Protect, Enhance, and Save Lives - 1 - Characterization of a new Photon counting Detector with XRF (X-ray fluorecense) Francesca Bisello, IBA-Dosimetry
2
Protect, Enhance, and Save Lives - 2 - Dosepix: General overview of Detector Dosepix: Comparison of Calibration Methods Dosepix: Measuring Xray Spectra (preliminary results) Outline Outline Dosepix CalibrationMethods XraySpectra Conclusion
3
Protect, Enhance, and Save Lives - 3 - Hybrid Pixel Detector Single Photon Counting Detector : Time-over-Threshold Method Dosepix: General Overview Outline Dosepix CalibrationMethods XraySpectra Conclusion
4
Protect, Enhance, and Save Lives - 4 - kVp Mode (counting) Dosi Mode Dosepix: General Overview Outline Dosepix CalibrationMethods XraySpectra Conclusion
5
Protect, Enhance, and Save Lives - 5 - kVp Mode Dosi Mode (assessing energy) Dosepix: General Overview Outline Dosepix CalibrationMethods XraySpectra Conclusion
6
Protect, Enhance, and Save Lives - 6 - This is a Pixel Wise Calibration Analog Test-Pulse Calibration Methods I: Analog Test Pulse method Outline Dosepix CalibrationMethods XraySpectra Conclusion The ToT value for each Test Pulse is recorded
7
Protect, Enhance, and Save Lives - 7 - Pixel Wise Calibration Analog Test-Pulse Calibration Methods I: Analog Test Pulse method Outline Dosepix CalibrationMethods XraySpectra Conclusion The ToT value for each Test Pulse is recorded
8
Protect, Enhance, and Save Lives - 8 - Calibration Methods I: Analog Test Pulse method Outline Dosepix CalibrationMethods XraySpectra Conclusion
9
Protect, Enhance, and Save Lives - 9 - XRF Calibrations Calibration Methods II: Energy Calibration with Fluorescence line Outline Dosepix CalibrationMethods XraySpectra Conclusion
10
Protect, Enhance, and Save Lives - 10 - XRF Lines Calibration Methods III: Energy Calibration with Fluorescence line Outline Dosepix CalibrationMethods XraySpectra Conclusion Response of one pixel at different fluorescence lines
11
Protect, Enhance, and Save Lives - 11 - Interpolation and Fitting of the response of one pixel Calibration Methods IV: Determine the DOSEPIX KeV – TOT response curve Outline Dosepix CalibrationMethods XraySpectra Conclusion MaterialEnergy (KeV) Pb11.41 Ag22.10 Sn25.07 Cu8.06 W9.67
12
Protect, Enhance, and Save Lives - 12 - Interpolation and Fitting of the response of one chip (256 pixels) Calibration Methods IV: Determine the DOSEPIX KeV – TOT response curve Outline Dosepix CalibrationMethods XraySpectra Conclusion MaterialEnergy (KeV) Pb11.41 Ag22.10 Sn25.07 Cu8.06 W9.67
13
Protect, Enhance, and Save Lives - 13 - Calibration Methods V : From Analogue Test Pulse to Energy Calibration Outline Dosepix CalibrationMethods XraySpectra Conclusion Analog Test Pulse Calibration Energy Calibration
14
Protect, Enhance, and Save Lives - 14 - Calibration Methods V : From Analogue Test Pulse to Energy Calibration Outline Dosepix CalibrationMethods XraySpectra Conclusion Analog Test Pulse Calibration Energy Calibration
15
Protect, Enhance, and Save Lives - 15 - Energy Calibrated Spectra at Xray device Overlapping Method for Energy Binning Pixel 1 Pixel 2 Xray Spectra: Preliminary results Outline Dosepix CalibrationMethods XraySpectra Conclusion
16
Protect, Enhance, and Save Lives - 16 - Detector Characterization: Thr: 20-30 An TP 5-7 KeV Second order proportionality between An. Test Pulse and Energy Energy Resolution: expected to be better than 1 kV, however preliminary results not conclusive Future Work Clinical Application Conclusion Outline Dosepix CalibrationMethods XraySpectra Conclusion
17
Protect, Enhance, and Save Lives - 17 - Collaborators and partners This research project has been supported by a Marie Curie Early Initial Training Network Fellowship of the European Community’s Seventh Framework Programme under contract number (PITN-GA-2011-289198- ARDENT). I.Ritter 2, F. Tennert 2, T. Gabor 2,5, M. Campbell 3, W.S. Wong 3, S. Wölfel 1, G.Anton 2, N.Michel 4 and T. Michel 2 1- IBA Dosimetry GmbH, Bahnhofstraße 5, 90592 Schwarzenbruck, Germany 2-Erlangen Centre for Astroparticle Physics, Radiation and Detector Physics, Erwin-Rommel-Str. 1, 91058 Erlangen, Germany 3- Medipix Team, Microelectronics Group, CERN, 1211 Geneva, Switzerland 4- Customized Microelectronic Solutions, Bruckwiesenstr. 3,91220 Schnaittach,Germany 5- Fraunhofer Institute for Solar Energy Systems ISE, Heidenhofstraße 2, 79110 Freiburg im Breisgau, Germany
18
Protect, Enhance, and Save Lives - 18 - Thank you fo your attention