Shortwave and Microwave Diathermy Chapter 10. Diathermy Application of High-Frequency Electromagnetic Energy Used To Generate Heat In Body Tissues Heat.

Slides:



Advertisements
Similar presentations
Physiological Properties of Thermal Modalities (1)
Advertisements

The Science of Therapeutic Modalities
MAGNETIC EFFECTS OF ELECTRIC CURRENT
Shortwave and Microwave Diathermy
Supriya Babu. Topics Physical basis of heat and cold Measurement of temperature: Thermometry Mapping of body’s surface temperature: Thermography Heat.
Physics for Electricity
Shortwave Diathermy Prof. Dr. Gehan Mosaad.
SMAW Equipment (Shielded Metal Arc Welding- commonly called “stick welding”, or “arc welding”)
Shortwave and Microwave Diathermy part(2). Induction Electrodes (Drum Electrode) One Or More Monopolar Coils Rigidly Fixed In A Housing Unit May Use More.
Heat in Therapy Class 3 Supriya Babu.
Chapter 1: How are Therapeutic Modalities Related to One Another? The Basic Science Jennifer Doherty-Restrepo, MS, LAT, ATC FIU Entry-Level ATEP Therapeutic.
Cutnell/Johnson Physics 8th edition
Shortwave Diathermy ESAT 3640 Therapeutic Modalities.
By B. Nelson. Definition of IR radiation Radiation: the process of emitting energy in waves from a source Portion of the electromagnetic spectrum with.
K L University 1 By G.SUNITA DEPARTMENT OF PHYSICS.
1. NameRoll No Athar Baig10EL40 Muhammad Faheem10EL38 Tassawar Javed10EL44 Tayyaba Abbas10EL09 Sadia Imtiaz10EL37 2.
Prof.Dr. Gehan Mosaad. At the end of this lecture the student must be able to:  Define electrotherapy, electric current and electromagnetic spectrum.
Basic Principles of Electricity Jennifer L. Doherty-Restrepo, MS, LAT, ATC FIU Entry-Level ATEP PET 4995: Therapeutic Modalities.
Infrared Radiation Prof.Dr. Gehan Mosaad.
Electrotherapy & Its Instruments
NE 110 – Introduction to NDT & QA/QC
CHAPTER - 13 MAGNETIC EFFECTS OF ELECTRIC CURRENT
Chapter 7 Notes.
Chapter 29 Electromagnetic Induction and Faraday’s Law
Chapter 22 Alternating-Current Circuits and Machines.
AP Physics C Montwood High School R. Casao
Shortwave Diathermy Chapter 9.
Diathermy.
At the end of the lecture, the students should be able to:  Define diathermy  Identify the parts of the SWD/MWD machine  Describe how heat is produced.
Physiological Properties of Thermal Modalities (2)
Microwave Diathermy By Prof.Dr. Gehan Mosaad.
By Ass.Prof.Dr. Gehan Mosad
Electromagnetic Induction
Fall 2001ENGR201 Capacitance & Inductance1 Capacitor, also called electrical condenser, device for storing an electrical charge. In its simplest form a.
Electricity and Magnetism
INTRODUCTION Microwave diathermy الميكروويف إنفاذ حراري(MWD), is a form of electromagnetic radiations lying between shortwave and infrared waves. Where.
Shortwave Diathermy Chapter 9.
Therapeutic Ultrasound
Shortwave and Microwave Diathermy
© 2005 – FA Davis Shortwave Diathermy Clinical Application.
Chapter 29 Electromagnetic Induction and Faraday’s Law
Microwave Diathermy Mohammed Taher Ahmed Ph.D PT, Rehabilitation Health Science College of Applied Medical Science.
Shortwave.
Chapter 9: Waves and Light. Lesson 1: Waves of the Electromagnetic Spectrum Key Questions: – 1. How does the Sun’s energy arrive on Earth? – 2. How do.
Shortwave Diathermy Clinical Application. Treatment Dosages Pulse DoseSensationWidthRate (pps) NTNo warmth65 µsec100 – 200 1Mild warmth100 µsec800 2Moderate.
Therapeutic Modalities
Ultrasound 10cm 2 Ultrasound (US) Therapeutic ultrasound –Uses acoustic energy, delivered at high frequencies for therapeutic purposes Electrical current.
Ultrasound. History: Available in 19th century. Available in 19th century. Was for sonar (SONAR  Sound Navigation and Ranging) Was for sonar (SONAR 
Unit 8 Electricity and Magnetism. Page 9: Essential Question 1 What causes charged objects to push and pull on each other?
The Different Types of Inductors and Their Affecting Factors
Microwave Diathermy ELECTROTHERAPY
Cardiovascular instrumentation. Function of Electrodes The electrical activity of the nervous system detected by using electrodes which transfer the biochemical.
Microwave diathermy.
Chapter 21 Magnetic Induction and Chapter 22.9: Transformers.
TITEL OF PRESENTATION :- ANALOG MEASURING INSTRUMENTS PREPARATION NAME ENROLLMENT NO.:  MR.SAVALIYA SURESH T  MR. SAVALIYA VISHAL H
DIATHERMY Diathermy was once one of the most popular modalities used in rehab. Became less popular with development of more modern tech like US and because.
Physics 2 Stage show Learning outcomes
DIATHERMY MISS HIRA JABEEN.
Deep Heat Ultrasound & Diathermy
1435 – 1436H – 1ST SEMESTER - PHT 221 – SECTION –
Electric Charge.
SURGICAL DIATHERMY.
SURGICAL DIATHERMY.
Shortwave Diathermy By Ms.B. Nelson.
Diathermy Presented by: Hesamoddin Bagheri Supervisor: Dr. Salamat.
CHAPTER - 13 MAGNETIC EFFECTS OF ELECTRIC CURRENT
INDUCTORS, CAPACITORS AND ALTERNATING CURRENT
CHAPTER - 13 MAGNETIC EFFECTS OF ELECTRIC CURRENT
Presentation transcript:

