1
Whatman New automated sample preparation system for HPLC L.P. Raman Technical Marketing Manager (Analytical Chemistry) Whatman, Inc, 9 Bridewell Place, Clifton, NJ 07013 Ingo Christ General Manager Chromsys LLC, PO Box 15131, Alexandria, VA 22309 Angelito Mendoza Chemistry Manager Gibraltar Labs, 122 Fairfield Road, Fairfield, NJ 07004
2
Whatman What is Mini-UniPrep™? Filtration & Storage device –Cuts sample preparation by 3x Unique, innovative design –500 µL Standardized –Compatible with autosamplers
3
Whatman How is Mini-UniPrep used
4
Whatman PrepBase automation platform Mini-UniPrep Rack (84 Slots) Diluter Compressor Stirrer Injection port
5
Whatman Automated sample prep used for... QC of Pharmaceutical tablets –Content uniformity (20) –Dissolution testing (12) –Blending (6) –QC of Raw materials (3) Assay of Cyclobenzaprine HCl Tablets
6
Whatman Automated Mini-UniPrep used for... Cosmetics QC of make-ups (lipstick, foundation) Sun-screen Pharmaceutical Topical Creams Arthritis Cream (Capsaicin) R&D Solubility studies Food and Beverage Juice Analysis of sugar content Label conformity Vitamin analysis
7
Whatman Results from our study Regular USP Vs Mini-UniPrep Automated Sample Prep –Assay of Cyclobenzaprine HCl Tablets »Performance results similar 116 % 25 % 12 % - Lab productivity - Labor costs - Consumables
8
Whatman Mini-UniPrep - HPLC Extractables
9
Whatman Cyclobenzaprine HCl - USP 27 NF:22 Finely powdered Tablets Transfer 10 mg Cyclobenzaprine to a flask Add 150 mL of 0.1 N HCl Stir for 15 minutes Dilute with 0.1 N HCl to 200 mL Filter Inject 10 µL into HPLC Take 5 mL in Syringe Place in standard 2 mL Vial In Autosampler Finely powdered Tablets Transfer 2.5 mg Cyclobenzaprine to MUP Add 0.5 mL of 0.1 N HCl Stir for 5 minutes Inject 10 µL into HPLC Compress MUP Place in standard in Autosampler
10
Whatman Mini-UniPrep automation platform (1) - Add HCl (3) Compress (2) Mix
11
Whatman Results (1) - Precision of method Objective - Mini-UniPrep Sample Prep provides precise results 5 vials of the standard prepared Mini-UniPrep Peak area of each injection was measured.
12
Whatman Results (2) - Linearity of method Acceptance: Correlation coefficient < 0.99; RSD - < 4% Objective - Establish linearity of Mini-UniPrep Sample 4 known dilutions of the standard prepared Mini-UniPrep Peak area of each injection was measured.
13
Whatman Results (3) - Sensitivity of method Objective: Mini-UniPrep Sample preparation provides good recovery at varying concentrations Tested at 3 spiked levels Acceptance criteria Recovery97-103% RSD < 4%
14
Whatman Results (4) - Automated Vs Regular
15
Whatman Results (4) - Automated Vs Regular Regular Method Automated Method
16
Whatman Automated Mini-UniPrep - Performance Data Summary Robust Scalable Sensitive Comparable to manual method
17
Whatman Benefits of automated Mini-UniPrep
18
Whatman Economic analysis …
19
Whatman Conclusions Mini-UniPrep™-PrepBase automated system Comparable to regular USP method Reduces sample preparation time Increases productivity To see Automated Min-UniPrep System in action come to Whatman Booth # 3479