SVD Slow Control Markus Friedl (HEPHY Vienna) DAQ/SC Kickoff Meeting, 8 November 2012
2
8 November 2012M.Friedl (HEPHY Vienna): SVD SC2 Introduction Slow Control Summary
3
SVD 8 November 2012M.Friedl (HEPHY Vienna): SVD SC3 APV25 chips Cooling pipe Origami ladder Sensor underneath flex circuit Pitch adapter bent around sensor edge End ring (support)
4
Belle II SVD Key Parameters 4 layers (L3…L6) of double-sided silicon strip detectors Straight barrel part with one slanted forward sensor (L4…L6) 1.14 m 2 silicon 223,744 channels (twice as many as in Belle I SVD) Cluster Signal/Noise = 14 (p-side), 24 (n-side) Time resolution = 2 (n-side), 3 (p-side) ns RMS Occupancy < 5% 30kHz maximum trigger rate with dead time <0.5% CO 2 cooling 0.55% X 0 overall material budget per layer 8 November 2012M.Friedl (HEPHY Vienna): SVD SC4
5
Opened SVD 8 November 2012M.Friedl (HEPHY Vienna): SVD SC5 Shown: Fully populated L3 One ladder for each L4…L6
6
8 November 2012M.Friedl (HEPHY Vienna): SVD SC6 Introduction Slow Control Summary
7
General Assumptions (open for discussion) Run Control (initialize, start run, stop run, internal calibration, …) Handled internally by SVD following global signals presumably issued through NSM Data quality monitoring (checking pedestals, noise, hit rates etc. during run) Handled by DAQ and/or SVD internal monitoring in parallel “Classical” Slow Controls (voltages, currents, temperatures, … – typically done with LabView on a smaller scale) Handled by EPICS 8 November 2012M.Friedl (HEPHY Vienna): SVD SC7
8
Slow Controls – Power Kenwood Power Supplies (GPIB interface + PC) 96 LV + 18 HV channels each with Set voltage, measured voltage, current limit, measured current Need to turn on/off LV and HV, read and clear error DOCK Box (controlled/measured through FADCs) 384 DC/DC converters need to be enabled/disabled Monitoring each converter output voltage and power good Measuring bias currents of each sensor (384 channels) Optional: measuring one temperature on each of the 48 FADCs 8 November 2012M.Friedl (HEPHY Vienna): SVD SC8
9
Slow Controls – Temperature CO 2 cooling plants Cooling plants are systems including their own sensors Manifolds Are the valves and sensors (including mass flow meters?) controlled/read out by CO 2 plants or do we install another PLC (or something else)? Front-End – not really defined yet Will need rad-hard temperature sensors (presumably Pt100[0] or thermocouples) for Cooling lines Environment 8 November 2012M.Friedl (HEPHY Vienna): SVD SC9
10
Monitoring – Front-End Not defined yet, but can assume a O(10) for each sensor type Presumably also interesting for PXD, thus we should make common effort (as already started with humidity) Humidity Problem: radiation hardness – maybe optical fiber? Backup solution: sensors in exhaust pipes outside detector Radiation Instantaneous dose rate? – PIN diodes? Total dose – RADFETs Distance (Omegas) between PXD and SVD Anything else? 8 November 2012M.Friedl (HEPHY Vienna): SVD SC10
11
8 November 2012M.Friedl (HEPHY Vienna): SVD SC11 Introduction Slow Control Summary
12
EPICS for the Belle II slow control I assume that Run Control and Data Quality Monitoring are not part of EPICS Tasks for EPICS slow control LV Power & bias switching and monitoring Temperature monitoring (CO 2 plants, valves, front-end) Humidity Radiation Distance (Omegas) Anything else? 8 November 2012M.Friedl (HEPHY Vienna): SVD SC12