Spectroscopic Surface Analysis and Cleaning Verification with SpotView ® Systems.

Similar presentations


Presentation on theme: "Spectroscopic Surface Analysis and Cleaning Verification with SpotView ® Systems."— Presentation transcript:

1 Spectroscopic Surface Analysis and Cleaning Verification with SpotView ® Systems

2 Overview © Remspec Corp. 2014

3 SpotView® and SpotView-SL Easy-to-use, portable system. Robust, transferable calibration. Realtime, in-situ detection and quantitation down to microgram levels. CFR 21/Part 11 compliant software. © Remspec Corp. 2014

4 FTIR Spectroscopy FTIR in the mid-IR range is the most sensitive optical measurement technique – extinction coefficients are larger than either in the NIR or the UV-visible range. Almost all organic compounds, including oils, lubricants, and cleaning agents have distinctive mid-IR spectra. © Remspec Corp. 2014

5 Grazing Angle FTIR Spectroscopy © Remspec Corp. 2014

6 Interchangeable with grazing-angle SpotView Shaft diameter 6 mm Direct reflectance is less sensitive than grazing angle, but can be used in tight places and on small objects. SpotView-SL © Remspec Corp. 2014

7 SpotView-SL SpotView-SL can be configured with a single cable, up to 3 m long. Ideal for tablet presses etc. © Remspec Corp. 2014

8 SpotView-SL © Remspec Corp. 2014

9 SpotView-SL The SL probe is the best choice for contamination levels above about 3 mg/ft 2 (approx. 3 μg/cm 2 ), equivalent to a film thickness of about 3 μm. The grazing-angle method breaks down for films above about 4 μm in thickness. © Remspec Corp. 2014

10 ValveView™ for Examining Restricted Interior Surfaces Like SpotView®, uses the grazing-angle method, this time inside a small hollow object –In use for cleaning validation by manufacturers of semiconductor gas- handling equipment –Used on interior surfaces of small ball valves Valve holder Fiber-optic cable Output Mirror Input mirror Detector © Remspec Corp. 2014

11 ValveView™ for Examining Restricted Interior Surfaces ValveView has been in use at a Taiwanese valve company for about ten years. –It is used as a QC tool to ensure an oil-free surface inside each valve. – ( 1 ) CLASS IV Max.1mg/m 2 – ( 2 ) CLASS V : Max.0.1mg/m 2 Valve holder Fiber-optic cable Output Mirror Detector © Remspec Corp. 2014

12 Calibration © Remspec Corp. 2014

13 Sample Preparation Smear Method –Dispense small amount of known API solution onto surface –Smear evenly –Collect spectra after drying –Estimate coverage by: Calculation from solution concentration. –Suitable method for demonstrations and qualitative analysis. © Remspec Corp. 2014

14 Sample Preparation Spray Method –Spray API solution on surface with air brush –Collect spectra after drying –Wash entire surface –Analyze washing by, e.g., HPLC –Calculate surface coverage © Remspec Corp. 2014

15 Sample Preparation Quantitative Printing –Developed from ink-jet printing technology –Provides known surface concentrations without requiring secondary analysis method. –Available as a service from Remspec. © Remspec Corp. 2014

16 SpotView-SL Calibration: some simple rules Smear Method for quick feasibility studies Use sample of the same material as the target reactor or mixer (metal, glass, plastic, etc.) Consider the range of surface finishes that you may encounter (mirror polish, matte, etc.) Spray or Print Method better for developing working calibrations and determining true detection limits. Spray or Print Method essential for multi-compound calibration. Secondary analysis (for Spray method) Is the analyte UV-active? How easy will it be to wash off the sample? © Remspec Corp. 2014

17 Set of API spectra –BioPharm 15(5), May 2002, pp. 36-42 Air brush /weight difference method. Unknown API from Pfizer Puerto Rico plant. © Remspec Corp. 2014

18 SpotView/PLS1 Results versus Swab/HPLC Second, independent set of coupons loaded with API. SpotView® spectra collected. Pfizer in-house swabbing method used to obtain “conventional” loading data. © Remspec Corp. 2014

19 SpotView-SL Calibration: API on stainless steel Intensity varies with concentration Can be visually estimated down to about 3 μg/cm 2. At lower surface concentrations, peak shape changes are often observed. Chemometric calibrations can be established down to about 1 μg/cm 2. © Remspec Corp. 2014

20 SpotView-SL Calibration: API on stainless steel Results are shown for a cross-validation test of a PLS1 model. Chemometric calibrations can be established down to about 1 μg/cm 2. © Remspec Corp. 2014

