Presentation is loading. Please wait.

Presentation is loading. Please wait.

Case report: The application of BIS (bispectral index monitor) during the operation Present by: Ri 陳賢典 Ri 詹宗諭 Supervisor: CR 郭書麟.

Similar presentations


Presentation on theme: "Case report: The application of BIS (bispectral index monitor) during the operation Present by: Ri 陳賢典 Ri 詹宗諭 Supervisor: CR 郭書麟."— Presentation transcript:

1 Case report: The application of BIS (bispectral index monitor) during the operation Present by: Ri 陳賢典 Ri 詹宗諭 Supervisor: CR 郭書麟

2 Brief history Gender: male Age: 25 y/o OP & anesthetic history: 92/01 in NTUH, Left lung pneumothorax s/p chest tube insertion. Present illness: Left lung pneumothorax, came for VATS and pleurodesis this time.

3 a b c d e f g h i j k l mno pq r abcdefghijklmnopqr

4 Morning Meeting BISPECTRAL INDEX MONITOR (BIS) 日期: 2003.09.19 報告者: Ri 陳賢典 Ri 詹宗諭 指導者: CR 郭書麟

5 Deep (or Adequate) of anesthesia

6 Bispectral index Index for hypnotic status manufactured by Aspect Medical Systems (Newton, USA) a continuous processed EEG parameter that provides a measure of the state of the brain computed real-time using a combination of three analysis steps BIS value lags 5 to 10 seconds behind the clinical change in the patient. a score number between 0 and 100 100 → awake ; 0 → Isoelectric EEG

7 BIS

8 BIS: Electrode and location

9 BIS development

10 BIS Range Guideline

11 BIS vs EEG feature

12 idealized pattern of changes in the EEG

13 Bispectral Index in Response to Drugs and Stimulation Oscillations in BIS can potentially provide a direct measure of the brain's response to stimulation and an alternative measure of responsiveness / reactiveness similar to other vital sign endpoints.

14 Physiological factors may also alter the hypnotic state Ischemia or severe hypoxia, resulting in a decrease in the BIS if severe enough to cause global EEG slowing or outright suppression Hypothermia (<33 ℃ ) will generally result in a corresponding decrease in BIS levels as brain processes slow Artifacts (EMG … ) BIS monitoring usually will not detect episodes of focal ischemia caused by embolic events.

15 The Benefit of BIS use (1)

16 The Benefit of BIS use (2) More Precise Titration of Anesthetic Agents More appropriate use of Adjuvant Drugs Faster, More Consistent Emergence and Extubation Better OR Utilization (i.e. less time spent waiting for extubation / transport to PACU) Shorter PACU Stays Reduces the Risk of Intraoperative Awareness

17 Awareness during general anesthesia

18 Incidence of awareness during elective general anesthesia ★ elective and emergency surgery:0.2~0.4% ★ BIS use: 0.003% most awareness with BIS : BIS number>65 So BIS reduce the risk of awareness? 1. It is difficult to prove that BIS can reduce this risk 2. for high risk group use, such as cardiac surgery, C/S

19 BIS and Immobility Initially, the BIS was evaluated “ as a measure of adequate anesthesia defined by patient movement in response to skin incision. ” However,when sevoflurane was administered alone, the BIS did not predict movement after skin incision better than chance alone. BIS has not been derived to predict movement

20 BIS For Pediatric Use The influence of neuronal and physiologic maturation of the brain on BIS, as well as its correlation to drug effects and anesthetic outcome, is unknown in pediatric patients an approximately linear relation between BIS and end-tidal sevoflurane concentration in infants and children BIS values obtained before anesthesia, during maintenance, and at emergence will be similar to those reported for adults

21 Current Limitation of BIS monitor Drug exception: nitrous oxide, Ketamine insufficient data to evaluate the use of BIS in patients with neurologic disease Data are currently lacking regarding opioid dose – responses and interaction of opioids with hypnotics on BIS Could not detect cerebral ischemia EMG contamination (artifacts)

22 Monitored Anesthesia Care and Intensive Care BIS values of 65 – 80 define an acceptable loss in conscious information processing and recall during sedation – hypnosis preliminary data from the ICU suggest that oversedation is common The influence of multi-system failure on hypnotic pharmacokinetic – pharmacodynamic response is unpredictable in these patients It would probably benefit patients and speed recovery from long-term sedation to accurately monitor and titrate hypnosis in the ICU

23 Intra-OP response,BIS and management

24 The END ………….

25 Combination of three analysis steps 3 steps: First step – an EEG pre-processor, which breaks the EEG signal down second by second and marks those segments containing artifact First step Second step - the calculation of the hypnosis /sedation index by combining selected EEG features using the algorithm Second step -hypnosis /sedation index Third step – The index is modified to better reflect the level of suppression in the EEG. The suppression ratio (SR) is computed as the percentage of suppressed EEG in the non-artifact data Third step

26 Hypnotic state/sedation level scoring criteria

27 The Bispectral Index development process


Download ppt "Case report: The application of BIS (bispectral index monitor) during the operation Present by: Ri 陳賢典 Ri 詹宗諭 Supervisor: CR 郭書麟."

Similar presentations


Ads by Google