Fluorometric determination of riboflavin

Slides:



Advertisements
Similar presentations
An introduction to Ultraviolet/Visible Absorption Spectroscopy
Advertisements

Raman Spectroscopy A) Introduction IR Raman
Atomic Absorption Spectroscopy (AAS) The Visible Spectra.
Molecular Fluorescence Spectroscopy
1 Molecular Luminescence Spectroscopy Lecture 31.
Molecular Luminescence Spectrometry
Atomic Absorption Spectroscopy (AAS)
Spectroscopy Chapter 7.
Chapter 15 Molecular Luminescence Spectrometry Molecular Fluorescence  Optical emission from molecules that have been excited to higher energy levels.
Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry Absorption measurements based upon ultraviolet and visible radiation.
Spectroscopy (continued) Last time we discussed what spectroscopy was, and how we could use the interaction of light with atoms and molecules to measure.
Biochemical instrumental analysis-2
Introduction to Instrumental Analysis - Spectrophotometry
COLORIMETRYCOLORIMETRY Prepared By Michigan Department of Environmental Quality Operator Training and Certification Unit Note: A printed description of.
Fluorescence Spectroscopy 1Dr. Nikhat Siddiqi. Principles Interaction of photons with molecules results in promotion of valence electrons from ground.
DETERMINATION OF THE RATE OF AN ELECTRON TRANSFER REACTION BY FLUORESCENCE SPECTROSCOPY Presenter:Sandor Kadar, Ph.D.
Colorimetry & Spectrophotometry.
SPECTROPHOTOMETRY. Determines concentration of a substance in solution by Measures light absorbed by solution at a specific wavelength by using spectrophotometer.
Fluorometric determination of riboflavin
CHAPTER 15: MOLECULAR LUMINESCENCE. Chapter LUMINESCENCE TECHNIQUES Emission of light is used to determine certain properties,e e.g.structure and.
Fluorometric determination of riboflavin. Introduction.
 Absorbs EM radiation from a source, like a D 2 and Tungsten lamp  Electrons relax back to ground state.
Spectroscopy Light in chemistry. The Nature of Light Acts as both a particle and a wave Photoelectric effect only explained through waves Interference,
Fluorometric Analysis
Chapter 15 Molecular Luminescence Spectrometry Three types of Luminescence methods are: (i) molecular fluorescence (ii) phosphorescence (iii) chemiluminescence.
1. Principles  Interaction of photons with molecules results in promotion of valence electrons from ground state orbitals to high energy levels.  The.
Fluorescence Spectroscopy
Spectrophotometry.
Spectrofuorometer lab. 1. Fluorescence spectroscopy or (spectrofluorometry) Is a type of electromagnetic spectroscopy which analyzes fluorescence from.
Asa Arjoon U6 Chemistry Presentation on Visible and Ultraviolet Spectroscopy Name : Form : Subject :
Fluorimetric Analysis. Absorption of energy Emission of energy Spectroscopy.
Spectrophotometer.
SPECTROPHOTOMETRY. Spectrophotometry Determines concentration of a substance in solution –Measures light absorbed by solution at a specific wavelength.
SPECTROPHOTOMETRY. Principle : there is interaction between the light and sample particle, spectrophotometer is employed to measure the amount of light.
MOLECULAR FLUORESCENCE SPECTROSCOPY. Fluorescence is a form of photoluminescence; and this later is a type of luminescence that occurs when certain molecules.
CLS 332 CLINICAL INSTRUMENTAL ANALYSIS. A VISIBLE ABSORPTION SPECTROMETER.
ATOMIC ABSORPTION SPECTROSCOPY (AAS) Atomization: It is the conversion of molecules to their component atoms in gaseous state using a source of heat (flame).
Determination of Concentration Using Spectrophotometry
Infrared Spectroscopy
SPECTROPHOTOMETRY PRACTICAL 213 PHC INSTUMENTAL ANALYSIS.
Summary: (Last lecture) Absorption spectroscopy definition electromagnetic spectroscopy matter absorption spectroscopy fundamental terms (transmittance,
Atomic-absorption spectroscopy Lab3 Atomic-absorption spectroscopy.
Atomic Absorption Spectroscopy.  Qualitative test for metals in solution  Can also be used for coloured lights  This emission of light is called an.
Instrumentation. Chromatography Chromatography is the collective term for a set of laboratory techniques for the separation of mixtures. It involves.
COLORIMETRY & SPECTROPHOTOMETR
Absorption spectrum and spectrophotometric determination of concentration Experiment 1.
Molecular Fluorescence Spectroscopy
Lab1 A VISIBLE ABSORPTION SPECTROMETER. -One of the simplest and most widely used methods to determine concentration of a substance in solution -Measures.
Spectrophotometry Measuring Concentrations of Substances in Body Fluids.
SPECTROFLUORIMETRY Prepared By: HIMANSHU SHARA 2010B2A2196P.
INSTRUMENTAL METHODS OF ANALYSIS (CHM 303)
UV/VIS SPECTROSCOPY.
Lab1 A VISIBLE ABSORPTION SPECTROMETER
Chem. 133 – 3/14 Lecture.
Beer-Lambert law Gihan Gawish.Dr.
Principles and practice of Spectrophotometer
Spectroscopy Chapter 7.
Spectrophotometer Dr . S. Jayakumar.
Atomic Absorption and Emission
Light and Matter Main Concept:
SPECTROPHOTOMETRY Applied Chemistry.
Fluorescence.
Molecular Luminescence Spectrometry
Spectroscopy Uses emission and absorption of light by electrons moving between ground and excited state configuration, hence electronic configuration.
Colourimetry The absorption of light by a coloured solution is directly related to the concentration of the solution. Colourimetry determines the concentration.
Spectrophotometry A method to determine concentration of a species exploiting the absorption of EMR.
FLAME SPECTROSCOPY The concentration of an element in a solution is determined by measuring the absorption, emission or fluorescence of electromagnetic.
Clinical instrumental analysis
Spectrophotometric Analysis
Presentation transcript:

