Carbons on Catalyst CAUTION: Samples are hot when removed from the oven and furnace. Use heat resistant gloves when handling hot samples. 1Place at least.

Similar presentations


Presentation on theme: "Carbons on Catalyst CAUTION: Samples are hot when removed from the oven and furnace. Use heat resistant gloves when handling hot samples. 1Place at least."— Presentation transcript:

1 Carbons on Catalyst CAUTION: Samples are hot when removed from the oven and furnace. Use heat resistant gloves when handling hot samples. 1Place at least 100 granules of lift engager catalyst into beaker. 2Pour approximately 10 mls of pentane into beaker. 3Swirl pentane around the catalyst at least 30 seconds. Let stand for 15 minutes. 4Dump out pentane and place beaker in brown oven for 15 minutes to dry the catalyst. 5Remove and place in desiccator for 1 hour. 6Proceed to run test. PPE Use Standard (SSP-0011) Y Full Face Shield / Goggles Hood Clothing—Chemical Protection Gloves—Chemical W Personal Fall Protection Respirator / Filter Mask Supplied Air Job Hazards Y es Job Hazards Y es Access and EgressMoving Parts Activities of OthersNoise Adverse WeatherRail Movement Chemical ExposureSharp Edges ElectricitySlips, Trips, and Falls FatigueStatic Electricity Fire and ExplosionTrip Injuries Ground ContaminationTrapped Pressure Manual HandlingWorking at Height Other: References / Documents  MSDS on Catalyst  Unit Safe Upper and Lower Limits Spreadsheet AreaUnitOperating Procedure #Certified byRevision #DatePage 3LabOPS-LB-NM-005B. Gereighty39/10/071 of 1 Carbons on Catalyst Environmental  N/A Quality Assurance  N/A Special Instructions  This procedure is to be used to run Carbon on Catalyst.  Any problems, notify Shift Supervisor or Duane Chaisson. W A R N I N G C A U T I O N Leak Check Procedure 1Turn valve toward back. 2Lower bottle and place as low as possible. 3Place rubber cork in end of furnace tube. 4Raise bottle and place in lower rack. 5Wait at least 30 seconds and if layer does not stop rising, recheck seal of cork on tube. 6If this does not stop layer from rising, the oven furnace tube must be replaced. 7After no leak is determined, place bottle back on upper rack.  Procedure Completed Date: Prepared By: Approved By:

2 AreaUnitOperating Procedure #Certified byRevision #DatePage 3LabOPS-LB-NM-005B. Gereighty39/10/072 of 2 Carbons on Catalyst (continued) Weighing of Samples 1Make sure that the samples boats are clean. 2Take out sample boat on balance (zero). 3Weigh 1.0000 grams of regenerator catalyst or 0.1000 grams of lift engager catalyst. 4Proceed to run test. Running Test 1Turn on oxygen cylinder in rear of lab. 2Make sure that valve is pointed towards back. 3Place boat in end of furnace tube. 4With left hand, take rod and slide the boat into the red zone of furnace. 5With right hand, place the cork with the oxygen tube on the end of furnace. 6Use gloves or towels and make sure cork forms a good seal on furnace tube. 7Open oxygen valve and start purging sample with oxygen. 8 Watch the pink side (tube) until the level gets even with the black line on the tube and then shut off the oxygen (approximately 1 minute). 9Place bottle on lower rack. 10Pull out cork from furnace tube. 11Level in the tube and the bottle should be even and on 0.00 (if not, adjust with the wheel on lower level until even). 12Move valve to the left and raise bottle and place into upper rack. 13Make sure the bubble locks in the pink tube and then move valve to front position. 14Wait 15 minutes. 15Grab bottle and lower in the left hand, then move valve back to the left position. 16Lower bottle with left hand until the bubble in the clean tube locks, then move valve to front position. 17Place bottle on lower rack and let the pink liquid in the tube stop rising (15 seconds).  Procedure Completed Date: Prepared By: Approved By:

3 AreaUnitOperating Procedure #Certified byRevision #DatePage 3LabOPS-LB-NM-005B. Gereighty39/10/073 of 3 Carbons on Catalyst (continued) Running Test (continued) 18Raise bottle until the level in the bottle is equal to the level in the tube. 19Take reading. 20Open valve to the right and let oxygen vent out. 21Make sure the bubble locks in the pink tube. 22Remove boat from furnace with rod. 23Shut off oxygen when finished.  Procedure Completed Date: Prepared By: Approved By: