Download presentation
Presentation is loading. Please wait.
Published byColleen O’Connor’ Modified over 8 years ago
1
Development of the X-ray Detectors for XTP Cao Xue-lei Institute of High Energy Physics, CAS 2014.02.21 IHEP, Beijing
2
Outline Payload configuration Primary design of the HE collimating telescope Detector architecture Performance of the CZT detectors with the readout chips. Development of the Focal Plane Detectors Characteristics Tests of the SPD Performance of the SDD
3
3 XTP payload configuration
4
Architecture of Detectors of the Collimator Telescope The telescope consists of lots of modules 25 Units In each unit there are 25 modules, with the HV power for the detectors, the ADC and the FPGA controllers. 25 Modules in one Unit. 16 CZT detectors used one ASIC for readout in one module. So there are 10000 channels for the hard X-ray collimated telescope. Collimator Telescope Unit1 (ADC, FPGA, HV) Unit2 (ADC, FPGA, HV) Unit3 (ADC, FPGA, HV) …. Module-1 16-ch ASIC CZT detector and PreA -1 CZT detector and PreA -2 CZT detector and PreA -3 CZT detector and PreA -16 …. Module-2 16 –ch ASIC Module-3 16-ch ASIC …. Module-25 16-ch ASIC Unit25 (ADC, FPGA, HV)
5
Characteristics of the High energy X-ray Collimating Telescope FOV : 1°×1° 。 Active area : 2500 cm 2 @30keV Detector : CZT Energy resolution : FWHM 3keV@59.5keV Time resolution : 200μs Energy coverage : 8-100keV Single detector size: 5mm*5mm*2mm Total number of the DET: 10000 Total mass: ~200kg Power dissipation: ~200W CZT detectors Front-to-end electronics Collimator
6
Characteristics Prototype (From 09-2012 to 07-2015) of the Collimator Detectors DetectorsCZT Single detector size5mm*5mm*2mm Number of detectors192 Energy resolution FWHM 2keV@59.5keV, -10 ℃2keV@59.5keV Time resolution200µs Power dissipation45W
7
The CZT detectors The CZT detectors are made by the Northwestern Polytechnical University; The efficiency is about 87% at 100 keV for a 2mm thickness CZT. Thickness (mm) Efficiency The CZT Crystal and the electrode preparation CZT detectors with different size Pixels electrodes Spectrums of the CZT detector Vs scintillator Spectrum of the LaBr 3 CZT has better energy resolution than LaBr 3
8
CZT detectors for the prototype
9
Test of the CZT chips code Volt-Ampere characteristic Am-241 source test -100v (nA) 100v (nA) -200v (nA) 200v (nA) -300v (nA) 300v (nA) V Energy resolution 040502065-2.032.05-4.064.11-6.136.27350v4.68% 040502075-1.291.22-2.472.38-3.623.54350v4.89% 010502017-1.411.35-2.792.63-4.183.88500v5.30% 080502139-1.051.05-2.122.06-3.213.09400v5.55% 020502224-2.812.57-6.125.34-9.778.30300v6.37% 190502235-1.441.50-2.933.02-4.464.63400v6.47% Size : 5mm×5mm×2mm >200 pieces have already made
10
Performance of the CZT detector Single channel readout system of the CZT detector Spectrum of the Am-241 source used the CZT detector prototype Functional block diagram of the prototype Energy resolution : FWHM 1.7keV@17.8keV 。 CZT detector size : 5mm*5mm*1mm Temperature : 23 ℃, Humidity : 18% Preamplifier CZT HV Power Amplifier MCA
11
The Chips made in IHEP Include three parts: Pre-Amplifier, Amplifier, Comparator and Peak-Holder Some results of the readout chips Preamp: Charge-sensitive amplifiers Resistor -capacitor feedback Noise : ENC ~60e , for Si detector, the Cd=0.1pF Peaking Time : 0.3μs, 1μs, 6μs, 12 μ s Power : ~5 mW/per channel Input:: +/- signals Process: 0.35 um N-well CMOS
12
The test board for the readout chips Pre- Amp Si-Detector Am p Comparator and Peak-Holder power
13
Bonding wire for Pre-Amp
14
Simulated result Result from the test board Input signals, blue Signal from the preamp ( red ) Signals from the amplifiers ( green ) Peak holding signals
15
Test with the Si-PIN detector Detector and Pre-Amplifier connect by bonding wire directly 5mm 2 Si - PIN detector Am-241 radioactive source Commercial MCA room temperature FWHM is 1.166KeV@5.9Ke V
16
A Single-channel test board for CZT with readout chips
17
Nine-channel prototype 9 CZT detectors with the size of 5mm*5mm*2mm Nine channels wire bonded-chip
18
Spectrum of the Am-241 of the 9-channel prototype Energy resolution about FWHM 3KeV @59.5KeV
19
ASIC for CZT pixel detectors Type: VA32TA6 Made by: Ideas company Power: <1mW /Channel Channel: 32 Noise: 39+12 e/pF ADC: 10 bit SEL : 18 MeV Process: 0.35 um N-well CMOS VA32TA6 are Used in HXMT ME telescope The VA-TA principle The architecture of VA parts
20
Some tests on the VA32TA6 A prototype of the Si-PIN with the VA32TA6 Arrival time intervals(us) Counts Spectrum of the Am-241 source by the 6mm 2 Si-PIN with the VA32TA6 ASIC, with the energy resolution of FWHM 1.1keV@59.5 keV This fig shows the arrival time interval distribution of the VA32TA6 prototype. The dead time is shorter than 180 us.
