IABP- Instrumentation, Indications and Complications

Slides:



Advertisements
Similar presentations
Assisted Circulation MEDICAL MEDICAL  Drugs  EECP MECHANICAL  IABP ( Introaortic balloon pump)  VAD (Ventricular assist device)
Advertisements

Jugular Venous Pulse and Carotid Arterial Pulse
Associated Prof. of physiology
Advanced Nursing Concepts Part 1: Hemodynamic Monitoring
Circulatory Adaptations to Exercise
Arterial Catheters Systemic arterial blood pressure is most accurately measured by placing a catheter directly into a peripheral artery. Peripheral arterial.
Intra-aortic Balloon Pump (IABP)
By: ABDULRAHMAN ALSALMI INTRA AORTIC BALLOON PUMP CHIEF CARDIAC PERFUSIONIST KFMMC.
Cardioanaesthesia. Coronary artery disease O 2 delivery Coronary blood flow = directly related to coronary perfusion pressure (CPP) CPP = aortic diastolic.
Innovative Minimally Invasive Circulatory Assist Device.
5/24/ HEMODYNAMIC MONITORING. OBJECTIVE 5/24/ Describe the three attributes of circulating blood and their relationships. 2. Identify types.
Cvs module phase II MBBS CARDIAC CYCLE PHYSIOLOGY CARDIOVASCULAR MODULE PHASE II
Cardiogenic Shock and Hemodynamics. Outline Overview of shock – Hemodynamic Parameters – PA catheter, complications – Differentiating Types of Shock Cardiogenic.
Chapter 15 Assessment of Cardiac Output
The Recording of Jugular Venous & Carotid Arterial Pulses.
Hemodynamic Monitoring By Nancy Jenkins RN,MSN. What is Hemodynamic Monitoring? It is measuring the pressures in the heart.
THE CARDIAC CYCLE.
Myocardial Ischemia, Injury, and Infarction
Cardiovascular System Assessments
Hemodynamic Monitoring By Nancy Jenkins RN,MSN. What is Hemodynamic Monitoring? It is measuring the pressures in the heart.
Ventricular Assist Devices Brian Schwartz, CCP February 25, 2003.
HEART PHYSIOLOGY & CONDUCTION SYSTEM
2nd Edition / 2002 The information contained on the following pages, in conjunction with the slides provided at the back of this binder, complete the Introduction.
IABP, VAD, ICD & Pacemaker Therapies Diane E. White RN CCRN PhD.
Intra-Aortic Balloon Pump Counterpulsation
Arrow Intra-Aortic Balloon Pump
Bio& 242 Unit 3 / Lecture 2 1. Position of the Heart and Associated Structures Coronary trivia Pumps blood through 60,000 miles of blood vessels Pumps.
1 © 2013 Cengage Learning. All Rights Reserved. May not be copied, scanned, or duplicated, in whole or in part, except for use as permitted in a license.
1 Arterial Lines Set Up & Monitoring Union Hospital Emergency Department.
Chapter 16 Assessment of Hemodynamic Pressures
Cardiogenic Shock Dr. Belal Hijji, RN, PhD October 12 & 15, 2011.
PWV & PWA Parameter Extraction / Transfer Function.
EKG Ventricular Pressure Aortic Pressure Ventricular Volume P Mm Hg ml A The P wave denotes atrial depolarization.
Copyright © 2008 Lippincott Williams & Wilkins. 1 Assessment of Cardiovascular Function Hemodynamic Monitoring.
Cardiac Output. Cardiac output The volume of blood pumped by either ventricle in one minute The output of the two ventricles are equal over a period of.
Position of the Heart and Associated Structures Coronary trivia Pumps blood through 60,000 miles of blood vessels Pumps about 3,600 gal per day 2.6 million.
Chapter 9 Heart. Review of Structure and Function The heart is divided into the systemic (left) and pulmonary (right) systems –The pulmonary system has.
Exercise Management Atrial Fibrillation Chapter 9.
Intra-Aortic Balloon Pump What it is and what it does
Pre-ICU training. 工作態度 會客時主動告知病情 病歷每天書寫 2 次 主動反應問題 接觸病人前後洗手.
Cardiac Cycle Dr. Wasif Haq. Introduction Cardiac events that occur from beginning of one heartbeat to the beginning of the next. Inversely proportional.
Intra-aortic Balloon Pump Counterpulsation
AB 1/03 Non-Coronary Intervention Circulatory Support Advanced Angioplasty 2003 Andreas Baumbach Bristol Royal Infirmary.
Interventions for Clients in Shock. Shock Can occur when any part of the cardiovascular system does not function properly for any reason Can occur when.
Ihab Alomari, MD, FACC Assistant professor – Interventional Cardiology University of California, Irvine Division of Cardiology Cath Lab Essentials : LV.
Terapie chirurgiche dell’Insufficienza Cardiaca
Intra-aortic baloon pumps what, who, why why why? Daniel Lovric Fellow, CVICU Auckland City Hospital Auckland Region ICU Study Day 30th October 2014.
Hemodynamic Monitoring John Nation RN, MSN Thanks to Nancy Jenkins.
Cardiac Cycle By Dr. Khaled Ibrahim Khalil By Objectives: By the end of this lecture, you should : By the end of this lecture, you should :  Describe.
FIRST AID AND EMERGENCY CARE LECTURE 4 Vital Signs.
Conflict of Interest Baxter Research Grant Medtronic Research Grant
Date of download: 7/6/2016 Copyright © The American College of Cardiology. All rights reserved. From: Manifestations of Cardiac Disease in Carotid Duplex.
Cardiac Cycle- 1 Mechanical events, Volume & Pressure changes in cardiac chambers & the great vessels during the cardiac cycle.
Indication Contraindication Preparation
IABP Review and Competency
Mechanical Support of the Failing Cardiorespiratory System
Cardiac Function in Disease Robert A. Augustyniak, PhD
What are the events that occur during one heart beat?
Cardiac Cycle.
Intra Aortic Balloon Pump (IABP)
The Electrocardiogram
Extracorporeal Life Support (ECLS)
Intra-Aortic Balloon Pumps
Dr. Arun Goel Associate professor Department of Physiology
Mechanical Circulatory Support Devices HOSEIN PASANDI.
Cardiovascular system- L3
Chapter 19: Physiology of the Cardiovascular System
CARDIC CYCLE Dr. Haseeb Sattar.
Intra-Aortic Balloon Pumps
Cardiac Cycle.
Presentation transcript:

