1 The study of new reconstruction method
for Brain SPECT on dual detector system 사진 속도 느리게,, Hyung Jin Lee

2 Brain spect procedure(1day protocol)
First source acetazolamide Basal acquisition Second source 애니메이션 클릭하면서 설명 TST : 40~45min Diamox acquisition

3 Brain SPECT Dual detector Short exam time Parallel collimator OSEM 2D
Triple detector Short exam time Fan-beam collimator Spatial resolution FBP-Metz filter Optical contrast FLASH 3D Resolution De-noise Contrast 1day protocol Metz filter: resolution recovery=contrast 2번째 사각형 컬러 변화,화살표 색깔 변환,.그물 빼고 Brain SPECT

4 Equivalent total exam time
1 Equivalent total exam time 2 Setting acquisition parameter 3 Parallel collimator 4 New reconstruction method 5 Reasonable image quality

5

6 Hoffman 3D Brain Phantom
Specifications - Cylinder inside diameter and height : 20.8 and 17.5 cm 100% bubble을 제거할 필요가 없는것은 머리안에 defect라고 가정하고 동일하게 acquire하였다. - Fillable volume: ~ 1.2 liter Source activity - Basal/Diamox brain SPECT : 99mTc-HMPAO 15 mCi (15/30 mCi) - 99mTc : mCi/L (=15mCi*0.05/1.2L, 투여량의 5% 뇌조직에 분포를 가정함)

7 Experiment for Volunteer
99mTc-HMPAO : 15 mCi Total count 1) Normal scan : 50 Kcounts/view (120 views : 6x106, 128 views : 6.4x106) 2) Short scan : 35 Kcounts/view (60% of Normal scan) (120 views : 4.2x106, 128 views : 4.5x106) Experiment summary *TST : total scan time Date Scanner Collimator cps Scan Time Total Count 08/08 (pm 7:10) ECAM 3 (Normal) LEHR 1.5 32sec TST : 34 min 5,931,948 08/12 (pm 8:00) PRISM Fan-beam (FB) 2.7 20sec TST : 13.3 min 6,079,541 (pm 8:20) (Short) LEAP 2.3 15sec TST : 16 min 4,116,391 08/22 (am 9:30) 1.8 TST : 21.3 min 4,699,939 *Unit : Kcount/sec (cps), sec/view (scan time)

8 *LEHR : low energy high resolution
Add Your Title 256% TST Similar TC PRISM-FB (Normal) 63% TST 79% TC Add Your Title ECAM3-LEHR (Normal) 32 sec/frame (TST=34 min) TC : 5.9×106 (5,931,948) 20 sec/frame (TST=13.3 min) TC : 6.1×106 (6,079,541) Add Your Title ECAM3-LEHR (Short) 20sec/frame (TST=21.3 min) TC : 4.7×106 (4,699,939) 75% TST Similar TC Add Your Title ECAM3-LEAP (Short) 15sec/frame (TST=16 min) TC : 4.1×106 (4,116,391) 변화조건을 더 강조 *TST : total scan time *FB : fan-beam collimator *LEAP : low energy all purpose *TC : total count *LEHR : low energy high resolution

9 What is FLASH3D? Flash 3D is an iterative reconstruction method for SPECT imaging with three dimensional resolution recovery and scatter correction Benefits: They can be used to shorten SPECT and planar acquisition times, or to improve image quality.

10 Result Drawing ROI on 6 cold spheres and 4 background regions Contrast
To do… Drawing ROI on 6 cold spheres and 4 background regions Contrast 결과 그래프 애니메이션 효과로 삽입

11 Result Drawing ROI on 4 hot spheres and 3 background regions
To do… Drawing ROI on 4 hot spheres and 3 background regions SNR=lesion average divided by background standard deviation Sphere S/R

12 Result

13 PRISM3000 vs ECAM (Hoffman 3D Phantom)

14 Hoffman-3D 비교 PRISM-FB-30sec *FBP-METZ ECAM-LEHR-15sec *FLASH3D-S8I8
**GAUSS-7mm 토탈 검사스캔타임 쓰고 이미지 강조, ECAM-LEHR-25sec *FLASH3D-S8I8 **GAUSS-7mm

15 Subset8에 iteration 변화 이미지(FLASH3D)
Non-filter OSEM2D 이미지 삽입(노말로) Post-filtering Gaussian 5mm

16 PRISM3000 vs ECAM (Volunteer image)

17 Why it has to be 20sec! Diamox 1day protocol Tc-99m dosage
15~20sec/frame Tc-99m dosage 기준카운터 즉 50,000카운터 의 65~70% 65~70% raw data

18 Result e.Cam fbp/osem/flash

19 Result E.cam/LEAP/15sec/FLASH3D E.cam/LEHR/20sec/FLASH3D
이미지 배경화면 다시,,,택스트 상자 편집 및 글자 색깔 강조 콜리메이터 E.Cam/LEUHR/20sec FLASH3D

20 Result PRISM/Fan/20SEC/METZ E.cam/LEAP/15sec/FLASH3D
E.cam/LEHR/32sec/FLASH3D 8/8/5…마지막 20SEC 영상을 강조하게 테두리 효과 E.camLEHR/20sec/FLASH3D

21 Conclusion

22 Thank You !