Presentation is loading. Please wait.

Presentation is loading. Please wait.

Processing the Image Chapter 8

Similar presentations


Presentation on theme: "Processing the Image Chapter 8"— Presentation transcript:

1 Processing the Image Chapter 8

2 Learning Objectives Understand the purpose of the various chemicals used in x-ray processing Distinguish between manual and automated processing Be able to list the steps of film processing Understand labeling of radiographs

3 Before Processing: Markers
Right (R) and Left (L) markers are essential to identify size and position of image on radiograph. May need to label both the front and rear in large animals If label was forgotten, can add later with marker, or scratch into film while still wet.

4 Before Processing: Film Identification
Proper labeling is vital All labels should include: Name/address of hospital or veterinarian. Date radiograph was taken Patient identification including name of the owner and patient name, age, sex, and breed. Can be done via lead lettering, lead impregnated tape, or photo- imprinting labels.

5 Film Processing Turning the latent image on exposed x-ray film
into an actual image Two ways to process film: Manually Automatically Both require the same steps: Developing Washing Rinsing

6 Film Processing Techniques
Manual Automatic

7 Film-Processing Steps
1. Developing 2. Rinsing (stop bath) 3. Fixing 4. Washing 5. Drying

8 1. The Developer Converts latent image to visible image
Chemicals convert exposed silver halide to black metallic silver. Amount of metallic silver on film is amplified Little to no effect on the unexposed silver halide crystals Developer temperature affects developing time Consists of mostly water with 2 developing agents Also, activators, preservative, restrainers, hardener Should be kept around 68° F Alkaline solution

9 2. The Rinse Bath Water Purpose: Stop the developing process
Rinses excess developer from the film Prevents carryover contamination to the fixer *If film went straight from developer to fixer, then the alkaline developer would neutralize the acid of the fixer.

10 Over-rinsed film

11 3. The Fixer Purpose: Mostly water with a fixing agent Acidic solution
Removes unexposed silver halide crystals from the film Leaves developed image intact & permanent Hardens the image to the base Mostly water with a fixing agent Acidifier, hardener and a preservative Acidic solution If exhausted fixer is used: Leaves a white/pinkish, blotchy appearance to the film

12 4. The Wash Bath Flowing water Removes all waste chemicals from film
____________ – fixer is diffused into the wash water Removes all waste chemicals from film If not washed properly, image will eventually discolor and fade. Referred to as _________________. Wash in circulating water so both surfaces of film receive water continuously.

13 Under washed film

14 5. Drying Must be completely dry before storage
Drying completes the process Emulsion will not be 100% dry Would become brittle and crack May use a film drying rack or a drying cabinet Drying rack should be dust free. Drying cabinet is a heated forced- air unit that hastens the drying process Must be completely dry before storage Two wet films will permanently stick together Drying Cabinet

15 Manual Processing Low volume/low cost clinics Requirements:
Stainless steel stir sticks Protective eye wear Film hangers

16 Manual Processing: Steps
Developer: 4 minutes total Should be agitated at the beginning, middle, and end Wash: Agitated several times Fixer: 8 minutes total Can turn on the light after 2 minutes Agitate again throughout Wash: 12 minutes with flowing water Dry: Several hours Film hangers must be immersed in each solution!

17 Automatic Processing Advantages over manual method:
Standardized & consistent Produces radiograph faster Reduces labor time Safer for staff Smaller darkroom space required Film is transported through processor by rollers Rollers can cause damage to film

18 Automatic Processing The film is placed in the _________ tray and advanced into the processor slowly It must be evenly placed Once excepted, do not pull the film back out! If air is dry, can build up static electricity The film will be guided through the rollers via the __________ system The fewer the rollers, the better Work in an undulating pattern

19 Automatic Processing The chemicals are constantly refreshed via the ____________ system Based on the amount/size of film being processed Developer and fixer have their own pumps Machine dries the film too! Solutions are kept at ________ Dryer is around 131°F

20 Silver Recovery _________ removes unexposed silver halide
Placed along the waste lines 1. Glass jar with a steel wool cartridge Metallic silver will fuse to the iron in the wool 2. Electrolysis Metal plates with current flowing through them Metallic silver binds to the plates

21 Automatic Processor Chemicals need to be checked daily
Processor takes time to warm up VTI time = _________ Feed films one at a time Will be signaled by the processor to feed the next film Don’t forget to turn on vent fan!

22 Stuck in rollers of the processor

23 Film Filing Organized film filing system is needed in veterinary practice Must be placed in individual envelope and filed appropriately Should be labeled, numbered if necessary and filed correctly for future reference Should be stored ___________ X-ray logs can also help with film identification

24 Chemicals Usually concentrated and need to be diluted
Correct preparation matters, or resulting solution may adversely affect the image. Maintain at 68°F Use caution: One drop of fixer splashing into your eye can cause blindness!

25 Solution Replacement Used/old chemicals are primary cause of poor-quality radiographs. “Exhausted” Exposure to air causes oxidation Check chemicals PRIOR to development of film. Developer may turn from brown to green. In general, chemicals should be changed q 4-6 weeks.


Download ppt "Processing the Image Chapter 8"

Similar presentations


Ads by Google