 2006.
6-Aug-02Itzhak Tserruya PHENIX Upgrade mini-Workshop1 Boris Khachaturov, Alexander Kozlov, Ilia Ravinovich and Itzhak Tserruya Weizmann Institute, Israel.
Dilepton Radiation Measured in PHENIX probing the Strongly Interacting Matter Created at RHIC Y. Akiba (RIKEN Nishina Center) for PHENIX Collaboration.
Itzhak Tserruya, BNL, May13, HBD R&D Update: Demonstration of Hadron Blindness A. Kozlov, I. Ravinovich, L. Shekhtman and I. Tserruya Weizmann Institute,
1 PUTTING TWO AND TWO TOGETHER To accumulate things You can’t possibly use Is a burdensome Form of conceit. Or in other words: Never have two pairs of.
HBD Status Report C. Woody For the HBD Crew DC Meeting June 13, 2007.
TPC/HBD R&D at BNL Craig Woody BNL Mini Workshop on PHENIX Upgrade Plans August 6, 2002.
The RICH Detectors of the LHCb Experiment Carmelo D’Ambrosio (CERN) on behalf of the LHCb RICH Collaboration LHCb RICH1 and RICH2 The photon detector:
HBD Status for Run 10 C.Woody For the HBD Group Collaboration Meeting January 15, 2010.
Pe Collection Efficiency in CF4 Revisited HBD Meeting 04/15/08 B.Azmoun, A.Caccavano BNL.
Mauro Iodice INFN Roma (Italy) Hall A Collaboration Meeting - JLAB May 18, 2004.
Future Possibilities for Measuring Low Mass Lepton Pairs in Christine Aidala for the Collaboration Quark Matter 2002, Nantes.
Update on the HBD Craig Woody BNL DC Meeting May 11, 2005.
Collection of Photoelectrons from a CsI Photocathode in Triple GEM Detectors Craig Woody Brookhaven National Lab B.Azmoun 1, A Caccavano 1, Z.Citron 2,
JPS/DNPY. Akiba Single Electron Spectra from Au+Au collisions at RHIC Y. Akiba (KEK) for PHENIX Collaboration.
Nov 1, 2007 IEEE NSS & MIC 2007 by C. Y. Chi 1 A Faster Digitizer System for the Hadron Blind Detector in the PHENIX Experiment Cheng-Yi Chi Nevis Lab.
Plans to Test HBD Prototype in Run 6 Craig Woody BNL DC Meeting March 8, 2006.
1 HBD R&D: Update Itzhak Tserruya (for A. Kozlov, I. Ravinovich and L. Shekhtman) Weizmann Institute, Rehovot DC meeting Feb. 14, 2003.
PHENIX Safety Review Overview of the PHENIX Hadron Blind Detector Craig Woody BNL September 15, 2005.
HBD Status Report from Run 9 Craig Woody BNL DC Meeting July 8, 2009.
HBD Report Craig Woody BNL DC Meeting January 7, 2009.
1 HBD Commissioning Itzhak Tserruya DC meeting, BNL December 13, 2006 Progress from October 3 to November 28, 2006.
Collection of Photoelectrons from a CsI Photocathode in Triple GEM Detectors C. Woody B.Azmuon 1, A Caccavano 1, Z.Citron 2, M.Durham 2, T.Hemmick 2, J.Kamin.
January 13, 2004A. Cherlin1 Preliminary results from the 2000 run of CERES on low-mass e + e - pair production in Pb-Au collisions at 158 A GeV A. Cherlin.
1 Sasha Milov Phenix focus HBD June 20, 2006 Hadron Blind Detector for low mass di-leptons in PHENIX Sasha Milov (PHENIX focus on HBD) Jun 20, 2006.
Il Ring Imaging CHerenkov (RICH) detector a Jefferson Lab – Hall A.
1 HBD Update Itzhak Tserruya DC meeting, May 7, 2008 May7, 2008.
Measurement of photons via conversion pairs with the PHENIX experiment at RHIC - Torsten Dahms - Master of Arts – Thesis Defense Stony Brook University.
Preparations to Install the HBD for Run 6 Craig Woody BNL PHENIX Weekly Meeting January 26, 2006.
Measurements of low mass e + e - pairs in p+p and Au+Au collisions with the HBD upgrade of the PHENIX detector Mihael Makek Weizmann Institute of Science.
Initial Performance of the PHENIX Hadron Blind Detector at RHIC Craig Woody Brookhaven National Lab For the PHENIX Collaboration N IEEE Nuclear.
P HOTON Y IELD DUE TO S CINTILLATION IN CF4 Bob Azmoun, Craig Woody ( BNL ) Nikolai Smirnov ( Yale University )
Update on HBD Activities at BNL Craig Woody BNL HBD Working Group Meeting November 16, 2004.
R&D on Hadron Blind detector, recent results Issues addressed: - gain limits in CF 4 with heavily ionizing particles - operation.
A Fast, Compact TPC for PHENIX and STAR with Enhanced Electron ID Capabilities Craig Woody BNL RHIC Detector Workshop Gas Detector Working Group November.
6/29/2016I. Ravinovich1 HBD status in Run-10 I. Ravinovich WIS.
Low Mass Lepton Pairs in PHENIX Craig Woody BNL Upgrades Plenary Meeting June 14, 2002.
Ilia Ravinovich for the PHENIX Collaboration
RICH 2010, Cassis, May 2-7, 2010 Itzhak Tserruya
Presentation transcript:

The PHENIX Hadron Blind Detector Craig Woody PHENIX Focus February 9, 2010

C.Woody, PHENIX Focus, Feb 9, The HBD Team n Weizmann Institute of Science A. Milov, M. Naglis, I. Ravinovich, D.Sharma, I.Tserruya n Stony Brook University Z.Citron, M.Connors, M. Durham, T.Hemmick, J.Kamin, B.Lewis, V. Pantuev, M.Prossl, J.Sun n Brookhaven National Lab B.Azmoun, R.Pisani, T.Sakaguchi, A.Sickles, S.Stoll, C.Woody n University of California Riverside A.Iordanova, S.Rolnick n Columbia University (Nevis Labs) C-Y. Chi

C.Woody, PHENIX Focus, Feb 9, Motivation To Measure Low Mass Dilepton Pairs in Heavy Ion Collisions Low mass dilepton pairs are the best probe for studying chiral symmetry restoration and in- medium modifications of light vector mesons ρ (m = 770MeV t ~ 1.3 fm/c)  e + e - ω (m = 782MeV t ~ 20fm/c)  e + e - φ (m =1020MeV t ~ 40fm/c)  e + e - Dilepton pairs are also a probe of thermal radiation from the QGP and Hadron Gas and can produced enhancements in the dilepton pair spectrum QGP: qq   *  e + e - HG:  +  -     *  e + e - - Rapp nucl-th/

C.Woody, PHENIX Focus, Feb 9, Low Mass Pairs in p-p Collisions Excellent agreement between measured data and cocktail predictions PPG088 – Fig. 25

C.Woody, PHENIX Focus, Feb 9, Low Mass Pairs in Au-Au Collisions Measurement limited by systematic errors due to large combinatorial background PPG088 – Fig. 26 PPG088 – Fig. 27

C.Woody, PHENIX Focus, Feb 9, Measuring Low Mass Electron Pairs in Heavy Ion Collisions Measurement of low mass pairs and vector mesons in heavy ion collisions is limited by large combinatorial background from  0 Dalitz decays and  conversions Would like to reduce this background by a factor of ~ 20 or better Need rejection factor > 90% for:  e + e - and    e + e – Requires single electron efficiency > 90% (pairs > 80%) Electrons pairs measured in PHENIX without HBD in Run 4 (200 GeV Au-Au) PPG088 – Fig 16

C.Woody, PHENIX Focus, Feb 9, Strategy ~12 m e+e+ e+e+ e-e- e-e- Vector Meson Dalitz pair or conversion Problem: Low momentum electrons from Dalitz pairs and conversions are easily wrapped up by the central field and do not reach the PHENIX tracking detectors. Electrons from vector mesons generally have higher momentum and larger opening angles. Solution: Create an essentially field free region with the central part of PHENIX. Then both electrons from Dalitz pairs and conversions travel together and appear as two overlapping electrons as they exit this region, where electrons from vector mesons appear as singles

C.Woody, PHENIX Focus, Feb 9, Separating Signal from Background π φ pads relativistic electrons Photon conversions peaked around 2m e Dalitz decays peaked around 2m e Heavier meson decays have large opening angles Identify electrons with P > 200 MeV/c in Central Arms, project back and match to HBD. Reject if there is another electron within  < 200 mrad