IABP- Instrumentation, Indications and Complications Dr Sajeer KT Senior Resident Dept.of Cardiology, MCH Calicut

Intra aortic balloon counter pulsation( IABP): Temporary support for the left ventricle by mechanically displacing blood within the aorta Most common and widely available methods of mechanical circulatory support Concepts: - Systolic unloading - Diastolic augmentation Traditionally used in surgical and non surgical patients with cardiogenic shock

Indications for IABP 1. Cardiogenic shock: - Associated with acute MI - Mechanical complications of MI - MR , VSD 2. In association with CABG : Preoperative insertion - Patients with severe LV dysfunction - Patients with intractable ischemic arrhythmias Postoperative insertion - Postcardiotomy cardiogenic shock 3. In association with nonsurgical revascularization: Hemodynamically unstable infarct patients High risk coronary interventions - severe LV dysfunction, LMCA, complex coronary artery disease 4. Stabilization of cardiac transplant recipient before insertion of VAD Post infarction angina Ventricular arrhythmias relathed to ischemia

Contraindications to IABP Severe aortic insufficiency Aortic aneurysm Aortic dissection Limb ischemia Thrombo embolism

Cardiac cycle LV contraction: - Isovol. Contraction (b) - maximal ejection (c) LV relaxation: - start of relaxation and reduced ejection (d) - isovol.relaxation (e) LV filling: - LV filling , rapid phase (f) - slow LV filling (g) - atrial systole( a)

Interpreting IABP waveforms

IABP –instrumentation and techniques

The IAB Counter pulsation system - two principal parts A flexible catheter -2 lumen first - for distal aspiration/flushing or pressure monitoring second - for the periodic delivery and removal of helium gas to a closed balloon. A mobile console system for helium transfer computer for control of the inflation and deflation cycle

HEMODYNAMIC EFFECTS — Inflation and deflation of the balloon Blood is displaced to the proximal aorta by inflation during diastole. Aortic volume ( afterload) is reduced during systole through a vacuum effect created by rapid balloon deflation

Expected changes with IABP support in hemodynamic profile in patients with Cardiogenic shock - Decrease in SBP by 20 % Increase in aortic Diastolic Press. by 30 % ( raise coronary blood flow) Increase in MAP Reduction of the HR by 20% Decrease in the mean PCWP by 20 % - Elevation in the COP by 20%

IABP catheter: 10-20 cm long polyurethane bladder 25cc to 50cc capacity Optimal 85% of aorta occluded (not 100%) The shaft of the balloon catheter contains 2 lumens: - one allows for gas exchange from console to balloon - second lumen - for catheter delivery over a guide wire - for monitoring of central aortic pressure after installation.

Intra Aortic Balloon

IABP sizing chart

Benefits of larger volume IABs 􀁑 More blood volume displacement 􀁑 More diastolic augmentation 􀁑 More systolic unloading

IABP Kit Contents Introducer needle • Guide wire • Vessel dilators • Sheath • IABP (34 or 40cc) • Gas tubing • 60-mL syringe • Three-way stopcock

Intraaortic balloon

STEP BY STEP- IABP insertion Connect ECG Set up pressure lines Femoral access – followed by insertion of the supplied sheath(7.5 F) 0.030 inch supplied J-shaped guide wire to the level of the aortic arch (LAO view)

Before taking the Catheter out of Tray

Take the entire catheter and T handle as one unit (DO NOT disconnect one-way valve when removing the extracorporeal tubing from the tray.) Pull out the T- handle only as shown

Inserting the Balloon catheter • Remove stylet/aspirate/Flush • Insert the balloon only over the guide wire • Hold the catheter close to skin insertion point • Advance in small steps of 1 to 2 cm at a time and stop if any resistance. • The IABP should advance freely Many vascular complications occur during insertion itself Resistance during insertion either indicates PVOD, or dissection - Kinking of IABP » improper inflation/deflation

Positioning The end of the balloon should be just distal (1-2 cm) to the takeoff of the left subclavian artery - Position should be confirmed by fluoroscopy or chest x-ray

Connecting to console: - Connect helium gas tube to the console via a long extender Open helium tank. The central lumen of the catheter is flushed and connected to pressure tubing with 3 way and then to a pressure transducer to allow for monitoring of central aortic pressure. - Zero the transducer Initial set-up: - Once connected properly the console would show ECG and pressure waveforms. - Check Basal mean pressure - Make sure the setting is at “auto” - Usually IABP started at 1:1 or 1:2 augmentation - Usually Augmentation is kept at maxim

Trigger modes Trigger : - Event the pump uses to identify the onset of cardiac cycle (systole) - Pump must have consistent trigger in order to provide patient assist - If selected trigger not detected, counter pulsation will interrupted 1.ECG - uses the slope of QR segment to detect triggering point 2. AP(Arterial pressure wave) - Systolic upstroke of the arterial pressure wave form is the trigger 3. IN(Internal trigger)

ECG signal – most common • Inflation - middle of T wave • Deflation – peak of R wave • Pacer (v/a) • Arterial waveform • An intrinsic pump rate (VF, CPB)

Initial settings Auto Operation Mode  Automatic lead and trigger selection  Automatic and continuous inflation and deflation timing management - User has ability to fine-tune deflation timing  Automatic management of irregular rhythms Semi-Auto Operation Mode  Operator selects most appropriate lead and trigger source

The “normal” augmented waveform Increased coronary perfusion

Not all Sub optimal augmentation is due to Timing errors/kinks

Factors affecting diastolic augmentation Patient - Heart rate - Mean arterial pressure - Stroke volume - Systemic vascular resistance Intra aortic balloon catheter - IAB in sheath - IAB not unfolded - IAB position - Kink in the IAB catheter - IAB leak - Low helium concentration Intra aortic balloon pump - Timing - Position of IAB augmentation control

How to check waveform is acceptable ? First change from 1:1 to 1:2 augmentation

How to check waveform is acceptable ? First change from 1:1 to 1:2 augmentation Check the dicrotic notch See if augmentation starts at that point This should produce a sharp “V” at inflation.

How to check waveform is acceptable ? First change from 1:1 to 1:2 augmentation Check the dicrotic notch See if augmentation starts at that point This should produce a sharp “V” at inflation. Check if diastolic augmented wave is › systolic wave

How to check waveform is acceptable ? First change from 1:1 to 1:2 augmentation Check the dicrotic notch See if augmentation starts at that point This should produce a sharp “V” at inflation. Check if diastolic augmented wave is › systolic wave Confirm if end diastolic wave following the augmented wave is less than an non augmented wave. Is Deflation slope ok

Late Inflation Inflation of the IAB markedly after closure of the aortic valve. Waveform Characteristics: • Inflation of IAB after the dicrotic notch. • Absence of sharp V. Sub optimal diastolic augmentation

Early Deflation Premature deflation of the IAB during the diastolic phase.

Late Deflation Late deflation of the IAB during the diastolic phase. Waveform Characteristics: • Assisted aortic end diastolic pressure may be equal to the unassisted aortic end diastolic pressure. • Rate of rise of assisted systole is prolonged. • Diastolic augmentation may appear widened

Normal Balloon Pressure Waveform Once Arterial waveform is ok, check balloon waveform Normal Balloon Pressure Waveform

Variation in balloon pressure wave forms Decreased duration of plateau due to shortened diastolic phase Increased duration of plateau due to longer diastolic phase

Variation in balloon pressure wave forms Varying R-R intervals result in irregular plateau durations

Variation in balloon pressure wave forms Increased height or amplitude of the waveform Decreased height or amplitude of the waveform

Variation in balloon pressure wave forms Gas leak Leak in the closed system causing the balloon pressure waveform to fall below zero baseline.. due to a loose connection a leak in the IAB catheter - H2O condensation in the external tubing a patient who is tachycardiac and febrile which causes increased gas diffusion through the IAB membrane

Catheter Kink Rounded balloon pressure waveform - Loss of plateau resulting from a kink or obstruction of shuttle gas - Kink in the catheter tubing Improper IAB catheter position Sheath not being pulled back to allow inflation of the IAB - IAB is too large for the aorta IAB is not fully unwrapped H2O condensation in the external tubing

“Balloon too large” syndrome

Patient Management During IABP support Anticoagulation-- maintain apTT at 50 to 70 seconds CXR daily – to R/O IAB migration Check lower limb pulses - 2 hourly. - If not palpable » ? - vascular obstruction - thrombus, embolus, or dissection (urgent surgical consultation) Prophylactic antibiotics --?? Hip flexion is restricted, and the head of the bed should not be elevated beyond 30°.

Patient Management During IABP support Never leave in standby by mode for more than 20 minutes > thrombus formation Daily – Haemoglobin (risk of bleeding or haemolysis) – Platelet count (risk of thrombocytopenia) – Renal function (risk of acute kidney injury secondary to distal migration of IABP catheter) Wean off the IABP as early as possible as longer duration is associated with higher incidence of limb complications

Weaning of IABP Timing of weaning: - Patient should be stable for 12 – 24 hours - Decrease inotropic support - Decrease pump ratio – From 1:1 to 1:2 or 1:3 - Decrease augmentation - Monitor patient closely – If patient becomes unstable, weaning should be immediately discontinued

IABP Removal Discontinue heparin six hours prior to removal Disconnect the IAB catheter from the IAB pump - Patient blood pressure will collapse the balloon membrane for withdrawal - Withdraw the IAB catheter through the introducer sheath until resistance is met. NEVER attempt to withdraw the balloon membrane through the introducer sheath. Remove the IAB catheter and the introducer sheath as a unit - Check for adequacy of limb perfusion after hemostasis is achieved.

Thank you

1. Major physiological effects of counter pulsation include? A) increased coronary artery perfusion, increased preload, decreased after load, decreased myocardial oxygen consumption B) increased coronary artery perfusion, increased preload, increased after load, decreased myocardial oxygen consumption C) increased coronary artery perfusion, decreased preload, decreased after load, increased myocardial oxygen consumption D) increased coronary artery perfusion, decreased preload, decreased after load, decreased myocardial oxygen consumption

2. the dicrotic notch on the arterial wave form reflects A) aortic valve opening B) aortic valve closure C) isovolumetric contraction D)rapid ejection

3. Expected changes with IABP support in hemodynamic profile in patients with Cardiogenic shock include all except? A) Decrease in SBP by 20 % B) Increase in aortic DP by 30 % C) Decrease in MAP by 10% D) Reduction of the HR by 20% E)Decrease in the mean PCWP by 20 %

4. late inflation of the balloon can result in? premature augmentation increased augmentation C) decreased augmentation D) increased coronary perfusion

5. A rounded balloon pressure wave form indicate 5. A rounded balloon pressure wave form indicate? A) helium leak B) power failure C) hypovolemia D) balloon occluding the aorta

6. width of balloon pressure wave form corresponds to A) length of systole B) length of diastole C) arterial pressure D) helium level

7. true statement Dicrotic notch- land mark used to set deflation Deflation is timed to occur during period of iso volumetric contraction Most common trigger used is arterial pressure wave method Internal trigger mode is acceptable to use in a patient with normal sinus rhythm

8. true statement A) pacing spikes are automatically rejected in ECG triggered modes B) pacing trigger modes can be used in a patient of 50% paced rhythm C) Varying R-R interval result in regular plateau durations in Balloon pressureWave form

9. Identify the tracing abnormality

10. Identify the tracing abnormality

1. Major physiological effects of counter pulsation include? A) increased coronary artery perfusion, increased preload, decreased after load, decreased myocardial oxygen consumption B) increased coronary artery perfusion, increased preload, increased after load, decreased myocardial oxygen consumption C) increased coronary artery perfusion, decreased preload, decreased after load, increased myocardial oxygen consumption D) increased coronary artery perfusion, decreased preload, decreased after load, decreased myocardial oxygen consumption

2. the dicrotic notch on the arterial wave form reflects A) aortic valve opening B) aortic valve closure C) isovolumetric contraction D)rapid ejection

3. Expected changes with IABP support in hemodynamic profile in patients with Cardiogenic shock include all except? A) Decrease in SBP by 20 % B) Increase in aortic DP by 30 % C) Decrease in MAP by 10% D) Reduction of the HR by 20% E)Decrease in the mean PCWP by 20 %

4. late inflation of the balloon can result in? A) premature augmentation B) increased augmentation C) decreased augmentation D) increased coronary perfusion

5. A rounded balloon pressure wave form indicate? A) helium leak B) power failure C) hypovolemia D) balloon occluding the aorta

6. width of balloon pressure wave form corresponds to A) length of systole B) length of diastole C) arterial pressure D) helium level

7. true statement Dicrotic notch- land mark used to set deflation Deflation is timed to occur during period of iso volumetric contraction Most common trigger used is arterial pressure wave method Internal trigger mode is acceptable to use in a patient with normal sinus rhythm

8. true statement A) pacing spikes are automatically rejected in ECG triggered modes B) pacing trigger modes can be used in a patient of 50% paced rhyth C) Varying R-R interval result in regular plateau durations in Balloon press. Wave form

9.