AGATA Detector Meeting March 16th, 2006, Cologne Detector Acceptance Tests: delivered: three symmetric detectors S001 properties excellent S002 properties excellent S003 properties excellent three asymmetric detectors A001 does not meet specification B001 does not meet specification C001 test ongoing, one segment missing
3
Cologne test cryostat Modified HeKo Preamplifiers: BNC output connectors Acceptance test includes: energy resolution at 60 keV, 1.3 MeV with analogue electronics, energy resolution and crosstalk with digital electronics (XIA DGF4C, Rev. E)
4
Segment specifications Guaranteed FWHMat 1.3MeV: <2.30keV, mean < 2.1keV at 60keV: <1.35keV, mean < 1.15keV Expected FWHM at 1.3MeV: mean < 2.05keV at 60keV: <1.00keV FWHM at 1.3MeV FWHM at 60keV Mean(1.3MeV)=1.99keV Mean(60keV)=1.14keV Core specification Guaranteed at 1.3MeV: 2.35keV at 122keV: 1.35keV Core FWHM: at 1.3MeV : 2.10keV at 122keV : 1.20keV IKP Cologne 001
5
Segment specifications Guaranteed FWHMat 1.3MeV: <2.30keV, mean < 2.1keV at 60keV: <1.35keV, mean < 1.15keV Expected FWHM at 1.3MeV: mean < 2.05keV at 60keV: <1.00keV FWHM at 1.3MeV FWHM at 60keV Mean(1.3MeV)=1.98keV Mean(60keV)=1.07keV Core specification Guaranteed at 1.3MeV: 2.35keV at 122keV: 1.35keV Core FWHM: at 1.3MeV : 2.08keV at 122keV : 1.19keV GSI 002
6
Segment specifications Guaranteed FWHMat 1.3MeV: <2.30keV, mean < 2.1keV at 60keV: <1.35keV, mean < 1.15keV Expected FWHM at 1.3MeV: mean < 2.05keV at 60keV: <1.00keV FWHM at 1.3MeV FWHM at 60keV Mean(1.3MeV)=2.01keV Mean(60keV)=1.03keV Core specification Guaranteed at 1.3MeV: 2.35keV at 122keV: 1.35keV Core FWHM: at 1.3MeV : 2.13keV at 122keV : 1.10keV INFN Padova 003
7
Detector 1 Detector 2 Detector 3 Crosstalk < 10 for all three detectors -3
8
FWHM at 1.3 MeV FWHM at 60 keV
9
FWHM at 1.3 MeV Resolution of A001-73892
10
Sample spectrum of segment A1 (2.7keV) Sample spectrum of segment A3 (2.33keV) Sample spectrum of segment A5 (3.66keV) Detector B001
11
Examples of spectra at V det = 5000V: Core: FWHM = 2.17 kev at 1.3 MeV Segment C2: FWHM = 1.92 keV at 1.3 MeV Segment D5: FWHM = 2.60 keV at 1.3 MeV Segment D6: FWHM = 2.63 keV at 1.3 MeV Detector A001
12
The three symmetric detectors are being mounted in test cryostats for scanning: cryostat 1: delivered to Liverpool cryostat 2: almost ready for delivery to Orsay cryostat 3: detector to be mounted Two of the asymmetric detectors were returned to Canberra for repair. Acknowledgement: symmetric detectors tested by D. Weißhaar analogue tests of asymmetric detectors by H. Hess, A. Wiens crosstalk measurement of test cryostats by N. Warr