C.Woody, PHENIX Focus, Feb 9, The PHENIX Hadron Blind Detector Cherenkov blobs e+e+ e-e-  pair opening angle ~ 1 m Radiator gas = Working gas Gas volume filled with pure CF 4 radiator ( n CF4 = , L RADIATOR = 50 cm) Proximity Focused Windowless Cherenkov Detector Electrons produce Cherenkov light Pions with P < 4 GeV/c do not However, we don’t want to detect the ionization signal from the hadrons ! Radiating tracks form “blobs” on an image plane  max = cos -1 (1/n)~36 mrad  Blob diameter ~ 3.6 cm)  Key to the measurement is distinguishing one blob from two Dalitz pairs & conversions make two overlapping blobs, single electrons from vector meson decays make only one

C.Woody, PHENIX Focus, Feb 9, The Hardon Blind Concept Primary ionization is drifted away from GEM and collected by a mesh UV photons produce photoelectrons on a CsI photocathode and are collected in the holes of the top GEM Triple GEM stack provides gain ~ few x10 3 Amplified signal is collected on pads and read out Mesh CsI layer Triple GEM Readout Pads e-e- Primary ionization g HV Primary ionization signal is greatly suppressed at slightly negative drift field while photoelectron collection efficiency is mostly preserved Collection efficiency for photoelectrons and ionization Z.Frankel et.el., NIM A546 (2005) A.Kozolov et.al. NIM A523 (2004)

C.Woody, PHENIX Focus, Feb 9, Challenges as a Cherenkov Detector  High purity gas radiator (deep VUV cutoff)  GEMs  Operation in pure CF 4  CsI photocathodes (w/GEMs)  Detect single photoelectrons  Be essentially blind to hadrons

C.Woody, PHENIX Focus, Feb 9, Gas Electron Multipliers V ~ V E ~ KV/cm 70  m 50  m 5  m 140  m Achieve gas gains ~ per foil ~ or higher in triple GEM structures Invented by F.Sauli at CERN NIM A386 (1997) Cylindrical GEM Tracker for KLOE2

C.Woody, PHENIX Focus, Feb 9, Gain Uniformity and Stability Single pad is irradiated with a 8 KHz 55 Fe source for ~ 20 min. Then all other pads are measured, and the source is returned to the starting pad. 1.5 Initial Rise Secondary rise 20% 5% HBD GEMs Gain increases range from a few % to ~ 2 J.Kamin (Stony Brook) Most GEMs show an initial charge-up effect where the gain increases with time with voltage on and then levels out. Effect not completely understood, but probably due charge build up on inner kapton holes

C.Woody, PHENIX Focus, Feb 9, Gain Stability vs Pressure and Temperature Large intrinsic gain variation due to P/T P/T variation at BNL over Run 9 HV was adjusted whenever P/T changed by more than ~ 1% (five P/T HV settings ) Gain generally remained stable to < 10% for all properly working modules Gain change ~ 20% per 1% change in P/T

C.Woody, PHENIX Focus, Feb 9, Photoelectron Yield for the HBD Yield = convolution of:  N  produced in Cherenkov radiator (50 cm CF 4, N  /d ~ 1/ 2 )  Absorption in gas (cutoff ~ 108 nm, ppms of O 2 and H 2 O)  Transparency of mesh (0.9) and GEM (0.80)  CsI QE (~ 1/ from 200 nm  108 nm)   C =  ext (E, ) x  trans (E)  Pad threshold (readout electronics and cluster reconstruction) Expected ~ p.e in Run 9 and measured ~ 22 p.e. Figure of Merit : N 0 ~ 380 cm -1 ( Wow !) N pe = N 0 L

C.Woody, PHENIX Focus, Feb 9, Production and Collection of Cherenkov Light Cherenkov ~ 1/ 2 Intrinsic wavelength cutoff in CF 4 Absorption in gas due to impurities Loss of primary photoelectrons due to O 2 and H 2 O absorption ~ 110 nm deep VUV