Fluorometric determination of riboflavin Lab 5 Fluorometric determination of riboflavin

Fluorescence Fluorescence is a kind of a luminescence, which is the emission of photons from electronically excited states. Fluorescence occurs when the electron is transferred from a lower energy state into an "excited" higher energy state to form fluorescence.

The electron will remain in this state for 10⁻⁸ sec The electron will remain in this state for 10⁻⁸ sec. then the electron returns to the lower energy state and it releases the energy in form of fluorescence.

Fluorescence -In ultraviolet absorption spectroscopy when molecule absorbs UV radiation at one wavelength and its immediately re-emission, usually in a longer wavelength. -Some molecules fluoresce naturally and others can be modified to make fluorescent compounds. 

First excited state with four vibrational energy levels excitation fluorescence Ground state with four Vibrational energy levels

Fluorescent compounds have two characteristic spectra: Fluorescent compounds have two characteristic spectra: *an excitation spectrum (the wavelength and amount of light absorbed) and an emission spectrum (the wavelength and amount of light emitted). These spectra are often referred as a compound's fluorescence signature or fingerprint. -No two compounds have the same fluorescence signature. This principle makes fluorometry a highly specific analytical technique.

*What is fluorometery? Fluorometry is the measurement of fluorescence. It’s measured by a fluorometer or fluorimeter. ,

A fluorometer involves using a beam of light, usually ultraviolet light. A fluorometer generates the wavelength of light required to excite the analyte of interest. it selectively transmits the wavelength of light emitted, then it measures the intensity of the emitted light. The emitted light is proportional to the concentration of the analyte being measured .

Fluorescence compounds: 1- aromatic and heterocyclic compounds. 2- compounds with multiple conjugated groups. 3- compounds containing electron donating groups as OH, NH2 ,OCH3... 4- poly cyclic compounds like vit K, purines, nucleosides, vit A .

