Installation of the ACS Guido De Marchi – TIPS meeting – 21.3.2002 Guido De Marchi - ESA/STScI - ACS Group The ACS was installed on board the HST on the.

Slides:



Advertisements
Similar presentations
Theory of Special Relativity
Advertisements

1 ACS Status Update STScI TIPS Meeting 19-October-2006 Ken Sembach.
Application of the SIDECAR ASIC as the Detector Controller for ACS and the JWST Near-IR Instruments Markus Loose STScI Calibration Workshop July 22, 2010.
SDW Marco Sirianni Marco Sirianni (ESA/STScI) Max Mutchler (STSci) Radiation Damage in HST Detectors.
Following the Photons… Empirical, Pixel-Based Corrections for CTE Jay Anderson STScI October 12, 2011 Back-Tracking the Electrons.
Echelle spectra reduction with IRAF*
Timepix Studies: Medipix Collaboration Summary and More Timewalk Plots Alessandra Borgia Marina Artuso Syracuse University Group Meeting – Thursday 20.
CCDs. CCDs—the good (+)  Linear response  photometry is “simple” +High efficiency, compared to other detectors +Sensitive to many wavelengths +2-D arrays.
XMM-Newton 1 M.Stuhlinger/P.Calderon, ESAC Re-appearance of the MOS1 CCD1 meteorite column during revs Madrid, 23. March 2010.
SDW20051 Vincent Lapeyrère LESIA – Observatoire de Paris Calibration of flight model CCDs for CoRoT mission.
Signal vs Noise: Image Calibration First… some terminology:  Light Frame: The individual pictures you take of your target.  Dark Frame: An image taken.
We have the first, direct measure of photometric loss due to imperfect CTE on ACS.
1 TIPS 2011 May Persistence in the WFC3 IR detector Knox S. Long.
AST3 detector properties
Pipeline calibrations of ACS data Max Mutchler Hubble Space Telescope Calibration Workshop October 2005.
1 ACS Side-2 Performance Update STScI TIPS Meeting 20-July-2006 Ken Sembach.
NGC 6217 in DSI mode with F658N Mosaic before new superbias why we don’t release images until after SMOV! ACS CCD monitoring and pipeline calibration review.
Printed by ACS 2 Gyro Mode Data Analysis Cheryl Pavlovsky, Marco Sirianni, Ken Sembach, ACS Instrument Team and the 2 Gyro Mode Team.
Understanding Persistence: A 3D Trap Map of an H2RG Imaging Sensor
CCDs in space: the effects of radiation on Hubble’s Advanced Camera for Surveys (ACS) Max Mutchler, David Golimowski (Space Telescope Science Institute),
The B-V colors and photometric variability of Nix and Hydra, Pluto’s two small satellites Max Mutchler (STScI) S. Alan Stern (SwRI) Hal Weaver (JHU/APL)
SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA TIPS / COS TV Update 21 December 2006 Agenda Agenda Testing Setup and Status Testing Setup.
Page 1. Page 2 SMOV Status SMOV is proceeding on-track Key programs executed –ERO program –Geometric distortion program completed –ACS to FGS alignment.
CCD Detectors CCD=“charge coupled device” Readout method:
NIRSpec Operations Concept Michael Regan(STScI), Jeff Valenti (STScI) Wolfram Freduling(ECF), Harald Kuntschner(ECF), Robert Fosbury (ECF)
May 21, 2009John MacKenty, STScI1 WFC3 SM4 Events and Current Status.
10/26/20151 Observational Astrophysics I Astronomical detectors Kitchin pp
ACS Drizzling Overview J. Mack; DA Training 10/5/07 Distortion Dither Strategies MultiDrizzle ‘Fine-tuning’ Data Quality Photometry.
TIPS/JIM 20 Aug ACS Status Update Norman Grogin (TIPS/JIM 20 Aug 2009) WFC Post-SM4 Performance Recap Day 174 Suspend : Determinations and Corrective.
D. Golimowski2010 STScI Calibration Workshop1 ACS after SM4: New Life for an Old Workhorse David Golimowski Space Telescope Science Institute 22 July 2010.
1 Black and White Photography. 2 3 Film Rewind This allows the exposed 35mm film to be rewound back into its cartridge. Lens The picture is formed and.
ACS STATUS Linda Smith (thanks to David Golimowski for providing the slides)
NICMOS Status Roelof de Jong (STScI) and the NICMOS team: Santiago Arribas, Elizabeth Barker, Eddie Bergeron, Ilana Dashevsky, Anton Koekemoer, Sangeeta.
TIPS May 22, 20081H. Bushouse. TIPS May 22, 20082H. Bushouse TV3 Science Team Cast of “Characters” Instrument Scientists: –Sylvia Baggett –Howard Bond.
Observational Astrophysics I
The B-V colors and photometric variability of Nix and Hydra, Pluto’s two small satellites Max Mutchler (STScI) S. Alan Stern (SwRI) Hal Weaver (JHU/APL)
UPDATED CTE CORRECTION FORMULAE FOR ACS Marco Chiaberge Pey Lian Lim, Vera Kozhurina-Platais, Marco Sirianni Ron Gilliland, Jennifer Mack.
Paul Derwent 14 Dec 00 1 Stochastic Cooling Paul Derwent 14 Dec 00
Torsten Böker TIPS May 2, 2002 NICMOS post-cooldown detector performance 1) SMOV status 2) Dark current 3) Quantum efficiency 4) “First light”, focus and.
START NICMOS COOLING SYSTEM ACTIVITY SUMMARY. 1/25/01Ken Pulkkinen2 START NCS ACTIVITY Applicable SMOV Requirements: J NICMOS Cooling System.
The WFPC2 WF4 Anomaly John Biretta 15 Dec Overview Discovery of the Anomaly Historical Evolution Imaging Impacts GO Science Impacts Image Repair.
"Data contained herein is exempt from ITAR regulations under CFR 125.4(13) -- data approved for public disclosure." 19 Sept, 2002TIPS – WFC3 Status 1 -UVIS.
TIPS - Oct 13, 2005 M. Sirianni Temperature change for ACS CCDs: initial study on scientific performance M. Sirianni, T. Wheeler, C.Cox, M. Mutchler, A.
SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA SMOV4 Requirements Review Cosmic Origins Spectrograph Scott D. Friedman STScI 30 July 2003.
NICMOS Calibration Challenges in the Ultra Deep Field Rodger Thompson Steward Observatory University of Arizona.
The Second Generation of IR detectors for WFC3 Massimo Robberto European Space Agency and Space Telescope Science Institute.
1 WFC3 Science Data Volume  Data Volume Estimates for WFC3 Operations ISR WFC ; Lisse, Henry, Knezek, Hanley; 3/27/2001 ISR WFC ; Lisse,
14 January Observational Astronomy SPECTROSCOPIC data reduction Piskunov & Valenti 2002, A&A 385, 1095.
WFC3 SMOV Report for TIPS 16 July 2009 John MacKenty This presentation contains material that is embargoed until after the SM4 ERO press conference. Images/spectra.
Introduction NICMOS (Near-Infrared Camera and Multi-Object Spectroscopy) was installed on the Hubble Space Telescope during SM2 in 1997 and has been the.
STIS Status TIPS July 16, STIS Status STIS MIE anomaly –What is known STIS CCD status before the anomaly STIS MAMA SMOV status.
February 21, 2002TIPS meeting1 "Data contained herein is exempt from ITAR regulations under CFR 125.4(13) -- data approved for public disclosure." TIPS.
STIS STATUS Charles Proffitt TIPs May 21, STIS Status - Status of Repair Results of FT Amp B Bias Anomaly Changes to SMOV Plan.
Image Stability  ACS SMOV image stability test (prop 9017) went from a cold soak (anti-solar pointing) for 8 orbits to a hot attitude (including off-nominal.
ACS WFC Flat-Field Changes Temperature change from -77 C to -81 C on July 4, 2006 leads to expected changes for flat fields. Are L-flat measures stable.
Marco Sirianni TIPS 12/18/2003 Anomalous scattered light in ACS/WFC Data Marco Sirianni D. Van Orsow, A. Welty, T. Wheeler R. Gilliland, R. van der Marel.
STIS Status TIPS, September 17, 2009 Charles Proffitt for the STIS team.
SMOVSMOV  ACS has now completed SMOV Program –Transitioned to interim calibration plan  Final SMOV programs –Further coronagraph calibration –Coronagraph.
HST Cal Conf -- Oct 27, Calibration Status and Results for Wide Field Camera 3 – R. Kimble/GSFC, Calibration Status and Results.
Application of a Charge Transfer Model to Space Telescope Data Paul Bristow Dec’03
CCD Calibrations Eliminating noise and other sources of error.
Announcements After a short lecture we will adjourn to the Farm. Tonight is a Dark Sky make-up night. Class will meet Thursday night. If clear: meet at.
Relative Spectral Response and Flat Fields with Internal Calibration Lamps Luisa M. Lara IAA-CSIC Granada (SPAIN)
Charge Transfer Efficiency of Charge Coupled Device
Update on Noise Analysis of the DHCal Test Beam at Fermilab
New static DQ masks for NICMOS
Photometric Analysis of Asteroids
Announcements HR Diagram lab will be extended for one week. I’ll talk about it today. Homework: Chapter 9 # 1, 2 & 3 Next week is a Dark Sky Night. If.
ACS PSF Variations Over The Field of View & ACS Coronagraph Status
Detecting cosmic rays using CMOS sensors in consumer devices
Presentation transcript:

Installation of the ACS Guido De Marchi – TIPS meeting – Guido De Marchi - ESA/STScI - ACS Group The ACS was installed on board the HST on the 6th of March replacing the ageing FOC No problems encountered by astronauts during installation Aliveness test proved that the camera can boot properly Functional test run soon after to verify ability to provide science grade data

ACS functional test Guido De Marchi – TIPS meeting – After switch on, both CCD cooled rapidly to -67 C Both filter wheels moved properly to position A set of bias and dark frames was taken for both CCD with several combinations of amplifiers and lamps The ACS IDT and ACS group at STScI analysed the data simultaneously but independently: same results!

ACS functional test: bias frames Guido De Marchi – TIPS meeting – Read noise found to be at nominal values: WFC: ~5 e- rms; HRC: ~4 e- rms Fourier analysis reveals no correlated noise Large number of cosmic ray hits seen in bias frames, non uniformly distributed (gradient). Reason: long read out time! One might benefit from CR-SPLIT for exposures shorter than a few minutes [Sirianni] Cosmic ray events show no tails (i.e. no deferred charge) thus proving good CTE [Riess]

ACS functional test: flat frames Guido De Marchi – TIPS meeting – No cosmetic degradation apparent (WFC, HRC), no new large dust particles nor shifting of existing ones [Riess] Illumination levels consistently higher (~20%) than on the ground for all lamps. Could be due to higher operating temperature in vacuum. Planned internal flat fields will need shorter exposure times [Hartig] Shadow cast by coronographic finger suggests a shift occurred during flight. Similar shifts seen when moving the camera on the ground [Hartig] Cycle HRC/SBC fold mirror to ensure nominal position

ACS SMOV programme Guido De Marchi – TIPS meeting – On-board memory tested for bit flips during SAA passa- ges: only 8 bits flipped in 6x34 million bytes [De Marchi] Both CCD (WFC, HRC) cooled rapidly to operating tem- peratures and remain stable there (–77 C, –80 C) [Cox] GSFC/Ball are reviewing the thermal models for opportu- nity to further lower operating temperatures Dark rate stable to within +/-1 at ~7.5 e-/px/hr, well below specifications (< 25 e-/px/hr) [Sirianni, Mutchler, Cox]

ACS SMOV programme: WFC bias Guido De Marchi – TIPS meeting – After CCD activation, bias level gracefully decays and stabilises at ~20 e- down after a few hours [Pavlovsky] Level of WFC amplifier B shows larger scatter, with variations of up to ~ +/- 10 e-, but RN still 5 e- [Sirianni] This is not a problem because the offset (i.e. the differen- ce between the bias level in the image and in the physi- cal overscan) is constant to within +/- 0.3 e- [Sirianni] Correct bias level can be determined from overscan