21
SEU/SEL test Single event test in the tandem accelerator line. The single particle lock threshold is about 18 MeV.cm 2 /mg Experiment with VA32TA6 work in the temperature -50 ℃, 14 days Electrostatics discharge test of the VA32TA6, it is very impressible for the ESD, <50V
22
Architecture of the Focal Plane Detectors 3 focal plane detectors 256 pixels CZT detectors are used. For one focal plane detectors, 4 64-channels ASIC are used for the read-out electronics. with the HV power for the detectors, the ADC and the FPGA. 16*16 pixel CZT Pixel size: 1mm*1mm*1mm
23
Characteristics of the focal plane detectors of the HE telescope Detector : CZT pixel detector DET Number : 3 Pixel Number per DET : 256 CZT pixel size : 1mm*1mm*1mm Power dissipation : ~80W Energy resolution : FWHM 2keV@20keV2keV@20keV Time resolution : 10us Work temperature : -10 ~ +20 ℃ Energy coverage: 8-30keV
24
Prototype of LE Focal Plane Detectors based on SPD Detector Si-PIN pixel detectors, SPD Pixel size 1mm*1mm*0.5mm Number of the pixels 100 Energy resolution FWHM 200eV@5.9keV, -40 ℃下200eV@5.9keV Time resolution 20µs Power 35W
25
SPD detectors Pixel size : 0.1-10mm 2 Thickness : 300μm-2mm Pixel structure : P-I-N Superiority : 1 ) High detection efficiency in 0.5-30keV, ~90% @15keV; 2 ) Good energy resolution , FWHM 170eV@5.9keV ; 3 ) Acceptable spatial resolution: 300μm-3mm ; 4 ) High time resolution , <1us
26
SPD chips The 100 pixels SPD packaged on PCB The leakage current of SPD in +25 ℃ Parameters of the SPD Pixel size : 1mm*1mm Thickness : 500um Pixel number : 10*10 Leakage current : < 0.4nA depleted The first prototype
27
The Prototype Based on the SPD SPD prototype block diagramThe testing photo of the SPD
28
Si-PIN 像素探测器样机(3) The spectrum of the Am -241 source by the 20 pixels of the SPD Am-241 source spectrum of one pixel, temperature 21 ℃, Results : Energy resolution: FWHM 1.1keV@59.5keV, at room temperature 1.1keV@59.5keV Dead time : 20us
29
Si-PIN 探测器低温测试 Temperature control system in vacuum, for SPD detector test Peltier device Results : Si-PIN pixel detector at -15 ℃, for the Am- 241 source, the energy resolution, FWHM 278eV@13.8keV The spectrum of the Am-241 used SPD, in -15 ℃
30
7-channel SDD detectors(2010) 7 SDD detectors. Active area 1cm^2 FWHM 151eV@5.9keVFWHM 151eV@5.9keV, -35 ℃
31
Summary SDD and SPD is suitable for the low energy X-ray focal plane detector, the energy resolution better than FWHM 200eV@5.9keV.200eV@5.9keV The high energy collimated telescope uses the CZT detectors, it has an active area of 2500 cm 2, and the energy resolution is better than 3keV@59.5keV. The CZT pixel detectors are used for the HE focal plane detectors. The energy resolution should better than 2 keV@59.5 keV.keV@59.5
Similar presentations
© 2024 SlidePlayer.com. Inc.
All rights reserved.