5- NADH fluorescence. 6- non fluorescence compounds when converts to fluorescent derivatives like: Steroids Metals by chelating. Antibodies.

Fluorescence spectra Fluorescence is used primarily for quantitative analysis in ppm (part per millions) . F=Kфp₀ (2.3 abc) Where F is fluorescence intensity K instrument constant Ф is the quantum efficiency p₀ intensity of excitation a molar absorptive, b cell path, c molar concentration,so F=K.c

Advantages of fluorometer: - Very specific - Very sensitive Advantages of fluorometer: - Very specific - Very sensitive. - Wide Concentration Range - Simplicity and Speed - Low Cost

Disadvantages: *The fluorescence is very sensitive to environmental changes which include PH, temperature, solvent contamination and UV light used for excitation can make photochemical change

Clinical uses of instrument Uses to measure fluorescence for qualitative and quantitative analysis Used to measure any substance that exists in small conc. for example: vitamins from blood sample.

Instrumentation 1-Light Source: The lamp or light source provides the energy that excites the compound by emitting light. Light sources include xenon lamps, high pressure mercury vapor lamps.

2-Excitation Filter /monochromator: The excitation filter is used to screen out the wavelengths of unabsorbed light by the compound being measured.

3-Sample Cell/Cuvette: 3-Sample Cell/Cuvette: *the cuvette material must allow the compound's absorption and emission light energy to pass through. *the size of cuvette affects the measurement. The greater the pathlength (or diameter) of the cell, the lower the concentration that can be read. Cuvettes are made from borosilicate or quartz glass.

4-Emission Filter(monochromator ): Stray light emitted from the sample.

5-Light Detector: The light detector is most often a photomultiplier tube, so photodiodes are increasingly being used. The light passing through the emission filter is detected by the photomultiplier or photodiode. The light intensity, which is directly proportional (linear) to the compound's concentration, is registered as a digital readout.

Schemating drawing

Precautions -pH -Cleanliness of glassware. -High standards of experimental technologist are necessary to prevent quenching. -Temperature. -Cuvette Size (Pathlength)

Riboflavin Experiment: Riboflavin (vitamin B2) is strongly fluorescent in 5% acetic acid solution.

Chemical and solutions required 1- 5 % Acetic acid: The addition of a few drops of glacial acetic acid to the solution will insure an acid pH and help to stabilize it. 2- Riboflavin stock solution . 3- Riboflavin standards solution 10( ppm) 4- Unknown.

Preparation of standards Blank Std .1 Std.2 Std.3 Std.4 Std.5 Riboflavin stoc.std(10 ppm) ............. 0.2 ml 0.4 ml 0.6 ml 0.8 1 ml 5% acetic acid 9.8 ml 9.6 ml 9.4 ml 9 ml

Procedure 1- switch on the machine. 2- install the excitation and emission filtered in their holders. 3- start lamp (UV) by pressing 3-4 sec. the start switch and release it. 4- leave to warm up ,15 mins. 5- open the door of the instrument which acts as sample holder.

6- adjust the range select or at x1 (excitation filter) for less conc 6- adjust the range select or at x1 (excitation filter) for less conc. sample (ppm) and the higher range for higher conc. sample. 7- Put reagent blank cuvette in the sample holder, close the door and with the help of blank knob set read out to zero. 8- Remove reagent blank and insert the std. highest conc. And with help of attenuator knob set the read out to 100 and record this reading.

9- Check twice for zeroing and 100 by inserting appropriate samples 9- Check twice for zeroing and 100 by inserting appropriate samples . If this not ok select higher range . 10- Take the reading for other std. and given unknown .

blank Std. 1 std 2 3 4 5 Conc. In ppm volume 0.2 0.4 0.6 0.8 1.0 reading 15.5 24. 7 52.9 92.4 100

F 11- plot the graph and from the graph find out the conc. of unknown. 5 CONC.