Download presentation
Presentation is loading. Please wait.
Published byJustin Thomas Modified over 8 years ago
1
Basis of Mammography F. Milano Dept. Clinical Physiopathology University of Florence Email: f.milano@dfc.unifi.it
2
Diagnostic sign characteristics - 1 Breast tissue typeDensity (g/cm3) Linear Attenuation Coefficient at 20 keV (cm-1) Mammary tissue Fibroglandular tissue 1.0350.80 Adipose tissue0.930.45 Skin1.090.82 Average breast (50% glandular – 50% fat) 0.980.62 Lesion Carcinoma1.0450.85 Calcifications2.212.5
3
Diagnostic sign characteristics - 2 Low contrast visible mass centimeter in diameter Calcifications 100 microns
4
The mammography image
5
Variation of the linear attenuation coefficient with energy
6
Linear attenuation coefficient and interaction components
7
The photoelectric effect in breast The photoelectric effect is the dominant interaction at energy below 22 keV IMAGE CONTRAST The photoelectric effect is responsible for most of the energy imparted to the breast DOSE
8
Scatter radiation The scattered photons reduce the contrast in the image grid compression of the breast
9
Scatter / primary ratio in mammography
10
Spatial resolution Focus dimension Geometry Breast compression Detector characteristics
11
Mammography equipment Dedicated x-ray unit Dedicated detectors Dedicated image processing
12
The x-ray unit
13
X-ray spectral considerations for mammography The quality of the image and the dose with which it is produced are critically dependent on the x-ray spectrum used to form the image The shape of the spectrum is determined by the anode material, the applied voltage to the tube and the type of added filtration to the tube
14
Molibdenum k edge lines: 17.4 – 19.6 keV Molibdenum anode and filtration
15
28 kVp 25 kVp X-ray tube applied voltage and image quality
16
Contrast – Noise - Unsharpness
17
Signal-to-Noise ratio and photon energy
18
Spectrum suited to largest breast thickness Optimal energy band and breast thickness 14-18 keV 2 cm 17-21 keV 4 cm 19-23 keV 6 cm 20.5-23.5 keV 8 cm K edge Rhodium 20.2 – 22.8 keV
19
Mo/Mo – Mo/Rh – Rh/Rh spectra
20
Scatter radiation The effect of scatter is to reduce contrast The fraction of the possible contrast imaged in the presence of scatter is given by the Scatter Degradation Factor (SDF) SDF = 1 / ( 1 + S/P ) In mammography S/P range from 0.33 to 1.5 Only 40% (S/P=1.5) to 75% (S/P=0.33) of the possible contrast is imaged unless scatter is controlled If scatter is controlled image contrast can be improved by a factor of 1.4 – 2.5
21
Grid in mammography
22
Spatial resolution – Geometry – Focus size
23
Direct magnification No grid ( air gap ) Microfocus – 0.1 mm
24
Compression Breast compression is critical to optimizing image quality and dose Scatter Spatial resolution Movement artifact Absorbed dose
25
Compression A correct B incorrect Compression permits a breast uniform thickness
26
Effect of compression on scatter
27
Compression The breast compression makes the structures to be imaged closer to the detector improving geometrically the spatial resolution It also makes more visible the edge of the lesion
28
Compression plate and photon attenuation
29
Automatic Exposure Control (AEC)
30
AEC compensation for kV and breast thickness AEC is critical in the mammographyc practice
31
Film-screen system
32
Exposure mAs of 50 Exposure mAs of 100 Exposure mAs of 150 Image quality and AEC
33
AEC film-screen Film screen and AEC
34
Higher speed, darker image Higher gradient, greater contrast Effects of film-screen settings and processing
35
Gradient 101.9 mAs 128.7 mAs 148.5 mAs 2.523.32 4.72 0.67 63.9 mAs 80.7 mAs 93.1 mAs 1.07 46.5 mAs 58.7 mAs 67.8 mAs 1.47 Speed Higher gradient, higher contrast image. Higher speed, darker image.
36
AEC to backAEC to front AEC and image quality degradation
37
Contrast and film sensitometry Gradient are different for different exposure
38
Patient dose It is important: To evaluate the risk to the patient To compare imaging systems To assess performances of the imaging equipment To comply patient radioprotection regulations
39
Dose and exposure parameters Exposure at breast entrance Dose to the entrance surface Dose to the midline Mean dose to the glandular tissue
40
Mean Glandular Dose (MGD) It provides the best indicator of the risk to the patient It cannot be measured directly It must be calculated from the results of simple measurements and tabulated values
41
Mean Glandular Dose (MGD) D = K p g K is the incident air kerma without backscatter to a 4 cm PMMA phantom p converts the air kerma from PMMA phantom to that for standard breast g converts the air kerma for the standard breast to MGD
42
p/g table and PMMA/breast thickness equivalence
43
MGD – kV – Breast thickness – Half Value Layer
44
Mean Glandular Dose
45
MDG – Entrance Skin Exposure (ESE) – kVp – Subject Contrast (Cs) – Signal to Noise ratio (S/N) t is the imaging time for a 5-mm cubic inflitrating ductal carcinoma Table 1 - Mo/Mo Table 2 - W/Mo
46
Dose – Mo/Mo – Rh/Rh
47
The END Thanks for the attention !
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.