Calibration of an Ionisation Chamber for use in Megavoltage Dosimetry

Slides:



Advertisements
Similar presentations
MOSFET Dosimeters for 50 kVp Electronic Brachytherapy with the Xoft Axxent Source Steve Axelrod, PhD Tom Rusch, PhD Xoft, Inc., Fremont, CA
Advertisements

Chapter 4 Radiation Dosimeters
RADIATION PROTECTION IN RADIOTHERAPY Part 10: Optimization of protection in External Beam Radiotherapy PRACTICAL EXERCISE IAEA Post Graduate Educational.
Diamond Detectors for Medical Dosimetry Jan U. Würfel, PTW-Freiburg, CARAT Workshop, 13–15 December 2010.
Ionization Chamber Array for External Beam Radiotherapy
RapidArc plan verification using ArcCHECK™
The Tomotherapy Experience at Advocate Good Samaritan Hospital
Radiation Units & Quantities
Photon Beam Monitor-Unit Calculations
Dose Calculations A qualitative overview of Empirical Models and Algorithms Hanne Kooy.
10-1 CHEM 312 Lecture 10: Part 1 Radiation Reactions: Dosimetry and Hot Atom Chemistry Readings: §Reading: Modern Nuclear Chemistry, Chap. 17; Nuclear.
Measurement of Absorbed Dose (6)
DOSIMETRY PROTOCOLS.
Kilovoltage X-ray Dosimetry
NPL’s activities dosimetry for proton and ion therapy Hugo Palmans Radiation Dosimetry Team, National Physical Laboratory, Middlesex, UK
TRAINING COURSE ON RADIATION DOSIMETRY: Instrumentation 1 – Gas detectors / Part 1 Anthony WAKER, University of Ontario Instutute of Technology Wed. 21/11/2012,
At the position d max of maximum energy loss of radiation, the number of secondary ionizations products peaks which in turn maximizes the dose at that.
Electron Beams: Physical Principles and Dosimetry
THE EXTERNAL HAZARD Radiation Protection.
Photon and Energy Fluence
Measuring dose Ionisation chamber. Ionization Chambers Thimble chambers 600cc chamber.
Beam Therapy Equipment 3 Patient Treatment and Accessories.
Dose Distribution and Scatter Analysis
Michele Togno ARDENT ESR 11 – 2D Ionization Chambers Array for Clinical Applications Michele Togno – ARDENT midterm review preparatory meeting, CERNJune,
TWO FIELD BREAST PLAN VS. OPTIMIZED CONFORMAL BREAST PLAN: COMPARISON OF PLAN PARAMETERS Authors: Borko Basarić, Ozren Čudić, Milan Teodorović, Borislava.
CENTRAL LABORATORY FOR RADIOLOGICAL PROTECTION IAEA PROJECT RER/9/117 IAEA PROJECT RER/9/117 Regional Workshop on Regulatory Control of Radioactive Discharges.
Part VIII:Medical Exposures in Radiotherapy
In vivo dosimetry Eirik Malinen Eva Stabell Bergstrand Dag Rune Olsen.
輻射劑量學之品質保證 講者:蕭安成 物理師 參考資料: 1. The Physics of Radiation Therapy Faiz M. Khan 2. Introduction to Radiological Physics and Radiation Dosimetry.
CT physics and instrumentation
Ye, Sung-Joon, Ph.D. Ove, Roger, M.D., Ph.D.; Shen, Sui, Ph.D.
Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics.
Use of the IC Profiler detector array for comprehensive machine QA ESTRO QA & Dosimetry Satellite Symposium Steve Morgan, Medical Physics Dept,
APHY398C 6/4/ Dosimetry   Quantifying the incidence of various biological changes as a function of the radiation dose.   Exposure Ratio of total.
AIR CORE SCINTILLATION DOSIMETER SUMMARY We have shown that Cerenkov light can be reduced to a negligible level in scintillation dosimetry by using an.
Determining Radiation Intensity
Ion recombination Robert Brackenridge, The Princess Royal Hospital, Hull & East Yorkshire NHS Trust.
What are the dose quantities in CT ? IAEA/RCA Kampala Kampala.
TLD POSTAL DOSE QUALITY AUDIT FOR 6MV AND 15MV PHOTON BEAMS IN RADIOTHERAPY CLINICAL PRACTICE Sonja Petkovska 1, Margarita Ginovska 2, Hristina Spasevska.
Interaction of x-ray photons (and gamma ray photons) with matter.
Qing Liang, PhD Medical Physicist Mercy Health System, Janesville, WI
1 A protocol for the determination of absorbed dose from high energy photon and electron beams AAPM TG-21 Protocol (Med Phys 10: , 1983) 和信治癌中心醫院.
Commissioning of a commercial treatment planning system for IMAT and Dose Painting treatment delivery. G. Pittomvils 1,,L. Paelinck 1, F. Crop 2, W. De.
The Effects of Small Field Dosimetry on the Biological Models Used In Evaluating IMRT Dose Distributions Gene Cardarelli,PhD, MPH.
Characteristics of the iba In-Vivo Dosimetry System. Dawn Broadhead Sheffield 9 th October 2010.
Adapting A Clinical Medical Accelerator For Primary Standard Dosimetry
AAPM TG-51 Protocol (Med Phys 26: , 1999)
Introductions Dr Emma Bowers Training Coordinator Talk about: NTC Blended learning Master’s Programme Joe Spoor 1 st year STP trainee Lauren Slater 1 st.
Qing Liang, PhD Medical Physicist Mercy Health System, Janesville, WI
CHAPTER 3 DOSE DETERMINATION FOR EXTERNAL BEAMS
Beam quality correction factors for linear accelerator with and without flattening filter Damian Czarnecki1,3, Philip von Voigts-Rhetz1, Björn Poppe3,
EURADOS Working Group 9: Radiation Dosimetry in Radiotherapy
Khan, The Physics of Radiation Therapy, Ch7
Dr: Mohamed Afifi By Lecturer Radiological Science
Very High Energy Electron for Radiotherapy Studies
Evaluation Of RTOG Guidelines For Monte Carlo Based Lung SBRT Planning
Absolute Dose Measurement
A system of dosimetric calculations
Linac Commissioning Overview
David Sutton or Colin Martin But Borrowed from Jerry Williams
Template Matching Can Accurately Track Tumor Evaluation of Dose Calculation of RayStation Planning System in Heterogeneous Media Huijun Xu, Byongyong Yi,
Chapter 9: Calibration of Photon and Electron Beams
Ch 10. A System of Dosimetric Calculations
Chapter 8 (Part 1) Measurement of Absorbed Dose
Ch6 Measurement of Ionizing Radiation
Dosimetry Standards at NMIJ/AIST Ionizing Radiation Standards Group
Dosimetric Measurements
Intercomparison on Personal Dose Equivalent (Hp(10))
Hp(10) irradiations for personal dosimetry traceability
GHG meeting at ESTRO36 May, 2017
Presentation transcript:

Calibration of an Ionisation Chamber for use in Megavoltage Dosimetry An insight into the Radiotherapy Placement undertaken by a Part 1 Trainee Clinical Scientist Will Mairs Princess Royal Hospital - Hull

Overview of talk Why we do dosimetry What we use Method of calibration Need to know dose delivered What we use Thimble ion chambers Hierarchy traceable to NPL Method of calibration Following IPSM COP Chamber factor in practice How linked to routine QA

Why do we carry out dosimetry? Radiation causes biological damage Need to quantify dose delivered Delivery of prescription

Bragg-Gray Principle Tissue Ξ water Absorbed dose in a given material can be deduced from the ionisation produced in a small gas-filled cavity within that material Tissue Ξ water

What do we use to carry out MV photon dosimetry? Thimble ionisation chambers Small volume of air (typically <1ml) Potential difference of a few hundred volts Used with a calibrated electrometer Outer electrode (graphite wall) Central Al electrode Basic design of FARMER thimble chamber Insulator

What do we use to carry out MV photon dosimetry? Hierarchy of chambers Secondary standards calibrated against primary every two years at NPL NPL provides calibration factor (ND) in terms of absorbed dose to water based on graphite calorimeter Field chambers calibrated against secondary standard (in house) at least once a year Primary Standard Secondary Standard Field Chamber

Summary of calibration method Described in IPSM COP Intercompare chambers – using same radiation qualities as will be measured by the field chamber Correct for Ion recombination losses Measure Q.I. to find NPL correction factor Calculate chamber factor to convert electrometer reading to absorbed dose to water

Calibration Factor Ms/Mf - ratio of secondary standard reading to field instrument reading Is/If - ion recombination factor ratio ND - NPL secondary standard calibration factor

Method for MV calibration Chambers placed in intercomparison jig in water tank Depth of 5cm in 10x10cm field 95cm TSD 100mu delivered Interchanged Ratio calculated Lasers Secondary standard Field chamber Thin window in plotting tank

Measuring Quality Index Measure of radiation quality Ratio of ion recombination corrected instrument readings – 20cm and 10cm depths 10x10cm field Constant TCD of 100cm Used to find NPL calibration factor from certificate

Results Nominal beam energy 6MV 10MV Is 1.008 If 1.020 1.022 Q.I. 0.667 0.734 ND (cGy/nC) 10.190 10.082 Ms/Mf (nC/div) 8.997 9.034 CF (Gy/nC) 0.906 0.898 Summary of the data required to calculate the calibration factor for field instrument NE 2571 s/n 3234.

Calibration factors in practice LINAC calibrated to give 1Gy 5cm depth 10x10cm field 100mu TSD of 95cm Calibration in Perspex rather than water Requires PBF

Calibration factors in practice CFlumped = CF x PBF x If Nominal beam energy 6MV 10MV PBF 1.014 CFlumped (Gy/nC) 0.937 0.931 Previous CFlumped (Gy/nC) 0.943 0.930 Error 0.6% 0.1% Found to be consistent with previous factors

Summary It is necessary to carry out dosimetry Field chambers are traceable to national standards Calibration carried out in house following COP Chamber used in the field for QA

Thank you for listening