TLD POSTAL DOSE QUALITY AUDIT FOR 6MV AND 15MV PHOTON BEAMS IN RADIOTHERAPY CLINICAL PRACTICE Sonja Petkovska 1, Margarita Ginovska 2, Hristina Spasevska.

Slides:



Advertisements
Similar presentations
بسم الله الرحمن الرحيم ﴿و قل رب زدنى علماً﴾ صدق الله العظيم.
Advertisements

IAEA Quality Audits in Radiotherapy
Ionization Chamber Array for External Beam Radiotherapy
RapidArc plan verification using ArcCHECK™
RADIOTHERAPY ACCIDENT IN COSTA RICA - CAUSE AND PREVENTION OF RADIATION ACCIDENTS IN HOSPITALS Module XIX.
Esteban Fabián Boggio 1 Lucas Provenzano 2 Sara Gonzalez 2,3 Sara Gonzalez 2,3 Juan Manuel Longhino 1 1 Bariloche Atomic Center, Atomic Energy National.
Energy deposition and neutron background studies for a low energy proton therapy facility Roxana Rata*, Roger Barlow* * International Institute for Accelerator.
Maria Skłodowska-Curie Memorial Cancer Centre and Institute of Oncology, Warsaw, Poland 8th ECMP, Athens, Dosimetry audits in radiotherapy.
Evaluation of the characteristics of TLD LiF:Mg.Ti-100 Powder: A Measure of Consistency Between Multiple Batches of Powder Paola Alvarez,Jose Francisco.
Radiation Therapy (RT). What is cancer? Failure of the mechanisms that control growth and proliferation of the cells Uncontrolled (often rapid) growth.
Quality Assurance: Manufacturer & Clinical Aspects  Alan Cohen, M.S. DABR  Paul Naine, MSc. MIPEM  Jim Schewe, PhD, DABMP Accuray Incorporated Elekta.
IAEA International Atomic Energy Agency IAEA quality audits in radiotherapy Joanna Izewska Head, Dosimetry Laboratory Dosimetry & Medical Radiation Physics.
IAEA International Atomic Energy Agency PREVENTION OF ACCIDENTAL EXPOSURE IN RADIOTHERAPY Part 3: Analysis of causes and contributing factors IAEA Training.
Radiation Protection in Radiotherapy
Clinical application of TCP ・ NTCP. Lyman JT. Complication probability as assessed from dosevolume histograms. Radiat Res Suppl 1985; 8:S13–S19. Because.
Dose Distribution and Scatter Analysis
Dosimetric evaluation of a new design MOSFET detector Per H. Halvorsen* & Stephanie Parker University of North Carolina.
Results The measured-to-predicted dose ratio criteria used by the RPC to credential institutions is , however for this work, a criteria of
11. – , Athens 8th European Conference on Medical Physics DOSIMETRY AUDITS IN RADIOTHERAPY IN THE CZECH REPUBLIC Irena Koniarová Daniela Ekendahl.
TWO FIELD BREAST PLAN VS. OPTIMIZED CONFORMAL BREAST PLAN: COMPARISON OF PLAN PARAMETERS Authors: Borko Basarić, Ozren Čudić, Milan Teodorović, Borislava.
Shannon M MacDonald 1, Salahuddin Ahmad 2, Stefanos Kachris 3, Betty J Vogds 2, Melissa DeRouen 3, Alicia E Gitttleman 3, Keith DeWyngaert 3, Maria T Vlachaki.
Evaluation of New Pre-Treatment In-Air Patient Specific QA Software for TomoTherapy Treatments Lydia L. Handsfield¹, Quan Chen¹, Kai Ding¹, Wendel Renner²,
Measurement of Dose to Critical Structures Surrounding the Prostate from Intensity-Modulated Radiation Therapy (IMRT) and Three Dimensional Conformal Radiation.
Test of the proposed method Introduction CCD Controller CCD Illuminator gel Filter 585nm Assembling the phantom before its irradiation. The phantom, ready.
AUTHORS (ALL): Huang, Xiaoyan 1, 2 ; Kuan, K M 2 ; Xiao, G L 2 ; Tsao, S Y 3, 2 ; Qiu, X B 2 ; Ng, K 2. INSTITUTIONS (ALL): 1. Radiation Oncology, Sun.
IAEA International Atomic Energy Agency PREVENTION OF ACCIDENTAL EXPOSURE IN RADIOTHERAPY Part 5: Reporting, investigating and preventing accidental exposures.
In vivo dosimetry Eirik Malinen Eva Stabell Bergstrand Dag Rune Olsen.
Comparison of Clinical Parameters for Proton Therapy in the United States Paige Summers, MS.
The RPC Proton Therapy Approval Process