C.Woody, PHENIX Focus, Feb 9, CsI Quantum Efficiency and Photoelectron Extraction Efficiency Monte Carlo Simulation  ext (E, ) : Depends strongly on the extraction field Quickly rises to 100% in vacuum Slower rise to lower efficiency in gas due to backscatter of photoelectrons off of gas molecules (small dependence) Plateau value depends on gas Measured on VUV Spectrometers at BNL and WIS  trans (E) ~ 0.8 B.Azmoun et.al., IEEE Trans. Nucl. Sci. Vol 36, No. 3 (2009) CsI QE( )

C.Woody, PHENIX Focus, Feb 9, Detector Construction 20 Triple GEM Detectors (10 modules per side) Mesh electrode Top gold plated GEM for CsI Two standard GEMS Kapton foil readout plane One continuous sheet per side Hexagonal pads (a = 15.6 mm) Honeycomb panels Mylar entrance window HV panel Pad readout plane HV panelTriple GEM module with mesh grid Very low mass (< 3% X 0 including gas) Detector designed and built at the Weizmann Institute

C.Woody, PHENIX Focus, Feb 9, HBD Detector Parameters Acceptance |  | ≤0.45,  =135 o GEM size ( ,z) 23 x 27 cm 2 Segmentation 26 strips (0.80 x 27 cm) 2 strips (0.65 x 27 cm) Number of detector modules per arm 10 Frame 5 mm wide, 0.3mm cross Hexagonal pad size a = 15.6 mm Number of pads per arm 960 Dead area within central arm acceptance 6% Radiation length within central arm acceptance box: 0.92%, gas: 0.54%, preamps+sockets: 0.66% Total: 2.12% Weight per arm (including accessories) <10 kg

C.Woody, PHENIX Focus, Feb 9, History  The HBD was originally installed into PHENIX in the fall of 2006 for Run 7 ( ) at RHIC  The detector encountered several severe high voltage problems which ultimately damaged many of the GEM foils Minor GEM sparks induced larger, more damaging mesh to GEM sparks due to large stored energy in filter capacitors Spark in one module would induce sparks in other modules due to photon propagation in CF 4 Problem found with LeCroy 1471N modules that allowed HV to be reapplied after a trip causing spurious HV spikes  Some very useful data was taken with the detector during Run 7, but the entire detector was subsequently rebuild and reinstalled for Run 9 (Feb - July 2009, 200 & 500 GeV pp)

C.Woody, PHENIX Focus, Feb 9, GEM Foil Damage Run 7 Voltage Divider GEMs segmented on one side to limit stored energy Filter Caps Mesh-GEM Spark Spark Damage On Top GEM

C.Woody, PHENIX Focus, Feb 9, Improvements and Lessons Learned  We believe the main problem with GEM sparking was due to dust !  New Rapid Method procedure was instituted to build the detector Most of the assembly is done near laminar flow table (clean) Only CsI coated GEMs were installed inside glove box (dusty) No testing of individual stacks inside glove box (only after assembly) Used only for the West Detector for Run 9. East Detector was rebuilt almost entirely inside the glove box and did not perform as well. East was rebuilt again using the Rapid Method during ’09 shutdown and reinstalled for Run 10  New voltage divider Zeners installed to limit mesh to GEM voltage Increased field in 1 st transfer gap. Enabled us to achieve same gain (G ~ 10 3 ) at lower dV across GEMs  HV relay box installed between LeCroy HV PS and detector Discharges both mesh and GEM divider chain if either one trips  New sophisticated HV control program  Increased gas flow through detector to improve transparency

C.Woody, PHENIX Focus, Feb 9, “Clean Tent” at Stony Brook High Vacuum GEM Storage Container CsI Evaporator and quantum efficiency measurement Large glove box O 2 < 5 ppm H 2 O < 10 ppm Laminar Flow Hood Class ( N < 0.5 m m particles/m 3 )

C.Woody, PHENIX Focus, Feb 9, Rebuilding the East Detector (Aug ’09) Standard GEMs installed near laminar flow table (cleanest area inside tent) CsI coated GEMs installed inside glove box (minimize exposure time)

C.Woody, PHENIX Focus, Feb 9, HBD East Reinstalled for Run 10 HBD East in PHENIX 10/1/09 after refurbishing at Stony Brook

