Finger Plethysmograph

Similar presentations


Presentation on theme: "Finger Plethysmograph"— Presentation transcript:

1 Finger Plethysmograph
For Measuring Blood Resistivity Client: John G. Webster, Ph.D Advisor: Thomas Yen, Ph.D. Team Tim Balgemann- Team Leader Lucas Vitzthum- Communications Josh White-BSAC Sarah Offut-BWIG

2 Outline Background Problem Statement Theory Design Components
Electrical Biological Concepts Design Components Circuit Design Mechanical Future Work

3 Diabetes Body’s inability to manage glucose levels
Prevalence is on the rise 26.3 million in U.S. $116 billion excess medical expenditures annually No reliable non-invasive monitoring Hypothesized that blood sugar levels may be correlated with blood resistivity What is diabetes? Marked by irregular blood glucose levels, this is caused by either the pancreas’s inability to produce enough insulin, or the inability to effectively use insulin produced 13.5% increase in last 3 years 26.3 million in US (8% of total US population) Worldwide: 171 million individuals living with diabetes Overall economic burden much higher

4 Problem Statement Design a finger plethysmograph to measure blood resistivity Goals: Build a mechanical device to secure finger Design a circuit that allows data acquisition of blood impedance

5 Biological Theory High blood flow Low blood flow
RBCs align, current meets little resistance Low blood flow RBCs misalign, greater resistance Impedance (ΔZ) results from this Change in volume of artery Saline solution cancels signal High Flow Low Flow

6 Electrical Theory Four electrode impedance plethysmography
2 electrodes pass current through finger 2 electrodes measure voltage Signal processing Calculate impedance and resistivity from voltage output

7 Design Specifications
Circuit Sensitivity Adaptable measurement range (i.e. autoreset) Digital filtering and signal processing Mechanical Device Motion artifacts Minimal electrical signature

8 Circuitry Measuring change in blood resistivity Sample and Hold
Filtering Current amplitude and frequency Data interpretation

9 Past Semester Results Occluded artery test Δ Impedance not visible

10 Current Work- Circuit Testing
Circuit Changes Constant current source Reduced ripple Verified circuit performance Phantom circuit to mimic expected change 1 part in 1000 signal change

11 Circuit Testing

12 Past Semester Design PVC Shell Rubber Elastomeric Foam Tin electrodes
0.9% NaCl solution 10 kHz input signal, .9 mA

13 Current Work – Finger-holder
Parameters to Vary Finger-holder volume Insulator placement Electrode placement Verify signal Cuff Design I V

14 Future Work – Finger-holder
Arm-rest Generalization Hinges Electrode and insulator placement Insulator material and geometry Frequency testing versus skin impedance Phantom for calibration

15 Questions??

16 References Webster J. G., Measurement of flow and volume of blood Medical Instrumentation: Application and Design, 3rd Edition ed J G Webster (New York: John Wiley & Sons). Rosenthal, J. H., An inductive plethysmograph for ambulatory ventilation measurement. American Diabetes Association: World Health Organization: