MIMO 3 Preliminary Test Results
MIMOSTAR 2 16/05/2007 MimoStar3 Status Evaluation of MimoStar2 chip Test in Laboratory + Beam test at DESY + Irradiation test 100% yield with RadTol pixels on 15 chips MimoStar2 performances meet STAR requirements End of June 2006: Submission engineering run Two types of substrates (14, 20 µm EPI) Reception (Oct. 2006) test (up to Feb. 2007) Missing HRES layer (option) in fabrication High diode leakage current Re fabrication with HRES (Jan. 2007) VTFPP out of spec (discussion ~ 1 month) Reception (April 2007) wafer test High diode leakage current
MIMOSTAR 3 16/05/ st Run: MimoStar3 Test MimoStar3 with 20 µm EPI substrate: ~100% pixel work 2 chips tested on PCB 25 & 50 MHz ) 5 on probe station 60 KHz) High ENC, Ex. ENC ~ 20 e 20 °C, t r.o. = 4 ms Confirmed by static diode leakage current measurements: 4-7 times higher than room T (I Leak < 5 fA/diode) M*3 M*2 (14 µm EPI)
MIMOSTAR 4 16/05/ st Run: MimoStar3 Test MimoStar3 with 14 µm EPI substrate: Seems less noise Ex. ENC ~ 18 e 20 °C, t r.o. = 4 ms Confirmed by static diode leakage current measurements: Diode leakage current 3-6 times higher than M*2
MIMOSTAR 5 16/05/ st Run: MimoStar3 Test MimoStar3 with 14 µm EPI substrate: Seems less noise Ex. ENC ~ 18 e 20 °C, t r.o. = 4 ms Confirmed by static diode leakage current measurements: Diode leakage current 3-6 times higher than M*2 "Dead" pixel zone for all tested chips: yield? 7 chips tested on PCB 25 & 50 MHz )
MIMOSTAR 6 16/05/2007 Wafer Map (2 nd Run) Diode Leakage Current: M*2 as ref Static I Leak < 5 fA/diode M*3 I Leak should be < 1 nA for a pixel array of 320x640 Even in the centre of wafer, I Leak = 2 x I Leak of M*2 Wafer 20 µm EPI Wafer 14 µm EPI
MIMOSTAR 7 16/05/2007 MimoStar3 (1 st & 2 nd Run) Problems: High diode leakage current Dead pixel zone (1st Run) Is there any relationship between these problems? Why? no explanation for the moment Notice! Only radtol pixel on M*3 M18 on same reticule use standard diodes 10 chips (14&20 µm) tested without yield & leakage current problems Next steps: More extensive tests on the wafer & PCB for all chips of reticule Establish contact with AMS, propose design/techno review Is there any impact of pixel geometries on the yield for ~reticule size?
Ultimate MIMO
MIMOSTAR 9 16/05/2007 Ultimate Chip: MimoStarXXL Geometries: Pitch: 18.4 µm radiation hard consideration Pixel Array: 1024 x 1088 Active surface: x mm 2 ~ 1.1 x 10 6 pixels / frame Read out speed: t R.O. : 200 µs / frame 200 ns /line >~ 5 x 10 9 pixels / s Need zero suppression system According to STAR analysis Hit density rate (inner barrel) 2.4x10 5 /cm 2 /s < 200 hits/frame (200 µs) ~ 2000 pixels / frame Pixel Data: 21 bits (addresses) 1-3 bits (ADC) bits / pixel Effective data stream: ~ 240 MBit/s
MIMOSTAR 10 16/05/2007 MimoStarXXL: Proposed Architecture Pixel array: 1024x1088 (18.4 µm) Pixel with CDS Readout: row by row Matrix divided in 17 groups ADC (1 bit) at the bottom of each column Find N hits for each group Find M hits for each row (with N & M determined by pixel array occupancy rate) Memory which stores M hits (2 frames) + Serial transmission
MIMOSTAR 11 16/05/2007 MimoStarXXL: Status Radtol Pixel with integrated CDS: Functionality tested in M8, M16 18.4 pitch pixels will be implemented in M22 Submission Sept. - Oct Test at beginning of the next year 1088 discriminators: Tested in a pitch of 25 µm in M8, M16 Implemented in M22 Zero suppression circuit: Prototype Suze-01 will be submitted at end of June 2007 Test from September
MIMOSTAR 12 16/05/2007 MimoStarXXL: Power Management Using trigger To reduce mean power consumption To decorrelate acquisition and readout Proposition Pixel Pitch (µm) PixelDiscri.Group Hits Finder Line Hits Finder (mW) Sq. controller (Pix+Dig) Memory (mW) LVDS receiver (mW) LVDS trans. (mW) Total (mW) Power (mW/cm 2 ) mW 0.2 x mW 0.3 x mW 15 x mW 0.2 x mW 0.3 x mW 15 x