Diamond Sensor Tests for the CMS BCM Alexandr Ignatenko FCAL collaboration meeting May 7, 2008
2
2 FCAL collaboration meeting2 Overview BRM system overview Standard set of measurements in Zeuthen Thickness Capacitance I-V characteristic Charge collection distance Results Usual behavior Abnormal cases Conclusion
3
May 7, 2008FCAL collaboration meeting3 The components:
4
May 7, 2008FCAL collaboration meeting4 BCM1, 1.8m BCM2+BSC2, 14.4m
5
5May 7, 2008FCAL collaboration meeting5 Thickness and capacitance measurements Thickness: mechanically with 10 µm precision Values 400-430 µm A sensor assembled in the frame Capacitance: with LCR meter HP 4263A 10.7-12.1 pF
6
6 I-V characteristic N 2 atmosphere 0-500-0 V 50V/30 sec steps Voltage source and picoamperemeter: Keithley 487 May 7, 2008FCAL collaboration meeting
7
7 Charge collection efficiency = Q meas /Q ind Charge collection distance CCD = × d May 7, 2008FCAL collaboration meeting Detection mechanism traps Charge collection distance (CCD)
8
May 7, 2008FCAL collaboration meeting8 CCD setup
9
99 ~ CCD CCD = (Mean-Ped)×calib×corr/N exp May 7, 2008 FCAL collaboration meeting Typical spectrum CCD = (MPV-Ped)*calib/N exp
10
1010 May 7, 2008FCAL collaboration meeting CCD setup Source holder with collimator Sensor box Trigger box Low voltage power supply HV supply and picoampearmeter
11
Results Normal behavior P28 depumped pumped negative positive May 7, 200811FCAL collaboration meeting
12
1212 CCD vs dose, pumping May 7, 2008FCAL collaboration meeting The current in the sensor circuit was monitored CCD dependence on dose, irradiation @ 0.7 Gy/h @ 0.7 Gy/h irradiation started
13
13 13 CCD dependence on time for the fully pumped sensors May 7, 2008FCAL collaboration meeting The current in the sensor circuit was monitored After additional 60 mGy 200 V applied 400 V applied
14
Abnormal cases P18 Depumped statePumped state Currents up to hundreds of nA at high bias voltages May 7, 200814FCAL collaboration meeting
15
P 45 depumped pumped May 7, 200815FCAL collaboration meeting
16
1616 Currents had changed in time even without irradiation May 7, 2008FCAL collaboration meeting
17
1717May 7, 2008FCAL collaboration meetingConclusion 25 samples have been investigated The characterization of each sample has been given 6 sensors have shown unstable currents currents The parameters of the sensor were reproducible