Download presentation
Presentation is loading. Please wait.
Published byLambert Bridges Modified over 8 years ago
1
Optimization of Phase Contrast Imaging Luke Powers Chris Weaver Jonathan Fermo Alfred Luk BME 273, Group 22 02/23/2005
2
Phase-Contrast Radiography Traditional radiography uses differences in absorption to develop images Phase-contrast Radiography (PC-R) uses differences in refraction and diffraction in the tissue Enhanced Edge Effects
3
Potential Has potential to detect objects that are invisible on conventional radiography Edge enhancements and early detection of breast cancer Monochromatic beam reduces excess radiation dosage to patient
4
Objectives To build a device that aids in testing PC-R parameters Requires controlled movement of the object and detector
5
Importance of Position A: Radiograph of plastic vial taken with.01mm focal spot B: Radiograph of plastic vial taken with.045mm focal spot C: Phase-extracted image created from the first two images
6
Our Device Main Uses: Optimizing distances/spaces for PC imaging of specific objects/tissues Being able to reproduce images No false edge effects Additional Uses: Rotational movement for CT images Images using scattering (defines resolution ~280 microns)
7
Programming Inputs: LabView GUI: positions, energy Outputs: Time/Date Image # Angle and x, y, z positions Energy Used
8
Advisors Advisors: Frank E. Carroll, M.D. Gary Shearer Robert Traeger Principal Investigator: Edwin Donnelly, M.D., Ph.D.
9
Facilities W.M. Keck Free Electron Laser Center at Vanderbilt Vanderbilt BME Department
10
Resources Monochromatic X- ray source at FEL LabVIEW and Virtual Instruments Stages, controllers, etc… for design construction provided by FEL & outside contractors
11
Design Schematic -Detector has z-stage movement (1m) -Object has x (6cm), y (5cm), rotational (360 degrees) movement
12
Machine Shop Work Optical Flags: x, y, and z Flag and Interrupter for x-y stages
13
Machine Shop Work (cont.) Mounting Blocks Mounting Rotational Stage on x-y stage
14
Components
15
Selected Components Microcontroller Linear Translation Stage (z)
16
Questions?
Similar presentations
© 2025 SlidePlayer.com. Inc.