Recent development of gaseous position-sensitive thermal neutron detectors for the IBR-2 spectrometers A.V.Belushkin, A.A.Bogdzel, A.N.Chernikov, A.V.Churakov,

Similar presentations


Presentation on theme: "Recent development of gaseous position-sensitive thermal neutron detectors for the IBR-2 spectrometers A.V.Belushkin, A.A.Bogdzel, A.N.Chernikov, A.V.Churakov,"— Presentation transcript:

1 Recent development of gaseous position-sensitive thermal neutron detectors for the IBR-2 spectrometers A.V.Belushkin, A.A.Bogdzel, A.N.Chernikov, A.V.Churakov, T.L.Enik, S.A.Kutuzov, F.V.Levchanovski, Li En Che, E.I.Litvinenko, A.S.Nikiforov, Tz.Panteleev, V.I.Prikhodko, V.N.Shvetsov, V.V.Zhuravlev NEOKS FLNP JINR 2007

2 MWPC with delay line readout High efficiency (~80 %) High efficiency (~80 %) High count rate (~ MHz) High count rate (~ MHz) Acceptable position resolution (1.5mm) Acceptable position resolution (1.5mm) Excellent γ-background rejection (10 -8 ) Excellent γ-background rejection (10 -8 ) Two dimensional capability Two dimensional capability Long term stability (~5 yrs) Long term stability (~5 yrs) Design flexibility Design flexibility

3 Infrastructure Clean room Clean room – ISO 14644-1 Class 7 clean room Radiation stand Radiation stand – Includes Roentgen source, neutron source and electronics Wire winding machine Wire winding machine – Designed and manufactured new wire winding machine, making possible manufacturing of MWPC with active area up to 600х600 мм 2 Gas stand Gas stand Radiation stand. Photo of the operators workplace

4 2D monitor PSD 100x100 mm2 Direct neutron beam monitor Designed and constructed for FRM-2 (Germany)

5 IBR-2 measurements Spectra of the 2D monitor, measured at IBR-2 reactor Profile of the channel №10 Mask (Cd) Profile of the channel №6b Profile of the channel №10 with mask

6 IBR-2 measurements Profile of the IBR-2 channel №7a Profile of the channel №7a Size of the beam 50x160 мм 2 ± 4 mm 2 Average intensity ≈ 1,8 * 10 6 n/sm 2 *sec. Sum on Y-axis

7 1D PSD 200x80 mm 2 Multi-purpose instrument for solid-state measurements

8 IBR-2 measurements Diffraction spectra of the (La 0.1 Pr 0.9 ) 0.7 Ca 0.3 Mn 0.3 sample, measured at the channel №5. At low temperatures it separated at FM-metallic and AFM-CO-insulating mesoscopic phases. Spectrum of the MgO/(4.7nm)Fe/(4.7nm)V]10/[(1ML)Fe/(1ML)V]17/(36.5nm)V/(2nm)Pd multilayer sample measured at the refletometer REFLEX (channel №9) T=10˚K T=290˚K

9 2D PSD 225x225 mm 2 Developed with support from MinPromNauki (grant №02.452.11.7044 from 12.04.06) Testing are started

10 2D PSD 225x225 mm 2 Test measurement of 2D PSD 225x225 mm 2 Gas mix: 50 mbar 3 He+3450 mbar 4 He+1500 mbar CF 4 Anode voltage+4300 V; Drift voltage -1500 V

11 Electronics Scheme of the electronics Crate NIM with analog electronics Pre-amplifiers

12 PCI-DAQ board PCI DAQ board Two main operating modes are provided: histogram (on-line sorting of data and building of spectra) and "list" (accumulation of raw data with subsequent off-line processing). In the regime of the accumulation of histograms the effectiveness of the DAQ system is 10 5 event/s, and in the raw data mode effectiveness is up to 810 5 event/s. Created in cooperation with HMI, Berlin. It comprises 8-channel TDC of F1 type (Acam), FIFOs of different types, CPLD, FPGA, 256 Mbyte histogram memory and DSP TMS 320C6711. The DAQ board has a PCI- interface and is installed directly in the case of PC.

13 New PCI-DAQ board Projected PCI-DAQ board A new DAQ board has been designed for 1D and 2D MWPC detectors with delay line data readout. This board contains the NIM-TTL signal level converter; 8-channel integrated time-digital converter (TDC- GPX); programmed logic matrix (FPGA) comprising about 6К of logic elements; 1Gbyte histogram memory, which makes it possible to accumulate three- dimensional spectra of size of up to 512×512×1024 of 32-digit words, and high-speed interface with an optical communication link to a personal computer. Logic and time simulation of work of the board has been performed using the Quartus II package. As a result, it has been demonstrated that the rate of data reception, filtration and accumulation in the board runs as high as 2 million events/s. Real registration rate (taking into account data transfer and recording into a computer) is no less than 1 million events/s. Now a new board is in the pilot production stage.

14 Software Software complex based on SONIX+ integrated in SONIX+ system integrated in SONIX+ system allows fully automates experiment allows fully automates experiment based on the Python script language based on the Python script language Software complex based on ROOT real-time data observation real-time data observation currently used for detectors testing and in BER-II reactor (together with CAERESS) currently used for detectors testing and in BER-II reactor (together with CAERESS) based on the ROOT library based on the ROOT library

15 Results Done: Developed and tested 2D 100x100 mm 2 monitor PSDDeveloped and tested 2D 100x100 mm 2 monitor PSD Developed and tested 1D 200x80 mm 2 PSDDeveloped and tested 1D 200x80 mm 2 PSD Developed 2D 225x225 mm 2 PSDDeveloped 2D 225x225 mm 2 PSD Planned: Finished testing of the 2D 225x225 mm 2 PSD Finished testing of the 2D 225x225 mm 2 PSD Continue equipment IBR-2 spectrometers with PSD Continue equipment IBR-2 spectrometers with PSD Further improve infrastructure in order to create large-area PSDFurther improve infrastructure in order to create large-area PSD Research and development of the individual readout system for the neutron PSD Research and development of the individual readout system for the neutron PSD


Download ppt "Recent development of gaseous position-sensitive thermal neutron detectors for the IBR-2 spectrometers A.V.Belushkin, A.A.Bogdzel, A.N.Chernikov, A.V.Churakov,"