Shortwave and Microwave Diathermy Chapter 10

Diathermy Application of High-Frequency Electromagnetic Energy Used To Generate Heat In Body Tissues Heat Produced By Resistance of Tissues Also Used For Non-Thermal Effects

Physiologic Responses To Diathermy Not Capable of Producing Depolarization and Contraction of Muscles –Wavelengths Too Short

Physiologic Responses To Diathermy Physiologic Effects Are Those of Heat In General –Tissue Temperature Increase –Increased Blood Flow (Vasodilation) –Increased Venous and Lymphatic Flow –Increased Metabolism –Changes In Physical Properties of Tissues –Muscle Relaxation –Analgesia

Diathermy Heating Doses Are Not Precisely Controlled Thus The Amount of Heating Cannot Be Accurately Measured –Basically means amount of heating patient receives cannot be directly measured Heating= Current 2 X Resistance

Non-Thermal Effects Pulsed SWD Used To Treat Soft Tissue Injuries and Wounds Related To Depolarization of Damaged Cells –Loss of Cell Division –Loss of Proliferation –Loss of Regenerative capabilities Repolarization Corrects Cell Dysfunction Generates A Magnetic Field To Increase Na Pump Activity

Possible Shortwave Diathermy Unit A=Power Switch B=Timer C=Power Meter(monitors current from power supply not current entering patient-volume control) D=Output Intensity(%max power to patient) E=Tuning Control(tunes output from RFO)

Shortwave Diathermy Unit Power Output Should Provide Energy To Raise Tissue Temp To Therapeutic Range (40-45 deg C) ( watts) Should Exceed SAR-Specific Absorption Rate (rate of energy absorbed /unit area of tissue mass)

Shortwave Diathermy Unit Generates Both an Electrical and a Magnetic Field Ratio Depends on Characteristics of Both The Generator and the Electrodes –SWD Units at MHz= Stronger Magnetic Field –SWD Units at MHz = Stronger Electrical Field

SWD Electrodes Capacitor Electrodes Inductor Electrodes Selection of Appropriate Electrodes Can Influence The Treatment

Capacitor (Condenser) Electrodes Create Stronger Electrical Field Than Magnetic Field Ions Will Be Attracted Or Repelled Depending on the Charge of the Pole

Capacitor Electrodes Electrical Field Is The Lines of Force Exerted on Charged Ions That Cause Movement From One Pole To Another Center Has Higher Current Density Than Periphery

Capacitor Electrodes u Patient Is Between Electrodes and Becomes Part of Circuit u Tissue Is Between Electrodes in a Series Circuit Arrangement

Electrical Field The Tissue That Offers The Greatest Resistance To Current Flow Develops The Most Heat –Fat Tissue Resists Current Flow –Thus Fat Is Heated In An Electrical Field –Precaution: electrical field may overheat area with large fat content

Capacitor Electrodes (Air Space Plates) Two Metal Plates Surrounded By Plastic Guard Can Be Moved 3cm Within Guard Produce High-Frequency Oscillating Current

Air Space Plate Electrodes Area To Be Treated Is Placed Between Electrodes Becoming Part of Circuit

Air Space Place Electrodes Sensation Of Heat In Direct Proportion To Distance Of Electrode From Skin Closer Plate Generates More Surface Heat Parts Of Body Low In Subcutaneous Fat Best Treated

