Gusev K., JINR 27.06.2005, Gerda meeting Spectrometric performance of Kurchatov Institute detectors
2
Gusev K., JINR 27.06.2005, Gerda meeting Detector 1Detector 2Detector 3Detector 4Detector 5 Full mass, kG 0.982.9062.4462.42.781 Depletion voltage, V 3000320029001900 Bias voltage, V (recommended) 4000 35002500 FWHM, keV at 1332 keV (specifications) 1.981.911.972.06 FWHM, keV at full HdMo Set-up 2.222.432.712.142.55 Test at February 2005 FWHM, keV at 1332 keV 2.72.272.532.352.75 New measurements Bias voltage, V 37004000 35002500 FWHM, keV at 1332 keV 2.882.53.02.763.05
3
Gusev K., JINR 27.06.2005, Gerda meeting Detector 1 Problems: Detector not reaches the recommended voltage Detector has comparatively poor energy resolution Detector had a big cross-talk Reasons: Detector has big leakage current (even at 3700 V, I > 10 -10 A in my estimation) Preamplifier electronics are not work correctly Possible solutions: Change the preamplifier Measure V-I characteristic of detector and choose the optimal voltage Pump and maybe heat
4
Gusev K., JINR 27.06.2005, Gerda meeting “Wandering” = contact or electronic problems Detector 2 Strange instability of peaks (doubling effect) Now this effect is absent Leakage current at operation voltage is normal
5
Gusev K., JINR 27.06.2005, Gerda meeting Detector 3 Detector has poor energy resolution FWHM = 3.0 keV Possible reason – big current (> 10 -10 A) Possible solution is similar to detector 1
6
Gusev K., JINR 27.06.2005, Gerda meeting Detector 4 No problems!
7
Gusev K., JINR 27.06.2005, Gerda meeting Detector 5 Poor resolution (FWHM = 3.05 keV), but leakage current is good Need further investigations
8
Gusev K., JINR 27.06.2005, Gerda meeting Conclusions All detectors are prepared for work, all crystals are good but their spectrometric performance are not ideal (especially for detectors 1, 3 and 5) For achieving the best performance of KI detectors we need: Good measurement conditions; Enough time for determination of various detector parameters (at least V-I characteristics); Precision pulse generator; Devices for pump and heat processes. We have to make all available procedures before any other activities (such as repacking, etc.)