FDA Public Meeting: Device Improvements to Reduce the Number of Under-doses and Misaligned Exposures from Therapeutic Radiation.
Application of a 2-D ionization chamber array for dose verification of dynamic IMRT with a micro-MLC Fujio ARAKI, PhD 1, S. TAJIRI 2, H. TOMINAGA 2, K.
Institute for Advanced Radiation Oncology
Commissioning of an Optically-Stimulated Luminescence (OSL) system for remote dosimetry audits J.F. Aguirre, P. Alvarez, C. Amador, A. Tailor, D. Followill,
Medical Accelerator F. Foppiano, M.G. Pia, M. Piergentili
Radiological Physics Center David Followill, Ph.D. and RPC Staff.
Integrating the Health Care Enterprise- Radiation Oncology Use Case: In Vivo Patient Dosimetry Editor: Juan Carlos Celi - IBA Reviewer: Zheng Chang – Duke.
Đ. Milković, M. Ranogajec-Komor, S. Miljanić, Ž. Knežević and K
Identification of a 3D dosimeter best-suited for use by the RPC M. Heard, G. Ibbott, D. Followill, R. White, E. Jackson, M. Salehpour.
Investigation of 3D Dosimetry for an Anthropomorphic Spine Phantom R. Grant 1,2, G. Ibbott 1, J. Yang 1, J. Adamovics 3, D Followill 1 (1)M.D. Anderson.
TPS & Simulations within PARTNER D. Bertrand, D. Prieels Valencia, SPAIN 19 JUNE 2009.
Introduction The Radiological Physics Center (RPC) anthropomorphic quality assurance (QA) phantom program is one tool the RPC uses to remotely audit institutions.
Analysis by Monte-Carlo simulations of the characteristics of nano and micro dosimeters for real time measurements in radiotherapy and medical physics.
Institute of Nuclear Engineering and Science National Tsing Hua University Institute of Nuclear Engineering and Science National Tsing Hua University BNCT.
Thermoluminescent dosimetry at the IFJ Krakow P. Bilski, B. Obryk INSTITUTE OF NUCLEAR PHYSICS (IFJ), KRAKÓW, POLAND RADMON, 15 February 2006.
Introduction to Radiation Therapy
IAEA Regional Meeting June 07 National Workshop June 07 Expert Visit to Facilities 29 June-1 July 07 Orientation 15 July 07 Competency Training.
Purpose N-isopropylacrylamide (NIPAM) polymer gel dosimeters were employed to verify the dose distribution of clinical intensity modulated radiation therapy.
Reducing excess imaging dose to cancer patients receiving radiotherapy Adam Schwertner, Justin Guan, Xiaofei Ying, Darrin Pelland, Ann Morris, Ryan Flynn.
Three-dimensional Dose Verification Using Normoxic Polymer Gel Dosimeters for Tomotherapy Tung-Hsin Wu 1, Chien-Yi Hsiao 2, Mu-Bai Chang 2, Geoffrey Zhang.
Commissioning of a commercial treatment planning system for IMAT and Dose Painting treatment delivery. G. Pittomvils 1,,L. Paelinck 1, F. Crop 2, W. De.
The Effects of Small Field Dosimetry on the Biological Models Used In Evaluating IMRT Dose Distributions Gene Cardarelli,PhD, MPH.
Rapid Arc Treatment Verification: post evaluation on Delta-4 and proposal of a new verification protocol G. Pittomvils 1,,L. Paelinck 1, T. Boterberg 1,
Adapting A Clinical Medical Accelerator For Primary Standard Dosimetry
Radiological Protection in Ion Beam Radiotherapy
CHAPTER 3 DOSE DETERMINATION FOR EXTERNAL BEAMS
IAEA E-learning Program
EURADOS Working Group 9: Radiation Dosimetry in Radiotherapy
Very High Energy Electron for Radiotherapy Studies
Introduction Materials & Methods Results Conclusion
Linac Commissioning Overview
Left Posterior Superior Right Anterior Inferior
A. Nisbet 1,2, A. Dimitriadis 1,2,3, A.L. Palmer 1,4, C.H. Clark 2,3
أجهزة العلاج الإشعاعي Clinical Radiation Generator
Fig. 4. Percentage of passing rate between clinical and 544 plans.
Median Volume (cc) of GTV Receiving Dose
Ch 10. A System of Dosimetric Calculations
Hot and cold spots are common problems associated with planning:
GHG meeting at ESTRO36 May, 2017
Average Dose-Volume Ratio
Presentation transcript:

TLD POSTAL DOSE QUALITY AUDIT FOR 6MV AND 15MV PHOTON BEAMS IN RADIOTHERAPY CLINICAL PRACTICE Sonja Petkovska 1, Margarita Ginovska 2, Hristina Spasevska 2, Nadica Dimitrovska 1, 1 University Clinic for Radiotherapy and Oncology, Skopje, RM 2 Faculty of Electrical Engineering and Information Technologies, Ss Ciril and Metodious University, Skopje, RM RAD 2012, April, Nish

Radiation therapy is the modality of choice for about half of the cancer patients at some stage during their treatment. Radiotherapy aim:  delivering as high dose as possible at the tumor volume  spearing the health tissue as much as possible. INTRODUCTION RAD 2012, April, Nish

INTRODUCTION  Delivered dose should provide the highest probability to cure the patient without inducing severe complications in normal tissues.  Delivered dose should be equal to prescribed one TCP – Tumor Control Probability NTCP – Normal Tissue Complication probability TSP – Probability for Succesfful Treatment RAD 2012, April, Nish

INTRODUCTION QA programme  Mechanical checks provide geometrical accuracy  Dosimetry checks provide dose distribution accuracy  Beam calibrations provide prescribed dose accuracy RAD 2012, April, Nish

INTRODUCTION External comparison with an independent dosimetry laboratory  Check the dose delivered by radiotherapy unit  Comparison in different beam parameters in clinical practice  Allows detecting dosimetric errors RAD 2012, April, Nish

ESTRO QUALity Assurance network (EQUAL) Compare beam parameters measured in the same conditions in different radiotherapy centers RAD 2012, April, Nish 2Gy

Thermoluminescent Dosimeters (TLDs) with lithium fluoride powder (LiF) MATERIALS AND MEASUREMENTS RAD 2012, April, Nish TLD holder Water phantom

MATERIALS AND MEASUREMENTS RAD 2012, April, Nish 4 TLDs per unit are irradiated one by one 2 for 6MV photon beam 2 for 15MV photon beam geometry setup for TLD irradiation

MATERIALS AND MEASUREMENTS RAD 2012, April, Nish D = 2Gy at reference depth 6MV photon beam299MU 15MV photon beam251MU

MATERIALS AND MEASUREMENTS RAD 2012, April, Nish Dcalc is TPS calculated dose Dm is measured dose with ionisation chamber beam (MV) Dcalc (Gy)Dm (Gy)  (%)

RESULTS RAD 2012, April, Nish The levels of deviation between measured and stated quantities (Qm/Qs) and the corresponding EQUAL actions are specified as follows: Qm/Qs < 3% optimal level 3% < Qm/Qs < 5% tolerance level 5% < Qm/Qs < 10% a level outside tolerance level Qm/Qs > 10% an emergency level

RESULTS RAD 2012, April, Nish beam (MV)  min (%)  max (%) Deviation for high energy photon beam s

CONCLUSION This evaluation shows significant absorbed dose reproducibility which confirmed the capability of the linear accelerator to deliver real treatments with an acceptable stability. RAD 2012, April, Nish The results from ML show that the relative deviation Qm/Qs for these analyzed 6MV and 15MV photon beams, do not exceed 3%.

Dose accuracy is just one segment of external quality procedures… CONCLUSION The purpose of this TLD audit is to check the dose delivered by radiotherapy unit, so this independent and trustworthy verification of dose delivery improve internal QA and help to detect some dosimetric errors. RAD 2012, April, Nish