1 Dept. of Radiology SHIN HEUNG College & University Ph. D. Sang-Chull Ma 장소 : 대전 컨벤션센터 일자 : 2008.10.28 주최 : 한국표준과학연구원

2 2. mimic ultrasonic interactions in the human body Structure, materials Structure, materials 1. simulate a body of tissue What is an ultrasound phantom?

3 U.S Q/A phantom Quality assurance of diagnostic u.s apparatus U.S previous(training) phantom Scan technique for human organ Demonstration Research & Development Applications of U.S Phantom

4 Phantom construction 1. Phantom construction U.S Q/A phantom Ⅰ. U.S Q/A phantom HousingTarget materialsTMM phantom

5 AgarZerdine TM Urethanes Others epoxies liquids Natural materials 1)Tissue Mimicking Materials(TMM) (1) materials

6 Urethanes (2) Urethanes TMM Hydrogel type dehydration Temperature Some durable Rubber-base type Artifact of propagated speed

7 Line targets (mono-filament wire) Gray scale target (cylinder type materials) 2) Target materials 0.1mm 0.12mm0.2mm 0.08mm

8 Serve to protect the TMM Thin films Saran polyurethane Rigid Containers ABS PVC Acryl 3) Housings

9 2. Phantom design Cylindric structure Cyst – solid echo TMM combination TMM (Gary scale target) Distance accuracy Axial resolution Lateral resolution Dead zone line target (mono- filament)

10 ATS Multipurpose phantom design

11 AIUM protocol image quality indicator suggested action level depth quality indicator ≥0.6cm vertical distance accuracy Error≥1.0% horizontal distance accuracy Error≥2% or 2mm image uniformity change from baseline axial resolution change from baseline lateral resolution Error >1.0mm dead zone change from baseline cyst image change from baseline 3. Measurement & Characteristics of phantom(TMM)

12 1) Penetration of TMM (depth quality) scatter materials type(graphite) particle size, hardness, air bubble 18cm/3.5MHz 6cm/7.5MHz ATS Devel.

13 2) Vertical & horizontal distance accuracy mono-filament wire : 1-2 cm interval reverberation > 0.1mm > low echo

14 3) Homogeneity(uniformity) of TMM scatter materials type, particle size(35-80 ㎛ ) milling ATS Devel. Liver

15 4) Axial & Lateral resolution 0.1mm mono-filament wire 1.0 mm interval ~ 10.0mm interval ATS

16 5) Dead zone ATS 0.1mm mono-filament wire 1.0 mm -10.0mm

17 6) Functional resolution gray scale target : scatter type, volume ± 15dB ATSDevel.

18 Ⅱ. P revious(training)Phantoms Biposy : breast, thyroid 3D : obstetricsDoppler U.S : vessels Brachytherap y Cryosurgery

19 1. Breast phantom Sonoguide biopsy & aspiration Scan technique

20 2. Thyroid gland phantom Human thyroid Thyroid phantom Sonoguide biopsy & aspiration Scan technique

21 3. 3D U/S Phantom 3D volume probe - scan training Fetal head

22 In Summary(TMM)  Increased durability & longevity  Phantoms contain various TMM  Keep pace with developments of ultrasound apparatus  Thin Film phantom targets  Smaller resolution point targets

23  Phantoms come in a wide variety of shapes and sizes  Phantoms are utilized for a variety of purposes such as QA, training and research  Phantoms are useful tools & protocol In Summary(phantom)