21 Amlodipine on Stainless Steel Spray method, with calibration by HPLC In this example, the drug is the only compound on the surface. Grazing-angle SpotView was used; note the lower RMS error compared to SpotView-SL © Remspec Corp. 2014

22 Amlodipine and CIP-90 on Stainless Steel Spray method, with calibration by HPLC In this example, the drug and a variable amount of CIP-90 (cleaning agent) is present on the surface. Note the increased scatter, larger estimated error, and higher rank of the model (it uses more factors). © Remspec Corp. 2014

23 Amlodipine and CIP-90 on Stainless Steel Graphing the average predictions against HPLC values highlights the increased scatter when CIP-90 is present. Amlodipine And CIP-90 © Remspec Corp. 2014

24 Case Study © Remspec Corp. 2014

25 Case Study: inside stainless steel production tanks The SpotView FTIR system for surface analysis was used to examine the interior surfaces of two tanks, T-2221 and T-2211, interior surfaces of the CIP recovery tank and pump, and of the CIP return pump P8211. A operator was lowered into the tank carrying the operating spectrometer. The SpotView head and fiber-optic cable were lowered separately and connected in situ. The computer was operated by a second person outside the tank. A total of 69 spectra was collected over a period of three days from selected sites on the walls of the tanks, on agitator shafts and spray balls, and from the manways and manway covers. Separately, spectra were collected from surfaces of the CIP recovery tank and pump. © Remspec Corp. 2014

26 Case Study: inside stainless steel production tanks Typical “100% line” taken from the clean background surface. Note (a) noise spike at approx. 2080-2280 cm -1, arising from residual Se-H in optical fiber and (b) minor features from atmospheric H 2 O and CO 2. © Remspec Corp. 2014

27 Case Study: inside stainless steel production tanks Two spectra from headspace surface (T-2221). Note features in upper spectrum: (a) pronounced shift in baseline, (b) centered near 2960 cm -1 and (c) spectral density between 1000 and 1770 cm -1. © Remspec Corp. 2014

28 Case Study: inside stainless steel production tanks T-2211, agitator shaft. Note spectral features in top trace, very similar to those found in a spectrum from T-2221 headspace on Day 1. © Remspec Corp. 2014

29 Case Study: inside stainless steel production tanks Two spectra from the manway lid of T-2211. These spectra are extremely noisy. The reflected signal is very low. © Remspec Corp. 2014

30 Case Study: inside stainless steel production tanks CIP supply pump. Note spectral features between 3600 and 2800 cm -1, and spectral density below 1800 cm -1. © Remspec Corp. 2014

31 Case Study: inside stainless steel production tanks Comparison with the spectrum from T-2221 headspace (olive trace) and from T2211 agitator shaft (green trace) suggests that the same or similar residue has been detected in all cases. © Remspec Corp. 2014

32 Case Study: inside stainless steel production tanks Of the 69 spectra collected, 65 showed no features attributable to organic residue on the surface. This indicates that the surfaces are free of spectroscopically detectable organic residue to a level better than 0.5 μg/cm 2. Four spectra—one each from tanks T-2221 and T2211 and two from the pump on the CIP recovery skip--showed a very small spectroscopic signature for an organic residue that was (a) the same or very similar in each case (b) highly localized in the two tanks (attempts to detect anything on immediately adjacent surfaces resulted in featureless spectra). © Remspec Corp. 2014

33 Software © Remspec Corp. 2014

34 SpotView ® uses CVal ™ Software for Routine Cleaning Validation Supervisor Mode for setting up procedures and Operator instructions. Operator Mode for routine cleaning validation Procedures are digitally locked and verified each time they are used. © Remspec Corp. 2014

35 CVal ™ Software for Routine Cleaning Validation Procedures and Results can be displayed using Viewers that verify each file as unchaged. Procedure and Results file are also saved in simple, human- readable format for long-term readability. © Remspec Corp. 2014

36 Conclusions © Remspec Corp. 2014

37 Expanding the Scope of Reflectance FTIR Spectroscopy SpotView® is ideal for surfaces that are: –Flat or moderately curved –Accessible with the SpotView® head –Interior surfaces of large reactors, for example SpotView®-SL is ideal for surfaces that are: Highly curved Restricted access Valves and nozzles, for example © Remspec Corp. 2014

38 In Summary SpotView can measure surface laodigns down to 0.2 μg/cm 2. SpotView-SL can measure loadings down to about 1-3 μg/cm 2. The results are quantitative and compare well with existing pharma industry methods like swabbing. In-situ results are obtained from a hand-held device in seconds. © Remspec Corp. 2014


Download ppt "Spectroscopic Surface Analysis and Cleaning Verification with SpotView ® Systems."