BY G.SHRAVANI 170213884010 ION EXCHANGE AND GEL PERMEATI ON CHROMATOGRAPHY 1G.SHRAVANI
2
Chromatography 2G.SHRAVANI
3
Definition Ion-exchange chromatography (or ion chromatography) is a process that allows the separation of ions and polar molecules based on the charge properties of the molecules. 3G.SHRAVANI
4
Ion-exchange chromatography The solution to be injected is usually called a sample, and the individually separated components are called analytes It can be used for almost any kind of charged molecule including large proteins, small nucleotides and amino acids. It is often used in protein purification, water analysis. 4G.SHRAVANI
5
PROPERTIES OF ION EXCHANGE RESIN Color Amount or cross linking Porosity Capacity Surface area Density Mechanical strength Size G.SHRAVANI5
6
Flow rates should be controlled due to the difference in rate of exchange Flow rate -0.5 to 5ml/minutes Capacity of ion exchange depends on the no. of sites available Cation resins(strong acid) are stable at 150degress & anions at 70 G.SHRAVANI6
7
Principle Ion exchange chromatography retains analyte molecules based on ionic interactions. The stationary phase surface displays ionic functional groups (R-X) that interact with analyte ions of opposite charge. This type of chromatography is further subdivided into: 1. cation exchange chromatography 2. anion exchange chromatography. 7G.SHRAVANI
8
Ion Exchangers 8G.SHRAVANI
9
Ion exchangers – Functional groups Anion exchanger Aminoethyl (AE-) Diethylaminoethyl (DEAE-) Quaternary aminoethyl (QAE-) Cation exchanger Carboxymethyl (CM-) Phospho Sulphopropyl (SP-) 9G.SHRAVANI
10
Cation exchange chromatography Cation exchange chromatography retains positively charged cations because the stationary phase displays a negatively charged functional groupcations 10 R-X C +M B R-X M + C + B - - + + + + - - _ _ + + + + - - G.SHRAVANI
11
Anion exchange chromatography Anion exchange chromatography retains anions using positively charged functional group: 11 R-X A +M B R-X B + M + A + + _ _ + + - - + + - - + + - - G.SHRAVANI
12
Procedure 1. A sample is introduced, either manually or with an autosampler, into a sample loop of known volume. 2. The mobile phase (buffered aqueous solution) carries the sample from the loop onto a column that contains some form of stationary phase material. 3. Stationary phase material is a resin or gel matrix consisting of agarose or cellulose beads with covalently bonded charged functional groups.agarosecellulosecovalently 12G.SHRAVANI
13
Procedure 4. The target analytes (anions or cations) are retained on the stationary phase but can be eluted by increasing the concentration of a similarly charged species that will displace the analyte ions from the stationary phase. For example, in cation exchange chromatography, the positively charged analyte could be displaced by the addition of positively charged sodium ions. 13G.SHRAVANI
14
Procedure 5. The analytes of interest must then be detected by some means, typically by conductivity or UV/Visible light absorbance. 6. A chromatography data system (CDS) is usually needed to control an IC. 14G.SHRAVANI
15
Procedure 15
16
Separating proteins Proteins have numerous functional groups that can have both positive and negative charges. Ion exchange chromatography separates proteins according to their net charge, which is dependent on the composition of the mobile phase. 16G.SHRAVANI
17
Affect of pH in the separation of proteins By adjusting the pH or the ionic concentration of the mobile phase, various protein molecules can be separated. For example, if a protein has a net positive charge at pH 7, then it will bind to a column of negatively-charged beads, whereas a negatively charged protein would not. 17G.SHRAVANI
18
Effect of pH in the separation of proteins Proteins are charged molecules. At specific pH, it can exist in anionic (-), cationic (+) or zwitterion (no net charge) stage. 18 cationic pH =pI anionic pH increase *pI isoelectric point G.SHRAVANI
19
Choosing your ion-exchanger: know your proteins 1. Stability of proteins stable below pI value, use cation-exchanger stable above pI value, use anion-exchanger 2. Molecular size of proteins <10,000 mw, use matrix of small pore size 10,000-100,000 mw, use Sepharose equivalent grade 19G.SHRAVANI
20
Important to consider the stability of proteins in choice of ion exchangers. Isoelectric focusing can be used to identify suitable ion-exchanger type 20G.SHRAVANI
21
Applications Determination of sodium and potasium in the mixture: column eluted with 0.1M HCL flow rate -0.6ml/sqmin Radiochemistry G.SHRAVANI21
22
Gel Permeation \GEL FILTRATION Chromatography 22G.SHRAVANI
23
DEFINATION Gel chromatography is a technique in which fractionation is based upon the molecular size & shape of the species in the sample. Gel chromatography is also called as gel permeation or exclusion or molecular sieve chromatography. 23G.SHRAVANI
24
TECHNIQUES IN GEL CHROMATOGRAPHY Gel chromatography is performed on a column by the elution method. The degree of retardation, depends upon the extent to which the solute molecules or ions can penetrate that part of the solution phase which is held within the pores or the highly porous gel like packing material. A series of resins with different pore sizes can be obtained by changing the amount of Epichlorohydrin.The resulting gel is called as SEPHADEX. 24G.SHRAVANI
