Chromatography is one of widely used separation techniques of chemical mixtures, exploiting differences in their physical/ chemical properties. All chromatography relies on the differential distribution of compounds between two phases, that are called stationary phase and mobile phase. Chromatography is used for qualitative and quantitative analysis and is also used for isolation.
5
Two Phases are involved in a adsorption LC process: Stationary Phase -Solid porous, surface-active material in small- particle form. Mobile Phase -Liquid Separation is achieved as a result of differences in the interaction between the different target compounds with the stationary phase
7
Normal Phase Mode Reverse Phase Mode Reverse Phase Ion Pairing Mode Ion Exchange Mode Chiral Separation Mode
8
Normal Phase Mode zColumn (Stationary Phase): Polar zSolvent (Mobile Phase): Non polar
9
Normal Phase HPLC Column zSilica gel Type: General use zCyano Type: General use zAmino Type: For Sugar Analysis zDiol Type: For Protein Analysis
10
Stationery Phase for Normal Phase (Silanols) HPLC Column
11
Mobile Phase for Normal Phae yPrimary solvents(non-polar) -Hydrocarbons (Pentane, Hexane, Heptane, Octane) -Aromatic Hydrocarbons (Benzene, Toluene, Xylene) -Methylene chloride -Chloroform -Carbon tetrachloride ySecondary solvents -Methyl-t-butyl ether (MTBE), Diethyl ether, THF, Dioxane, Pyridine, Ethyl acetate, Acetonitrile, Acetone, 2-propaol, ethanol, methanol 4A primary solvent is used as mobile phase. Addition of secondary solvents is to adjust retention time.
12
What is the Interaction
14
Hydrogen Bonding If the sample has -COOH, -NH 2 and -OH Hydrogen bonding become strong If the sample has balk group -due to stereo Hindrance, Hydrophobic group, non-polar part, long chain of hydrocarbon. -Hydrogen bonding become weak.
15
Retention Time and Hydrogen Bonding
16
Reverse Phase Mode zColumn (Stationary Phase): Non Polar zSolvent (Mobile Phase): Polar
17
Reverse Phase HPLC Column.C18 (ODS) Type.C8 (Octyl) Type.C4 (Butyl) Type.Phenyl Type.TMS Type.Cyano Type.Amino Type
18
Stationary Phase for Reverse Phase (C18)HPLC Column CH 3 | -0-Si-(CH 2 ) 17 -CH 3 | CH 3
19
Mobile Phase for Reverse Phase zWater (buffer) + Organic Solvents -When buffer is used, the concentration and pH are important factors -Methanol, Acetonitrile or THF are common organic solvents for RP HPLC 4Optimization of water (buffer) and organic solvents ratio is very important
20
What is the Interaction
21
Hydrophobicity zIf the sample has balk group -due to stereo Hindrance, Hydrophobic group, non-polar part, long chain of hydrocarbon. Hydrophobicity become strong. zIf the sample has -COOH, -NH 2 and -OH Hydrophobicity become weak.
22
Retention Time and Hydrophobicity
23
Ion Pairing Reverse Phase Mode
26
Effect of pH on analysis
30
Choice of LC mode
31
System Suitability Repeatability,%RSD (CV) Capacity factor, K’ Selectivity/Relative Retention Time Theoretical Plates, N (Efficiency) Resolution Asymmetry/Tailing Factor
32
Repeatability, CV (%RSD) X=(X1+X2……..Xn-1+Xn)/n N=: Number of Analysis X1..Xn: Retention time (or area or heights) X: Average C.V: Coefficient of Variation
33
Capacity Factor, K’
34
Selectivity/Relative Retention Time,
35
Theoretical Plate
36
Resolution
37
Tailing Factor,T T=W 0.05 /2f
38
HPLC Configuration Generally two types of HPLC configuration are possible *Isocratic System *Gradient system
39
Isocratic system
40
High Pressure Gradient system
41
Low Pressure Gradient system
42
Facilities of Gradient system zSome analysis required gradient mode zSave the solvents zSave labour zMethod development zAccuracy of Analysis
43
Instrumentation of HPLC Main ModulesAccessories zPump- Column Oven zInjector- Degasser zColumn zDetector zData Processor
44
Desirable HPLC High Sensitive -Low pressure fluctuation -High sensitive detector Reliable: Reproducibility - Precision of flow rate, injection volume, oven temperature, wavelength or applied potential and data transfer Flexible: Modular Type
45
Solvent delivery pump
46
Desirable Pump performance oHigh Pressure Resistance Shimadzu LC-10ATvp- 400Kgf oPrecise Flow Rate Shimadzu LC-10ATvp- Within 0.3 (%RSD<0.1) oLow Pressure Fluctuation Shimadzu LC-10ATvp- 0.5 Mpa oHigh-accuracy gradient performance oPulse-free solvent delivery oValidation and Productivity
47
Desirable Pump performance
50
Automatic Sample Injector
51
Desirable Auto Sampler
53
z Performs various pretreatment functions, including addition of sample or reagent, mixing and diluting z Performs derivatization, addition of internal standard or sample dilution z transfer reagents and sample to an empty vial for pretreatment mixing
54
Desirable Auto Sampler
55
HPLC Detectors Ultraviolet/Visible Detector (UV/Vis) Photodiode Array Detector (PDA) Fluorescence Detector (RF) Conductivity Detector (CDD) Refractive Index Detector(RID) Electrochemical Detector (ECD) Mass Spectrophotometer Detector (MS)
56
Desirable HPLC Detectors zHighly sensitive zLow Noise Level zWavelength Reproducibility zSimultaneous dual- wavelength measurement zWavelength programming zWavelength Scan mode
57
Desirable HPLC Detectors
60
Desirable Column Oven
61
Desirable On line Degasser
62
Desirable Chromatographic Workstation Software Windows based User friendly Excellent Graphical Interface GLP Compliance Automated Sequential Run System Suitability, QC Check Auto baseline Check Auto Validation Run Auto Shutdown Customize Report Automation of Analytical Works Lot of way to calculations Easy to Upgrade Easy to reinstall by the customer
63
Desirable Chromatographic Workstation Software