C.Woody, PHENIX Focus, Feb 9, Evaporator and QE Measurement Complete CsI evaporation station was given on loan to Stony Brook from INFN/ISS Rome (>$1M) Produces 4 photocathodes per evaporation Deposit 2400 – 4500 Å 2 nm /sec Vacuum ~ Torr Contaminants measured with RGA Measures photocathode quantum efficiency in situ from nm over entire area Photocathodes transported to glove box without exposure to air (virtually no water) Small “chicklets” evaporated at same time for full QE measurement ( nm )

C.Woody, PHENIX Focus, Feb 9, Quantum Efficiency Measuring Station GEM with CsI Molybinum boats GEM mounting box w/ wheels on track Harpoon for moving mounting box AC γ e-e- e-e- ~ 100 V ~ 2mm ampmeter GEM w/ CsI mirror reference PMT mesh (e - collection) D 2 lamp (λ=160,185,200 nm)  Uniform relative quantum efficiency across the GEM

C.Woody, PHENIX Focus, Feb 9, Scintillation Light Yield in CF 4 CF 4 has a strong scintillation emission at 160 nm Use this to measure and monitor CsI photocathodes in situ Scintillation Yield = dN  /dE 314 ± 15  / MeV B.Azmoun et.al., IEEE Trans. Nucl. Sci. April-May 2010 Setup to measure scintillation yield in CF 4 using a CsI photocathode GEM

C.Woody, PHENIX Focus, Feb 9, Scintillation Cube Lucite with Al/MgF 2 coating  particles from an 241 Am source traverse ~ 1 cm of CF 4 gas depositing several MeV Energy of the  particle is measured with a silicon surface barrier detector Light is collected by a reflecting cavity 55 Fe source mounted to base of cube allows simultaneous measurement of the gas gain One of these devices is installed in each of half of the HBD to monitor the QE of the photocathodes Use CF 4 scintillation as a calibrated light source

C.Woody, PHENIX Focus, Feb 9, CsI Photocathodes Photocathodes produced over one year ago have not deteriorated in terms of their QE New photocathodes have good QE Measured on VUV Spectrometer PC produced Aug ‘09

C.Woody, PHENIX Focus, Feb 9, New Voltage System LeCroy 1450 HV System 1471N HV Module Added Trip Detection Circuit Relay Board trips mesh + GEM whenever either trip M.Proissl

C.Woody, PHENIX Focus, Feb 9, HBD HV Control System M.Proissl

C.Woody, PHENIX Focus, Feb 9, Problem with EN2 GEMs held voltage well beyond operating, but first transfer gap would not Due to time constraints, we adopted a workaround solution: Go back to old voltage divider with lower voltage in 1 st transfer gap and compensate with higher gain in GEMs Initial testing and diagnosis resulted in some damage to top and middle GEMs (2 partial shorts), but seemed ok After going to full operation at the start of Run 10, EN2 developed several additional shorts Several attempts have been made to try and fix EN2, including installing a zener voltage divider, but it still has problems

C.Woody, PHENIX Focus, Feb 9, Gas Transmission Monitor Scanning VUV Monochrometer ( nm) B.Azmoun, R. Pisani, S.Stoll The HBD “Hutch” Recirculating gas system (~ 5 lpm to each detector) Measure absorption spectra to determine ppms of H 2 O and O 2 Input Outputs East & West D2 Lamp

C.Woody, PHENIX Focus, Feb 9, Gas Transmittance Output ~ 20 ppm H 2 O and ~ 1 ppm O 2 Input ~ 0 ppm H 2 O and O 2 Flow rate ~ 4.75 lpm to each detector B.Azmoun, S.Stoll 1/8/10

C.Woody, PHENIX Focus, Feb 9, New Reduced Data Format for Run 10 Old 12 sample format Data volume ~ KB/ev New 3 sample format Data volume ~ 7-8 KB/ev S.Bathe Rate: KHz Run Number C-Y. Chi

C.Woody, PHENIX Focus, Feb 9, Pedestals and Noise ~ 1.5 chs ~ 900 e’s ~ 0.3 pe New format T.Sakaguchi

C.Woody, PHENIX Focus, Feb 9, Gain Determination Using Scintillation in CF 4 Scintillation Ionization FORWARD REVERSE 1 p.e. For low multiplicity events, scintillation produces essentially single photoelectrons on each pad Fit single p.e. distribution to an exponential Gain ~ 1/slope Allows pad by pad calibration ADC channel Used as a gain monitor throughout the run

C.Woody, PHENIX Focus, Feb 9, Setting Gains Using Scintillation Select on most peripheral events where << 1 Ionization Scintillation 1 p.e. Forward Reverse S.Rolnick

C.Woody, PHENIX Focus, Feb 9, West Gains

C.Woody, PHENIX Focus, Feb 9, East Gains

C.Woody, PHENIX Focus, Feb 9, Setting Reverse Bias Voltages (dV Scan) Operating Point Need to set reverse bias voltage to optimize pe collection efficiency and minimize hadron response Requires setting voltage between mesh and top GEM to ~ 5V Applied voltage is ~ 4KV, so this requires high precision (~ 0.1%) and good stability in the HV PS S.Rolnick -15V -10V -5V -15V -5V -15V -10V -5V -10V WS1 WS2 WS3 WS4 WS5 ES1 ES2 ES3 ES4 ES5

C.Woody, PHENIX Focus, Feb 9, Single vs Double Electron Clusters (Run 9) Use reconstructed Dalitz pairs (M ee < 150 MeV/c) in PHENIX Central Arms Match to single or double clusters in HBD ~ 40 p.e. per two electron track I.Ravinovich ~ 22 p.e. per single electron track Agrees with our expected yield taking into account p.e. collection efficiency and transmission loss in the gas

C.Woody, PHENIX Focus, Feb 9, Hadron Rejection (Run 9) Hadron Rejection Factor Charge from dE/dx in the small (~ 100  m) region above the top GEM and in the first transfer gap is very small compared to the Cherenkov signal from electrons Hadron Response Measured using reconstructed tracks in the PHENIX Central Arm projected to the HBD Reverse Bias I.Ravinovich

C.Woody, PHENIX Focus, Feb 9, Electron Efficiency (Run 9) Measured using well identified electrons in low mass region (0.025 < m < GeV) measured in the PHENIX Central Arms and matched with hits in the HBD Single electron efficiency > 90% Pair efficiency ~ 80%. Meets requirements for achieving 90 % rejection of photon conversions and Dalitz decays I.Ravinovich “Opening angle”

C.Woody, PHENIX Focus, Feb 9, Electron Data from Run 10 ~ 22 p.e. per single electron track~ 40 p.e. per two electron track I.Ravinovich Low statistics Need  200 M peripheral events

C.Woody, PHENIX Focus, Feb 9, First Look at Low Mass Pairs in Run 10 Conversions I.Ravinovich HBD backplane Beam pipe Window, gas Pair spectrum w/o HBD Opening angle cut Matching to HBD Low statistics Need  200 M peripheral events

C.Woody, PHENIX Focus, Feb 9, Scintillation in Au+Au Estimate of Npe per pad due to scintillation in Au-Au using Run 7 data Occupancy is high in MB  central collisions ! I.Ravinovich

C.Woody, PHENIX Focus, Feb 9, Effect of Scintillation on Single vs Double Separation per pad with high centrality cut Monte Carlo Single vs Double Electrons with Scintillation Z.Citron =10 Double Single  single ~ 80%  double ~ 76% 3 pad clusters S.Rolnick = 8.8 = 9.3 = 10.3 = 9.3 = 7.5

C.Woody, PHENIX Focus, Feb 9, Improvements in S/B with HBD in Run 10 Z.Citron Fraction relative to Run 4 Run 4 = 0.8 x 10 9 MB events Run 7 = 5.4 x 10 9 MB Run 10  ~ 6.0 x 10 9 MB f ~ 7.5 ~ 25% decrease due to scintillation in central collisions Npe single = 22 pe Run 4 statistics

C.Woody, PHENIX Focus, Feb 9, Summary  The HBD is a first of its kind, very high performance Cherenkov counter that has been very challenging to build an operate.  The detector performed well in Run 9 in p-p and gave the expected level of performance in terms of p.e. yield, electron efficiency and hadron rejection  It seems to be performing well again in Run 10 and should give the required level of efficiency and rejection in Au+Au collisions  Given this level of performance and the added benefit of the HBD in terms of suppressing Dalitz pairs and photon conversions, it should enable a much more sensitive and precise measurement of low mass dilepton pairs in heavy ion collisions at RHIC  Come back next week for Tom’s talk on some of the great physics that we can do now that we have the HBD