25
In gel chromatography the granulated or beded gel material is called as packing material. The solutes which are distributed through the entire gel phase is called as stationary phase and the liquid flowing through the bed is called as mobile phase. The cross linked dextran (sephadex) & xerogels of the polyacrylamide (Bio-gel) originally used as stationary phases in gel permeation chromatography are semi rigid gels. 25G.SHRAVANI
26
These are unable to withstand the high- pressure used in HPLC. Hence modern stationary phases consist of micro particles of stryene-divinyl benzene copolymers (ultrastyragel) silica or porous glass. 26G.SHRAVANI
27
GEL PERMEATION CHROMATOGRAPHY A chromatographic method in which particles are separated based on their size, or in more technical terms, their hydrodynamic volume. Organic solvent as the mobile phase. The stationary phase consists of beads of porous polymeric material. 27G.SHRAVANI
28
OBJECTIVE Analysis of synthetic and biologic polymers Purification of polymers Polymer characterization Study properties like: Molecular weight Polydispersity Index Viscosity Conformation Folding 28G.SHRAVANI
29
Aggregation Branching Copolymer composition Molecular size 29G.SHRAVANI
30
PRINCIPLE 30G.SHRAVANI
31
PRINCIPLE Different sizes will elute (filter) through at different rates. Column 1. Consists of a hollow tube tightly packed with extremely small porous polymer beads designed to have pores of different sizes. 2. Pores may be depressions on the surface or channels through the bead. 3. Smaller particles enter into the pores, larger particles don't. 31G.SHRAVANI
32
The larger the particles, the less overall volume to transverse over the length of the column 32G.SHRAVANI
33
FEATURES Solvent 1. Should be kept dry 2. Should be degassed in some applications 3. The samples should be made from the same solvent 4. For GPC/light scattering the solvent should be filtered before it ever hits the pump 5. Common solvents for tetrahydrofuran (THF) & toluene. 33G.SHRAVANI
34
34G.SHRAVANI
35
Pump &Filters Pump: a. Designed to deliver very constant, accurate flow rates. b. At microprocessor-controlled rate. c. Designed not to produce any pressure pulses. Filters: Prevent major junk from getting into the columns 35G.SHRAVANI
36
Injector Loop Allows you to load the sample loop which is a piece of tubing precut for a precise volume. the output of the pump flushes through the loop Carries the sample to the columns. sends a signal to the detector to indicate that the sample has been loaded. 36G.SHRAVANI
37
Columns Columns Contain the beads through which the sample is allowed to pass. Reference column is also present Very expensive Never change the pumping rate by a large amount They are very delicate 37G.SHRAVANI
38
Detectors Detectors o Viscosity o Light Scattering o Ultraviolet detectors o Differential Refractive Index detector placed at the end to reduce pressure on it 38G.SHRAVANI
39
Analysis Spectroscopic Techniques 1. Refractive Index 2. Light Scattering 3. Ultraviolet Spectroscopy Viscometry Techniques 1) Viscosity 2) Flow rate 39G.SHRAVANI
40
Conventional GPC Analysis Molecules separated according to their hydrodynamic volume. Molecular weights (MW) and molecular weight distribution can be determined from the Measured retention volume (RV) A calibration curve (log MW against RV), using known standards RI signal = KRI. dn /dc. C 40G.SHRAVANI
41
KRI = apparatus-specific sensitivity constant dn /dc = the refractive index increment C = concentration. Limitation Their signals depend solely on concentration, not on molecular weight or polymer size. Not very reliable 41G.SHRAVANI
42
Molecular Mass Sensitive Detectors Detectors sensitive to molecular weight used to overcome limitations of Conventional GPC E.g., light scattering and viscosity detectors Advantages over Conventional GPC I. True molecular weight distributions can be obtained II. Structural information III. Size distribution molecular weight can be directly determined without a calibration curve 42G.SHRAVANI
43
Characterization Light Scattering LS signal = KLS. (dn/dc)2. MW. KLS = sensitivity constant dn/dc = refractive index increment MW = molecular weight C = concentration dn/dc depends on the Polymer Solvent combination and if it is low, then proper analysis cannot be done. 43G.SHRAVANI
44
Applications Polymer characterization Molecular weight Polydispersity Index Viscosity Folding Aggregation Branching Copolymer composition Molecular size 44G.SHRAVANI
45
Proteomics Purification Conformation Hydrodynamic volume 45G.SHRAVANI
46
Advantages Can be used to find shape also Rapid, routine analysis Identify high mass components even in low concentration Can analyze polydisperse samples Branching studies can be done Absolute molecular weights can be obtained 46G.SHRAVANI
47
Drawbacks There is a size window Bad response for very small molecular weights Standards are needed. Sensitive for flow rate variation. Internal standard should be used whenever possible. High Investment cost 47G.SHRAVANI
48
48
49
REFERENCES Instrumental methods of chemical analysis.B.K Sharma p.g C123-170 www.gel permeation chromatography www.gel www.ion exchange chromatography. www.ion G.SHRAVANI49