Capacitor Electrodes (Pad Electrodes) Greater Electrical Field Patient Part of Circuit Must Have Uniform Contact (toweling) Spacing Equal To Cross- sectional Diameter of Pads Part To Be Treated Should Be Centered

Pad Electrodes Increasing The Spacing Will Increase The Depth Of Penetration But Will Decrease The Current Density Capacitive Method Good for Treating Superficial Soft Tissues

Induction Method Creates A Stronger Magnetic Field Than Electrical Field A Cable Or Coil Is Wrapped Circumferentially Around An Extremity Or Coiled Within n Electrode

Induction Electrodes Passing Current Through A Coiled Cable Creates A Magnetic Field By Inducing Eddy Currents (small circular electrical fields) That Generate Heat

Induction Electrodes Patient In A Magnetic Field Not Part Of A Circuit Greatest Current Flow Through Tissue With Least Resistance –Tissues react like a parallel circuit –Fat does not provide as much resistance to electromagnetic energy Tissue High In Electrolytic Content Respond Best To A Magnetic Field –Muscle and blood Precaution: Not as much superficial sensation of heat with electromagnetic

Induction Electrodes (Cable Electrode) Two Arrangements: –Pancake Coils –Wraparound Coils Toweling Is Essential Pancake Coil Must Have 6” in Center Then 5-10cm Spacing Between Turns Best Frequency

Induction Electrodes (Drum Electrode) One Or More Monopolar Coils Rigidly Fixed In A Housing Unit May Use More Than One Drum Depending On Area Treated Penetration –Deeper Soft Tissues Toweling Important

Heating With Continuous SWD Patient Sensation Provides Basis For Recommendations Of Continuous SWD Dose I (Lowest) (<38 W) - No Sensation of Heat Dose II (Low) (~80 W)- Mild Heating Sensation Dose III (Medium) ( W) - Moderate or Pleasant Heating Sensation Dose IV (Heavy) (>300 W) -Vigorous Heating Within Pain Threshold

Pulsed SWD Referred To By Different Names –Pulsed Electromagnetic Energy (PEME) –Pulsed Electromagnetic Field (PEMF) –Pulsed Electromagnetic Energy Treatment (PEMET)

Pulsed SWD

PSWD Mean Power Calculations Pulse Period (Pulse on + off time) = –Peak Pulse Power (W) / Pulse Rep Freq (Hz) Percentage On Time = –Pulse Duration (msec) / Pulse Period (msec) Mean Power = –Peak Pulse Power (W) / Percentage on Time

Pulsed SWD Interrupted Output Delivered In Series Of High-Frequency Bursts ( usec) Pulse Rate Selected With Pulse Frequency Control (1–1000 Hz) Off-Time Longer Than On-Time Low Mean Power Output Uses Drum Electrode

Shortwave Diathermy vs.Ultrasound n Pulsed SWD Produces The Same Magnitude And Depth Of Muscle Heating as 1MHz Ultrasound (Draper, JAT 1997)

Treatment Time Most Typically SWD Treatments Last For Minutes Remember As Skin Temperature Rises Resistance Falls

When Should Diathermy Be Used? l If The Skin Or Some Underlying Soft Tissue Is Tender And Will Not Tolerate Pressure l In Areas Where Subcutaneous Fat Is Thick And Deep Heating Is Required l Induction method l When The Treatment Goal Is To Increase Tissue Temperatures Over A Large Area l If The Skin Or Some Underlying Soft Tissue Is Tender And Will Not Tolerate Pressure l In Areas Where Subcutaneous Fat Is Thick And Deep Heating Is Required l Induction method l When The Treatment Goal Is To Increase Tissue Temperatures Over A Large Area

Microwave Diathermy Two FCC Assigned Frequencies-2456 MHz and 915 MHz MWD Has Higher Frequency and Shorter Wavelength Than SWD Generates Strong Electrical Field and Relatively Little Magnetic Field Advantage: better focus wave on body, thereby more local heating affects Disadvantage: Depth Of Penetration Is Minimal In Areas With Subcutaneous Fat > 1 cm

Microwave Diathermy Unit A= Power Switch B=Timer C=Output Meter (indicates relative output in watts D= Power Output Level Knob E= Amber Light-Warming up / Red Light- Ready

MWD Applicators (Electrodes) Circular Shaped Applicators –4” or 6” –Maximum Temperature At Periphery Rectangular Shaped Applicators –4.5 x 5” or 5 x 21” –Maximum Temperature At Center

Best Treated areas for Microwave Tendons of foot, hand and wrist AC and SC joints Patellar tendon Distal tendons of hamstrings Achilles tendon Other areas of low subcutaneous fat