Download presentation
Presentation is loading. Please wait.
Published byBriana Jenkins Modified over 8 years ago
1
Feb 12 2007 C.Smith UVA EC energy calibration – g13 pass0 For pass0 data were cooked with CALDB calibration constants reset to nominal 10 channels / MeV –HV have not been changed since 2004 –Good opportunity to check absolute calibration of PMT without software corrections. In the early years absolute calibration based on cosmic muon runs, while minimum ionizing pions from data used to cross-check and make run dependent adjustments. More recently pion data used for absolute energy calibration Need to check consistency between pion and muon calibrations.
2
Feb 12 2007 C.Smith UVA Wrong pedestals for ADC in slot18 ADC swapped out after Run 53296 New pedestal run performed not until Run 53329 - two days later ! This means between 53296 - 53329 ADC in slot 18 was loaded with incorrect sparcification tables EC energy calibration – g13 pass0 On-call detector experts should insure that new hardware is updated promptly before data-taking resumes.
3
Feb 12 2007 C.Smith UVA Gain matching is reasonable but absolute calibration is low. Part of this is continued PMT gain drift since HV last adjusted in 2004. Another component is due to outdated attenuation length corrections (actual attentuation lengths are smaller than assumed). Currently Avto recooking several runs at my request to iterate software gain adjustments. EC energy calibration – g13 pass0
4
Feb 12 2007 C.Smith UVA Flash ADC SIS3320 8 Channel 200 MHz 12-bit ADC / Wave Form Digitizer Sector 5 inner V13 - V20 3:1 UVA splitter For details see Serguei Pozdniakov, CLAS NOTE 2006-023 2/3 1/3 EC Cosmics Test Setup
5
Feb 12 2007 C.Smith UVA Flash ADC Studies – Digitized PMT Pulses 200 time samples per event per channel Signal summed over time samples 13-29 Remaining ~ 182 samples used to calculate baseline (pedestal) and subtracted event-by-event. Direct hit on single pixel Sector 5 Inner V13-16
6
Feb 12 2007 C.Smith UVA Two day cosmic run in January Cosmic run 53882 Saved FADC raw banks 2 GB file ~ 10 minutes 272 files…almost 2 days Missed events in FADC ? 1881 sparcification threshold
7
Feb 12 2007 C.Smith UVA Possible non-linearity Comparison of LC1881 and SIS3320 – charge
8
Feb 12 2007 C.Smith UVA 1881 ADC Channel SIS 3320 SUM Comparison of LC1881 and SIS3320 - noise 1881 noise (ac coupled) SIS 3320 noise Main issue for calibration purposes is noise relative to mip peak. Use pixel cut to isolate mip peak in cosmic runs
9
Feb 12 2007 C.Smith UVA Pixel cut for minimum ionizing muon peak u v w inner outer Acceptance fraction ~0.0009
10
Feb 12 2007 C.Smith UVA Pulse height versus distance to PMT - LC1881 and SIS3320 sigma / centroid = 0.30 sigma / centroid = 0.27
11
Feb 12 2007 C.Smith UVA Pulse height versus distance to PMT - LC1881 and SIS3320 sigma / centroid = 0.32 sigma / centroid = 0.33 Shapes of mip peaks similar – somewhat noisier FADC makes negligible contribution to sigma of mip distribution.
12
Feb 12 2007 C.Smith UVA Absolute light yield EC laser pulses attenuated to match ADC range of cosmic muons Measured sigma of pulse height distributions Two components of sigma: –Laser fluctuations –Poisson statistics Poisson component (and number of photoelectrons) could be separated using linear fit to sigma/centroid data Yield of 3-4 p.e. / MeV obtained
Similar presentations
© 2024 SlidePlayer.com. Inc.
All rights reserved.