1
MuTr Chamber properties K.Shoji Kyoto Univ.
2
Measurement of MuTr raw signal Use oscilloscope & LabView Read 1 strip HV 1850V Gas mixture Ar:CO 2 :CF 4 =50%:30%:20% Total resistance 10kohm
3
Typical Pulse Shape
4
Fit to these region Ground Level Noise Subtract Event-by-event ground level noise subtract Use linear function y=gradient*x+intercept 3000 samples
5
Decay time Detector Capacitance C Total Resistance R=9800ohm ~ 10Kohm Exp(-t/CR) Fit Error <300nsec CR ~ 2.8usec ⇒ C ~ 300pF (?)
6
Pulse Height and Charge distribution Distribution of 1 Strip Typical Charge is 200fC But I used narrow scintillator for trigger, These distribution has dependence of place Pulse height charge Pulse HeightCharge
7
Correlation of Pulse Height & Charge There is clear correlation. But little bit rise where pulse height is low Because of the Noise?
8
Simple Model Cross mark is Profile of pulseheight&charge Red is Model Line
9
ASD Chip for MuTr LL1 Trigger 4 Channels for 1 Chip Input Impedance 370ohm Preamp Integration Time 80nsec Preamp gain 0.8mV/fC ENC 2000electrons for 300pF ⇒ Analog output Main amp gain 7 Comparator with LVDS output Threshold voltage : common for all 4 channels ⇒ Digital output
10
Expected Response of ASD Chip Input Impedance 370ohm Preamp Integration Time 80nsec Preamp gain 0.8mV/fC C ~ 300pF ⇒ CR ~ 110nsec Analog output Rise time : few ~ 10nsec
11
Linearity of ASD Chip 1Kohm load Linearity of Analog output is wrong where charge>500fC
12
From now… Correlation with neighbor strips Try to Change HV or Gas mixture Time Response Leading edge fluctuation from ion drift time<100nsec ENC 2000electrons for 300pF ~ 0.3fC Requirement is less than 1% Can we get 100um resolution with ASD Chip?